PT - JOURNAL ARTICLE AU - Gregory Reychler AU - Anne-Sophie Aubriot AU - Virginie Depoortere AU - François Jamar AU - Giuseppe Liistro TI - Effect of Drug Targeting Nebulization on Lung Deposition: A Randomized Crossover Scintigraphic Comparison Between Central and Peripheral Delivery AID - 10.4187/respcare.03068 DP - 2014 Aug 26 TA - Respiratory Care PG - respcare.03068 4099 - http://rc.rcjournal.com/content/early/2014/08/26/respcare.03068.short 4100 - http://rc.rcjournal.com/content/early/2014/08/26/respcare.03068.full AB - BACKGROUND: Recent technological advances in nebulization permit researchers to target specific parts of the lungs by modifying delivery method. The aim of this study was to compare the central and peripheral targeted modalities of administration. METHODS: Lung and regional deposition of inhaled technetium-99m diethylene triamine penta-acetic was measured by scintigraphy after peripheral and central targeted modalities of administration with an Akita device in 6 healthy subjects. RESULTS: Drug targeting nebulization delivered a large amount of drug into the peripheral part of the lung independent of the modality (outer-to-inner deposition ratio of 1.24 ± 0.21 vs 1.22 ± 0.14 for central and peripheral modalities, respectively), but there was no difference in lung deposition (whole-body deposition, 83.3 ± 6.5% vs 82.8 ± 7.3%, P =.86) or regional deposition (P = .77) between both modalities. The extrathoracic deposition was < 20% of the whole-body deposition, without a difference between modalities (P = .86). CONCLUSIONS: This study shows for the first time that choosing 2 different specific drug targeting nebulization modes does not influence the amount of drug delivered into the lung in healthy male subjects. Moreover, the modes do not modify the site of deposition under the conditions of our study.