TY - JOUR T1 - Development and Testing of a Bubble Bi-Level Positive Airway Pressure System JF - Respiratory Care SP - 1131 LP - 1136 DO - 10.4187/respcare.05443 VL - 62 IS - 9 AU - Stephen C John AU - Joseph D Barnett AU - Nickolas D Habben AU - Hoa T Le AU - Eric Cheng AU - Sunil P John AU - Peter A Gustafson Y1 - 2017/09/01 UR - http://rc.rcjournal.com/content/62/9/1131.abstract N2 - BACKGROUND: Neonatal respiratory distress results in > 1 million annual deaths worldwide. Bubble CPAP is a simple, effective, and widely used therapy for infants in respiratory distress. In low-resource settings, more advanced respiratory support is limited by cost, technical expertise, and sporadic electricity. We sought to develop a safe, inexpensive, and simple solution to provide further respiratory support for these infants.METHODS: A standard bubble CPAP system was modified to provide 2 levels of positive airway pressure (bi-level positive airway pressure) by attaching a novel device. To demonstrate reliability, the system was run with continuous pressure monitoring on full-term and preterm neonatal mannikins with pressure targets of 8/5 cm H2O and 15/5 cm H2O to simulate 2 different modes of noninvasive ventilation (NIV).RESULTS: At a ventilation rate set between 30 and 45 cycles/min, by adjusting the leak rate of the device, the following mean pressures ± SD were demonstrated: term mannikin low-pressure NIV, 7.9 ± 0.2/5.3 ± 0.2 cm H2O; term mannikin high-pressure NIV, 15.1 ± 0.1/6.1 ± 0.1 cm H2O; preterm mannikin low-pressure NIV, 7.9 ± 0.2/5.3 ± 0.2 cm H2O; preterm mannikin high-pressure NIV, 16.5 ± 0.4/5.1 ± 0.1 cm H2O.CONCLUSIONS: The modified bubble CPAP system reliably provided alternating pressures similar to bi-level positive airway pressure modes of respiratory support in neonatal mannikins. The dual-pressure technology is a simple, single connection add-on that can readily be applied to existing bubble CPAP systems. ER -