Delivery of Inhaled Nitric Oxide During MRI to Ventilated Neonates and Infants

Respir Care. 2021 Aug;66(8):1254-1262. doi: 10.4187/respcare.08408. Epub 2021 May 18.

Abstract

Background: Many pediatric and neonatal ICU patients receive nitric oxide (NO), with some also requiring magnetic resonance imaging (MRI) scans. MRI-compatible NO delivery devices are not always available. We describe and bench test a method of delivering NO during MRI using standard equipment in which a NO delivery device was positioned in the MRI control room with the NO blender component connected to oxygen and set to 80 ppm and delivering flow via 12 m of tubing to a MRI-compatible ventilator, set up inside the MRI scanner magnet room.

Methods: For our bench test, the ventilator was set up normally and connected to an infant test lung to simulate several patients of differing weight (ie, 4 kg, 10 kg, 20 kg). The NO blender delivered flows of 2-10 L/min to the ventilator to achieve a range of NO and oxygen concentrations monitored via extended tubing. The measured values were compared to calculated values.

Results: A range of NO concentrations (12-41 ppm) and FIO2 values (0.67-0.97) were achieved during the bench testing. The additional flow increased delivered peak inspiratory pressure and PEEP by 1-5 cm H2O. Calculated values were within acceptable ranges and were used to create a lookup table.

Conclusions: In clinical use, this system can safely generate a range of NO flows of 15-42 ppm with an accompanying FIO2 range of 0.34-0.98.

Keywords: critical care; infant; magnetic resonance imaging; mechanical ventilators; newborn child; nitric oxide.

MeSH terms

  • Child
  • Humans
  • Infant
  • Infant, Newborn
  • Magnetic Resonance Imaging
  • Nitric Oxide*
  • Oxygen
  • Ventilators, Mechanical*

Substances

  • Nitric Oxide
  • Oxygen