Contamination of a multiple-use suction catheter in a closed-circuit system compared to contamination of a disposable, single-use suction catheter

Respir Care. 1986 Nov;31(11):1086-91.

Abstract

Multiple-use (M-U) closed-system endotracheal suction catheters are effective in preventing arterial oxygen desaturation in patients on positive end-expiratory pressure (PEEP) and may lessen the frequency of bradycardia and hypotension in unstable patients who are prone to these complications of suctioning. However, because M-U catheters remain attached to the ventilator circuit and are reintroduced repeatedly into the patient's airway over 24 hours or longer, they could become heavily contaminated with pathogens. We hypothesized a risk of autocontamination to the patient by re-inoculation of the respiratory tract with organisms that flourished on the M-U catheter while it was isolated from the patient's immune defenses or antibiotic therapy.

Methods: We tested this hypothesis in 30 mechanically ventilated adult patients with positive sputum cultures. We measured and compared the amount of bacteria present on an M-U catheter at the end of a 24-hour use period, the amount of bacteria present in the patient's sputum at that time, and the amount of bacteria present on a single-use (S-U) catheter at that time, after it had made one pass into the patient's airway. Organisms recovered and colony counts were also compared to results of a sputum culture obtained before the study began.

Results: Fourteen different pathogens or potential pathogens were recovered, in numbers of 2 x 10(1) to 2 x 10(7)colony-forming units. The greatest number of colonies was most often recovered from the sputum specimen, and statistical analysis showed no differences in rate or magnitude of contamination between M-U and S-U catheters.(ABSTRACT TRUNCATED AT 250 WORDS)

Publication types

  • Comparative Study

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Catheters, Indwelling*
  • Disposable Equipment*
  • Equipment Contamination*
  • Female
  • Hospital Bed Capacity, 300 to 499
  • Humans
  • Male
  • Middle Aged
  • Respiration, Artificial / instrumentation*
  • Sputum / microbiology*
  • Suction / instrumentation*
  • Washington