Article Figures & Data
Tables
- IPV Bench Model Comparisons
ADULT CIRCUIT Phasitron Tee deadspace= 27mL, Spring Tee deadspace= 20mL Total PEEP (cmH20) PIP (cmH20) MV (L/min) Vti (mL) Vte (mL) Servo I Control 10 22 1.9 112 104 Position A (IPV at wye w/ phasitron tee) 12 (STD 1.41) 17.3 (STD 3.32) 7.5 (STD 3.95) 201 (STD 62.93) 163.7 (STD 42.21) Position B (IPV at heater w/ spring tee) 14.3 (STD 3.04) 23.3 (STD 0.91) 9.9 (STD 5.65) 371.7 (STD 193.63) 322.3 (STD 154.36) Position C (IPV at heater w/ phasitron tee) 12 (STD 1.41) 17.3 (STD 3.32) 7.1 (STD 3.67) 182.6 (STD 49.9) 136.3 (STD 22.83) INFANT CIRCUIT Adapters used from circuit, NIF-T, Bag Adapter- total deadspace= 45mL Servo I Control 10 14 1.9 115 96 Position A (IPV at wye w/ phasitron tee) 14.6 (STD 3.25) 18 (STD 2.82) 7.9 (STD 4.24) 166 (STD 36.06) 65.6 (STD 21.49) Position B (IPV at heater w/ spring tee) 22 (STD 8.48) 28 (STD 9.89) 9.3 (STD 5.23) 334.3 (STD 155.06) 216.3 (STD 85.06) Position C (IPV at heater w/ phasitron tee) 11.3 (STD 0.91) 15 (STD 0.70) 6.2 (STD 3.04) 39.6 (STD 53.31) 63 (STD 23.33) Each value shows the average over three identical trials.