Ventilator Settings | 731 | SAVe II | T1 | VOCSN | *VOCSN Int |
RR30/VT250/PEEP5/CL80/1POC | 51.0 ± 0.7 | 49.7 ± 0.6 | 43.0 ± 2.3 | 35.5 ± 0.1 | 34.0 ± 0.4 |
RR30/VT250/PEEP5/CL80/2POC | 84.7 ± 0.5 | 72.7 ± 0.2 | 65.8 ± 7.2 | 51.2 ± 0.3 | 45.3 ± 0.5 |
RR30/VT250/PEEP20/CL20/1POC | 50.7 ± 0.5 | 45.5 ± 0.3 | 43.1 ± 3.7 | 35.9 ± 0.1 | 32.0 ± 0.1 |
RR30/VT250/PEEP20/CL20/2POC | 84.4 ± 0.5 | 72.7 ± 0.3 | 64.2 ± 10.3 | 52.3 ± 0.1 | 43.2 ± 0.1 |
Measured FIO2 for the 250 mL VT settings at all conditions. *VOCSN internal oxygen concentrator with pulse dose settings equivalent to 1 and 2 SAROs POCs.