Automatic control of tracheal tube cuff pressure in ventilated patients in semirecumbent position: a randomized trial

Crit Care Med. 2007 Jun;35(6):1543-9. doi: 10.1097/01.CCM.0000266686.95843.7D.

Abstract

Objective: The aspiration of subglottic secretions colonized by bacteria pooled around the tracheal tube cuff due to inadvertent deflation (<20 cm H2O) of the cuff plays a relevant role in the pathogenesis of ventilator-associated pneumonia. We assessed the efficacy of an automatic, validated device for the continuous regulation of tracheal tube cuff pressure in preventing ventilator-associated pneumonia.

Design: Prospective randomized controlled trial.

Setting: Respiratory intensive care unit and general medical intensive care unit.

Patients: One hundred and forty-two mechanically ventilated patients (age, 64 +/- 17 yrs; Acute Physiology and Chronic Health Evaluation II score, 18 +/- 6) without pneumonia or aspiration at admission.

Interventions: Within 24 hrs of intubation, patients were randomly allocated to undergo continuous regulation of the cuff pressure with the automatic device (n = 73) or routine care of the cuff pressure (control group, n = 69). Patients remained in a semirecumbent position in bed.

Measurements and main results: The primary end point variable was the incidence of ventilator-associated pneumonia. Main causes for intubation were decreased consciousness (43, 30%) and exacerbation of chronic respiratory diseases (38, 27%). Cuff pressure <20 cm H2O was more frequently observed in the control than the automatic group (45.3 vs. 0.7% determinations, p < .001). However, the rate of ventilator-associated pneumonia with clinical criteria (16, 22% vs. 20, 29%) and microbiological confirmation (11, 15% vs. 10, 15%), the distribution of early and late onset, the causative microorganisms, and intensive care unit (20, 27% vs. 16, 23%) and hospital mortality (30, 41% vs. 23, 33%) were similar for the automatic and control groups, respectively.

Conclusions: Cuff pressure is better controlled with the automatic device. However, it did not result in additional benefits to the semirecumbent position in preventing ventilator-associated pneumonia.

Publication types

  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • APACHE
  • Equipment Design
  • Female
  • Hospital Mortality
  • Humans
  • Intensive Care Units
  • Length of Stay
  • Male
  • Middle Aged
  • Pneumonia, Ventilator-Associated / prevention & control*
  • Prospective Studies
  • Respiration, Artificial / instrumentation
  • Respiration, Artificial / methods*
  • Trachea*