RT Journal Article SR Electronic T1 High-Frequency Assisted Airway Clearance JF Respiratory Care FD American Association for Respiratory Care SP 1224 OP 1237 VO 52 IS 9 A1 Robert L Chatburn YR 2007 UL http://rc.rcjournal.com/content/52/9/1224.abstract AB High-frequency airway clearance assist devices generate either positive or negative transrespiratory pressure excursions to produce high-frequency, small-volume oscillations in the airways. Intrapulmonary percussive ventilation creates a positive transrespiratory pressure by injecting short, rapid inspiratory flow pulses into the airway opening and relies on chest wall elastic recoil for passive exhalation. High-frequency chest wall compression generates a negative transrespiratory pressure by compressing the chest externally to cause short, rapid expiratory flow pulses, and relies on chest wall elastic recoil to return the lungs to functional residual capacity. High-frequency chest wall oscillation uses a chest cuirass to generate biphasic changes in transrespiratory pressure. In any case (positive or negative pressure pulses or both), the general idea is get air behind secretions and move them toward the larger airways, where they can be coughed up and expectorated. These techniques have become ubiquitous enough to constitute a standard of care. Yet, despite over 20 years of research, clinical evidence of efficacy for them is still lacking. Indeed, there is insufficient evidence to support the use of any single airway clearance technique, let alone judge any one of them superior. Aside from patient preference and capability, cost-effectiveness studies based on existing clinical data are necessary to determine when a given technique is most practical.