The accuracy of non-invasive carbon dioxide monitoring: a clinical evaluation of two transcutaneous systems

Anaesthesia. 2007 Apr;62(4):394-9. doi: 10.1111/j.1365-2044.2007.04987.x.

Abstract

We determined the accuracy of two transcutaneous carbon dioxide monitoring systems (SenTec Digital Monitor with V-Sign Sensor and TOSCA 500 with TOSCA Sensor 92) for the measurement of single values and trends in the arterial partial pressure of carbon dioxide in 122 adult patients during major surgery and in 50 adult patients in the intensive care unit. One or several paired measurements were performed in each patient. The first measurement was used to determine the accuracy of a single value of transcutaneous carbon dioxide; the difference between the first and the last measurements was used to analyse the accuracy and to track trends. We defined a 95% limit of agreement of <or=1 kPa as being clinically useful. There was insufficient agreement between transcutaneous carbon dioxide partial pressure values derived from the two systems and arterial carbon dioxide values for both single values and trends as defined by our suggested limit of agreement. We conclude that these systems cannot replace conventional blood gas analysis in the clinical setting studied.

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • Aged
  • Blood Gas Monitoring, Transcutaneous / instrumentation
  • Blood Gas Monitoring, Transcutaneous / methods*
  • Carbon Dioxide / blood*
  • Critical Care / methods
  • Female
  • Humans
  • Male
  • Middle Aged
  • Monitoring, Intraoperative / instrumentation
  • Monitoring, Intraoperative / methods*
  • Monitoring, Physiologic / methods
  • Partial Pressure
  • Prospective Studies
  • Reproducibility of Results

Substances

  • Carbon Dioxide