Pilot study of a new device to titrate oxygen flow in hypoxic patients on long-term oxygen therapy

Respir Care. 2011 Apr;56(4):429-34. doi: 10.4187/respcare.00983. Epub 2011 Jan 21.

Abstract

Background: The O(2) Flow Regulator (Dima, Bologna, Italy) is a new automated oxygen regulator that titrates the oxygen flow based on a pulse-oximetry signal to maintain a target S(pO(2)). We tested the device's safety and efficacy.

Methods: We enrolled 18 subjects with chronic lung disease, exercise-induced desaturation, and on long-term oxygen therapy, in a randomized crossover study with 2 constant-work-load 15-min cycling exercise tests, starting with the patient's previously prescribed usual oxygen flow. In one test the oxygen flow was titrated manually by the respiratory therapist, and in the other test the oxygen flow was titrated by the O(2) Flow Regulator, to maintain an S(pO(2)) of 94%. We measured S(pO(2)) throughout each test, the time spent by the respiratory therapist to set the device or to manually regulate the oxygen flow, and the total number of respiratory-therapist titration interventions during the trial.

Results: There were no differences in symptoms or heart rate between the exercise tests. Compared to the respiratory-therapist-controlled tests, during the O(2) Flow Regulator tests S(pO(2)) was significantly higher (95 ± 2% vs 93 ± 3%, P = .04), significantly less time was spent below the target S(pO(2)) (171 ± 187 s vs 340 ± 220 s, P < .001), and the O(2) Flow Regulator tests required significantly less respiratory therapist time (5.6 ± 3.7 min vs 2.0 ± 0.1 min, P = .005).

Conclusions: The O(2) Flow Regulator may be a safe and effective alternative to manual oxygen titration during exercise in hypoxic patients. It provided stable S(pO(2)) and avoided desaturations in our subjects.

Publication types

  • Randomized Controlled Trial

MeSH terms

  • Aged
  • Automation
  • Cross-Over Studies
  • Exercise Test
  • Humans
  • Hypoxia / physiopathology
  • Hypoxia / therapy*
  • Oxygen Inhalation Therapy / instrumentation*
  • Pilot Projects
  • Pulmonary Disease, Chronic Obstructive / physiopathology
  • Pulmonary Disease, Chronic Obstructive / therapy*