PneuView 3 Measurements for Bench Model Comparison
 VT (mLs)PIP (cmH20)PEEP (cmH20)
Control (NIV w/ Non-Vented Mask)11.49.84.5
 14.7105
 13.810.25.1
Control Average13.3104.86
    
Bench Model A (Metaneb at dry side of heater) 24.211.22.2
 24.7132
 23.811.52.1
Bench A Average24.211.92.1
    
Bench Model B (Metaneb between mask and circuit)17.99.81.9
 159.92.3
 16.99.32.4
Bench B Average16.69.62.3
    
Bench Model C (Direct to mask)1013.52.2
 9.713.32.3
 9.613.42.2
Bench C Average9.713.42.2