Abstract
Objective
We designed a new endotracheal tube (ETT) cuff that does not form the folds that cause leakage of colonized subglottic secretions upon inflation within the trachea: a standard high-volume low-pressure cuff was draped with a second, highly elastic cuff made of a low-protein guayule natural latex rubber with 0.5 ml gel between the cuffs. We compared this prototype ETT cuff with four commercially available ETTs for efficacy in the prevention of fluid leakage across the cuff.
Design
In vitro study.
Measurements and results
We compared fluid leakage in our prototype cuff with that in four commercially available ETTs. Three cylindrical glass tubes 16, 20, and 22 mm in diameter were used as model tracheas, and five different intracuff pressures (20, 25, 30, 40, and 50 cmH2O) were tested. Each test was repeated three times with new ETTs. The guayule latex ETT cuff showed an average fluid leakage of 0.0007 ± 0.002 ml/min which was significantly lower than that in any of the other ETTs (Microcuff 0.07 ± 0.09, Mallinckrodt/Hi-Lo Evac 5 ± 5, Euromedical 7 ± 4, Sheridan/CF 41 ± 69).
Conclusions
Our prototype ETT cuff significantly reduced fluid leakage in this bench-top study. In vivo testing and evaluation is to follow.
Similar content being viewed by others
Introduction
Endotracheal tube (ETT) cuffs were initially made of thick (500 μm) Hevea latex rubber and required a high inflation pressure (200–400 cmH2O) to form an adequate tracheal seal [1]. The pressure transmitted to the tracheal wall was difficult to estimate from the intracuff pressure, and consequently overinflation of the cuff was common. In some studies tracheal wall pressures were as high as 200 cmH2O, greatly above the tracheal mucosal capillary pressure about 30 cmH2O [2, 3]. Widespread use of these low-volume high-pressure cuffs resulted in frequent tracheal injury [2, 3]. In the 1970s disposable ETT cuffs made of polyvinylchloride (PVC), termed high-volume low-pressure (HVLP) inflatable cuffs, were introduced to overcome this problem, and these have remained the standard until today. PVC cuffs are inelastic and 1.5–2 times larger than the internal diameter (ID) of the trachea. The HVLP cuff fills the trachea without being stretched, transmitting the intracuff pressure entirely to the tracheal wall. When inflated at 30 cmH2O, the HVLP cuff permits mechanical ventilation and preserves tracheal capillary perfusion but invariably forms multiple longitudinal folds. Bacteria colonize oropharyngeal secretions, and gastric contents can leak along the folds into the lower airways and the lungs [4–6], a major risk factor for ventilator-associated pneumonia (VAP) [7].
We have designed, manufactured, and tested a prototype ETT cuff, created by covering a standard HVLP ETT cuff with a thin, high-compliance, low-protein guayule latex rubber cuff. As with traditional HVLP cuffs this prototype requires low-inflation pressure to fill the trachea; however, the stretching of the compliant latex covering during inflation eliminates folds, assuring a perfect seal. This cuff design may therefore prove beneficial in the prevention of ischemic damage to the trachea and the leakage of colonized secretions into the lower airways.
Materials and methods
Latex prototype ETT cuff
The guayule latex cuff is cylindrical in shape (13 mm diameter) and 50–60 μm thick (Yulex, Maricopa, AZ, USA; Fig. 1). We draped the guayule latex cuff over the Mallinckrodt Hi-Lo Evac (ID 8 mm) ETT cuff (Mallinckrodt, NY, USA) and introduced 0.5 ml sterile gel (Surgilube, Altana Inc., Melville, NY, USA) between the two cuffs to facilitate a homogeneous distribution of inflation pressure and to reduce friction between the cuffs. The outer latex cuff was secured on both ends using silk ligatures.
Leakage test
We tested fluid leakage across the cuff using the guayule latex ETTs and four commercially available ETTs: Euromedical (Euromedical, Malaysia), Mallinckrodt Hi-Lo Evac, Microcuff (Kimberly Clark, GA, USA), and Sheridan/CF (Hudson, NC, USA). Tests were performed using vertically positioned, cylindrical glass tubes, 20 cm long, of three internal diameters (16, 20, and 22 mm), matching the broad range of adult human tracheas [6]. All ETTs (ID 8 mm) were inflated at intracuff pressures of 20, 25, 30, 40, and 50 cmH2O. A small reservoir was positioned below the model trachea to collect water leakage. Then 15 ml water was poured above the cuff and observed until all water was collected or until the 2-h test period had ended. Leakage is reported as average flow across the cuff, calculated by dividing the volume of water collected by either 120 min or the time at which all 15 ml water had leaked. Three new ETTs of each type were tested by three different investigators. Three guayule latex prototype ETTs and three microcuff ETTs were similarly tested for 24 h, using an inflation pressure of 20 cmH2O and a 20 mm diameter model trachea.
Wall pressure measurement
Three new ETT cuffs of each type were inflated inside a glass tube (ID 22 mm) at four pressures (10, 20, 30, and 40 cmH2O). The pressure exerted against the internal wall of the glass tube was measured with a manometer connected to a flat, small (4 cm long, 1 cm wide), inelastic, partially inflated PVC cuff inserted between the ETT cuff and the glass tube.
Statistical analysis
Since we performed three observations per pressure/diameter block, we obtained an F approximation to the Friedman's test based on the generalized linear model using the within block ranks to study the overall effect of the different cuffs. The four pairwise comparisons of interest were performed only if the overall F test was significant at p < 0.05. Interactions of main effects were also investigated. Bonferroni's correction was applied. The volume of fluid leakage during the 24 h test was compared by the Wilcoxon test.
Results
Assessment of folds
The guayule latex prototype cuff inflated in all model tracheas showed no folds; however, folds always occurred with the four tested HVLP cuffs.
Assessment of leakage
The average leakage was 6.6 × 10−4 ± 2.5 × 10−3 ml/min with the guayule latex prototype ETT, 7.3 × 10−2 ± 9.3 × 10−2 ml/min with the Microcuff, 5.0 ± 4.7 ml/min with the Mallinckrodt Hi-Lo EVAC, 7.2 ± 4.4 ml/min with the Euromedical, and 41 ± 69 ml/min with the Sheridan/CF. The primary analysis results showed that the overall test of equality between the five cuffs is statistically significant (p < 0.0001). In all four paired comparisons the guayule latex cuff outperformed the other ETT cuffs (p < 0.0001). In a secondary analysis, adjusted for the significant variables (intracuff diameter and pressure), the guayule latex cuff also outperformed the other four cuffs (p < 0.0001). The latter analysis suggests that the interaction of ETT brand and diameter was significant. However, stratification by diameter analyses did not change the above conclusions. Figure 2 shows average leakage flow as a function of the intracuff pressure. The latex prototype cuff performed significantly better (p < 0.0001) than all others, followed closely only by Microcuff.
Twenty-four hour tests
The average volume of water leaked (ml) across the guayule latex prototype cuff was 0.9 ± 5 ml and across the Microcuff 14.1 ± 2.2 ml (p = 0.03).
Tracheal wall pressure
Figure 3 shows the relationship between intracuff pressure and the pressure transmitted to the model tracheal wall for each cuff. The guayule latex cuff exerted on average a wall pressure 7.0 ± 1.9 cmH2O lower than the intracuff pressure.
Discussion
We designed a novel ETT cuff which forms no folds when inflated in a model trachea and substantially highly reduced, almost eliminated, fluid leakage at inflation pressure as low as 20 cmH2O. In comparison, all commercial HVLP cuffs showed multiple folds and did not prevent fluid leakage, even at 50 cmH2O. The Microcuff cuff, made of thin (7 μm) polyurethane film, formed only small folds and thus performed significantly better than the PVC HVLP cuffs. Dullenkopf et al. [8] previously tested the Microcuff and reported leakage lower than that in our study. This difference may be due to the larger volume of fluid that we injected above the cuffs (15 ml vs. 5 ml) resulting in higher hydrostatic pressure. We also tested the cuff in three model tracheas with different diameters vs. the single 20-mm diameter model trachea in the Dullenkopf et al. study. Young et al. [6] developed a silicone rubber ETT cuff (LoTrach, UK) that produces no folds upon inflation and thus capable of preventing fluid leakage. We did not include the LoTrach in our report since it behaves differently than standard HVLP cuffs: to seal the trachea requires intracuff pressures above 90 cmH2O, which is partially transmitted to the tracheal wall [9] and is much higher than the inflating pressure used in our study.
Our ETT cuff differs from the others in that it uses a very thin outer layer of guayule latex to completely enclose a traditional HVLP cuff. Guayule latex rubber cuff is highly compliant, tear resistant [10, 11], requires low pressure to be stretched, allowing an almost complete transmission of the intracuff pressure to the tracheal wall, and relies on the mechanical support of the internal HVLP cuff to be uniformly expanded. Therefore our cuff creates a tight seal and prevents fluid leakage even at low pressures. Although 1–8% of the population has hypersensitivity to latex proteins, the guayule latex material used in our prototype cuff is expected to be well-tolerated by latex-sensitive persons. Latex produced from the guayule shrub contains very little protein and no epitopes that cross-react with type I latex allergy [12–14]. Additionally, guayule latex also provides a strong barrier against blood-borne pathogens [15].
In conclusion, our results confirm previous reports [4–6] that traditional HVLP cuffs do not adequately seal the trachea and contribute to leakage of bacteria-colonized oropharyngeal secretions and gastric contents into the lower airways, a major pathogenic pathway of VAP [7, 16]. The development of a leak-proof ETT is therefore a major step towards the prevention of VAP. While in vitro tests never perfectly simulate physiological conditions, we believe that our results are a valid reflection of the performance of these cuffs. Long exposure to stress, biological secretions, heavy friction, and low pH may cause the thin guayule latex rubber to deteriorate and thus reduce the efficacy of this prototype ETT cuff in preventing fluid leakage. Prolonged in vivo testing is required to evaluate this issue.
References
Bernhard WN, Yost L, Turndorf H, Danziger F (1982) Cuffed tracheal tubes-physical and behavioral characteristics. Anesth Analg 61:36–41
Guyton D, Banner MJ, Kirby RR (1991) High-volume, low-pressure cuffs. Are they always low pressure? Chest 100:1076–1081
Cooper JD, Grillo HC (1969) The evolution of tracheal injury due to ventilatory assistance through cuffed tubes: a pathologic study. Ann Surg 169:334–348
Spray SB, Zuidema GD, Cameron JL (1976) Aspiration pneumonia; incidence of aspiration with endotracheal tubes. Am J Surg 131:701–703
Seegobin RD, van Hasselt GL (1986) Aspiration beyond endotracheal cuffs. Can Anaesth Soc J 33:273–279
Young PJ, Pakeerathan S, Blunt MC, Subramanya S (2006) A low-volume, low-pressure tracheal tube cuff reduces pulmonary aspiration. Crit Care Med 34:632–639
Rello J, Sonora R, Jubert P, Artigas A, Rue M, Valles J (1996) Pneumonia in intubated patients: role of respiratory airway care. Am J Respir Crit Care Med 154:111–115
Dullenkopf A, Gerber A, Weiss M (2003) Fluid leakage past tracheal tube cuffs: evaluation of the new Microcuff endotracheal tube. Intensive Care Med 29:1849–1853
Young PJ, Young WH (2003) Inflation of a pressure-limited cuff inside a model trachea. Med Eng Phys 25:465–473
Cornish K, Williams J (2006) Guayule latex: effective substitute for natural rubber latex to produce medical goods. Global handbook and directory on NR and SR lattices, pp 71–74
Cornish K, Williams J, Hall JL, McCoy III RG (2005) Production and properties of guayule latex—the natural solution to latex allergy. Proceedings of the 168th Technical Meeting of the Rubber Division, American Chemical Society
Siler DJ, Cornish K, Hamilton RG (1996) Absence of cross-reactivity of IgE antibodies from subjects allergic to Hevea brasiliensis latex with a new source of natural rubber latex from guayule (Parthenium argentatum). J Allergy Clin Immunol 98:895–902
Siler DJ, Cornish K (1994) Hypoallergenicity of guayule rubber particle proteins compared to Hevea latex proteins. Ind Crops Prod 2:307–313
Carey AB, Cornish K, Schrank PJ, Ward B, Simon RA (1995) Absence of cross-reactivity of IgE antibodies from Hevea brasiliensis latex allergic subjects with a new source of natural rubber latex from guayule (Parthenium argentatum). J Allergy Clin Immunol 98:895–902
Cornish K, Lytle CD (1999) Viral impermeability of hypoallergenic, low protein, guayule latex films. J Biomed Mater Res 47:434–437
Safdar N, Crnich CJ, Maki DG (2005) The pathogenesis of ventilator-associated pneumonia: its relevance to developing effective strategies for prevention. Respir Care 50:725–739
Acknowledgements
We thank Drs. Martha Vaughan, Danielle Springer, and Katrina Cornish for reviewing this manuscript. We thank Dr. Katrina Cornish and Jali Williams (Yulex Corporation, Maricopa, AZ, USA) for manufacturing and providing the guayule latex cuff to the authors without charge and without additional financial support.
Author information
Authors and Affiliations
Corresponding author
Electronic supplementary material
Rights and permissions
About this article
Cite this article
Zanella, A., Cressoni, M., Epp, M. et al. A double-layer tracheal tube cuff designed to prevent leakage: a bench-top study. Intensive Care Med 34, 1145–1149 (2008). https://doi.org/10.1007/s00134-008-1016-9
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00134-008-1016-9