Research
Pneumatic Tube Delivery System for Blood Samples Reduces Turnaround Times Without Affecting Sample Quality

https://doi.org/10.1016/j.jen.2005.11.013Get rights and content

Study objectives

In this study, blood samples from ED patients that were delivered to the laboratory by a pneumatic tube delivery system and by a human courier were compared for timeliness and quality of results.

Methods

We studied all consecutive measurements of serum hemoglobin and potassium ordered from 2 emergency departments of a multisite tertiary care hospital system, one with a pneumatic tube system and the other using human couriers. Turnaround time was measured from the time that the test was ordered by the physician to the time the result was reported on the hospital information system. Hemolysis was measured with use of a standardized, validated method.

Analysis

Times were normalized by log transformation (ln [minutes + 1]), and a comparison of sites was conducted using analysis of variance. Hemolysis rates of the 2 delivery systems were compared by χ2.

Results

There was no significant difference in hemolysis rate between the 2 methods of delivery (7/121 [5.79%] with a pneumatic tube system and 20/200 [10%] with a human courier). When delivered with a pneumatic tube system, the mean turnaround times (with ranges) for both hemoglobin (33 minutes [4-230]) and potassium (64 [34-208]) were shorter than those delivered by a human courier (43 minutes [3-150] and 72 [28-213], respectively).

Conclusion

The use of a pneumatic tube delivery system for transporting blood samples from the emergency department to the laboratory can significantly reduce the turnaround times of results without a reduction in sample quality.

Section snippets

Ethics

This quality assessment study did not involve human subjects or experiments and therefore was deemed by our hospital and university Review Ethics Board to be exempt from formal approval. Identification of blood samples was removed as the data for each sample were entered into a computer database for analysis. This process maintained the confidentiality of patient information by eliminating all patient identifiers from each sample.

Setting

The setting was 2 emergency departments of a multisite, tertiary

Results

Table 1 illustrates that the turnaround time for hemoglobin was significantly less than that for serum potassium (F[1,660] = 406, P < .001). More importantly for the current discussion, the same table illustrates that the turnaround time for the PTS was significantly less than that of a human courier system (F[1,66] = 136, P < .001). Finally, the hemolysis rate of 7/121 (5.8%) observed with a PTS did not differ from the rate of 20/200 (10%) observed with a human courier (χ2 = .174, P > .15).

Discussion

The results of this study demonstrate that transportation of blood specimens from the emergency department to the laboratory via a PTS can be time effective without compromising the quality of the samples. A previous study simultaneously comparing TATs via a human courier and a PTS found no statistical difference between the two.6 However, students were employed specifically for the purpose of carrying the specimens to the laboratory in that study, presumably as soon as the specimens were

Conclusion

The use of a pneumatic tube delivery system for transporting blood samples from the emergency department to the laboratory can significantly reduce the TATs of results without a reduction in sample quality.

Christopher M.B. Fernandes is Professor and Head of Emergency Medicine, McMaster University, Hamilton, Ontario, Canada.

References (15)

There are more references available in the full text version of this article.

Cited by (60)

  • Causes, consequences and management of sample hemolysis in the clinical laboratory

    2017, Clinical Biochemistry
    Citation Excerpt :

    Transport of blood samples from the location of blood collection to the laboratory may also be associated with hemolysis. Fernandes et al. reported a significantly lower percentage of troublesome hemolysis when centrifugation took place at the site of collection rather than in the laboratory [46]. Exposure of blood samples to extreme temperatures during transport and long transport duration may cause hemolysis [5,46].

  • Effects of one directional pneumatic tube system on routine hematology and chemistry parameters; A validation study at a tertiary care hospital

    2017, Practical Laboratory Medicine
    Citation Excerpt :

    In this study, there were no statistical significant differences for CBC, WBC differential count, and ESR between both methods. Results of the present study correspond well with those of the earlier studies which reported that PTS does not affect hematology parameters [3,9–11]. A recent study found no statistical differences for CBC, and ESR between the PTS and hand-carried method [10].

View all citing articles on Scopus

Christopher M.B. Fernandes is Professor and Head of Emergency Medicine, McMaster University, Hamilton, Ontario, Canada.

Andrew Worster is Clinical Assistant Professor, Emergency Medicine, Hamilton Health Sciences and McMaster University, and Clinical Assistant Professor, Department of Clinical Epidemiology and Biostatistics, McMaster University, Hamilton, Ontario, Canada.

Kevin Eva is Assistant Professor, Department of Clinical Epidemiology and Biostatistics, McMaster University, Hamilton, Ontario, Canada.

Stephen Hill is Assistant Professor, Hamilton Regional Laboratory Medicine Program and Department of Pathology and Molecular Medicine, McMaster University, Hamilton, Ontario, Canada.

Catherine McCallum is Quality Manager, Hamilton Regional Laboratory Medicine Program, Hamilton, Ontario, Canada.

Presented at the American College of Emergency Physicians Research Forum, Boston, Massachusetts, October 2003.

View full text