Introduction

Use of mechanical ventilation is constantly increasing [1, 2] and is expected to continue rising in the near future as a consequence of an ageing population with increasingly complex disease states [3]. There has been a trend over the years towards use of ventilatory modes in which some degree of spontaneous respiratory activity is preserved, known as partial ventilator assist [4, 5]. These modes have been shown to reduce the adverse effects of prolonged sedation [68], neuropathy associated with use of neuromuscular blocking agents [9, 10], and ventilator-induced diaphragmatic dysfunction [1113], while also improving gas exchange [14].

The most often used partial ventilatory assist mode is pressure support (PS) [4, 5]. PS is often well tolerated, but the difference between the ventilatory profiles of the patient and the ventilator can lead to patient–ventilator asynchrony [1517]. Thille et al. [18] found that 24% of intubated patients undergoing PS had severe asynchrony. Asynchrony increases respiratory muscle load [19] and is associated with longer duration of mechanical ventilation [18, 20], its reduction therefore being of paramount importance. Neurally adjusted ventilatory assist (NAVA) could contribute to attaining that goal. Its principle is to record the diaphragmatic electrical activity (Eadi) at the distal esophageal level by means of electrodes embedded in a nasogastric tube and to use Eadi, after proper signal treatment, to control the ventilator [21]. With NAVA, the ventilator can be triggered by the Eadi signal, delivers inspiratory assistance in proportionality to the Eadi, and is usually cycled to exhalation by the Eadi signal.

Therefore, NAVA should theoretically allow near-perfect synchrony between the patient and the ventilator. The aim of our study is to explore if, in comparison with PS, NAVA could reduce trigger delay, improve expiratory synchrony, and reduce the number of patient–ventilator asynchronies in intubated intensive care patients. Partial results of the current study have been previously reported in abstract form [22].

Materials and methods

This was a prospective sequential interventional study conducted in the medicosurgical intensive care units (ICUs) of two university hospitals (Geneva, Switzerland and Brussels, Belgium). The study was approved by the Ethics Committee of both participating centers.

Patients

Patients admitted to the ICU, intubated for acute respiratory failure, and ventilated by PS were eligible for inclusion in the study if they had none of the exclusion criteria described in the Electronic Supplementary Material (ESM).

Ventilator

All patients were ventilated using a Servo-i® (Maquet Critical Care, Solna, Sweden) ventilator equipped with the standard commercial version of the NAVA module and software. With NAVA, the ventilator can be triggered and cycled off by the Eadi signal. The delivered pressurization is proportional to the Eadi, which is a direct expression of the patient’s inspiratory effort. Details of the ventilator’s functioning under NAVA are given in the ESM.

Both PS and NAVA were applied using the same machine.

Measurements

Respiratory parameters were acquired from the Servo-i® ventilator and recorded by Servo-tracker V 4.0® software (Maquet Critical Care, Solna, Sweden). Details about the measurements are given in the ESM. The percentage of respiratory cycles triggered by the Eadi and interrupted following the Eadi criteria were also recorded by Servo-tracker®. Recorded data were stored in a laptop computer for subsequent analysis.

Experimental protocol

After written informed consent was obtained, the patient’s standard nasogastric tube was replaced by the modified NAVA tube positioned according to the manufacturer’s recommendations [23]. Airway suctioning was performed before the beginning of the protocol. During the entire recording period, pressure support level, positive end-expiratory pressure (PEEP), FiO2, inspiratory trigger, and cycling off settings were maintained as set by the clinician in charge of the patient. Three sequential 20 min periods were recorded: (1) PS1, (2) NAVA with gain initially set to deliver the same peak pressure (comparable level of assist) as during the initial PS1 period and then kept at the same level during the 20 min recording, (3) PS2 with the same settings as those of period PS1. During NAVA, electrical signals recorded by the electrodes embedded in the NAVA nasogastric tube are processed automatically by the Servo-i® ventilator and used to control it [21]. Breath-by-breath analysis was performed on the recorded data by using Acqknowledge® software (Biopac Systems Inc., Goleta, CA, USA). Time parameters were determined from the flow and Eadi signals as illustrated in Fig. 1 and described in detail in the ESM. If there were artifacts on the Eadi signal or if the ventilator mode was automatically switched back from NAVA to PS (NAVA includes a built-in safety feature whereby, in case the Eadi signal became artifacted or lost, the machine automatically reverts to standard PS), the corresponding periods of time were excluded from the analysis. The total number of the five different types of patient–ventilator asynchronies described by Thille et al. [18], namely ineffective effort, autotriggering, premature cycling, delayed cycling, and double triggering, was determined for each 20 min recording period, based on flow and airway pressure signal as well as on the Eadi signal according to a previous publication from our group [24]. The five types of asynchronies are described in the ESM. During NAVA, we separated double triggering into two different groups, namely type 1, when double triggering was the consequence of an Eadi cycling off criterion reached too early because of a biphasic aspect of the Eadi signal, and type 2, when the cause was different. Figure 2 provides an illustration of both types of double triggering under NAVA. The total number of asynchronies was reported as the total number of events per minute as well as in terms of AI, namely the total number of asynchronies given as a percentage of the total respiratory rate, computed as the sum of the ventilator respiratory rate and of the number of ineffective efforts per minute [AI% = (number of asynchrony events/ventilator respiratory rate + ineffective efforts) × 100]. The number of each type of asynchrony was reported as the total number of each event per minute. The indexed tidal volume (VTi) was defined as the tidal volume in milliliters divided by the patient’s predicted body weight (PBW) in kilograms, calculated for each respiratory cycle from the tidal volume obtained from the Servo-tracker® software.

Fig. 1
figure 1

Airway pressure, flow, and electrical diaphragmatic activity curves in pressure support and in neurally adjusted ventilatory assist. PS pressure support, NAVA neurally adjusted ventilatory assist, PEEP positive end-expiratory pressure, Eadi electrical activity of the diaphragm, Tiv ventilator pressurization time, Tin neural inspiratory time, Td trigger delay, Tiex inspiratory time in excess

Fig. 2
figure 2

Double triggering under PS and NAVA. Paw airway pressure, Eadi electrical activity of the diaphragm

Statistics

Normality was assessed using the Kolmogorov–Smirnov test, which failed for all results; therefore, values are reported as median (IQR). Parameters reflecting patient–ventilator synchrony in the same subject were compared between PS1, NAVA, and PS2 using one-way repeated-measures analysis of variance (ANOVA) on ranks. Post hoc pairwise comparisons were performed using the Newman–Keuls procedure. Statistics were computed using SigmaStat 3.0 (SPSS, Chicago, IL, USA). For all statistical tests, the significance level was set to p < 0.05.

Results

Twenty-five patients were included in the study. Three patients were excluded at time of analysis, one because of a final diagnosis of neuromuscular disease (exclusion criterion) and two because of nonanalyzable Eadi recording under PS.

The main characteristics of the included patients and ventilator settings are summarized in Table 1. A typical recording under NAVA and PS is shown in Fig. 3. From the total recording time of 1,320 min, 24.3 min (1.8%) was discarded from the analysis because of artifacts on the Eadi signal. From the 440 min of recording under NAVA, 9.4 min (2.1%, corresponding to 20 events) was discarded from the analysis because of automatic reversion to PS. During NAVA, the Eadi was used to trigger 80% (58–89%) and cycle 100% (96–100%) of the breaths. No significant differences were found between the two PS ventilation periods in terms of trigger delay (Td), inspiratory time in excess (Tiex), total number of asynchronies, and number of each type of asynchrony. Detailed results of pairwise analysis are given in Table 2.

Table 1 Main clinical characteristics of the patients (n = 22) and ventilator settings
Fig. 3
figure 3

Typical recording in pressure support and in NAVA. Paw airway pressure, Eadi electrical activity of the diaphragm

Table 2 Number of total and specific asynchronies per minute as well as other respiratory parameters

Td and Tiex

As shown in detail in Table 2, Td was reduced in NAVA. In NAVA, Td was below 150 ms for every patient, whereas in PS 16/22 patients (72.7%) had Td >150 ms. Tiex was also reduced in NAVA.

Asynchronies

Results are summarized in Table 2, with p values for ANOVA and pairwise analysis. The total number of asynchronies was significantly reduced in NAVA compared with PS1 and PS2. The median value of AI index was reduced in NAVA compared with PS1 and PS2. AI was higher than 10% in 6/22 (27%) in NAVA and in 12/22 (54.5%) and 14/22 (63.6%) of patients in PS1 and PS2, respectively. There were neither ineffective efforts nor late cycling in NAVA. In contrast, there were ineffective efforts and late cycling in PS1 and PS2. There were fewer premature cyclings in NAVA compared with PS1 and PS2. There was no difference between NAVA and PS in terms of autotriggering. Finally, there was more double triggering in NAVA than in PS1 and PS2. Under NAVA, 63.1% of double triggerings were type 1 double triggering whereas only 36.9% were type 2 double triggering.

Other respiratory parameters

There was no difference in minute ventilation between NAVA and PS. VTi was lower in NAVA than in PS1 and PS2. Respiratory rate was higher in NAVA than in PS1 and PS2. Mean airway pressure (Pawm) was lower with NAVA than with PS1 and PS2. Main respiratory parameters are summarized in Table 2.

Discussion

Our results show that NAVA can improve patient–ventilator synchrony by reducing Td, Tiex, and the total number of asynchronies compared with PS. In particular, there were no ineffective efforts or delayed cycling under NAVA.

Before discussing these results, several limitations of this study must be addressed. First, the tracings were analyzed by one investigator from Geneva, rather than one from each center, which could lead to systematic bias. However, the methodology and the reading criteria had been strictly defined before the beginning of the analysis. Moreover, in case of doubt, questionable tracings were discussed among three investigators during data session meetings to find consensus. Second, artifacted tracings were not considered for the analysis. However, only 1.8% of the whole recording time was thus excluded. Third, the size of our patient population was fairly small. Fourth, even if the level of assistance in NAVA has been determined according to the usual procedure described by the manufacturer and to previous studies (NAVA gain or proportionality factor between Eadi and pressure delivered by the ventilator, set so as to obtain the same peak pressure in NAVA as in PS) [25, 26], the comparability of assistance levels in PS and NAVA is questionable. Indeed, the airway pressure curves have different profiles as illustrated in Figs. 1 and 3, and Pawm is significantly lower in NAVA than in PS, as demonstrated by our results and as previously described [26]. Finally the expiratory trigger criterion in PS (chosen by the clinician in charge of the patient) for the eight obstructive patients included in our study was between 25 and 30% of the maximal flow value. This means that this criterion had not been optimized for these eight obstructive patients [27], which could have influenced patient–ventilator synchrony parameters.

Td was significantly shorter under NAVA, which is easily explained by the fact that the Eadi, an expression of the respiratory center’s activity, is used to trigger the ventilator, rather than a pneumatic signal located at the airway opening or inside the ventilator. The significant reduction of Td in NAVA compared with PS should be attenuated by the fact that, under NAVA, according to a recent publication [28], not every respiratory cycle is triggered by the Eadi because of the first-come first-served algorithm used for triggering by the commercially available system. However, as in our study 80% (58–89%) of the respiratory cycles were triggered by the Eadi in NAVA, there was still a significant advantage for most patients. Under NAVA, no patient had mean Td above 150 ms, whereas it was above this threshold in 72.7% of patients under PS. This is likely of clinical importance, as this delay of 150 ms corresponds to the conscious threshold of perception, a potential source of discomfort during the triggering process [29]. In PS, a significant inspiratory effort (estimated from the pressure–time product) is needed to trigger the ventilator [30]. In contrast, with NAVA (when the ventilator’s delivered cycle is triggered by the Eadi), as the ventilator’s delivered pressurization immediately follows the increase in Eadi signal (without any depression on the pressure–time curve), no inspiratory effort is theoretically needed to trigger the ventilator. As obstructive patients must overcome the intrinsic PEEP due to air trapping to trigger the ventilator in PS, they often present major trigger asynchronies with long trigger delays and high number of ineffective efforts [20, 31] associated with increased work of breathing (WOB) [32]. Because in NAVA the triggering process is independent of air trapping, NAVA could be particularly attractive to improve trigger asynchronies in these patients, as demonstrated in a recent study by Spahija et al. [26]. However, the results of that study might overestimate the benefits of NAVA triggering compared with PS, as the expiratory trigger was set in PS at a low value of 5%, which could have increased hyperinflation under PS and thereby worsened patient–ventilator synchrony under PS [27, 33]. In contrast, our study was performed with usual (but, as previously mentioned, not optimized) clinical settings for expiratory trigger (set between 25% and 30%) and also shows that there was no ineffective effort under NAVA whereas there were ineffective efforts under PS.

Delayed cycling was improved with NAVA, with Tiex reduced by more than 50% compared with PS. Theoretically, patient–ventilator synchrony should be nearly perfect with NAVA. However, because in NAVA the ventilator’s delivered assist cycles off into exhalation when the Eadi decreases to 70% of the maximal Eadi value, there is an inherent delay between the end of the neural inspiratory time and the end of the assistance delivered by the ventilator. We found an expected residual Tiex of 126 ms (111–136 ms). Tiex in NAVA is, however, very short and under the threshold of consciousness of 150 ms. The significant reduction of Tiex under NAVA compared with PS is important, as the presence of Tiex is associated with increased WOB [34, 35] and promotes dynamic hyperinflation [27]. As delayed cycling occurs mostly in patients with obstructive pulmonary disease (because of changes in instantaneous flow in the airways due to obstruction that delays the moment at which the expiratory set criterion is reached) [31, 32], the NAVA technology should be, also for that reason, of particular interest in this subgroup of patients. It is important to note that, despite the improvement of delayed cycling in NAVA, we did not find more premature cycling than in PS. In fact, the number of premature cycling is even reduced in NAVA compared with PS.

With NAVA, in contrast to what Colombo et al. [25] have previously described, more double triggering was present than under PS. This increase in double triggering can probably be explained by technical reasons. Indeed, the filtered Eadi signal used by NAVA to command the ventilator sometimes has a biphasic aspect. In this case (63.1% of the total number of double triggering events in NAVA in our study), the initial decrease in the Eadi signal after the first peak is interpreted by the NAVA software as the cycling criterion which stops the ventilator’s delivered pressurization. A new increase in the Eadi signal immediately follows this premature expiratory cycling and induces a new pressurization. This probably does not have major clinical importance, as the presence of double triggering is probably not associated with increased WOB. However, it is possible that this phenomenon is a source of discomfort for some patients. As our study was not designed to evaluate patient comfort, this point should be further investigated.

Despite the fact that in NAVA every Eadi signal variation is immediately followed by pressurization from the ventilator, we found no increase in autotriggering. When autotriggering was present in NAVA, it was usually the consequence of artifacts on the Eadi signal.

Our study shows that there are overall fewer asynchronies in NAVA compared with PS, which confirms our hypothesis that NAVA can improve patient–ventilator interaction. The number of patients with AI >10% was also reduced by more than 50% with NAVA compared with PS. Of note, we found twice as many patients with AI >10% in PS than Thille et al. [18]. This is probably related to the fact that we detected asynchronies not only on the basis of flow and pressure signal as Thille et al. did, but also on a reliable Eadi signal. Our results are in line with those of Colombo et al. [25], who found, taking into account only ineffective efforts and double triggering in the AI, 36% of patients with AI >10% under PS.

Despite minute volume being comparable under NAVA and PS, it is interesting to note that VTi was lower with NAVA and in the lower part of the range defined as safe by the acute respiratory distress syndrome (ARDS) network [36], suggesting, as previously reported [25], a possible protective effect of NAVA against volutrauma in such patients. This could be related to the fact that in NAVA, in contrast to PS, the amount of pressure delivered by the ventilator is proportional to the Eadi, which is a direct expression of the patient’s inspiratory effort. NAVA could thereby prevent overassist. This hypothesis should be further investigated.

Conclusions

NAVA can improve patient–ventilator interaction in intubated patients by reducing Td, suppressing ineffective efforts, and reducing Tiex and the total number of asynchronies. These effects can be explained by the use of the Eadi, a direct expression of the respiratory center activity, to trigger and cycle off the ventilator as well as to the potential of NAVA to prevent overassist.

As a reduced number of asynchronies is associated with shorter duration of mechanical ventilation [18, 20], NAVA could prove beneficial in this patient population. Further studies should now aim to determine whether this improved synchrony can impact patient outcome.