Abstract
Using an electronic literature search for published articles indexed in PubMed between January 1990 and August 2011, the update of this clinical practice guideline is the result of reviewing 84 clinical trials, 54 reviews, 25 in vitro studies, and 7 evidence-based guidelines. The recommendations below are made following the Grading of Recommendations Assessment, Development, and Evaluation (GRADE) criteria: 1: It is recommended that selection of the appropriate aerosol generator and interface be made based on the patient's age, physical and cognitive ability, cost, and the availability of the prescribed drug for use with a specific device. 2: Nebulizers and pressurized metered-dose inhalers (pMDIs) with valved holding chambers are suggested for use with children ≤ 4 years of age and adults who cannot coordinate the use of pMDI or dry-powder inhaler (DPI). 3: It is suggested that administration of aerosols with DPIs be restricted to patients ≥ 4 years of age who can demonstrate sufficient flow for the specific inhaler. 4: For patients who cannot correctly use a mouthpiece, aerosol masks are suggested as the interface of choice. 5: It is suggested that blow-by not be used for aerosol administration. 6: It is suggested that aerosol therapy be administered with a relaxed and nondistressed breathing pattern. 7: Unit dose medications are suggested to reduce the risk of infection. 8: It is suggested that nebulizer/drug combinations should be used as approved by the FDA. 9: It is recommended that healthcare providers know the correct use of aerosol generators; they should teach and periodically re-teach patients about how to use aerosol devices correctly. 10: It is suggested that intermittent positive-pressure breathing should not be used for aerosol therapy. 11: It is recommended that either nebulizer or pMDI can be used for aerosol delivery during noninvasive ventilation.
- aerosol
- dry-powder inhaler
- metered-dose inhaler
- nebulizer
- patient education
- patient adherence
- noninvasive ventilation
- device selection
ADS 1.0 DESCRIPTION
Effective administration of aerosolized medications depends on the patient's age, physical and cognitive ability, the delivery system, and the patient-device interface.1–7 Physical ability means the patient's ability to use a specific device, based on factors such as inspiratory volumes and flows, hand-breath coordination, or ability to use a mouthpiece. Cognitive ability indicates the patient's understanding of how and when to use a device and medication.8–10 Airway size, respiratory rate, inspiratory flow rate, and breathing pattern create substantial challenges for effective aerosol delivery.4,5,8–12 While most aerosol generators can be used with all age groups, special consideration should be given to young children because they cannot master the complex steps required for adequate delivery of aerosol treatments.8,9,13 A mouthpiece may be used for patients ≥ 3 years who are able to cooperate,8,10,14 while a face mask is recommended for patients who cannot use a mouthpiece.2,3,8,9,14–17 Face masks should be properly fitted with minimal leak, particularly avoiding aerosol delivery into the eyes, to optimize inhaled dose.18–24 Aerosol generators are equally efficacious if they are age appropriate and used correctly.25–39 Regardless of age, patients need to demonstrate ability to seal the lips around the mouthpiece and ability to generate sufficient flow for the specific inhaler. Table 1 provides age guidelines for use of aerosol delivery device types.
This clinical practice guideline (CPG) addresses the selection of an aerosol delivery device for the administration of inhaled medications by small-volume nebulizer (SVN), large-volume nebulizer (LVN), pressurized metered-dose inhaler (pMDI), and dry-powder inhaler (DPI) to spontaneously breathing patients without an artificial airway. Studies on spontaneously breathing patients with tracheostomy were excluded from this CPG, due to lack of evidence in this area of research. Using an electronic literature search for published articles indexed in PubMed between January 1990 and August 2011, the update of this CPG is result of reviewing 84 clinical trials, 54 reviews, 25 in vitro studies, and 7 evidence-based guidelines. The search terms used in this CPG include nebulizer, dry-powder inhaler, metered-dose inhaler, aerosol, face mask, mouthpiece, hood, blow-by, valved holding chamber (VHC), spacer, patient education, patient adherence, noninvasive ventilation, and device selection.
1.1 Small-Volume Nebulizer. SVNs are divided into 3 types: jet nebulizer, mesh nebulizer, and ultrasonic nebulizer. The jet nebulizer is powered by a compressed gas source, in the form of a compressor or hospital pressurized gas source such as piped medical gas or high pressure tanks. The medication is displaced up a capillary tube from the nebulizer's reservoir and is dispersed continuously as aerosolized particles.14,40–43 There are 4 types of jet nebulizer: jet nebulizer with reservoir tube, jet nebulizer with collection bag, breath-actuated jet nebulizer, and breath-enhanced jet nebulizer. Jet nebulizer with reservoir tube is the most commonly used and generates continuous aerosol during the entire breathing cycle. Jet nebulizer with a reservoir bag collects aerosol leaving the jet nebulizer when patient is not inhaling. Thus, it increases dose efficiency by delivering aerosol in the reservoir bag with the next inspiration. While the breath-actuated jet nebulizer generates aerosol only during inspiration, which reduces loss of medication during expiration, the breath-enhanced jet nebulizer uses one-way valves to prevent the loss of aerosol to the environment and to increase aerosol delivery to the patient. Mesh nebulizer is powered by electricity to create aerosol by way of an aperture plate or an ultrasonic horn. The diameter of the mesh or aperture determines the size of the particle generated.14,44–46 Ultrasonic nebulizer is also powered by electricity to create high frequency vibrations in a piezo crystal, which are transferred to the surface of the solution, creating a standing wave that generates aerosol.14
The aerosolized particles are inhaled by the patient or delivered in conjunction with positive-pressure breaths such as intermittent positive-pressure breathing, noninvasive ventilation (NIV), and intrapulmonary percussive ventilation. Whereas aerosol delivery via SVN with intermittent positive-pressure breathing is less efficient than SVN alone,47 using NIV for aerosol therapy is feasible and effective in improving bronchospasm.48–52 However, it should be noted that aerosol delivery with NIV is influenced by the NIV settings, the leak port position, the type of aerosol generator, and the interface used during the treatment.53–55 For instance, increasing the inspiratory pressure during NIV increases aerosol deposition, as opposed to increased expiratory pressure, which decreases aerosol delivery.53 The efficiency of an aerosol generator is influenced by the location of the leak port. If the leak port is distal to the device, drug delivery is greater than at the proximal location, regardless of the type of aerosol generator used.54,55 Consequently, drug delivery via mask with a leak port is less than that without. If the leak port is in the mask, the efficiency of pMDI is better than that of nebulizer because aerosol loss with pMDI during expiration is less, compared to the nebulizer. According to a recent in vitro comparison, aerosol deposition with mesh nebulizer is significantly greater, compared to the jet nebulizer during NIV.55 Also, the use of nebulization with NIV is better than nebulization alone in some patients.52 The intrapulmonary percussive ventilator is primarily an airway clearance device, but it is usually combined with a nebulizer, although its physiologic and clinical effects have not been studied extensively. Studies have shown that aerosol deposition with the intrapulmonary percussive ventilator is significantly lower than with a standard nebulizer alone and has large inter-individual variability.56,57
1.2 Large-Volume Nebulizer. LVNs with a fill volume > 10 mL, powered by a compressed gas source, are utilized to administer continuous aerosol delivery over a prolonged period of time. Due to changes in drug concentration over time, LVNs should be emptied and refilled at 5 hours.58 An alternative approach for continuous nebulization is small volume jet or mesh nebulizer with infusion pump. Regardless of the type of nebulizer utilized, a face mask is typically used as the interface for continuous nebulization. Previous studies reported that continuous inhaled bronchodilator administration is safe, effective, and less time consuming, when compared with intermittent nebulization in patients with severe asthma.59–61 Continuous bronchodilator delivery is the most common aerosol treatment for continuous nebulization.62,63
1.3 Pressurized Metered-Dose Inhaler. There are 2 types of pMDI: conventional and breath-actuated. Both include a canister, propellants, drug formulary, metering valve, and actuator. Actuation of the pMDI results in the ejection of one dose of aerosolized medication.64 The conventional pMDI has a press-and-inhale design, whereas the breath-actuated pMDI eliminates the need for hand-breath coordination. A simple spacer attached to the pMDI enhances aerosol delivery by decreasing the velocity of the particles and the number of large particles, thus reducing oropharyngeal deposition.65,66 A spacer without valves requires coordination between inhalation and actuation and is not suitable for patients with poor hand-breath coordination; a VHC is more desirable in this setting. Electrostatic charge can decrease aerosol delivery unless it is washed with soap and water, or, alternatively, a non-electrostatic device can be used without the need for pretreatment.66–69 Use of breath-actuated pMDI in children ≥ 5 years old may result in better asthma control, less hospitalization, and less use of relief medication prescription.70
1.4 Dry-Powder Inhaler. A DPI is a breath-actuated device that requires sufficient inspiratory flow to inhale the medication from the device. The patient's inspiratory flow disperses the dry particles and draws them from the device into the lower airways. Patients unable to demonstrate sufficient inspiratory flow for the inhaler, particularly those in respiratory distress or children < 4 years of age, may not reliably use DPIs.3,11,12,14,26,71,72 At 4 years of age and above, patients may be able to use the DPI if they are shown how to use it effectively and to generate sufficient inspiratory flow required by the device.73,74 Current DPIs are available in 3 categories: single-dose, multiple single-dose, and multiple-dose. Like the pMDI, and unlike nebulizers, the DPI is always provided as a drug/device combination.14
1.5 Interfaces Used with Aerosol Generators. Mouthpieces, masks, hoods, and spacers are the most common interfaces used between the aerosol generator and the patient. Evidence is lacking for better clinical response with one or another interface (eg, mask vs mouthpiece). Selection of interface is dependent on age, ability to use a mouthpiece, and patient preference. When a nebulizer is used, a mouthpiece is preferred, but a mask can be used if the patient cannot effectively hold the mouthpiece between the lips. Also, a face mask should be avoided in the delivery of corticosteroids, due to the side effects of steroid administration to the facial skin and eyes. A VHC is preferred when a pMDI is used in a patient with poor hand-breath coordination, and it can be used with a mask in patients unable to use a mouthpiece effectively. For a VHC with mask, a larger dead volume reduces aerosol delivery to infants.75,76 The lack of mask seal leads to a substantial decrease in aerosol delivery,21,75–78 and aerosolized medications delivered with a mask may inadvertently deposit in the eyes.23,24,76,79 Aerosolized drug delivery to children < 3 years old should be through a mask or hood. For an infant, a hood is as efficient as a mask, its use results in better therapeutic index with minimal deposition at the infant's eyes, and it is preferred by parents for aerosol drug administration.80–83
ADS 2.0 PATIENT PREPARATION
2.1 Identify patient and assess the need for inhaled medication.
2.2 Describe the procedure to be performed, how it will be performed, what the patient is expected to do, and how frequently it will be performed.
2.3 Utilize age-appropriate strategies.
2.4 Clear the airways as needed.
2.5 Position the patient appropriately.
ADS 3.0 PROCEDURE
It is recommended that techniques for using aerosol delivery devices follow the procedural steps from A Guide to Aerosol Delivery Devices for Respiratory Therapists.14 In some cases, the FDA has approved a drug-device combination with specific nebulizers identified on the drug label. Table 2 shows approved nebulizers for specific drug formulations.14 The use of specified nebulizers is recommended.
ADS 4.0 FOLLOW-UP CARE
4.1 Monitor the patient for adverse response.
4.2 Assess whether or not the patient is using the device correctly.
4.3 Assess response to therapy and document findings.
ADS 5.0 SETTING
Aerosolized medications can be administered by properly trained healthcare providers in a number of settings that include (but are not limited to):
5.1 Hospital
5.2 Extended care facility
5.3 Out-patient clinic
5.4 Physician's office
5.5 Transport vehicle
5.6 Home
ADS 6.0 INDICATIONS
6.1 SVN
6.1.1 Delivery of aerosolized medications available as a solution64,84,85
6.1.2 Need to modify drug concentration or combine compatible nebulized solutions14
6.1.3 Device of choice for patients who are unable to operate, coordinate, cooperate, or perform the necessary inspiratory maneuvers required for the use of other devices (eg, infants, small children, and the elderly)
6.2 SVN With Mouthpiece
6.3 SVN With Mask
6.4 SVN With Hood
6.5 LVN
6.6 pMDI: General Indications
6.6.1 Delivery of medications that are available in pMDI form86
6.6.2 Convenience of being small and portable
6.7 Breath-actuated pMDI
6.7.1 Delivery of inhaled bronchodilator for patients with poor hand-breath coordination who can use a mouthpiece
6.8 pMDI with VHC and mouthpiece
6.8.1 Patients who are able to hold the mouthpiece during treatment
6.9 pMDI with VHC and mask
6.10 pMDI with spacer (non-valved pMDI accessory device)
6.10.1 Patients who can coordinate inspiration and actuation
6.10.2 Reduces oropharyngeal impaction, particularly with the delivery of corticosteroids88
6.11 DPI
ADS 7.0 ASSESSMENT OF OUTCOME
Appropriate selection of an aerosol generator is reflected by the following evidence:
7.1 A positive clinical outcome after aerosol therapy:
7.2 Use of proper technique in applying aerosol delivery systems:
7.3 Patient adherence with application of aerosol delivery systems:
ADS 8.0 CONTRAINDICATIONS
8.1 No contraindications exist to the administration of aerosols by inhalation. Aerosol therapy is contraindicated when there is a known hypersensitivity or history of an allergic reaction to a specific pharmacologic agent, its preservatives, and/or its excipients.97
8.2 Contraindications associated with specific medications being delivered may exist. Pharmaceutical information in the package insert should be consulted for relative contraindications.
8.3 Aerosol generators should not be used for patients with known allergies to medication preservatives and other excipients.
8.4 A pMDI or DPI should not be used for patients unable to perform the respiratory maneuver required to deliver the drug.
ADS 9.0 HAZARDS/COMPLICATIONS
9.1 Aerosol Delivery
9.1.1 When aerosol generators are contaminated with bacteria, they increase the risk of infection in patients with respiratory diseases.98–105
9.1.2 Care providers and bystanders have the risk of infection due to the inhalation of pathogens and second hand aerosols during aerosol therapy.106 If indicated, negative pressure rooms and personal protective equipment should be used.107–109
9.1.3 Workplace exposure to aerosol may increase the risk of asthma-like symptoms and cause occupational asthma.14,110–112
9.1.4 Malfunction of device and/or improper technique may result in underdosing or overdosing.41–43,96,113–116
9.1.5 Specific pharmacologic agents can produce adverse side effects such as headache, insomnia, tachycardia, tremor, and nervousness with adrenergic agents; local topical effects with anticholinergics; airway reactivity with antibiotics, hypertonic saline, inhaled corticosteroids, and bronchodilators; systemic/local effects with corticosteroids; and bad taste with mucolytics and hypertonic saline.14,117
9.1.6 Bronchospasm may be induced due to a cold and high-density aerosol administration in patients with pulmonary diseases.117–119
9.1.7 The prescription of aerosol delivery devices for use in the home can lead to misuse if the user has not been properly trained.95,96,120
9.2 SVN
9.3 LVN
9.3.1 There may be an increase in the drug concentration in the nebulizer cup when jet and ultrasonic nebulizers are used.
9.3.2 Side effects may occur at any time during continuous nebulization, and frequent assessment is required.62,87,125
9.3.3 The nebulizer may become contaminated and can be a source of infection.
9.3.4 Drug concentration increases over time, and the solution might need to be changed after 5 hours of operation.58
9.4 pMDI
9.4.1 Inappropriate patient use may result in underdosing or overdosing.7,66,95,126
9.4.2 Reaction to propellants and other additives such as coughing and wheezing may occur.64,117,126
9.4.3 Oropharyngeal impaction of corticosteroid may result in local side effects such as candidiasis.64,126
9.4.4 Immersion of the canister in water may result in valve blockage.127
9.5 DPI
ADS 10.0 LIMITATIONS OF METHOD
10.1 SVN
10.1.1 Deposition of medication into the lungs ranges from 1–15% of the dose128–133 and may vary from brand to brand of SVN, and unit to unit of the same brand.65,115,128,134
10.1.1.1 Drug delivery varies with different nebulizer types.42,65,115,134 If specified, only the nebulizer cited on the drug label should be used.
10.1.1.2 Examples of drugs that require approved nebulizers include budesonide, tobramycin, dornase alfa, pentamidine, iloprost, treprostinil, and aztreonam (see Table 2).
10.1.2 Patients with smaller tidal volumes and rapid respiratory rates, particularly neonates130,135 and dyspneic patients with shallow breathing, may inhale less of the aerosolized agent and receive less of the dose when nebulization is continuous.40,130
10.1.3 Patients with airways of smaller diameter may receive a smaller fraction of the total particles produced.10,130,136
10.1.4 Crying children receive virtually no aerosol drug to the lungs.137,138
10.1.4.1 Most of the inhaled dose deposits in the upper airways and is then swallowed.
10.1.4.2 Inhaled drugs should be given to infants when they are breathing quietly.139
10.1.5 Fill volume in the SVN affects output.134 Since filling volumes may be different, it is suggested to follow the drug label or device manufacturer's product insert to cover those devices that are not labeled for specific use with a particular medication.
10.1.6 Approximately 25–50% of the initial solution volume remains on the internal walls and reservoir of the jet nebulizer after aerosol therapy is completed.41,43,129
10.1.7 Aerosol is lost during the expiratory phase of breathing unless a breath-actuated design or a design with a reservoir bag is used.41,43,126
10.1.8 Duration of treatment is variable and may be prolonged. A prolonged treatment time may be associated with reduced patient adherence to prescribed therapy.
10.1.10 The need for a power source makes the SVN less portable, particularly in the ambulatory setting, outside of the home.11,14,64,84
10.2 SVN and LVN With Face Mask
10.2.1 Cold, wet mist may be irritating to children and may limit the time that the treatment is tolerated.15,85
10.2.2 Aerosol deposition is reduced because of upper airway impaction.136
10.2.3 Use of LVN for bronchodilator delivery is limited to use in a critical care setting and not appropriate for home use.
10.2.4 Leak between the mask and the face decreases the amount of aerosol inhaled by the patient.17,18
10.3 pMDI
10.3.1 Patients who cannot perform hand-breath coordination or proper inhalation technique should not use a pMDI.7,10,35,36,126,143,144
10.3.2 Failure to shake a pMDI before each use can interfere with correct drug release.7,66,95,126,144
10.3.3 Failure to prime a pMDI can also affect correct drug release.7,66,126
10.3.4 Use of a pMDI without a spacer device results in greater oropharyngeal impaction and a reduction in airway deposition.64,84,126
10.3.5 Inadequate or inaccurate instruction and technique may result in misuse and reduced aerosol deposition.10,85,126
10.3.6 Propellants, excipients, and drugs may cause bronchospasm in some patients with hyper-reactive airway diseases.
10.3.6.1 The breath-actuated pMDI (Autohaler) uses chlorofluorocarbon (CFC) as a propellant.
10.3.7 Average aerosol deposition in the lungs ranges from 1% to 40% of the total dose, depending on size, age, device and interface.4,67,68,145–147
10.3.8 Lack of Built-In Dose Counter.
10.4 DPI
10.4.1 The efficiency of DPI is dependent upon patient's inspiratory flow. In clinical situations where inspiratory flow is < 40–60 L/min, such as respiratory disease exacerbations or children < 4 years old,72,85 use of DPI is associated with reduced lung deposition.64,72,130
10.4.2 Vulnerable to ambient humidity or exhaled humidity.10,64,84,85
10.4.4 If a single-dose DPI is used, additional time is needed to load the dose, and the patient must be capable of loading the dose before using the DPI.
10.4.5 Patients are less aware of delivered dose unless the DPI has a built-in dose counter.
10.4.7 Wide range of designs with different ways to load the dose may lead to incorrect use. This can be best resolved by proper patient and caregiver education and return demonstration.
10.5 Device Interface
10.5.1 Face Mask
10.5.1.1 Face masks with larger dead space volume reduce aerosol delivery to infants and children.75,148
10.5.1.2 Leaks between the mask and the face decrease the amount of aerosol inhaled by the patient.18,22,23,75,77,78,149–154
10.5.1.2.1 In infants and small children, a small leak decreases drug inhaled by > 50%.
10.5.1.3 Aerosolized medications delivered with a face mask may inadvertently deposit in the eyes, resulting in eye irritation.23,76,79
10.5.2 Blow-by. Blow-by significantly decreases aerosol delivery as the distance from the device and the patient's face is increased. It is ineffective and should be discouraged.133,155,156
10.5.3 Mouthpiece
10.5.3.1 Inappropriate patient use may result in underdosing.
10.5.4 Spacer/VHC
10.5.4.1 Using the spacer/VHC with the pMDI increases cost.
10.5.4.2 Assembly is necessary.
10.5.4.3 All spacers do not eliminate coordination problems.
10.5.4.3.1 Open tube spacer/accessory devices (non-valved) require coordination.
10.5.4.3.2 A spacer/accessory device with one-way valve (ie, VHC) eliminates coordination problems.
10.5.4.4 Dose delivery can be affected in some spacer designs if the device does not fit the pMDI properly.6,14,126
10.5.4.5 Electrostatic charge decreases aerosol delivery.148
10.5.4.6 The spacer/VHC is larger and more cumbersome than the pMDI alone.10,126
10.5.4.6.1 VHCs with large volume are disadvantageous for infants and small children, as it is difficult to empty larger VHCs with fewer inhalations.148
10.5.4.7 All spacer/VHCs may not fit all pMDIs.
10.5.4.8 Improper use of VHC results in inconsistent aerosol delivery.148
10.5.4.9 Possible contamination with inadequate cleaning.
10.5.4.10 Valve malfunction in the VHC may decrease drug delivery.
10.5.4.11 The inspiratory flow and the number of inhalations required by children < 3 years to effectively use VHC with mask is not completely understood.2,10,85,135
10.6 Drug pharmacokinetics and pharmacodynamics are markedly altered in neonates and may require dose adjustments.15,93,130,147
ADS 11.0 RESOURCES
11.1 Equipment
11.1.1 Power source such as electricity, hospital compressed oxygen or air, portable oxygen or air cylinder, or domiciliary air compressor capable of producing a flow of 6–8 L/min.64,84
11.1.2 Aerosol Generators
11.1.2.1 SVN capable of producing a high drug output, short nebulization time, aerosol particles with a mass median aerodynamic diameter (MMAD) of 1–5 μm, and with a low residual volume.11,12,41,129,157 Characteristics of nebulizers may vary among brands, and among units of the same brand.115,128,134,158
11.1.2.2 LVN capable of producing aerosol particles with an MMAD of 1–5 μm, with face mask
11.1.2.3 pMDI, which includes the medication canister and actuator, with appropriate accessories for the patient's ability and circumstances
11.1.2.4 DPI with accompanying medication capsule/blister and dispenser
11.1.3 Interface
11.1.3.1 Face mask for small children unable to utilize a mouthpiece.2,16,86
11.1.3.2 Mouthpiece, with or without extension reservoir, depending on the type of nebulizer used.84
11.1.3.3 VHC or spacer with mouthpiece or mask, depending on the patient's age and physical and cognitive ability5
11.1.3.4 Hood, depending on the patient's age and tolerance as well as parents' preference81–83
11.1.4 Medication and Diluent
11.1.4.1 Selection of an aerosol generator is limited by drug availability for a specific device or type of aerosol generator.
11.1.5 Cost
11.1.5.1 Selecting the least expensive aerosol generator for the patient is essential. The cost of an aerosol generator depends upon the type of aerosol device, its brand, drug formulation, and dosage. While jet nebulizers are low-cost, newer and more efficient aerosol generators are more expensive.14 Table 3 shows the cost of nebulizers.
View this table:11.1.5.2 Since aerosol treatment with pMDI reduces the treatment time at the bedside and increases the productivity of respiratory therapists, it is considered an economical alternative to nebulizer for aerosol delivery to patients with pulmonary diseases.140,141,159–164 Previous studies reported success in substituting a pMDI for a nebulizer by documenting a 30–50% reduction in the annual cost of aerosol therapy.140,142,165,166 However, success in this substitution depends on proper planning, careful implementation, and comprehensive education programs that were directed to physicians, respiratory therapists, and other healthcare professionals. Also, it must be noted that some medications are not available in the form of pMDI, and some patients are not capable of using pMDI correctly, due to their age and physical and cognitive abilities.
11.2 Most healthcare providers do not know how to use aerosol generators correctly.167–171 Healthcare providers responsible for delivery of aerosolized medications should have demonstrated and documented knowledge and skills related to:
11.2.1.1 Aerosol delivery devices and their limitations94,95,114
11.2.1.2 Assembly, care, and use of aerosol delivery devices94,95,114
11.2.1.3 Provision of comprehensive patient and lay caregiver instruction86,172,173
11.2.1.4 Medications being delivered, including contraindications, potential side effects, and desired response
11.2.1.5 Incompatibility of drugs if combined in the nebulizer cup174–176
11.2.1.6 Recognition and response to adverse reactions during medication delivery, and modification of treatment accordingly
11.2.1.7 Performance of the necessary subjective and objective assessments in order to determine medication efficacy and the patient's ability to properly utilize aerosol delivery devices94,95,177
11.2.2 Healthcare providers should train and re-train patients about how to use aerosol generators correctly.94,95,178
11.2.3 Most patients do not use their inhalers well enough to benefit from their prescribed medication.120,179–181 Patients and/or family members or lay caregivers should:
11.2.3.1 Demonstrate proper use and understanding of the aerosol delivery device and delivery technique.94,95,113,173,177,182–184
11.2.3.2 Demonstrate proper assembly, cleaning, and care of the aerosol delivery device, and aseptic medication preparation.95,113
11.2.3.3 Demonstrate an understanding of medication purpose, dosage, indications, and side effects, be able to alter medication as needed, and know when to report to physician or surrogate.95
ADS 12.0 MONITORING
12.1 Observe delivery technique of spontaneously breathing patients who are able to self-administer aerosolized medications.
12.1.1 A periodic slow deep inhalation with an inspiratory pause/hold is performed during SVN treatments.157 Hyperventilation should be avoided, and the patient should be observed to ensure that aerosol mist is being inhaled.
12.1.2 pMDI actuation occurs at the beginning of inhalation, followed by a slow inspiration and breath-hold for up to 10 seconds.157
12.1.3 Patient is able to produce a rapid inhalation in order to fully activate and discharge DPI.
12.2 Observe patient and/or patient's family member following instructions and demonstration.94,95
12.2.1 Proper understanding and return-demonstration of delivery device and accompanying equipment is observed.
12.2.2 Proper understanding and preparation of medication is observed.
12.3 Observe response to medication by performing subjective (eg, physical examination) and objective (eg, pulmonary function measurements) assessments and other diagnostic techniques that are appropriate for the specific medication being delivered.
12.3.1 Ensure that medication volume is nebulized over desired amount of time when using LVN.
12.4 Documentation
12.4.1 Successful training of patients and/or patient's family member is documented in the patient's medical record.95
12.4.2 Treatments administered in a clinical setting are documented in the patient's medical record. Information on medication dose, frequency, response, and adverse reactions are included.
ADS 13.0 FREQUENCY
Aerosol delivery devices are used according to the frequency of the prescribed medication.
ADS 14.0 INFECTION CONTROL
14.1 Standard precautions and measures to limit the transmission of airborne pathogens must be adhered to at all times.103,124
14.2 SVN
14.2.1 Jet nebulizers are for single patient use and should be changed every 24 hours, or at a frequency determined in collaboration with infection control, based on local data, when used in the hospital.37,124,185–187
14.2.2 Jet nebulizers should be cleaned, rinsed with sterile water, and air-dried between treatments on the same patient.14,124,186
14.2.3 At home, parts of aerosol generators should be washed with soap and hot water after each treatment, with care not to damage any parts of the aerosol generator.103
14.2.4 Mesh and ultrasonic nebulizers should be cleaned and disinfected based on the manufacturer's recommendations. The mesh should not be touched during the cleaning of mesh nebulizers, in order to prevent the damage of the unit.14
14.3 LVN
14.3.1 LVNs are for single patient use.
14.4 pMDI
14.4.1 The plastic boot of pMDIs should be cleaned according to the manufacturer recommendation.
14.4.2 When a spacer is used with a pMDI, it should be cleaned before first use, and then periodically cleaned based on the manufacturer's suggestions.
14.5 DPI
14.5.1 DPI and accessories are for single patient use only. Clean or replace when they appear dirty.
14.5.2 DPIs should not be submerged in water. Also, they should be kept dry, as moisture will decrease drug delivery.14
14.5.3 Although there is no clear evidence about the cleaning practice for DPI, each manufacturer has recommendations for periodic cleaning and suggests wiping the mouthpiece of the DPI with a clean dry cloth.14
14.6 Aerosol Solutions
14.7 Patients should be instructed to rinse the mouth with water following each administration of inhaled steroids.
ADS 15.0 RECOMMENDATIONS
The recommendations below are made following the Grading of Recommendations Assessment, Development, and Evaluation (GRADE)189,190 criteria.
15.1 It is recommended that selection of the appropriate aerosol generator and interface be made based on the patient's age, physical and cognitive ability, cost, and the availability of the prescribed drug for use with a specific device. (1B)
15.2 Nebulizers and pMDIs with VHCs are suggested for use with children ≤ 4 years of age and adults who cannot coordinate the use of pMDI or DPI. (2B)
15.3 It is suggested that administration of aerosols with DPIs be restricted to patients ≥ 4 years of age who can demonstrate sufficient flow for the specific inhaler. (2B)
15.4 For patients who cannot correctly use a mouthpiece, aerosol masks are suggested as the interface of choice. (2B)
15.5 It is suggested that blow-by not be used for aerosol administration. (2B)
15.6 It is suggested that aerosol therapy be administered with a relaxed and nondistressed breathing pattern. (2B)
15.7 Unit dose medications are suggested to reduce the risk of infection. (2C)
15.8 It is suggested that nebulizer/drug combinations should be used as approved by the FDA. (2A)
15.9 It is recommended that healthcare providers know the correct use of aerosol generators; they should teach and periodically re-teach patients about how to use aerosol devices correctly. (1A)
15.10 It is suggested that intermittent positive-pressure breathing should not be used for aerosol therapy. (2B)
15.11 It is recommended that either nebulizer or pMDI can be used for aerosol delivery during NIV. (1B)
16.0 ADS CPG IDENTIFYING INFORMATION
16.1 Adaptation
Original Publication: Respir Care 1995;40(12):1325-1335
16.2 Guideline Developers
American Association for Respiratory Care Clinical Practice Guidelines Steering Committee.
Arzu Ari PhD RRT PT CPFT FAARC (Member), Division of Respiratory Therapy, Georgia State University, Atlanta, Georgia.
Ruben D Restrepo MD RRT FAARC (Chair), The University of Texas Health Sciences Center at San Antonio, San Antonio, Texas.
16.3 Source(s) of Funding
None
16.4 Financial Disclosures/Conflicts of Interest
The authors have disclosed no conflicts of interest.
Footnotes
- Correspondence: Arzu Ari PhD RRT PT CPFT FAARC, Division of Respiratory Therapy, School of Health Professions, Georgia State University, PO Box 4019, Atlanta GA 30302-4019. E-mail: arzuari{at}hotmail.com.
The authors have disclosed no conflicts of interest.
- Copyright © 2012 by Daedalus Enterprises Inc.
References
- 1.↵
- 2.↵
- 3.↵
- 4.↵
- 5.↵
- 6.↵
- 7.↵
- 8.↵
- 9.↵
- 10.↵
- 11.↵
- 12.↵
- 13.↵
- 14.↵
- 15.↵
- 16.↵
- 17.↵
- 18.↵
- 19.
- 20.
- 21.↵
- 22.↵
- 23.↵
- 24.↵
- 25.↵
- 26.↵
- 27.
- 28.
- 29.
- 30.
- 31.
- 32.
- 33.
- 34.
- 35.↵
- 36.↵
- 37.↵
- 38.
- 39.↵
- 40.↵
- 41.↵
- 42.↵
- 43.↵
- 44.↵
- 45.
- 46.↵
- 47.↵
- 48.↵
- 49.
- 50.
- 51.
- 52.↵
- 53.↵
- 54.↵
- 55.↵
- 56.↵
- 57.↵
- 58.↵
- 59.↵
- 60.
- 61.↵
- 62.↵
- 63.↵
- 64.↵
- 65.↵
- 66.↵
- 67.↵
- 68.↵
- 69.↵
- 70.↵
- 71.↵
- 72.↵
- 73.↵
- 74.↵
- 75.↵
- 76.↵
- 77.↵
- 78.↵
- 79.↵
- 80.↵
- 81.↵
- 82.↵
- 83.↵
- 84.↵
- 85.↵
- 86.↵
- 87.↵
- 88.↵
- 89.↵
- 90.↵
- 91.↵
- 92.↵
- 93.↵
- 94.↵
- 95.↵
- 96.↵
- 97.↵
- 98.↵
- 99.
- 100.↵
- 101.
- 102.↵
- 103.↵
- 104.
- 105.↵
- 106.↵
- 107.↵
- 108.
- 109.↵
- 110.↵
- 111.
- 112.↵
- 113.↵
- 114.↵
- 115.↵
- 116.↵
- 117.↵
- 118.
- 119.↵
- 120.↵
- 121.↵
- 122.
- 123.↵
- 124.↵
- 125.↵
- 126.↵
- 127.↵
- 128.↵
- 129.↵
- 130.↵
- 131.
- 132.
- 133.↵
- 134.↵
- 135.↵
- 136.↵
- 137.↵
- 138.↵
- 139.↵
- 140.↵
- 141.↵
- 142.↵
- 143.↵
- 144.↵
- 145.↵
- 146.
- 147.↵
- 148.↵
- 149.↵
- 150.
- 151.
- 152.
- 153.
- 154.↵
- 155.↵
- 156.↵
- 157.↵
- 158.↵
- 159.↵
- 160.
- 161.
- 162.
- 163.
- 164.↵
- 165.↵
- 166.↵
- 167.↵
- 168.
- 169.
- 170.
- 171.↵
- 172.↵
- 173.↵
- 174.↵
- 175.
- 176.↵
- 177.↵
- 178.↵
- 179.↵
- 180.
- 181.↵
- 182.↵
- 183.
- 184.↵
- 185.↵
- 186.↵
- 187.↵
- 188.↵
- 189.↵
- 190.↵