Endotracheal Tube Seal and Suction Performance in a Novel Biorealistic Tracheal Model

Respir Care. 2015 Aug;60(8):1113-9. doi: 10.4187/respcare.03799. Epub 2015 May 5.

Abstract

Background: Endotracheal tube (ETT) cuffs create a seal to protect against secretion entry to the lungs. Cuff inflation currently is recommended at 20-30 cm H2O pressure. ETT designs have variable seal performance in bench studies using rigid tracheal models lacking the dynamic characteristics of the human trachea. We compared ETT designs within a new, biorealistic tracheal model to assess cuff and suction performance in the setting of a compliant trachea.

Methods: Three ETT designs (Mallinckrodt Hi-Lo, KimVent Microcuff, and Sheridan/HVT) were tested for performance by simulant leakage below the cuff and air leakage (measured as return tidal volume ≥ 80% delivered) over a range of cuff (5-25 cm H2O) and end-expiratory pressure (PEEP 0-15 cm H2O). Subglottic suction channel performance was tested in 2 ETTs (TaperGuard Evac [Covidien] and ISIS HVT [Teleflex]) as time to evacuate the simulant.

Results: All ETT cuffs provided effective seals at an inflation pressure of 12 cm H2O when PEEP was ≤ 5 cm H2O. The Microcuff ETT sealed at the lowest pressure of 6 cm H2O, whereas the Sheridan/HVT cuff sealed at 12 cm H2O (P = .01). With a PEEP of 15 cm H2O, a reciprocal increase in air leak occurred, requiring a cuff inflation up to 22 cm H2O to maintain a return tidal volume at ≥ 80% delivered. Suction channel performance improved in the lateral position compared with supine for both ETT designs during continuous 15 mm Hg suction pressure (P = .001).

Conclusions: Within a novel model with normal trachea compliance, we found all ETT designs tested to seal at lower than current recommended cuff pressures.

Keywords: compliance; endotracheal tube; insufflation pressure; mucus; suctioning; ventilator-associated pneumonia.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Equipment Design
  • Humans
  • Intubation, Intratracheal / instrumentation*
  • Models, Anatomic
  • Positive-Pressure Respiration
  • Pressure
  • Suction / instrumentation
  • Tidal Volume
  • Trachea* / physiology