Chest
Volume 118, Issue 5, November 2000, Pages 1441-1449
Journal home page for Chest

Laboratory and Animal Investigations
Nasal Continuous Positive Airway Pressure Devices Do Not Maintain the Set Pressure Dynamically When Tested Under Simulated Clinical Conditions

https://doi.org/10.1378/chest.118.5.1441Get rights and content

Study objectives

Nasal continuous positive airwaypressure (CPAP) is standard therapy for obstructive sleep apneasyndrome. The effective nasal mask pressure may be adversely affectedby factors that increase system resistance (eg, longtubing and/or water condensation) and by dynamic variables (breathingfrequency [f] and tidal volume [Vt]).The present study was conducted in order to assess the performance ofCPAP machines throughout a range of simulated clinical conditions.

Design

Four currently used CPAP machines were testedat settings of 5, 10, 15, and 20 cm H2O using a pulmonarywaveform generator to produce Vts of 0.4, 0.8, and 1.2 L atfrequencies of 10, 20, and 30 breaths/min. Machines were tested underfive conditions: 6-foot and 12-foot tubing, with and without an in-linehumidifier, and 12-foot tubing with humidifier and watercondensation.

Measurements

Maximum and minimum maskpressure measurements were obtained during five respiratory cycles foreach dynamic variable under each of the five conditions and CPAPsettings (180 experiments on each of four CPAP models).

Results

Using typical clinical parameters (Vt,0.4 L and 0.8 L; f, 10 breaths/min and 20 breaths/min;and CPAP, 5 to 15 cm H2O), mask pressure consistentlyvaried above and below the set point when additional tubing and/or ahumidifier were added to the system (0.7 to 2.9 cm H2Obelow and 0.5 to 1.0 cm H2O above the set pressure). Watercondensation caused large pressure deviations (inspiratory pressureranged from 3.5 to 5.6 cm H2O below set pressure, andexpiratory pressure ranged from 0.7 to 3.5 cm H2O above setpressure).

Conclusions

Therapy and compliance couldbe adversely affected because some CPAP machines in current use do notmaintain constant continuous mask pressure when tested using simulatedconditions, especially when water condenses in thetubing.

Section snippets

Materials and Methods

Testing was performed in the Pulmonary Laboratory at LDSHospital in Salt Lake City, UT. The average ambient barometric pressureat the altitude of 1,400 m is 647 mm Hg, average temperature was22°C, and average relative humidity was 37%.

A schematic of the experimental design is shown in Figure 1. A pulmonary waveform generator (PWG) was used to simulate tidalbreathing of a patient (PWG System S/N 714; MH Custom Design &Manufacturing; Midvale, UT). The PWG is precise and reproducible andhas been

Steady Flow Conditions

All four devices performed equivalently under steady flowconditions (ie, without simulated respiration). There was nosignificant pressure drop due to increasing resistance with any systemconfiguration during steady flow. The mean pressure decrease from thedevice to the mask was 0.36 cm H2O.

Dynamic Flow Conditions

Pressure measured at the mask was found to vary substantially inparallel with the respiratory cycle, with minimal phase delay and withvery little difference between machines. For example, in the

Discussion

The major conclusion from this study is that some currently usedCPAP machines do not maintain a constant continuous set mask pressurewhen tested using simulated ordinary clinical conditions. Differencesamong the four machines used in this study were not clinicallyimportant with either steady-state or dynamic tests; with all devices,the main deviation from the set pressure was a decrease during earlyinspiration. Various alterations of the apparatus and especially theaccumulation of water

References (23)

  • LG Olson et al.

    A community study of snoring and sleep-disordered breathing: prevalence

    Am J Respir Crit Care Med

    (1995)
  • Cited by (0)

    Financial support was provided by the Deseret Foundation, LDS Hospital,Salt Lake City, UT.

    Equipment was provided by the Devilbiss Corporation and IntermountainHome Health Care.

    Study data were presented at the American Thoracic Society Annualmeeting, San Diego, CA, April 23–28, 1999.

    View full text