Chest
Clinical Investigations in Critical CareComparing Two Heat and Moisture Exchangers With One Vaporizing Humidifier in Patients With Minute Ventilation Greater Than 10 L/min
Section snippets
Methods
The patients were included in a prospective, randomized, controlled, nonblinded study. With institutional approval and informed consent from the families, we studied tracheally intubated, mechanically ventilated patients sedated with sufentanil and paralyzed with vecuronium bromide. They needed mechanical ventilation for acute respiratory failure following multiple trauma. The ventilatory circuit consisted of inspiratory and expiratory lines connected by a Y-piece (Fig 1). The ventilator used
Results
Nine patients were included in the study: 7 men and 2 women with a mean age of 55 ± 8 years (17 to 71 years) and a mean weight of 67 ± 10 kg (44 to 75 kg). The minute ventilation was 11.9 ± 1.2 L/min (10 to 14 L/min), the tidal volume was 694±143 mL (400 to 840 mL), the respiratory rate was 17.6±1.9 cycles/min (15 to 23 cycles/min), the PEEP was 7.4±3.5 cm H2O (0 to 13 cm H2O), and the FI02 was 0.55 ±0.12 (0.4 to 0.7). The esophageal temperature was 37.5±1,4°C (36.0 to 39.5°C) at the beginning
Discussion
This prospective randomized controlled study clearly shows large variations in the ability of one VH and two HMEs to preserve heat and water of the ventilatory gases when used in patients with minute ventilation >10 L/min. The VH tested (Bennett Cascade 2) was significantly superior to the other two devices (Pall Ultipor filter and DAR Hygroster). The DAR Hygroster HME was also clearly superior, in terms of temperature and water preservation, to the Pall Ultipor HME. This was observed in each
References (35)
Temperature, humidity and mucus flow in the intubated trachea
Br J Anaesth
(1974)- et al.
Heat and moisture exchangers and vaporizing humidifiers in the intensive care unit
Chest
(1990) - et al.
Humidification in anaesthesia: a review of the present situation
Br J Anaesth
(1972) - et al.
Effect of dry anesthetic gases on tracheobronchial epithelium
Anesthesiology
(1972) - et al.
Adult body temperature and heated humidification during general anesthesia
Anesth Analg
(1981) - et al.
Effect of dry and humidified gases on the respiratory epithelium in rabbits
J Pediatr Surg
(1975) Humidification and mucus flow in the intubated trachea
Br J Anaesth
(1973)- et al.
A study of humidification in tracheostomized dogs
Br J Anaesth
(1973) - et al.
Humidity and the anesthetized patient
Anesthesiology
(1979) - et al.
Contaminated condensate in mechanical ventilator circuits: a risk factor for nosocomial infections
Am Rev Respir Dis
(1984)
Comparative evaluation of disposable humidifiers
Acta Anaesthesiol Scand
Hydrophobic versus hygroscopic heat-moisture exchangers
Anaesthesia
The Pall Ultipor breathing circuit filters—an efficient heat and moisture exchanger
Anesth Analg
A comparison of five heat and moisture exchangers
Anaesthesia
An evaluation of six disposable heat moisture exchangers
Anaesth Intensive Care
Laboratory investigation of six artificial noses for use during endotracheal anesthesia
Anesth Analg
Cited by (62)
Humidification performance of 48 passive airway humidifiers comparison with manufacturer data
2009, ChestCitation Excerpt :This may underline the specific issue of the hygrometric measurements. Given that the performance of some devices may decrease with prolonged use, one limitation of the present study was that the performances of the devices were not assessed after prolonged use.17,49 The impact of prolonged use on HME performance and clinical consequences remains poorly documented and should be assessed, especially if passive humidifiers are used as first-line humidification devices.
Performance of heated wire humidifiers: An in vitro study
2007, Journal of Critical CareCitation Excerpt :The dry probe directly measures the actual gas temperature, whereas the wet probe, which is coated with sterile water–wetted cotton, measures the dryness of the gas. The evaporation of the sterile water is proportional to the dryness of the gas; thus, the difference in temperature levels between the dry and wet probes is related to the dryness of the gas [20]. The temperature levels at the dry and wet probes were measured over 10 consecutive breaths, and the mean value was calculated.
Main Techniques for Evaluating the Performances of Humidification Devices Used for Mechanical Ventilation
2023, Humidification in the Intensive Care Unit: The EssentialsHumidification During Invasive Mechanical Ventilation
2023, Humidification in the Intensive Care Unit: The EssentialsDevelopment and characterization of triclosan coated heat and moisture exchange filter for ventilation therapy
2022, Indian Journal of Fibre and Textile Research