Clinical Implementation of Automated Oxygen Titration in a Tertiary Care Hospital ================================================================================= * Pierre-Alexandre Bouchard * Geneviève Parent-Racine * Cassiopée Paradis-Gagnon * Mathieu Simon * Yves Lacasse * François Lellouche * François Maltais ## Abstract **BACKGROUND:** When treating acute respiratory failure, both hypoxemia and hyperoxemia should be avoided. SpO2 should be monitored closely and O2 flows adjusted accordingly. Achieving this goal might be easier with automated O2 titration compared with manual titration of fixed-flow O2. We evaluated the feasibility of using an automated O2 titration device in subjects treated for acute hypoxemic respiratory failure in a tertiary care hospital. **METHODS:** Health-care workers received education and training about oxygen therapy, and were familiarized with an automated O2 titration device (FreeO2,). A coordinator was available from 8:00 am to 5:00 pm during weekdays to provide technical assistance. The ability of the device to maintain SpO2 within the prescribed therapeutic window was recorded. Basic clinical information was recorded. **RESULTS:** Subjects were enrolled from November 2020 to August 2022. We trained 508 health-care workers on the use of automated O2 titration, which was finally used on 872 occasions in 763 subjects, distributed on the respiratory, COVID-19, and thoracic surgery wards, and in the emergency department. Clinical information could be retrieved for 609 subjects (80%) who were on the system for a median (interquartile range) of 3 (2-6) d, which represented 2,567 subject-days of clinical experience with the device. In the 82 subjects (14%) for whom this information was available, the system maintained SpO2 within the prescribed targets 89% of the time. Ninety-six subjects experienced clinical deterioration as defined by the need to be transferred to the ICU and/or requirement of high flow nasal oxygen but none of these events were judged to be related to the O2 device. **CONCLUSIONS:** Automated O2 titration could be successfully implemented in hospitalized subjects with hypoxemic respiratory failure from various causes. This experience should foster further improvement of the device and recommendations for an optimized utilization. * respiratory failure * oxygen supplementation * automated oxygen titration * hypoxemia * hyperoxemia * oxygen saturation ## Introduction Oxygen supplementation is ubiquitous in hospitalized patients with hypoxemic respiratory failure. Traditionally, the main concern of clinicians was to alleviate hypoxemia with little concern for hyperoxemia, except in the neonatal population1 and in COPD, where avoidance of hyperoxia to protect against respiratory acidosis was recommended.2 However, the appreciation that hyperoxemia may also be harmful in other conditions, such as sepsis and myocardial infarction, and after emergency surgery,3 has led to recommendations that ![Formula][1] should be maintained within therapeutic zones that vary according to the cause of respiratory failure. For example, the British Thoracic Society has proposed ![Formula][2] targets of 94 to 98% for patients who are the most acutely ill or 88 to 92% in those at risk of hypercapnic respiratory failure.4 Although guidelines for oxygen therapy are not uniform,4,-,8 their implications are that ![Formula][3] should be monitored closely and O2 flows adjusted frequently to maintain patients within a relatively narrow therapeutic window. Achieving this goal with manual O2 titration is labor intensive and often not feasible in the context of routine care.9 As such, it is common that patients are found outside the desired range of ![Formula][4] , both in the hypoxemic and hyperoxemic ranges.10-12 Automated O2 titration with devices that are based on closed-loop algorithms has been developed with the objective of maintaining ![Formula][5] by providing continuous adjustment of O2 flow in the context of fluctuating oxygen requirements.13,14 These devices have been shown to increase the proportion of time spent in the desired ![Formula][6] range in various clinical situations, including hospitalized subjects with COPD exacerbation,11,14 subjects in the emergency department,12 subjects after surgery,15,16 and subjects in COVID-19–related hypoxemic respiratory failure.17 Automated O2 titration has also been shown to be effective in situations in which rapid adjustments of O2 flows is required, for example, during exercise.18-20 It might also accelerate weaning from oxygen and hospital discharge,11,12,21 with potential reduction in hospitalization costs.22 Automated O2 titration has been implemented with high-flow nasal cannula, showing efficacy of the system to maintain ![Formula][7] within the desired target zone during walking exercise in patients with COPD and in the context of hypoxemic respiratory failure.23,24 By reducing the requirement for direct contacts between hospital workers and patients because the O2 flows are automatically adjusted, the risk of transmission of contagious disease is reduced.25 This is an important consideration given the scarcity of hospital workers, overloaded health-care systems, and the high risk of hospital transmission of pathogens25,26 among health-care workers during the recent pandemic.27 The cumulative experience with automated O2 titration has been obtained in the research context, data also needs to be obtained in the clinical setting. We evaluated the feasibility of using this technology in the clinical care of subjects treated for acute hypoxemic respiratory failure in a tertiary care hospital. ### Quick look #### Current Knowledge It is recommended that ![Formula][8] be closely monitored to avoid hypoxemia and hyperoxia when treating patients with acute respiratory failure. Maintaining ![Formula][9] within a relatively narrow therapeutic window may be challenging with manual O2 titration, the standard of care in oxygen therapy. Automated O2 titration systems are currently being developed to reach this objective but their use is mostly limited to research settings. #### What This Paper Contributes to Our Knowledge Automated O2 titration was implemented safely in the clinical care of subjects treated for hypoxemic respiratory failure in a tertiary care hospital. We also found that the ability of automated O2 titration to maintain ![Formula][10] within the desired range in various diseases may also apply to “real life” clinical situations. ## Methods In 2019, a multidisciplinary committee advised hospital administrators on the possibility of implementing automated O2 titration within the context of clinical care. In 2020, 30 automated O2 titration devices were acquired by the hospital to treat patients in the emergency department or the hospital wards with various forms of hypoxemic respiratory failure, including COPD exacerbation, exacerbation of interstitial lung disease, viral (COVID-19) or bacterial pneumonia, and heart failure. This was done in the context of a prospective observational technological evaluation that took place between September 2020 and August 2022 at the Institut universitaire de cardiologie et de pneumologie de Québec, a 330-bed tertiary care, university-affiliated hospital with specialization in cardiology and respiratory medicine. When possible, clinical information, including etiology of respiratory failure, duration of O2 therapy, and hospital length of stay were noted. Prospectively collecting clinical data on the participants was exempted from ethics committee review by the institution because this was considered to be part of routine clinical care. The permission to use the clinical data anonymously for a scientific report was granted by the medical director of the hospital with a waiver of consent from the institutional ethics review board, when considering that this evaluation was done in the context of clinical care. The manufacturer had no role in data collection, interpretation, and presentation. The automated O2 titration device used in this technological evaluation (FreeO2, Oxynov, Québec City, Canada) relies on continuous ![Formula][11] recording to feed a closed-loop algorithm, which allows automatic O2 flow titration to maintain ![Formula][12] within a target window prescribed by the clinician.13 The device is coupled with a finger sensor linked to an embedded pulse oximeter (OEM III Module, Nonin Medical, Plymouth, Minnesota). The automated O2 titration device has 3 operating modes, the closed-loop mode being the primary operating mode, whereby O2 flow is changed based on measurement of ![Formula][13] . In the closed-loop mode, O2 flow is automatically titrated based on the difference between the real-time ![Formula][14] and the target value by using a proprietary proportional integral algorithm with an O2 flow command adjustment rate of once per second to achieve or maintain a pre-set ![Formula][15] level. O2 flow is limited to 0.1 to 20.0 L/min. O2 flow may increase or decrease so as to maintain a stable ![Formula][16] . The constant-flow mode is comparable with a standard O2 regulator, with the device only providing fixed O2 flow as set by the attending physician, between 0.1 and 20.0 L/min. In the acquisition mode, the device only monitors the oximeter readings (![Formula][17] , breathing frequency, and heart rate) without any oxygen being provided. Data are visible on the front screen and captured in the device memory, from which it can be retrieved as long as the corresponding subject has been appropriately identified in the system (name, date of birth) when initiating therapy. However, this was not mandatory to initiate treatment with O2 titration device, and this step was often overlooked by health-care workers. The implementation of automated O2 titration in clinical care of the hospital followed a multistep process to acclimate hospital workers to its use. A multidisciplinary implementation committee composed of nurses, respiratory therapists, physicians, a physiotherapist, and a patient representative oversaw the implementation of the devices and made recommendations about use (Table 1). A coordinator (PAB), helped by a senior nursing consultant (GPR) and one physician (FL), trained hospital workers (nurses, respiratory therapists, physicians) in various aspects of oxygen therapy and about the use of automated O2 titration. Training sessions were planned to be done in person but due to infection control measures in the COVID-19 pandemic, they were delivered remotely. The teaching material remained available for subsequent consultation if needed. The main topics covered were the following: (i) update on oxygen therapy, including the importance of avoiding hypoxemia and hyperoxemia; (ii) prescribing oxygen with lower and upper ![Formula][18] boundaries in a specific patient population and revision of the current guidelines; and (iii) practical use of automated O2 titration, including monitoring of subjects with the device, accurate patient selection, accurate ![Formula][19] target, and potential issues associated with improper use of the device. A key learning objective was the detection of clinical deterioration with automated O2 titration. With fixed O2 flow, clinical deteriorations are detected when ![Formula][20] worsens. With automated O2 titration, ![Formula][21] remains stable during disease instability. This is due to the intrinsic ability of automated O2 titration to maintain ![Formula][22] within the target zone as long as the maximum O2 flow allowed by the device is not surpassed. With automated O2 titration, clinical deteriorations, rather, are detected when requirements for O2 flow increase to maintain the target ![Formula][23] , which is a major change of practice for health-care workers. View this table: [Table 1.](http://rc.rcjournal.com/content/69/9/1081/T1) Table 1. Composition and the Role of the Members of the Multidisciplinary Team The coordinator (PAB) oversaw the use of automated O2 titration from 8:00 am to 5:00 pm during weekdays to address any concerns or technical questions with its use. In addition, 4 advanced practice nurses were available to accompany the hospital workers in the early weeks of the project when familiarity with the system had to be developed. A clinical protocol was developed by advanced practice nurses and respiratory therapists to address practical issues with the use of automated O2 titration (Table 2). This document covered the following topics: (i) contraindications to automated O2 titration, (ii) initiation of the device, (iii) setting of the clinical parameters of the device according to the clinical situation, (iv) how to mobilize subjects on the device, (v) clinical surveillance for nurses and respiratory therapists of patients on automated O2 titration, and (vi) weaning from automated O2 titration. View this table: [Table 2.](http://rc.rcjournal.com/content/69/9/1081/T2) Table 2. Clinical Care Protocol for the Installation of Automated O2 Titration Patients admitted to the emergency department, respiratory and thoracic surgery wards, or COVID-19 unit with a diagnosis of hypoxemic respiratory failure were potentially eligible to be treated with automated O2 titration when a device was available. Patients with one or more of the following characteristics were not considered for automated O2 titration: (i) requirement for >8 L/min of O2 to obtain a ![Formula][24] of ≥92% on the respiratory ward or emergency department or > 6 L/min of O2 to obtain an ![Formula][25] of ≥90% on the COVID-19 unit; (ii) requirement for noninvasive ventilation, high-flow nasal cannula, imminent endotracheal intubation, or cardiac arrest; (iii) impossibility of measuring ![Formula][26] (poor peripheral perfusion, Raynaud, scleroderma); or (iv) agitation and/or absence of collaboration. We did not collect information about patients who were not considered for automated O2 titration. The proposed ![Formula][27] targets were 88 to 90% for COPD and 90 to 92% for other causes of hypoxemia but the final decision belonged to the physician. These ![Formula][28] targets are lower than typically recommended4-8; they were selected based on our findings that the built-in oximeter that was used with the automated O2 titration device (OEM III Module; Nonin Medical) systematically underestimates arterial oxygen saturation (![Formula][29] )28 and with the objective of avoiding hypoxemia and hyperoxemia. Weaning from automated O2 titration was proposed when O2 flow was ≤ 1 L/min. O2 flow, ![Formula][30] , and breathing frequency alarms were set at 8 L/min, 85%, and 40 breaths/min, respectively. The use of nasal cannula was recommended when O2 flows were < 5 L/min; oxygen masks (OxyMask [Medline, Northfield, Illinois] or a simple O2 mask) were used when O2 flows were ≥ 5 L/min or if it was more comfortable. Health-care workers had the opportunity to answer a short anonymous survey about their experience with automated O2 titration. The survey was available for one day, covering the 3 working shifts. The following questions were asked: (i) would you consider using an automated O2 titration device for your patients on oxygen, rarely, occasionally, or often; (ii) did you receive sufficient technical support for the use of automated O2 titration, yes or no; and (iii) on a 0 to 10 scale, 0 being completely useless and 10 being the most useful, how do you rate the clinical utility of automated O2 titration? We report data to support the feasibility of using an automated O2 titration device in clinical practice, including the number of subjects for each hospital site (respiratory ward, COVID-19 unit, emergency department, thoracic surgery), etiology of respiratory failure, duration on automated O2 titration, hospital length of stay, ![Formula][31] targets, and ![Formula][32] data. Clinical deterioration, defined as the need for high-flow nasal oxygen and/or transfer to the ICU, was recorded. Whether this was associated with the use of automated O2 titration or to progression of the underlying disease was documented from the medical chart or by discussing with the attending physician. We did not pre-specify the number of participants; this was determined by the duration of the project and by the availability of O2 titration devices. Some subjects were treated twice during the same hospitalization; when this happened, we only reported the first use. Categorical variables are presented as absolute or relative frequencies and were analyzed by using the Fisher exact test. Continuous variables are expressed as mean ± SD or as median (interquartile range [IQR]) according to the variable distribution. The between-group comparison for continuous variables was to perform a one-way analysis of variance. The normality assumption was verified with the Shapiro-Wilk tests by using residuals from the statistical model. The Brown and Forsythe variation of the Levene test statistic was used to verify the homogeneity of variances. Length of hospital stay was log-transformed to fulfill the normality and variance assumptions. We used the Wilcoxon rank-sum test to compare groups when the normality and variance assumptions were rejected. Statistical significance was present with a 2-tailed *P* < .05. Analyses were performed by using SAS version 9.4 (SAS Institute Inc, Cary, North Carolina). ## Results The automated O2 titration system began to be used on the respiratory ward in November 2020, on the COVID-19 unit in January 2021, in the emergency department on April 2021, and, finally, on the thoracic surgery ward on March 2022, with the last subject enrolled on August 24, 2022 (Fig. 1). A total of 508 health-care workers were trained to the use of automated O2 titration, including 399 nurses and respiratory therapists, 16 physiotherapists, 28 nurse assistants, 27 physicians (14 emergency physicians and 13 pulmonologists), and 38 medical residents. The flow of subjects is presented in Figure 2. O2 therapy was administered with the automated O2 titration device on 872 occasions in 763 subjects distributed on the respiratory, COVID-19, and thoracic surgery wards, and in the emergency department (Fig. 3). Of this number, we could retrieve clinical information on 609 subjects who were in the system for a median (IQR) of 3 (2-6) d, which represented 2,567 patient-days of clinical experience with the device. The characteristics of these subjects are presented in Table 3. The mean ± SD age of the subjects was 72 ± 12 y, median (IQR) hospital length of stay was 9 (6-15) d. Automated O2 titration was started at a median (IQR) of 0 (0-1) d after hospitalization, most subjects being previously treated with fixed-flow O2 at 2 L/min (IQR: 1–3 L/min). ![Fig. 1.](http://rc.rcjournal.com/http://rc.rcjournal.com/content/respcare/69/9/1081/F1.medium.gif) [Fig. 1.](http://rc.rcjournal.com/content/69/9/1081/F1) Fig. 1. Chronology of the implementation of the automated O2 titration device. ![Fig. 2.](http://rc.rcjournal.com/http://rc.rcjournal.com/content/respcare/69/9/1081/F2.medium.gif) [Fig. 2.](http://rc.rcjournal.com/content/69/9/1081/F2) Fig. 2. Flow chart. ![Fig. 3.](http://rc.rcjournal.com/http://rc.rcjournal.com/content/respcare/69/9/1081/F3.medium.gif) [Fig. 3.](http://rc.rcjournal.com/content/69/9/1081/F3) Fig. 3. Distribution of subjects on the automated O2 titration device. View this table: [Table 3.](http://rc.rcjournal.com/content/69/9/1081/T3) Table 3. Subject Characteristics Three hundred two subjects of 609 subjects (49.6%) were weaned from oxygen while being on automated O2 titration, the remaining individuals had a few hours of fixed-flow O2 at low flows before O2 therapy was stopped. Due to disease worsening or instability, 42 subjects (7.0%) had to be transferred to the ICU and 81 subjects (13.3%) had to be treated with high-flow nasal cannula, including 54 who received this therapy outside the ICU. Thus, a total of 96 subjects (15.8%) reached the definition of clinical worsening. Under those circumstances, automated O2 titration was stopped according to the clinical care protocol. Attending physicians attributed these events to deterioration of the underlying condition and not to the use of the automated O2 titration device. With earlier versions of the software, the system was occasionally unstable and would stop automatically adjusting O2 flows for unclear reasons. When this occurred, the system automatically reverted to the constant-flow mode, delivering a O2 flow based on the analysis of the last 15 min of treatment. This functionality, which is also activated when the ![Formula][33] signal is interrupted or of poor quality, was created for safety purposes. These situations were immediately reported to the manufacturer and the software updated version 1.2.6, which was found to be reliable and working as expected. ![Formula][34] targets prescribed by clinicians were generally in agreement with those proposed by the implementation committee (Table 4). Based on the analysis of 387 subjects for whom the information could be retrieved from the device memory, the ![Formula][35] target was 88% to 90% in 306 subjects (79%), whereas ![Formula][36] target ≥ 93% was used in 8 subjects (2.1%). The ability of automated O2 titration to maintain ![Formula][37] within the desired range could be assessed in 82 subjects for whom detailed ![Formula][38] data could be retrieved from the device memory, 38 from the respiratory ward, 9 from the COVID-19 unit, and 35 from the emergency department. Baseline characteristics of these 82 subjects were similar to those of the 527 subjects for whom the oxygenation data were not available (Table 3). As can be seen in Table 5, the subjects were maintained within the ![Formula][39] target zone for 89% of the time. Hypoxemia (![Formula][40] < 85%) occurred in < 5% of recording time, whereas hyperoxemia (![Formula][41] ≥ 98%) was found < 1% of recording time. Thirty-six of 508 health-care workers (7.1%) who used the automated O2 titration device filled out the survey about its utilization. Twenty-two of 36 health-care workers (61%) would often consider using automated O2 titration for patients on O2 therapy, 10 of 36 (28%) would do so occasionally, whereas 4 of 36 (11%) reported that they would rarely use it. Twenty-eight of 36 respondents (78%) considered that the technical support was sufficient and that they felt comfortable with the use of the device. The respondents gave the following device utility scores on a scale of 0 to 10: 5 (*n* = 2), 6 (*n* = 1), 7 (*n* = 5), 8 (*n* = 10), 9 (*n* = 5), and 10 (*n* = 13), with a mean ± SD utility rating score for automated O2 titration of 8.5 ± 1.5. View this table: [Table 4.](http://rc.rcjournal.com/content/69/9/1081/T4) Table 4. Prescribed ![Formula][42] Targets in 387 Subjects on the Respiratory Ward, COVID-19 Unit, and Emergency Department View this table: [Table 5.](http://rc.rcjournal.com/content/69/9/1081/T5) Table 5. ![Formula][43] Data in 82 Admissions on the Respiratory Ward, COVID-19 Unit, and Emergency Department ## Discussion We report our clinical experience with the feasibility of using an automated O2 titration device in subjects with acute respiratory failure requiring O2 therapy as a part of their routine clinical care. The cumulative data during this evaluation extends previous clinical trials in showing that the ability of automated O2 titration to maintain ![Formula][44] within the desired range in various diseases10-12,14-17 may also apply to “real-life” clinical situations. Clinical worsening, defined as the requirement for high-flow nasal oxygen and/or transfer to the ICU, in 96 subjects (15.8%) was attributed to progression of the underlying disease. Although not a unanimous choice among 36 health-care workers who responded to a short survey, the automated O2 titration device was felt to be useful and a positive experience was reported by the majority of users. This evaluation of the use of automated O2 titration in routine clinical care provided several learning opportunities that helped to address some frequently overlooked practical issues related to (i) oxygenation measurements with pulse oximetry and definition of optimal oxygenation with corresponding ![Formula][45] targets; (ii) organizational factors, including health-care workers training and supervision with the use of automated O2 titration; (iii) optimal utilization of automated O2 titration systems, and (iv) limitations in the accuracy of ![Formula][46] readings with currently available oximeters. This technological evaluation provided an opportunity for nurses, physicians, respiratory therapists, and researchers from our institution to reflect and discuss various aspects of O2 therapy. The use of an automated O2 titration device, which forces clinicians to consider ![Formula][47] targets adapted to the clinical situation, may be helpful in implementing the recommendations that ![Formula][48] should be maintained in pre-specified therapeutic windows, avoiding both hypoxemia and hyperoxemia. Despite clear recommendations about the benefits of prescribing oxygen according to target ranges, clinical implementation of this approach is challenging and required numerous trainings, discussions, and feedback from clinicians. For example, in a recent audit of O2 therapy conducted in New Zealand, ![Formula][49] targets could be found in only 60% of hospitalized patients.29 ![Formula][50] monitoring with the automated O2 titration device made us even more aware of the fluctuations in O2 needs in patients with acute respiratory failure, especially in patients with COVID-19, something that is not typically addressed by only intermittent (and often infrequent) ![Formula][51] measurements. Automated O2 titration in response to ![Formula][52] fluctuations was more effective than manual O2 titration to maintain subjects in the desirable ![Formula][53] range as previously reported in a variety of clinical situations, such as in trauma subjects who are critically injured,10 in subjects admitted to the emergency department,12 after thoracic or abdominal surgery,15 those with COPD exacerbation,11,14 during exercise in subjects with chronic lung diseases,19 and in the pediatric population.30 Typically, ![Formula][54] is maintained in the therapeutic zone 80% of the time with automated O2 titration compared with 40 to 55% of the time with manual O2 titration.10-12,14-17 When considering the importance of avoiding both hyperoxemia and hypoxemia,31 the goal of maintaining subjects within a safe ![Formula][55] target zone is an argument in favor of automated O2 adjustment when caring for subjects with hypoxemic respiratory failure. Our goal was to test the clinical implementation of an automated O2 titration device and not to make recommendations about O2 targets in specific conditions, and our experience should be interpreted in this context. The suggested ![Formula][56] targets from the implementation committee (88 to 90% in the subjects with COPD, 90 to 92% in other situations) were lower than typically recommended, particularly in subjects other than those with COPD, with targets ranging between 90 and 98% have been recently proposed.4-8 The ![Formula][57] targets that were used took into consideration a study from our center that provided evidence that ![Formula][58] readings from the built-in Nonin oximeter, systematically underestimate ![Formula][59] and are lower than those of other oximeters.28 In this study, it was found that ![Formula][60] readings were, on average, 3% to 4% lower with the Nonin oximeter than with the Philips (Mississauga, Canada), Nellcor (Medtronic, Canada), or Masimo (Irvine, California) oximeters.28 One advantage of the Nonin oximeter is that an ![Formula][61] target of 90% allowed detection of all the hypoxemic episodes, whereas other tested pulse oximeters only detected 11 to 37% of these occurrences.28 We, therefore, were confident that proposing ![Formula][62] targets of 88% in COPD and 90% in other diseases with the system that was used in this technological evaluation would protect against hypoxemia and hyperoxemia while avoiding the risk of worsening hypercapnia in COPD and ensuring the safety of subjects.28 Although no blood gas data are available in the present report, a recent study from our group supports this practice by showing the an ![Formula][63] target of 90% with the Nonin oximeter is appropriate to protect against hypoxemia and hyperoxemia.32 Another consideration is that using higher targets could delay weaning of O2 and, therefore, unduly prolong the hospital stay, a situation that is more likely to occur when using an oximeter that systematically underestimates ![Formula][64] . Some clinicians showed some reluctance with our proposal for ![Formula][65] targets early in the technological evaluation, but, as clinical experience was gained and with appropriate teaching, most of them became comfortable with their use as shown in Table 4. The proposed ![Formula][66] targets were applied primarily in people with light skin pigmentation who constitute the vast majority of the population treated in our hospital, an important consideration given the possibility to overestimate ![Formula][67] with pulse oximetry in people with dark skin pigmentation.33 In theory, automated O2 titration, which allows for multiple adjustments of O2 flows, should be safer than relying on the current practice, in which it is challenging for health-care workers to precisely titrate O2 flows.9 However, we acknowledge that confirmation of this theory would require a formal randomized clinical trial. In patients with hypercapnic respiratory failure, the avoidance of hyperoxemia, which was infrequently observed (Table 5), should reduce the risk of worsening respiratory acidosis. For automated O2 titration to be used safely, the premise that ![Formula][68] is an accurate surrogate of ![Formula][69] should be fulfilled and, unfortunately, this is not always the case.28,34,35 Indeed, oximetry readings should be viewed as approximating ![Formula][70] rather than considered as an accurate vital sign. Although imprecision of oximetry readings was reported years ago,34,35 this finding was largely unheeded by the medical community, with potential clinical consequences. For example, using an oximeter that systematically underestimates ![Formula][71] may lead to unduly high O2 flows if the ![Formula][72] target is not adjusted accordingly.32 In people with dark skin pigmentation, overestimation of ![Formula][73] by pulse oximetry may lead to the occurrence of undetected hypoxemia and the risk of undertreatment based on certain ![Formula][74] thresholds.33 Low perfusion and motion artifact may also compromise reliable ![Formula][75] readings and thus any valid estimation of ![Formula][76] .35 These limitations are not specific to automated O2 titration devices because they apply to any situations in which O2 therapy is governed by ![Formula][77] readings. What remains to be seen is to what extent imprecision of oximetry readings influence clinical outcomes; but, increased hospital mortality has been associated with the presence of undetected hypoxemia.36 While we are awaiting better oximeter accuracy, the understanding of current limitations of pulse oximetry should help to provide safer medical care.37 When this project was conceived in 2019, we did not foresee the use of an automated O2 titration device in subjects hospitalized with severe COVID-19 pneumonia. It turned out that such a system was useful in this circumstance because of the ability to continuously adjust O2 flows without direct contact between the subjects and the nurses and respiratory therapists. This offered added safety for the subjects and for hospital workers, likely reducing the risk of transmission of contagious disease.27 Another advantage of automated O2 adjustment was the reduction in the use of personal protective equipment, which was a major issue early in the pandemic due to scarcity. Our experience with automated oxygen titration is consistent with that of Danish investigators who reported a similarly positive experience in 20 subjects hospitalized with COVID-19 and with mild-to-moderate hypoxemic respiratory failure.17 Implementing new technology in clinical practice is challenging. Changing medical practice is a slow process, and there are multiple barriers to the adoption rate of health-care solutions.38,39 The importance of education and continued support and feedback for the medical team cannot be overemphasized. We also observed some clinical situations during which the behavior of the system was difficult to understand by the clinical team. On some occasions, subjects with COPD required relatively high O2 flows (4-6 L/min) to maintain target ![Formula][78] . These situations were uncomfortable for clinicians who are generally trained to avoid high O2 flows in COPD without appreciating that it is hyperoxemia and not high O2 flows per se that are responsible for worsening hypercapnia. In this context, the use of automated O2 titration may offer additional protection to subjects because as long as ![Formula][79] is maintained within a safe therapeutic window, hyperoxemia and worsening hypercapnia should not occur. Some clinical situations were difficult to comprehend because of the time lag between oxygenation status (![Formula][80] ) and incremental (or decremental) adjustments in O2 flows. One such situation is illustrated in Figure 4. A recovering subject resumed mild physical activities during which O2 desaturation was observed. This provoked a dip in ![Formula][81] to which the automated O2 titration device responded by increasing O2 flow. On return to rest, ![Formula][82] remained higher than the target value for some time while O2 flows progressively returned to lower values. For example, at 1,000 s, an O2 flow of 4 L/min could appear unexpected because the ![Formula][83] was above the target. In this situation, instantaneous reading of the physiologic parameters displayed on the automated O2 titration device, showing O2 flows that were higher than expected from the ![Formula][84] value could have mistakenly led to the conclusion that the device was not operating as intended, whereas this was simply a reflection of a time lag between the ![Formula][85] correction and the O2 flow response. Inspecting the ![Formula][86] and O2 flow trend report from the device is crucial to understand the nature of the situation. Another example that requires some experience with the device is when the required O2 flow is low (approximately or < 1 L/min) and the patient is almost ready to be weaned from oxygen. Small fluctuations in ![Formula][87] related to physical activities could induce a transient increase in O2 flow, which may prevent or slow O2 discontinuation. This situation may generally be resolved by simply stopping oxygen and monitoring the patient to ensure that ![Formula][88] remains adequate. ![Fig. 4.](http://rc.rcjournal.com/http://rc.rcjournal.com/content/respcare/69/9/1081/F4.medium.gif) [Fig. 4.](http://rc.rcjournal.com/content/69/9/1081/F4) Fig. 4. Relationship between ![Formula][89] and O2 flow in one subject who was treated with the automated O2 titration device while transitioning from walking in the corridor to bed rest. Target ![Formula][90] was set at 90%. During walking, a sudden fall in ![Formula][91] , down to 80% provoked a rapid rise in O2 flow aiming to return ![Formula][92] in the target zone. An overshoot in ![Formula][93] up to 95% was noted. Returning to bed resulted in reduced O2 needs as seen by a progressive decline in O2 flow down to 0 L/min over a 10-min period while ![Formula][94] remained at the target value of 90%. There are limitations to the present report that should be considered in interpreting the findings and their generalizability. The most obvious is that data were available only for a fraction of the subjects. There was a shortage of resources to collect detailed information in a systematic fashion as we are accustomed to in clinical research. For example, detailed ![Formula][95] data could only be retrieved for 13% of subjects because many were not appropriately identified in the device, which made it impossible to match recorded oxygenation data with the corresponding subjects. Although the availability of detailed ![Formula][96] data in only a small portion of the subjects raises questions about the external validity of the findings, we were reassured that the proportion of time spent within the ![Formula][97] targets was similar to that previously observed in other clinical circumstances.10-12,14-17 Only a small number of nurses and respiratory therapists responded to the survey about the utilization of automated O2 titration. Therefore, its interpretation should be done cautiously. For example, it is possible that only those who had a positive experience with the device took the time to answer the survey. Lastly, the decision to use or not use the automated O2 titration system was made by the clinical team; we do not have data to estimate the proportion of hospitalized patients who require oxygen therapy treated with the device. ## Conclusions We found that automated O2 titration could be implemented safely in the context of routine clinical care in subjects with hypoxemic respiratory failure of various etiology. Based on this clinical experience and analysis of pilot data that suggest the possibility to reduce hospital length of stay with automated O2 titration,11 we encourage the conduct of randomized clinical trials that test the impact of automated O2 titration on hospital length of stay compared with the standard care in subjects with acute hypoxemia. ## Acknowledgments The authors thank the implementation team and health-care workers of our hospital for their contribution and who worked diligently to provide safe O2 therapy to the subjects. We also thank Serge Simard MSc, for his assistance with the statistical analyses. ## Footnotes * Correspondence: François Maltais MD, Centre de pneumologie, Institut universitaire de cardiologie et de pneumologie de Québec, 2725, chemin Ste-Foy, Québec, Canada G1V 4G5. E-mail: francois.maltais{at}med.ulaval.ca * Funding was provided by the Fondation de l’Institut universitaire de cardiologie et de pneumologie de Québec, Ville de Québec, Oxynov. * Drs Simon, Lacasse, Lellouche, and Maltais are shareholders of Oxynov, the maker of FreeO2; Dr Lellouche is co-founder of Oxynov. * See the Related Editorial on Page [1214](http://rc.rcjournal.com/lookup/doi/10.4187/respcare.12230) * Copyright © 2024 by Daedalus Enterprises ## REFERENCES 1. 1.1. Stenson BJ, 2. Tarnow-Mordi WO, 3. Darlow BA, 4. et al BOOST II United Kingdom Collaborative Group; BOOST II Australia Collaborative Group; BOOST II New Zealand Collaborative Group; Stenson BJ, Tarnow-Mordi WO, Darlow BA, et al. Oxygen saturation and outcomes in preterm infants. N Engl J Med 2013;368(22):2094-2104. [CrossRef](http://rc.rcjournal.com/lookup/external-ref?access_num=10.1056/NEJMoa1302298&link_type=DOI) [PubMed](http://rc.rcjournal.com/lookup/external-ref?access_num=23642047&link_type=MED&atom=%2Frespcare%2F69%2F9%2F1081.atom) [Web of Science](http://rc.rcjournal.com/lookup/external-ref?access_num=000319574200007&link_type=ISI) 2. 2.Austin MA, Wills KE, Blizzard L, Walters EH, Wood-Baker R. Effect of high flow oxygen on mortality in chronic obstructive pulmonary disease patients in prehospital setting: randomised controlled trial. BMJ 2010;341:c5462. [Abstract/FREE Full Text](http://rc.rcjournal.com/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiQUJTVCI7czoxMToiam91cm5hbENvZGUiO3M6MzoiYm1qIjtzOjU6InJlc2lkIjtzOjE3OiIzNDEvb2N0MThfMi9jNTQ2MiI7czo0OiJhdG9tIjtzOjI0OiIvcmVzcGNhcmUvNjkvOS8xMDgxLmF0b20iO31zOjg6ImZyYWdtZW50IjtzOjA6IiI7fQ==) 3. 3.Chu DK, Kim LH-Y, Young PJ, Zamiri N, Almenawer SA, Jaeschke R, et al. Mortality and morbidity in acutely ill adults treated with liberal versus conservative oxygen therapy (IOTA): a systematic review and meta-analysis. Lancet 2018;391(10131):1693-1705. [CrossRef](http://rc.rcjournal.com/lookup/external-ref?access_num=10.1016/S0140-6736(18)30479-3&link_type=DOI) [PubMed](http://rc.rcjournal.com/lookup/external-ref?access_num=29726345&link_type=MED&atom=%2Frespcare%2F69%2F9%2F1081.atom) 4. 4.O'Driscoll BR, Howard LS, Earis J, Mak V; BTS Emergency Oxygen Guideline Development Group; BTS Emergency Oxygen Guideline Development Group. BTS guideline for oxygen use in adults in healthcare and emergency settings. Thorax 2017;72(Suppl 1):ii1-ii90. [FREE Full Text](http://rc.rcjournal.com/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiRlVMTCI7czoxMToiam91cm5hbENvZGUiO3M6OToidGhvcmF4am5sIjtzOjU6InJlc2lkIjtzOjE0OiI3Mi9TdXBwbF8xL2lpMSI7czo0OiJhdG9tIjtzOjI0OiIvcmVzcGNhcmUvNjkvOS8xMDgxLmF0b20iO31zOjg6ImZyYWdtZW50IjtzOjA6IiI7fQ==) 5. 5.Siemieniuk RAC, Chu DK, Kim LH-Y, Guell-Rous M-R, Alhazzani W, Soccal PM, et al. Oxygen therapy for acutely ill medical patients: a clinical practice guideline. BMJ 2018;363:k4169. [FREE Full Text](http://rc.rcjournal.com/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiRlVMTCI7czoxMToiam91cm5hbENvZGUiO3M6MzoiYm1qIjtzOjU6InJlc2lkIjtzOjE3OiIzNjMvb2N0MjRfOS9rNDE2OSI7czo0OiJhdG9tIjtzOjI0OiIvcmVzcGNhcmUvNjkvOS8xMDgxLmF0b20iO31zOjg6ImZyYWdtZW50IjtzOjA6IiI7fQ==) 6. 6.Beasley R, Chien J, Douglas J, Eastlake L, Farah C, King G, et al. Thoracic Society of Australia and New Zealand Oxygen guidelines for acute oxygen use in adults: 'swimming between the flags'. Respirology 2015;20(8):1182-1191. [CrossRef](http://rc.rcjournal.com/lookup/external-ref?access_num=10.1111/resp.12620&link_type=DOI) [PubMed](http://rc.rcjournal.com/lookup/external-ref?access_num=26486092&link_type=MED&atom=%2Frespcare%2F69%2F9%2F1081.atom) 7. 7.Gottlieb J, Capetian P, Hamsen U, Janssens U, Karagiannidis C, Kluge S, et al. German S3 guideline: oxygen therapy in the acute care of adult patients. Respiration 2022;101(2):214-252. 8. 8.Piraino T, Madden M, Roberts KJ, Lamberti J, Ginier E, Strickland SL. AARC clinical practice guideline: management of adult patients with oxygen in the acute care setting. Respir Care 2022;67(1):115-128. [Abstract/FREE Full Text](http://rc.rcjournal.com/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiQUJTVCI7czoxMToiam91cm5hbENvZGUiO3M6ODoicmVzcGNhcmUiO3M6NToicmVzaWQiO3M6ODoiNjcvMS8xMTUiO3M6NDoiYXRvbSI7czoyNDoiL3Jlc3BjYXJlLzY5LzkvMTA4MS5hdG9tIjt9czo4OiJmcmFnbWVudCI7czowOiIiO30=) 9. 9.Cousins JL, Wark PAB, McDonald VM. Acute oxygen therapy: a review of prescribing and delivery practices. Int J Chron Obstruct Pulmon Disease 2016;11:1067-1075. 10. 10.Johannigman JA, Branson RD, Lecroy D, Beck G. Autonomous control of inspired oxygen concentration during mechanical ventilation of the critically injured trauma patient. J Trauma 2009;66(2):386-392. [PubMed](http://rc.rcjournal.com/lookup/external-ref?access_num=19204511&link_type=MED&atom=%2Frespcare%2F69%2F9%2F1081.atom) 11. 11.Lellouche F, Bouchard P-A, Roberge M, Simard S, L'Her E, Maltais F, Lacasse Y. Automated oxygen titration and weaning with FreeO2 in patients with acute exacerbation of COPD: a pilot randomized trial. Int J Chron Obstruct Pulmon Dis 2016;11:1983-1990. [PubMed](http://rc.rcjournal.com/lookup/external-ref?access_num=http://www.n&link_type=MED&atom=%2Frespcare%2F69%2F9%2F1081.atom) 12. 12.L'Her E, Dias P, Gouillou M, Riou A, Souquiere L, Paleiron N, et al. Automatic versus manual oxygen administration in the emergency department. Eur Respir J 2017;50(1):1602552. [Abstract/FREE Full Text](http://rc.rcjournal.com/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiQUJTVCI7czoxMToiam91cm5hbENvZGUiO3M6MzoiZXJqIjtzOjU6InJlc2lkIjtzOjEyOiI1MC8xLzE2MDI1NTIiO3M6NDoiYXRvbSI7czoyNDoiL3Jlc3BjYXJlLzY5LzkvMTA4MS5hdG9tIjt9czo4OiJmcmFnbWVudCI7czowOiIiO30=) 13. 13.Lellouche F, L'Her E. Automated oxygen flow titration to maintain constant oxygenation. Respir Care 2012;57(8):1254-1262. [Abstract/FREE Full Text](http://rc.rcjournal.com/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiQUJTVCI7czoxMToiam91cm5hbENvZGUiO3M6ODoicmVzcGNhcmUiO3M6NToicmVzaWQiO3M6OToiNTcvOC8xMjU0IjtzOjQ6ImF0b20iO3M6MjQ6Ii9yZXNwY2FyZS82OS85LzEwODEuYXRvbSI7fXM6ODoiZnJhZ21lbnQiO3M6MDoiIjt9) 14. 14.Hansen EF, Hove JD, Bech CS, Jensen J-US, Kallemose T, Vestbo J. Automated oxygen control with O2matic® during admission with exacerbation of COPD. Int J Chron Obstruct Pulmon Dis 2018;13:3997-4003. [PubMed](http://rc.rcjournal.com/lookup/external-ref?access_num=http://www.n&link_type=MED&atom=%2Frespcare%2F69%2F9%2F1081.atom) 15. 15.L'Her E, Jaber S, Verzilli D, Jacob C, Huiban B, Futier E, et al. Automated closed-loop versus standard manual oxygen administration after major abdominal or thoracic surgery: an international multicentre randomised controlled study. Eur Respir J 2021;57(1):2000182. 16. 16.Denault M-H, Ruel C, Simon M, Bouchard P-A, Simard S, Lellouche F. Evaluation of hyperoxia-induced hypercapnia in obese patients after cardiac surgery: a randomized crossover comparison of conservative and liberal oxygen administration. Can J Anaesth 2020;67(2):194-202. [PubMed](http://rc.rcjournal.com/lookup/external-ref?access_num=http://www.n&link_type=MED&atom=%2Frespcare%2F69%2F9%2F1081.atom) 17. 17.Hansen EF, Bech CS, Vestbo J, Andersen O, Kofod LM. Automatic oxygen titration with O2matic® to patients admitted with COVID-19 and hypoxemic respiratory failure. Eur Clin Respir J 2020;7(1):1833695. 18. 18.Lellouche F, L'Her E, Bouchard P-A, Brouillard C, Maltais F. Automatic oxygen titration during walking in subjects with COPD: a randomized crossover controlled study. Respir Care 2016;61(11):1456-1464. [Abstract/FREE Full Text](http://rc.rcjournal.com/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiQUJTVCI7czoxMToiam91cm5hbENvZGUiO3M6ODoicmVzcGNhcmUiO3M6NToicmVzaWQiO3M6MTA6IjYxLzExLzE0NTYiO3M6NDoiYXRvbSI7czoyNDoiL3Jlc3BjYXJlLzY5LzkvMTA4MS5hdG9tIjt9czo4OiJmcmFnbWVudCI7czowOiIiO30=) 19. 19.Vivodtzev I, L'Her E, Vottero G, Yankoff C, Tamisier R, Maltais F, et al. Automated O2 titration improves exercise capacity in patients with hypercapnic chronic obstructive pulmonary disease: a randomised controlled cross-over trial. Thorax 2019;74(3):298-301. [Abstract/FREE Full Text](http://rc.rcjournal.com/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiQUJTVCI7czoxMToiam91cm5hbENvZGUiO3M6OToidGhvcmF4am5sIjtzOjU6InJlc2lkIjtzOjg6Ijc0LzMvMjk4IjtzOjQ6ImF0b20iO3M6MjQ6Ii9yZXNwY2FyZS82OS85LzEwODEuYXRvbSI7fXM6ODoiZnJhZ21lbnQiO3M6MDoiIjt9) 20. 20.Kofod LM, Westerdahl E, Kristensen MT, Brocki BC, Ringbæk T, Hansen EF. Effect of automated oxygen titration during walking on dyspnea and endurance in chronic hypoxemic patients with COPD: a randomized crossover trial. J Clin Med 2021;10(21):4820. 21. 21.Denault M-H, Peloquin F, Lajoie A-C, Lacasse Y. Automatic versus manual oxygen titration in patients requiring supplemental oxygen in the hospital: a systematic review and meta-analysis. Respiration 2019;98(2):178-188. [PubMed](http://rc.rcjournal.com/lookup/external-ref?access_num=http://www.n&link_type=MED&atom=%2Frespcare%2F69%2F9%2F1081.atom) 22. 22.Poder TG, Kouakou CRC, Bouchard P-A, Tremblay V, Blais S, Maltais F, Lellouche F. Cost-effectiveness of FreeO2 in patients with chronic obstructive pulmonary disease hospitalised for acute exacerbations: analysis of a pilot study in Quebec. BMJ Open 2018;8(1):e018835. [Abstract/FREE Full Text](http://rc.rcjournal.com/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiQUJTVCI7czoxMToiam91cm5hbENvZGUiO3M6NzoiYm1qb3BlbiI7czo1OiJyZXNpZCI7czoxMToiOC8xL2UwMTg4MzUiO3M6NDoiYXRvbSI7czoyNDoiL3Jlc3BjYXJlLzY5LzkvMTA4MS5hdG9tIjt9czo4OiJmcmFnbWVudCI7czowOiIiO30=) 23. 23.Harper JC, Kearns NA, Maijers I, Bird GE, Braithwaite I, Shortt NP, et al. Closed-loop oxygen control using a novel nasal high-flow device: a randomized crossover trial. Respir Care 2021;66(3):416-424. [Abstract/FREE Full Text](http://rc.rcjournal.com/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiQUJTVCI7czoxMToiam91cm5hbENvZGUiO3M6ODoicmVzcGNhcmUiO3M6NToicmVzaWQiO3M6ODoiNjYvMy80MTYiO3M6NDoiYXRvbSI7czoyNDoiL3Jlc3BjYXJlLzY5LzkvMTA4MS5hdG9tIjt9czo4OiJmcmFnbWVudCI7czowOiIiO30=) 24. 24.Harper J, Kearns N, Bird G, Braithwaite I, Eathorne A, Shortt N, et al. Automatic versus manual oxygen titration using a novel nasal high-flow device in medical inpatients with an acute illness: a randomised controlled trial. BMJ Open Respir Res 2021;8(1):e000843. [Abstract/FREE Full Text](http://rc.rcjournal.com/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiQUJTVCI7czoxMToiam91cm5hbENvZGUiO3M6NzoiYm1qcmVzcCI7czo1OiJyZXNpZCI7czoxMToiOC8xL2UwMDA4NDMiO3M6NDoiYXRvbSI7czoyNDoiL3Jlc3BjYXJlLzY5LzkvMTA4MS5hdG9tIjt9czo4OiJmcmFnbWVudCI7czowOiIiO30=) 25. 25.Cohen B, Hyman S, Rosenberg L, Larson E. Frequency of patient contact with health care personnel and visitors: implications for infection prevention. Jt Comm J Qual Patient Saf 2012;38(12):560-565. [CrossRef](http://rc.rcjournal.com/lookup/external-ref?access_num=10.1016/S1553-7250(12)38073-2&link_type=DOI) 26. 26.Huang L, Lin G, Tang L, Yu L, Zhou Z. Special attention to nurses' protection during the COVID-19 epidemic. Crit Care 2020;24(1):120. [PubMed](http://rc.rcjournal.com/lookup/external-ref?access_num=http://www.n&link_type=MED&atom=%2Frespcare%2F69%2F9%2F1081.atom) 27. 27.Bandyopadhyay S, Baticulon RE, Kadhum M, Alser M, Ojuka DK, Badereddin Y, et al. Infection and mortality of healthcare workers worldwide from COVID-19: a systematic review. BMJ Glob Health 2020;5(12):e003097. [Abstract/FREE Full Text](http://rc.rcjournal.com/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiQUJTVCI7czoxMToiam91cm5hbENvZGUiO3M6NToiYm1qZ2giO3M6NToicmVzaWQiO3M6MTI6IjUvMTIvZTAwMzA5NyI7czo0OiJhdG9tIjtzOjI0OiIvcmVzcGNhcmUvNjkvOS8xMDgxLmF0b20iO31zOjg6ImZyYWdtZW50IjtzOjA6IiI7fQ==) 28. 28.Blanchet M-A, Mercier G, Delobel A, Nayet E, Bouchard P-A, Simard S, et al. Accuracy of multiple pulse oximeter brands in stable critically ill patients. Respir Care 2023;68(5):565-574. [Abstract/FREE Full Text](http://rc.rcjournal.com/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiQUJTVCI7czoxMToiam91cm5hbENvZGUiO3M6ODoicmVzcGNhcmUiO3M6NToicmVzaWQiO3M6ODoiNjgvNS81NjUiO3M6NDoiYXRvbSI7czoyNDoiL3Jlc3BjYXJlLzY5LzkvMTA4MS5hdG9tIjt9czo4OiJmcmFnbWVudCI7czowOiIiO30=) 29. 29.Harper J, Kearns N, Bird G, McLachlan R, Eathorne A, Weatherall M, Beasley R. Audit of oxygen administration to achieve a target oxygen saturation range in acutely unwell medical patients. Postgrad Med J 2022;98(1160):461-465. [Abstract/FREE Full Text](http://rc.rcjournal.com/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiQUJTVCI7czoxMToiam91cm5hbENvZGUiO3M6MTI6InBvc3RncmFkbWVkaiI7czo1OiJyZXNpZCI7czoxMToiOTgvMTE2MC80NjEiO3M6NDoiYXRvbSI7czoyNDoiL3Jlc3BjYXJlLzY5LzkvMTA4MS5hdG9tIjt9czo4OiJmcmFnbWVudCI7czowOiIiO30=) 30. 30.Roue J-M, Delpeut J, d'Hennezel A, Tierrie T, Barzic A, L'Her E, Cros P. Automatic oxygen flow titration in spontaneously breathing children: an open-label randomized controlled pilot study. Pediatr Pulmonol 2020;55(11):3180-3188. 31. 31.Munshi L, Ferguson ND. Evolving issues in oxygen therapy in acute care medicine. JAMA 2020;323(7):607-608. [PubMed](http://rc.rcjournal.com/lookup/external-ref?access_num=http://www.n&link_type=MED&atom=%2Frespcare%2F69%2F9%2F1081.atom) 32. 32.Lellouche F, Bouchard P-A, Branson RD. Impact of SpO2 targets and pulse oximeter brand on oxygen flow requirements and oxygenation. Respir Care 2023;69(1):110-113. 33. 33.Sjoding MW, Dickson RP, Iwashyna TJ, Gay SE, Valley TS. Racial bias in pulse oximetry measurement. N Engl J Med 2020;383(25):2477-2478. [CrossRef](http://rc.rcjournal.com/lookup/external-ref?access_num=10.1056/NEJMc2029240&link_type=DOI) [PubMed](http://rc.rcjournal.com/lookup/external-ref?access_num=33326721&link_type=MED&atom=%2Frespcare%2F69%2F9%2F1081.atom) 34. 34.Jubran A, Tobin MJ. Reliability of pulse oximetry in titrating supplemental oxygen therapy in ventilator-dependent patients. Chest 1990;97(6):1420-1425. [CrossRef](http://rc.rcjournal.com/lookup/external-ref?access_num=10.1378/chest.97.6.1420&link_type=DOI) [PubMed](http://rc.rcjournal.com/lookup/external-ref?access_num=2347228&link_type=MED&atom=%2Frespcare%2F69%2F9%2F1081.atom) [Web of Science](http://rc.rcjournal.com/lookup/external-ref?access_num=A1990DG62600029&link_type=ISI) 35. 35.Jubran A. Pulse oximetry. Intensive Care Med 2004;30(11):2017-2020. [CrossRef](http://rc.rcjournal.com/lookup/external-ref?access_num=10.1007/s00134-004-2399-x&link_type=DOI) [PubMed](http://rc.rcjournal.com/lookup/external-ref?access_num=15278272&link_type=MED&atom=%2Frespcare%2F69%2F9%2F1081.atom) [Web of Science](http://rc.rcjournal.com/lookup/external-ref?access_num=000224956000005&link_type=ISI) 36. 36.Wong A-KI, Charpignon M, Kim H, Josef C, de Hond AAH, Fojas JJ, et al. Analysis of discrepancies between pulse oximetry and arterial oxygen saturation measurements by race and ethnicity and association with organ dysfunction and mortality. JAMA Netw Open 2021;4(11):e2131674. 37. 37.Hess DR. Using SpO2: not as simple as it seems. Respir Care 2023;68(5):708-712. [FREE Full Text](http://rc.rcjournal.com/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiRlVMTCI7czoxMToiam91cm5hbENvZGUiO3M6ODoicmVzcGNhcmUiO3M6NToicmVzaWQiO3M6ODoiNjgvNS83MDgiO3M6NDoiYXRvbSI7czoyNDoiL3Jlc3BjYXJlLzY5LzkvMTA4MS5hdG9tIjt9czo4OiJmcmFnbWVudCI7czowOiIiO30=) 38. 38.Boonstra A, Broekhuis M. Barriers to the acceptance of electronic medical records by physicians from systematic review to taxonomy and interventions. BMC Health Serv Res 2010;10:231. [CrossRef](http://rc.rcjournal.com/lookup/external-ref?access_num=10.1186/1472-6963-10-231&link_type=DOI) [PubMed](http://rc.rcjournal.com/lookup/external-ref?access_num=20691097&link_type=MED&atom=%2Frespcare%2F69%2F9%2F1081.atom) 39. 39.Boyle LD, Husebo BS, Vislapuu M. Promotors and barriers to the implementation and adoption of assistive technology and telecare for people with dementia and their caregivers: a systematic review of the literature. BMC Health Serv Res 2022;22(1):1573. [1]: /embed/mml-math-1.gif [2]: /embed/mml-math-2.gif [3]: /embed/mml-math-3.gif [4]: /embed/mml-math-4.gif [5]: /embed/mml-math-5.gif [6]: /embed/mml-math-6.gif [7]: /embed/mml-math-7.gif [8]: /embed/mml-math-8.gif [9]: /embed/mml-math-9.gif [10]: /embed/mml-math-10.gif [11]: /embed/mml-math-11.gif [12]: /embed/mml-math-12.gif [13]: /embed/mml-math-13.gif [14]: /embed/mml-math-14.gif [15]: /embed/mml-math-15.gif [16]: /embed/mml-math-16.gif [17]: /embed/mml-math-17.gif [18]: /embed/mml-math-18.gif [19]: /embed/mml-math-19.gif [20]: /embed/mml-math-20.gif [21]: /embed/mml-math-21.gif [22]: /embed/mml-math-22.gif [23]: /embed/mml-math-23.gif [24]: /embed/mml-math-24.gif [25]: /embed/mml-math-25.gif [26]: /embed/mml-math-26.gif [27]: /embed/mml-math-27.gif [28]: /embed/mml-math-28.gif [29]: /embed/mml-math-29.gif [30]: /embed/mml-math-30.gif [31]: /embed/mml-math-31.gif [32]: /embed/mml-math-32.gif [33]: /embed/mml-math-33.gif [34]: /embed/mml-math-34.gif [35]: /embed/mml-math-35.gif [36]: /embed/mml-math-36.gif [37]: /embed/mml-math-37.gif [38]: /embed/mml-math-38.gif [39]: /embed/mml-math-39.gif [40]: /embed/mml-math-40.gif [41]: /embed/mml-math-41.gif [42]: T4/embed/mml-math-42.gif [43]: T5/embed/mml-math-43.gif [44]: /embed/mml-math-44.gif [45]: /embed/mml-math-45.gif [46]: /embed/mml-math-46.gif [47]: /embed/mml-math-47.gif [48]: /embed/mml-math-48.gif [49]: /embed/mml-math-49.gif [50]: /embed/mml-math-50.gif [51]: /embed/mml-math-51.gif [52]: /embed/mml-math-52.gif [53]: /embed/mml-math-53.gif [54]: /embed/mml-math-54.gif [55]: /embed/mml-math-55.gif [56]: /embed/mml-math-56.gif [57]: /embed/mml-math-57.gif [58]: /embed/mml-math-58.gif [59]: /embed/mml-math-59.gif [60]: /embed/mml-math-60.gif [61]: /embed/mml-math-61.gif [62]: /embed/mml-math-62.gif [63]: /embed/mml-math-63.gif [64]: /embed/mml-math-64.gif [65]: /embed/mml-math-65.gif [66]: /embed/mml-math-66.gif [67]: /embed/mml-math-67.gif [68]: /embed/mml-math-68.gif [69]: /embed/mml-math-69.gif [70]: /embed/mml-math-70.gif [71]: /embed/mml-math-71.gif [72]: /embed/mml-math-72.gif [73]: /embed/mml-math-73.gif [74]: /embed/mml-math-74.gif [75]: /embed/mml-math-75.gif [76]: /embed/mml-math-76.gif [77]: /embed/mml-math-77.gif [78]: /embed/mml-math-78.gif [79]: /embed/mml-math-79.gif [80]: /embed/mml-math-80.gif [81]: /embed/mml-math-81.gif [82]: /embed/mml-math-82.gif [83]: /embed/mml-math-83.gif [84]: /embed/mml-math-84.gif [85]: /embed/mml-math-85.gif [86]: /embed/mml-math-86.gif [87]: /embed/mml-math-87.gif [88]: /embed/mml-math-88.gif [89]: F4/embed/mml-math-89.gif [90]: F4/embed/mml-math-90.gif [91]: F4/embed/mml-math-91.gif [92]: F4/embed/mml-math-92.gif [93]: F4/embed/mml-math-93.gif [94]: F4/embed/mml-math-94.gif [95]: /embed/mml-math-95.gif [96]: /embed/mml-math-96.gif [97]: /embed/mml-math-97.gif