Ventilator Settings731SAVe IIT1VOCSN*VOCSN Int
RR30/VT250/PEEP5/CL80/1POC51.0 ± 0.749.7 ± 0.643.0 ± 2.335.5 ± 0.134.0 ± 0.4
RR30/VT250/PEEP5/CL80/2POC84.7 ± 0.572.7 ± 0.265.8 ± 7.251.2 ± 0.345.3 ± 0.5
RR30/VT250/PEEP20/CL20/1POC50.7 ± 0.545.5 ± 0.343.1 ± 3.735.9 ± 0.132.0 ± 0.1
RR30/VT250/PEEP20/CL20/2POC84.4 ± 0.572.7 ± 0.364.2 ± 10.352.3 ± 0.143.2 ± 0.1
  • Measured FIO2 for the 250 mL VT settings at all conditions. *VOCSN internal oxygen concentrator with pulse dose settings equivalent to 1 and 2 SAROs POCs.