RT Journal Article SR Electronic T1 Longitudinal Evaluation of Compressor/Nebulizer Performance JF Respiratory Care FD American Association for Respiratory Care SP 1053 OP 1061 DO 10.4187/respcare.02776 VO 59 IS 7 A1 Samah Awad A1 David K Williams A1 Ariel Berlinski YR 2014 UL http://rc.rcjournal.com/content/59/7/1053.abstract AB BACKGROUND: Inhaled medications are the mainstay of treatment for maintenance of lung health in patients with cystic fibrosis (CF). Compressor/nebulizer units are used an average of 100–120 min/d by patients with CF. Each compressor/nebulizer has unique flow/pressure characteristics that affect particle size distribution and drug output rate. Few data are available regarding longitudinal performance of compressor/nebulizers. We hypothesized that their use over a 24-wk period under conditions similar to those of patients with CF would affect their performance. METHODS: Four new units of compressor/reusable nebulizers from 3 brands (Pari Vios/Pari LC Plus, Pulmo-Aide 5650D/Viox, and Inspiration Elite/SideStream Plus) commonly used by patients with CF were tested. Compressor/nebulizers were operated for 1 h twice daily 5 d/wk for 24 wk. Compressor flow/pressure characteristics were measured every 6 wk. Maximal flow was recorded without and with the nebulizer (MF and MF/NEB, respectively). Pressure was recorded at zero flow (MP) and at MF/NEB (P/NEB). Particle size distribution, inhaled mass (IM), and IM in respirable range were evaluated at baseline and every 12 wk. RESULTS: Vios had statistically significant declines in MP and P/NEB at each measurement compared with baseline (45.8 and 32.6 psi for MP and 16.7 and 14.3 psi for P/NEB at wk 0 and 24, respectively, P < .05), but other compressors did not. MF and MF/NEB were stable over time but significantly varied among brands. Vios had the greatest slope of flow/pressure relationship (Vios > Pulmo-Aide > Inspiration Elite). Two Vios units stopped working at wk 11 and 24, respectively. All compressors maintained baseline IM, IM in respirable range, and aerosol characteristics. CONCLUSIONS: Long-term use of compressor/nebulizers in a regimen similar to that of patients with CF affected their performance. Pari Vios was the most affected brand, with declines in MP and P/NEB and 2 units that stopped working. Measurement of MF and MF/NEB could help identify compressors that are likely to fail.