PT - JOURNAL ARTICLE AU - Subat, Yosuf W AU - Guntupalli, Siva Kamal AU - Sajgalik, Pavol AU - Hainy, Matthew E AU - Torgerud, Keith D AU - Helgeson, Scott A AU - Johnson, Bruce D AU - Allison, Thomas G AU - Lim, Kaiser G AU - Niven, Alexander S TI - Aerosol Generation During Peak Flow Testing: Clinical Implications for COVID-19 AID - 10.4187/respcare.08731 DP - 2021 Aug 01 TA - Respiratory Care PG - 1291--1298 VI - 66 IP - 8 4099 - http://rc.rcjournal.com/content/66/8/1291.short 4100 - http://rc.rcjournal.com/content/66/8/1291.full AB - BACKGROUND: Peak flow testing is a common procedure performed in ambulatory care. There are currently no data regarding aerosol generation during this procedure. Given the ongoing debate regarding the potential for aerosol transmission of SARS-CoV-2, we aimed to quantify and characterize aerosol generation during peak flow testing.METHODS: Five healthy volunteers performed peak flow maneuvers in a particle-free laboratory space. Two devices continuously sampled the ambient air during the procedure. One device can detect ultrafine particles 0.02–1 μm in diameter, while the second device can detect particles 0.3, 0.5, 1.0, 2.0, 5.0, and 10 μm in diameter. Five different peak flow meters were compared to ambient baseline during masked and unmasked tidal breathing.RESULTS: Ultrafine particles (0.02–1 μm) were generated during peak flow measurement. There was no significant difference in ultrafine particle mean concentration between peak flow meters (P = .23): Respironics (1.25 ± 0.47 particles/mL), Philips (3.06 ± 1.22), Clement Clarke (3.55 ± 1.22 particles/mL), Respironics Low Range (3.50 ± 1.52 particles/mL), and Monaghan (3.78 ± 1.31 particles/mL). Ultrafine particle mean concentration with peak flow testing was significantly higher than masked (0.22 ± 0.29 particles/mL) and unmasked tidal breathing (0.15 ± 0.18 particles/mL, P = .01), but the ultrafine particle concentrations were small compared to ambient particle concentrations in a pulmonary function testing room (89.9 ± 8.95 particles/mL).CONCLUSIONS: In this study, aerosol generation was present during peak flow testing, but concentrations were small compared to the background particle concentration in the ambient clinical environment. Surgical masks and eye protection are likely sufficient infection control measures during peak expiratory flow testing in asymptomatic patients with well controlled respiratory symptoms, but COVID-19 testing remains prudent in patients with acute respiratory symptoms prior to evaluation and peak expiratory flow assessment while the community prevalence of SARS-CoV-2 cases remains high.