Average Values and Percent Discrepancy of Alarms vs. Measured Values
 Patient (Measured Values)Alarm Setting% Discrepancy
PIP high22.70 (+/- 6.45)41.00(+/- 4.74)93.25 (+/- 55.26)
VE high9.18 (+/-2.00)18.98 (+/- 2.44)113.92 (+/- 40.56)
VE low9.18 (+/-2.00)3.28 (+/- 0.67)63.18 (+/- 8.59)
RR high19.54 (+/- 4.63)39.50 (+/- 3.94)110.82 (+/- 43.67)