RT Journal Article SR Electronic T1 Oscillating Positive Expiratory Pressure Therapy May Be Performed Poorly by Children With Cystic Fibrosis JF Respiratory Care FD American Association for Respiratory Care SP 398 OP 405 DO 10.4187/respcare.06329 VO 64 IS 4 A1 Kevin J O'Sullivan A1 Louise Collins A1 Deirdre McGrath A1 Barry Linnane A1 Leonard O'Sullivan A1 Colum P Dunne YR 2019 UL http://rc.rcjournal.com/content/64/4/398.abstract AB BACKGROUND: Oscillating positive expiratory pressure devices aid removal of excess secretions and reduce gas trapping in patients with hypersecretory pulmonary diseases, for example, cystic fibrosis. Oscillating positive expiratory pressure works when the patient exhales actively against a fixed resistor, which generates mean intrapulmonary pressures of 10–20 cm H2O with rapid fluctuations of at least 1 cm H2O from the mean. In this study, we evaluated the performance of oscillating positive expiratory pressure therapy by pediatric subjects with cystic fibrosis to determine adherence to target therapeutic pressures.METHODS: Twenty-one pediatric subjects were recruited. Each had a history of using an oscillating positive expiratory pressure device twice daily and had received standardized training and instructions from the same specialist physiotherapist. Performance was evaluated by using a flow and pressure sensor placed in-line between the participant's mouth and the device. The participants performed expirations as per their normal routine.RESULTS: None of the participants achieved target therapeutic pressure ranges during expiration. The mean ± SD pressure generated was 16.2 ± 6.8 cm H2O, whereas mean ± SD flow was 31.3 ± 8.9 L/min. The mean ± SD expiration length was 2.5 ± 1.4 s.CONCLUSIONS: Despite standardized instruction, the results demonstrated considerable variation among the participants and overall poor technique during use. Outcomes of this study indicated that airway clearance effects of oscillating positive expiratory pressure were compromised due to poor technique.