Article Figures & Data
Figures
These pictures display the PneuView 3 graphics diaplyed at each of the three bench model comparisons.
Tables
- PneuView 3 Measurements for Bench Model Comparison
VT (mLs) PIP (cmH20) PEEP (cmH20) Control (NIV w/ Non-Vented Mask) 11.4 9.8 4.5 14.7 10 5 13.8 10.2 5.1 Control Average 13.3 10 4.86 Bench Model A (Metaneb at dry side of heater) 24.2 11.2 2.2 24.7 13 2 23.8 11.5 2.1 Bench A Average 24.2 11.9 2.1 Bench Model B (Metaneb between mask and circuit) 17.9 9.8 1.9 15 9.9 2.3 16.9 9.3 2.4 Bench B Average 16.6 9.6 2.3 Bench Model C (Direct to mask) 10 13.5 2.2 9.7 13.3 2.3 9.6 13.4 2.2 Bench C Average 9.7 13.4 2.2