Abstract
With current ventilator triggering design, in initiating ventilator breaths patient effort is only a small fraction of the total effort expended to overcome the inspiratory load. Similarly, advances in ventilator pressure or flow delivery and inspiratory flow termination improve patient effort or inspiratory muscle work during mechanical ventilation. Yet refinements in ventilator design do not necessarily allow optimal patient-ventilator interactions, as the clinician is key in managing patient factors and selecting appropriate ventilator factors to maintain patient-ventilator synchrony. In patient-ventilator interactions, unmatched patient flow demand by ventilator flow delivery results in flow asynchrony, whereas mismatches between mechanical inspiratory time (mechanical TI) and neural TI produce timing asynchrony. Wasted efforts are an example of timing asynchrony. In the triggering phase, trigger thresholds that are set too high or the type of triggering methods induces wasted efforts. Wasted efforts can be aggravated by respiratory muscle weakness or other conditions that reduce respiratory drive. In the post-triggering phase, ventilator factors play an important role in patient-ventilator interaction; this role includes the assistance level, set inspiratory flow rate, TI, pressurization rate, and cycling-off threshold, and to some extent, applied PEEP. This paper proposes an algorithm that clinicians can use to adjust ventilator settings with the goal to eliminate or reduce patients' wasted efforts.
- mechanical ventilation
- positive-pressure ventilation
- pressure-control ventilation
- patient-ventilator interaction
- ventilator triggering
- work of breathing
- time constant
Introduction
Nearly a quarter of a century ago, flow triggering was incorporated in intensive care unit (ICU) microprocessor-based ventilators with the goal to reduce patient triggering effort.1,2 Ventilator triggering designs have evolved such that the proportion of patient effort required to trigger the ventilator is only a small fraction of the total effort expended to overcome inspiratory muscle load.3 Recently, another ventilator triggering method, called shape-signal triggering, was introduced to further minimize patient-triggered effort.3 However, despite refinements in ventilator triggering design, any mismatch between the patient's triggering of the ventilator and the clinician-set ventilator settings will induce patient-ventilator asynchrony that remains problematic in ventilator management.4 Significant patient-ventilator asynchrony (using a threshold value of 10% or greater of breaths being asynchronous) is relatively common, occurring in approximately 25% of patients receiving conventional patient-triggered mechanical ventilation, and is associated with prolonged duration of mechanical ventilation.4,5 Of note, some degree of patient-ventilator asynchrony is ubiquitous.
Important determinants of patient-ventilator asynchrony include both patient and ventilator factors; however, the clinician's input cannot be overemphasized (Fig. 1). The clinician can either improve or worsen the interactions. For instance, reducing the pressure-support level when it has been set too high6 or administering a high sedation level7 can improve or worsen patient-ventilator asynchrony, respectively. Thus, the clinician is key in optimizing patient-ventilator interactions, and must understand the underlying pathophysiology or determinants and types of patient-ventilator asynchrony. This fundamental knowledge is important to take steps in improving patient-ventilator interaction. A patient-ventilator interaction without clinician input is essentially an automation (eg, SmartCare, Dräger Medical, Lübeck, Germany), a pressure-support mode that uses artificial intelligence for weaning.8,9
This paper will review the influence of ventilator factors on patient effort or inspiratory muscle work in the triggering and post-triggering phase, determinants of patient-ventilator interaction, and the impact of controlling ventilator factors on patient-ventilator interaction, and present a proposed algorithm to improve patient-ventilator interaction in patients receiving conventional mechanical ventilation.
Influence of Ventilator Factors on Patient Effort or Work of Breathing
The triggering phase—that is, from onset of patient effort to onset of flow delivery—comprises the trigger variable of a mechanical breath. The post-triggering phase—that is, from onset of flow delivery to termination of inspiratory flow—constitutes the variables controlling pressure or flow delivery and the cycling-off variable, also known as the expiratory triggering variable (Fig. 2).
Triggering Phase
Trigger Variable.
New designs have markedly improved ventilator performance. In bench testing of ICU ventilators released in 2000 or afterwards, in which the trigger variable was set to sufficient sensitivity but without auto-triggering, all but 3 ventilators tested had a triggering time delay of less than 60 ms (Fig. 3).10 In contrast, among those released prior to 2000, only 2 ventilators had a triggering time delay of less than 60 ms.11 Triggering time delay is the time from onset of triggering effort to onset of flow delivery. Patient effort to trigger, expressed as pressure-time product of the inspiratory muscles during triggering (PTPtrig), is relatively constant with increasing ventilatory support.3,12 A constant PTPtrig is caused by a decrease in the pressure-development rate when the subject triggers the ventilator, and a simultaneous increase in the triggering time delay when the ventilator support is increased (Table 1). In response to 3 levels of pressure-support ventilation (PSV)—a baseline of 18 cm H2O, low, and high levels (5 cm H2O less or greater than baseline, respectively)—PTPtrig was small: 5.2%, 4.2%, and 6.3%, respectively, of the total inspiratory effort or total PTP of the inspiratory muscles (PTPtot)3 (see Table 1).
Post-Triggering Phase
Variable Controlling Pressure or Flow Delivery.
For most ICU ventilators released since 2000, the performance of the pressurization rate for flow delivery in the PSV mode has been adequate, with only a few exceptions (eg, Avea, Viasys, Yorba Linda, California; and Centiva, GE Healthcare, Madison, Wisconsin).10 However, increasing the pressure-support level improves performance.10
With PSV, setting the pressurization rate allows alteration of the initial pressure ramp and the initial peak flow rate. Increasing the pressurization rate is associated with a reduced time to reach the pressure-support level and with an increased peak flow rate. The effects of the pressurization rate on inspiratory muscle work, expressed as work rate (Joules/min) and the sensation of dyspnea were studied in patients with acute lung injury, employing a 5 or 15 cm H2O pressure-support level.13 The pressurization rate—an arbitrary dimensionless scale of −9 (lowest) to +9 (highest)—was varied at 5 levels: a baseline or default level and 2 below and 2 above it. Overall work rate was lower with a high PS level. With either a low or a high PS level, the highest level of work rate was achieved at the lowest pressurization rate (Fig. 4).13 Compared with the baseline pressurization rate, the highest setting did not further reduce work rate.
The effect of pressurization rate on the perception of dyspnea, measured using the Borg perceived dyspnea scale, ranged from 0 to 10 cm, with 0 cm as no shortness of breath and 10 cm as extremely severe shortness of breath. As shown in Figure 4, the perception of dyspnea followed a U-shape pattern, with the worst dyspnea at the lowest and highest pressurization rates.13 At the lowest pressurization rate the sensation of dyspnea was perceived as “air hunger,” with the afferent signal arising from the chest. At the highest pressurization rate, the sensation of dyspnea was perceived as “airway discomfort,” with the afferent signal originating from the upper airways.14 Thus, in acute lung injury, increasing the pressurization rate or delivered flow rate above what is sufficient to meet patient flow demand does not provide further benefit in unloading inspiratory muscle work, and in fact, induces breathing discomfort.
Cycling-Off Variable.
Effects of varying the cycling-off variable on inspiratory muscle work are determined by the underlying lung disease or expiratory time constant. In patients with acute lung injury, in whom the expiratory time constant is short, when both pressurization rates and cycling-off thresholds were varied, for a given pressurization rate, cycling off the ventilator at the lowest or highest percentage of peak flow rate had no effect on work rate (Joules/min, Fig. 5A).15 Pressurization rate was the only determinant of work rate. Unloading of inspiratory muscle work was achieved with a high pressurization rate, irrespective of the PS level (see Fig. 5A). Conversely, in patients with COPD, in whom the expiratory time constant is prolonged, the cycling-off threshold has a significant impact on work rate.16 When 5 and 15 cm H2O PSV were applied, with and without PEEP, with the cycling-off threshold set at 5% or 40% of the peak flow rate, work rate was significantly reduced at a high cycling-off threshold (see Fig. 5B). The high cycling-off threshold results in early breath-termination, compared with that of the low cycling-off threshold, which results in decreased mechanical TI and tidal volume (VT), and, consequently, decreased intrinsic PEEP (PEEPi) and triggering time delay.16 At a PS of 15 cm H2O, varying the cycling-off threshold did not have any impact on work rate. A high PS level unloads inspiratory muscle work significantly, such that the high cycling-off threshold has no further impact on work (see Fig. 5B).
Ventilator Factors as Determinants of Patient-Ventilator Interaction
Figure 6 shows a simplified concept of patient-ventilator interaction during patient-triggered ventilation. Patient factors consist of patient neural intensity per breath and frequency, whereas ventilator factors consist of the ventilator driving pressure and frequency.17 Both neural and ventilator frequencies comprise inspiratory timing (neural TI and mechanical TI) and expiratory timing (neural TE and mechanical TE) components. Patient neural intensity per breath is translated into pressure generated by the respiratory muscles (Pmus). The ventilator provides the driving pressure (Paw) to deliver gas flow. Both Pmus and Paw make up the pressure applied to overcome the elastance and resistance of the respiratory system, including PEEPi (see Fig. 6). When ventilator-driving pressure is inadequate, flow delivery cannot meet patient flow demand, and flow asynchrony ensues.
In patient-ventilator interaction, feedback mechanisms may have little influence on the patient's neural (respiratory) controller. For instance, during patient-triggered volume-cycled ventilation (ie, volume-controlled continuous mandatory ventilation) if one doubles the ventilator VT and increases inspiratory flow rate to maintain mechanical TI constant, respiratory frequency decreases little, approximately 12%.18,19 Similarly, in a study with proportional assist ventilation (PAV) in which VT increased in response to CO2 challenge, Pmus decreased only slightly (14%).20 These observations demonstrate that neuromechanical inhibition, via increasing VT on respiratory rate and Pmus, is weak. As VT increases, the small decrease in respiratory frequency results in shortened TE, dynamic hyperinflation, a delay in ventilator triggering, and ultimately, ineffective triggering or wasted inspiratory effort, a form of timing or phase asynchrony.
Timing asynchrony occurs when the patient's and ventilator's timing components are mismatched. Timing asynchrony may exist in the form of:
Ineffective triggering or wasted effort, which can occur during either inhalation or exhalation
Auto-triggering, which happens when a ventilator breath occurs without patient effort, and is associated with low respiratory drive, prolonged exhalation time in the absence of PEEPi, cardiogenic oscillation, hiccup, low triggering threshold, water in the circuit, or circuit leak
Double-triggering, which happens when 2 consecutive patient-triggered breath cycles occur with an interval of less than one-half of the mean TI, and is associated with short mechanical TI relative to patient neural TI
Short cycling, which occurs when the ventilator breath cycle ceases abruptly, and is associated with high respiratory drive
Prolonged cycling, which occurs when the ventilator breath cycle is longer than the patient TI (mechanical TI > neural TI)
The most common form of timing or phase asynchrony is ineffective triggering or wasted effort(s)5; henceforth, the terms ineffective triggering and ineffective or wasted effort(s), will be used interchangeably. Contributing ventilator factors that are responsible for ineffective triggering can be analyzed within the triggering and post-triggering phases.
Triggering Phase
Trigger Threshold.
In the triggering phase, ineffective triggering is associated with an insensitively set trigger threshold and with the ventilator triggering method. An insensitively set trigger threshold induces wasted effort, which can be aggravated by respiratory muscle weakness, blunted respiratory drive,12 or conditions that suppress respiratory drive (eg, metabolic or respiratory alkalemia, heavy sedation), or dynamic hyperinflation.21
Triggering Methods.
Currently, patient-triggered ventilation can occur by 4 methods: pressure triggering, flow triggering, volume triggering, or shape-signal triggering. Replacing pressure triggering with flow triggering does not decrease wasted effort.22 Recently, Prinianakis and co-workers3 showed that shape-signal triggering results in less wasted effort than flow triggering does. However, because the shape-signal method has highly sensitive triggering, auto-triggering occurs more frequently than with flow triggering. Shape-signal triggering is available on the Vision ventilator (Philips Respironics, Andover, Massachusetts),3 which combines the volume-signal and shape-signal methods, whichever occurs first. Shape-signal triggering is activated when the ventilator generates a new shape-signal flow (that is, by offsetting the signal from the actual flow by 0.25 L/s and delaying it for 300 ms), then crosses the sudden decrease in actual expiratory flow as the patient generates an inspiratory effort (Fig. 7). Likewise, the shape-signal flow can be used to cycle off inspiration. If the shape-signal does not cross the patient's actual expiratory flow—for instance, due to a delay in patient effort—volume-triggering is activated once 6 mL of volume is accumulated above the baseline flow.3 The role of volume-triggering in decreasing wasted effort, in comparison with that of pressure or flow triggering, has not been evaluated.
Post-Triggering Phase
In the post-triggering phase, ventilator factors contributing to wasted effort include variables that influence the matching of mechanical TI and neural TI. These include the level of ventilatory assistance, the cycling-off variable in PSV, the inspiratory flow rate, and to a certain extent the applied PEEP.
Level of Ventilatory Assistance.
Leung et al12 demonstrated that with PSV the number of wasted efforts increases in proportion to the level of ventilatory assistance. Analysis of the VT revealed a higher VT in the breaths prior to the wasted efforts than in those of the triggered breaths. Furthermore, the breaths prior to the wasted efforts had an abbreviated mechanical TE and higher PEEPi than those of the triggered breaths. In chronically ventilator-dependent patients, decreasing the pressure-support level to an average of 11 cm H2O uniformly eliminated wasted efforts.22 The reduced PS level was associated with reduced VT and increased Pmus, but at the expense of the development of a rapid shallow breathing pattern and respiratory distress.22 Recently, Thille et al6 demonstrated that in patients with weaning difficulty, while maintaining the applied PEEP constant, decreasing the PS from baseline (average of 20 cm H2O) to optimum (defined as the level of PS that eliminates wasted efforts but does not induce respiratory intolerance) improved the asynchrony index from a median of 45% to 0% (Fig. 8A). The asynchrony index is the ratio of wasted efforts to the total number of wasted efforts plus the triggered mechanical breaths collected for 2 min,22 and expressed as a percentage.4 Wasted efforts were completely eliminated in two thirds of the patients. The optimum PS level (average of 13 cm H2O) is associated with a mean VT of 6.0 mL/kg, versus 10.2 mL/kg at baseline PS, a mechanical TI of 0.8 s versus 1.3 s, and a dynamic PEEPi of 1.1 cm H2O versus 1.8 cm H2O, respectively.
Cycling-Off Threshold.
To decrease wasted efforts, Thille et al6 also employed a different approach: by adjusting the cycling-off variable from a baseline of 25% of peak flow to a maximum value of 45%, mechanical TI was reduced until an optimal mechanical TI was achieved. If wasted efforts persisted, the insufflation time was further decreased by steps of 0.2 s until wasted efforts were eliminated or respiratory intolerance ensued. At optimal mechanical TI, the asynchrony index decreased by 38% (see Fig. 8B). VT decreased to an average of 7 mL/kg, mechanical TI decreased to 0.8 s, and dynamic PEEPi decreased to 1.5 cm H2O. Thus, decreasing the PS level or reducing mechanical TI by adjusting the cycling-off threshold is an effective method for reducing wasted efforts, and is associated with reduced VT, mechanical TI, and PEEPi, and therefore with a better match between patient and ventilator TI.
In selecting the cycling-off threshold it is important to take into account the respiratory system's time constant—that is, the time required to empty approximately two thirds of the expired volume—the product of resistance and compliance. In patients with COPD, the time constant is prolonged, and setting the cycling-off threshold high (as a percentage of the peak flow rate), improves matching of mechanical TI and neural TI, and reduces wasted efforts (Fig. 9A).23,24 VT, neural TI, triggering time delay, and PEEPi decrease significantly with the cycling-off threshold set high, as compared with low.23 Conversely, in patients with acute lung injury, the time constant is short; a high cycling-off threshold results in premature breath-termination and causes re-triggering (double-triggering) or premature opening of the exhalation valve (see Fig. 9B).24
Inspiratory Flow Rate.
With patient-triggered volume-control ventilation, a low set inspiratory flow rate that is less than patient ventilatory demand induces not only flow asynchrony (see Fig. 6) but also mismatching between mechanical TI and neural TI. Mechanical TI is prolonged relative to neural TI, encroaching into patient TE such that the next breath occurs at a high lung volume, above its resting end-expiratory volume and resulting in ineffective triggered breaths. This type of wasted effort can be corrected by increasing the inspiratory flow rate, or, when pressure-control time-cycled ventilation is employed, by shortening the TI.25
Applied PEEP.
Thille et al6 did not find PEEP application useful in reducing wasted efforts, because PEEPi was less than the applied PEEP. Their findings6 contrasted with those of Nava et al,21 in which wasted efforts decreased significantly, from a mean ± standard error of 19 ± 5.7% without PEEP to 4 ± 2.5% when a PEEP equivalent to 75% of the measured static PEEPi was applied. PEEPi ranged from 3 to 20 cm H2O. Applying PEEP to improve wasted efforts will have to be customized to the degree of static PEEPi. However, measurement of PEEPi is frequently not feasible without the administration of heavy sedation or paralytics.
Ventilator Modes for Improving Patient-Ventilator Interaction
Currently, only 2 ventilator modes—PAV26 and neurally adjusted ventilatory assist (NAVA)27—are capable of decreasing patient-ventilator asynchrony, and will be mentioned only briefly in this review. The trigger variable differs between these 2 modes. PAV employs pneumatic triggering, with either pressure or flow to initiate the breath. In contrast, triggering with NAVA employs diaphragmatic electrical activity, a signal near proximal to the respiratory controller, and, therefore, avoids triggering delay, as occurs with dynamic hyperinflation or circuit leaks.28 Diaphragmatic electrical activity signal is used not only to initiate but also to assist and cycle off inspiration. Triggering of the ventilator occurs when the diaphragmatic electrical signal, above a manually set trigger threshold, is detected. Setting the trigger threshold above the baseline diaphragmatic electrical activity noise level avoids auto-triggering.27,29
The efficacy of both PAV and NAVA in improving patient-ventilator synchrony was compared with PSV.26,27 In patients receiving PAV or PSV for 48 h, wasted efforts with an asynchrony index of greater than 10%, were observed in 1.3% (10/744 measurements) of patients receiving PAV, compared with 9.6% (67/696 measurements) of patients receiving PSV. In a short-term trial with 14 patients with 2 levels of NAVA and PSV, in which airway pressure with NAVA was matched to that with PSV (an average of 10.6 cm H2O and 17.4 cm H2O for low and high support levels, respectively), wasted efforts were observed in 6 of 14 patients. In the patients with wasted efforts, the mean ± SD wasted efforts with low and high PSV were 5 ± 4% and 31 ± 26%, respectively, while complete patient-ventilator synchrony was achieved with NAVA.
Proposed Algorithm to Improve Patient-Ventilator Synchrony
Figure 10 shows a proposed algorithm that can be applied at the bedside to decrease or eliminate wasted efforts, based on the studies discussed above. Commonly, a low level of PEEP (5 cm H2O) is applied at the onset of mechanical ventilation. In this case, PEEP can be maintained while taking the steps shown in the algorithm. When one can measure static PEEPi—recognizing that in patient-triggered breaths this may not be feasible—for a PEEPi greater than 5 cm H2O, apply PEEP of 75–80% of static PEEPi.21 For a PEEPi of less than 5 cm H2O, or immeasurable, maintain 5 cm H2O PEEP. If wasted efforts with asynchrony index of greater than 10% persist, one can increase applied PEEP by 1 cm H2O increments to a maximum of 8 cm H2O. Following this maneuver, if wasted efforts persist, the assist level, whether in the form of pressure or volume assist, is adjusted to achieve a VT of 6–8 mL/kg.6 Afterwards, one can increase the set inspiratory flow rate with volume-controlled continuous mandatory ventilation25 or increase the pressurization rate with pressure-control ventilation.15 In addition, with time-cycled pressure-control ventilation, TI can be decreased while maintaining patient comfort. With PSV, one can adjust the cycling-off threshold, taking into account the respiratory system's time constant.23,24 In patients with a prolonged time constant (ie, COPD), the flow cycling-off threshold is adjusted upward as a percentage of peak flow rate,23 while in patients with a short time constant (ie, acute lung injury or acute respiratory distress syndrome) the cycling-off threshold is adjusted downward.24 With a high-set cycling-off threshold, mechanical TI and VT may decrease. It may be necessary to fine-tune the cycling-off threshold or adjust mechanical TI and VT to avoid respiratory distress. The patient needs to be reassessed and the step may be halted at any level in the proposed algorithm when the asynchrony index has successfully been reduced to less than 10%. However, this algorithm remains to be prospectively tested for its efficacy.
Discussion
Kallet:
A lot of the asynchrony I see in ALI [acute lung injury] patients stems from having an inspiratory time that's too short. What I find is that when someone's managed with the ARDS [acute respiratory distress syndrome] Network low-tidal-volume protocol,1 clinicians jack up the inspiratory flow to 60 or 70 L/min to make the patient less dyspneic. Sometimes, with low tidal volumes we're cutting the inspiratory time down to 0.5 or 0.6 seconds. It end ups that the patient's neural inspiratory time is longer than the ventilator inspiratory time, and they start double-triggering.
I think some of the asynchrony we see in ALI is inescapable and you need to make a decision that is least harmful. For me, I'd rather keep a high inspiratory flow rate and extend the inspiratory time a bit by creating more of a pause time. Unfortunately, patients often continue to inspire during the pause phase, but it cuts down on their ability to double-trigger breaths and get a big tidal volume. The down side is that work of breathing is worse, and as we've found to our dismay, it can exacerbate pulmonary edema.2
I think the conundrum we face with asynchrony in ALI patients is that we don't have very good options to choose from: either we heavily sedate them, liberalize their tidal volume, or continue to try to restrain their tidal volume. That means that we try the strategies I just mentioned, or we reduce inspiratory flow rate to try to match their inspiratory time, which also increases work of breathing.3
Sassoon:
I agree that it depends on the patient's expiratory time constant, short or prolonged time constant: the latter in patients with COPD. In patients with short time constant, as in ARDS, to set the inspiratory time too short may not be a good idea.
Branson:
It strikes me again how important the patient is. In Scott's [Epstein] talk we were talking about the asynchrony index in COPD versus ARDS after trauma or surgery, and I think the presence of dynamic hyperinflation or auto-PEEP is what's really key to the asynchrony index. What's good in the pressure-support setting for the patient with COPD, in terms of both triggering and cycling, is not good for the patient with ALI. It goes back to the education issue that Bob [Kacmarek] raised: we can't treat everybody the same. People have to understand the underlying physiology and set the ventilator appropriately.
Kacmarek:
Well, I took it differently: it's the clinician that's important. I really liked your first slide where you added the clinician into the patient-ventilator diagram. Although the patients all respond differently, it's the clinician's recognition of the differences in those patients that truly makes the difference in how well the patient is managed.
I want to go back to Sai's [Parthasarathy] comments about the need for automated approaches to ventilatory support, particularly in patients breathing spontaneously. We could sit here and design the types of responses that were necessary, and by the time this symposium was done create the appropriate algorithms to manage, for example, pressure support. No offense to our co-workers, but all of us have been in many institutions, and the fact of the matter is that people don't pay attention to these subtleties. The majority of patients are ventilated by a standard approach, and you ask staff to look at a waveform and try to analyze it, staff have no idea of what they're looking at. So I am not sure that we can achieve the needed outcome by education. I believe it will take a better approach to closed-loop control to improve synchrony.
Parthasarathy:
It goes back to the education team that Dave [Pierson] was talking about: it's such a vital issue. I have fellows come to me and groan because I ask them what the compliance is today, what's the resistance today? They say, you're the only guy who asks that! And secretly I'm thinking, what the heck are they doing? I really think you have to subscribe to it and do it, or if you don't do it, you should just focus on the VAP [ventilator-associated pneumonia] in that patient, and I've got all these nurses chasing me as to whether I'm washing my hands and all of that.
My question is about double-triggering. There was one patient with ARDS on the ventilator and set to receive low tidal volume at 6 mL/kg ideal body weight, but the patient was double-triggering nearly every other breath. Does that really mean that the patient is receiving the low tidal volume set on the machine, when they're double-triggering? If they're set at 400 mL, are they not actually getting much more, such as 800 mL?
Sassoon:
I think the flow rate setting and the increase in respiratory drive have to be considered as well. In double-triggering, most of the time the neural inspiratory timing is greater than the mechanical inspiratory timing, so you have to adjust.
Parthasarathy:
That's the down side. If someone wasn't paying attention to double-triggering, you could actually change your patient outcome considerably.
Sassoon:
Yes. I think the most important thing when we return home is to try to establish better education for our RTs [respiratory therapists]. Even at my institution we have a Puritan Bennett 840 with PAV [proportional assist ventilation], and not a single RT knows how to use it, so that's a problem.
Younes:*
I'd like to add support to what Bob and Sai are saying. I'm the oldest one here by far, and I spent the first 20 years of my career trying to teach the fellows and RTs and ICU doctors how to think about mechanics and interactions, and after 20 years I gave up and I decided it has to be automated, because it's too complicated and everybody is so busy. You have to spend a couple of years just to figure it out.
Sassoon:
The down side of automation is that you still have to have the knowledge on what to do if something goes wrong.
Younes:
The problem with automation is how sophisticated it is. You can have all kinds of algorithms.
Pierson:
Every time I hear the analogy of the airline checks brought up and directed into our arena it bothers me. Is the critically ill patient or the home-care patient with a sleep disorder just a more complex system that takes us more time to figure out how to fully automate and implement a Toyota system or Boeing checklist, or is there something fundamentally different about the clinical setting that's not going to allow us to do this? Is it just a matter of logistics, or is there something fundamentally different about the system that we have?
Hurford:
The thing is that an airplane is defined, and an airplane is standard, and it's always an airplane. It doesn't turn into a satellite, or a helicopter, or a U-boat, whereas a patient may start out with COPD, then get pneumonia, then get bronchospasm: they change. We need algorithms that define diagnosis: not necessarily therapy. We aren't lumping here. We're not talking about an asthmatic versus someone with COPD versus a 500-pound guy who fell down the steps. Those patients are going to have very different patterns on the ventilator. I think it makes it much more difficult than to come up with a remote flight algorithm for a drone.
Branson:
I'd take it one step further, having flown a plane myself. If I check the plane and find the radio doesn't work and there's only a quarter tank of gas, I don't fly it. We don't have that luxury with a patient. The patient can't communicate, but we still have to deal with them. If the patient doesn't have enough fluid, from blood loss or whatever, we still have to ventilate them. I think it's disingenuous. I think we can learn from the airline industry as to preventing errors, but I think we have nothing to learn from the airline industry in how we deal with a patient as compared to an airplane.
Catherine, you didn't mention Du,1 who came up with an automated system for setting the flow cycle of pressure support. And the Newport ventilators actually do automated rise time and automated flow cycling. Is the issue there that that system hasn't been tested sufficiently, or is the issue that Newport has such a small market that it doesn't care if nobody knows if it works?
Sassoon:
Probably the Newport has not been sufficiently tested because of its small market. It would be interesting to see how the Newport compares to ventilators with manually adjustable settings.
Branson:
So the system essentially uses compliance and resistance to create the expiratory time constant, and changes the cycling percentage based on the expiratory time constant, and also pressure rise at the end of inspiration. Outside of their original paper I don't know if anybody has really looked at it seriously.
Kacmarek:
We have a paper that we're about submit in which we looked at default ventilator settings for triggering, rise time, and cycling criteria compared to what we could set optimally to eliminate these issues of asynchrony. Our manual settings were clearly better.
But let me go back to the airplane versus patient. Bill [Hurford] commented about how all patients are different, but all those differences translate to differences in measurable variables. I don't think the etiology of the problem has a big impact; what's important is the impact the problem has on cardiovascular-pulmonary interaction. Those interactions are programmable! It seems to me that if you can fly a plane with algorithms, we should be able to manage synchrony in most patients. If we are willing to put the time and money into developing the correct algorithms, we can better manage synchrony. I don't think it eliminates the need for RTs or physicians at the bedside; I think it intensifies the need. As Catherine indicated, they have to be smarter and better able to look at trends and how the particular algorithm is applied.
The automated pilot on an airplane has never eliminated the need for an actual pilot, and the same is true when we talk about closed-loop control of mechanical ventilation. I think closed-loop control is critical, because I, like Magdy, have been trying to teach the same thing over and over for 25 years. How many times have we given presentations about asynchrony—and the slides we're using today aren't that different from the slides we used 25 years ago. We just add more studies that show the same things that we knew 25 years ago about asynchrony. Everything we're talking about here, from trigger asynchrony to auto-PEEP, John Marini and others published back in the 1980s.
Parthasarathy:
I want to point out that the 100,000 Lives Saved Campaign from the healthcare system is derived from the airline industry. It was the fact that we found it appalling that we could kill 100,000 patients and that was not on the radar, whereas the airline industry kills a few hundred people and it's all over the news. They were, like, What is wrong with this picture? The airline industry already helped the healthcare industry, with the 100,000 Lives Saved Campaign.
Kallet:
To try to answer Dave's [Pierson] question, I think there's the cultural component in all of this. I think there's been a certain amount of arrogance in the medical field, where physicians or RTs are stubborn about how they deliver patient care, regardless of the evidence. I don't think society is going to accept our attitudes much longer. In listening closely to the recent healthcare debate, I was struck by how often major politicians cited the cost-cutting benefits of evidence-based medicine. In the coming years, clinicians are not going to be allowed to just ignore therapies that clearly improve outcomes. Whether some of these issues can be solved by closed-loop control of ventilation, or just some type of financial punishment for not implementing care that's been shown to improve outcomes, change is coming!
Branson:
The difference between airline process improvement and making sure the patients gets gastrointestinal prophylaxis, DVT [deep venous thrombosis] prophylaxis, making sure the bed is up—those things are the things that the airlines do right. But comparing the patient/ventilator to a plane/pilot is where I think the whole thing falls apart.
Epstein:
Even when you demonstrate improvement in outcomes, practitioners are slow to adopt the changes. Sai [Parthasarathy], you said that when you compared yourself to some of your colleagues you hoped you were doing a better job, but you have no data to show that you are. Until we show this is actually clinically important, it's going to be very difficult to get people to change. I would suggest we need to do that first, before we spend millions of dollars to change the process.
Sassoon:
Regarding automation, maybe, as we were saying yesterday, nobody knows what the static compliance is.
Parthasarathy:
I think that's always the down side, and we see surgeons lamenting about how my friend is now a unique surgeon who does robotic surgery and the rest of the group can't, so they call upon him to do these prostate surgeries, but then they're also worried about what those guys will do if there's malfunction. So this is a philosophical question, because we need a good approach and a long-term plan: not something that we just sit around and wait for, hoping that it happens. It needs to be methodically thought about, and I think when you, Scott [Epstein], say that there are no data about how we measure these things and look at these tracings, measure the compliance, and whether what we do will have a tangible difference in outcomes. We have a situation of equipoise; we really don't know that intervention can actually help improve lives, which would allow the institutional review board to approve such a study. What else do we need to develop something? There's an opportunity there for research.
Hess:
I think that a smart system has to look at a lot more than the ventilator, because the ventilator may not be the problem. The problem may be with the patient, so a smart system that I would envision would look at things like: What is the patient's pH? What is the patient's PCO2? Maybe the problem with asynchrony is not what's set on the ventilator; maybe it's a problem with the patient and it should prompt the user that this patient has metabolic acidosis that needs to be addressed. That's not a matter of just tweaking something on the ventilator.
Pierson:
That gets back to the way I phrased my question initially. Is it just an incredibly complicated system and it's just a matter of time before we've figured out how to do the cloud computing or whatever needs to be done to figure it all out and make it work—or is there something inherently different?
Hess:
I think it's incredibly complicated and just looking at the ventilator and signals is not going to be the answer.
Younes:
I think how complicated it is depends on your target. If you want to have a comprehensive system to manage the patient, that would obviously be extremely complicated, but if you focus on a specific issue like improving synchrony, then that is a manageable problem: it's not very complicated. I'd say we should have an open mind about automation. Taking the position that technology is bad because an RT or doctor knows best is not a good idea. You point out that patients are highly variable, but if you take some COPD patients with the same numbers and ask 5 RTs or 5 doctors how to manage them, they would all treat them in different ways. Now, who's right? There's always going to be errors. It's a question of who makes the errors: an automated system or an individual? We should accept that we should try an automated system but that the decision should be based on outcomes, along with human intervention.
Footnotes
- Correspondence: Catherine S Sassoon MD, Pulmonary and Critical Care Section, Veterans Affairs Long Beach Healthcare System, 5901 East 7th Street, 11/111P, Long Beach CA 90822. E-mail: csassoon{at}uci.edu.
The author has disclosed no conflicts of interest.