Abstract
BACKGROUND: Aspiration of colonized oropharyngeal secretions is a major factor in the pathogenesis of ventilator-associated pneumonia (VAP). A tapered-cuff endotracheal tube (ETT) has been demonstrated to reduce aspiration around the cuff. Whether these properties are efficacious in reducing VAP is not known.
METHODS: This 2-period, investigator-initiated observational study was designed to assess the efficacy of a tapered-cuff ETT to reduce the VAP rate. All intubated, mechanically ventilated patients over the age of 18 were included. During the baseline period a standard, barrel-shaped-cuff ETT (Mallinckrodt Hi-Lo) was used. All ETTs throughout the hospital were then replaced with a tapered-cuff ETT (TaperGuard). The primary outcome variable was the incidence of VAP per 1,000 ventilator days.
RESULTS: We included 2,849 subjects, encompassing 15,250 ventilator days. The mean ± SD monthly VAP rate was 3.29 ± 1.79/1,000 ventilator days in the standard-cuff group and 2.77 ± 2.00/1,000 ventilator days in the tapered-cuff group (P = .65). While adherence to the VAP prevention bundle was high throughout the study, bundle adherence was significantly higher during the standard-cuff period (96.5 ± 2.7%) than in the tapered-cuff period (90.3 ± 3.5%, P = .01).
CONCLUSIONS: In the setting of a VAP rate very near the average of ICUs in the United States, and where there was high adherence to a VAP prevention bundle, the use of a tapered-cuff ETT was not associated with a reduction in the VAP rate.
Footnotes
- Correspondence: David L Bowton MD, Department of Anesthesiology, Wake Forest Baptist Health, Medical Center Boulevard, Winston-Salem NC 27157-1009. E-mail: dbowton{at}wakehealth.edu.
This research was partly supported by Covidien. See Methods for details. Drs Bowton and Hite were partly supported by grants from the National Institutes of Health. Drs Martin and Sherertz have no disclosed no other conflicts of interest.
See the Related Editorial on Page 1704
- Copyright © 2013 by Daedalus Enterprises