Abstract
BACKGROUND: Mechanical insufflation-exsufflation (MI-E) devices increase expiratory air flow and thereby promote increased cough peak flow (CPF) in conjunction with a cough. There is little research looking at long-term use of MI-E in subjects with neuromuscular disease (NMD), and no long-term study has reported CPF, MI-E device settings, and adherence.
METHODS: We evaluated 181 patient records (130 adults, 51 children) of individuals who received a MI-E device from our center between February 2014 and February 2018. Median age (interquartile range [IQR]) was 27 (14–51) y. Duchenne muscular dystrophy (DMD), spinal muscular atrophy (SMA), and amyotrophic lateral sclerosis (ALS) were the 3 most common diagnoses.
RESULTS: MI-E devices were provided to the weakest subjects with a CPF < 160 L/min. Median (IQR) settings were insufflation, 25 (23–30) cm H2O, exsufflation −35 (−30 to −40) cm H2O, insufflation time 1.5 (1.3–1.7) s, exsufflation time 1.8 (1.5–2.0) s, and pause 1.5 (1.3–2.0) s. The inspiratory flow profile was set to high in all subjects, and no subject used supplemental oxygen with the MI-E device. When comparing insufflation pressures to exsufflation pressures, a greater negative pressure was used relative to positive pressure (P < .001). When comparing insufflation to exsufflation time, there was a significantly longer exsufflation duration (P < .001). Median (IQR) CPF at the start of MI-E was 60 (10–100) L/min. There was no correlation between either insufflation or exsufflation pressures and CPF. Median (IQR) usage for the group was 60% (13.5–100%) of days for the total days. Subjects with tracheostomies or SMA type I had the greatest adherence to treatment. Median (IQR) duration of MI-E use was 17 (8.5–32) months. Ninety-six percent of subjects were receiving ventilatory support.
CONCLUSIONS: Greater exsufflation pressures than insufflation pressures, together with a shorter insufflation time than exsufflation time, were used. Predicting good adherence among the subjects was difficult. Subjects who produced daily secretions were more likely to use MI-E every day.
Introduction
The aim of a mechanical insufflation-exsufflation (MI-E) device is to increase expiratory air flow and thereby promote, in conjunction with a cough, increased cough peak flow (CPF). By increasing CPF, secretions move more efficiently to the central airways to be cleared from the mouth or an artificial airway. Two recent publications by an international panel of experts in this area concluded that MI-E is an expensive but “ideal” technique to assist cough in individuals severely affected by neuromuscular disease (NMD).1,2 They also produced recommendations for individuals over the age of 12 years based on the available evidence of which cough-augmentation technique would be most effective, depending on the individual's CPF. MI-E devices are recommended in individuals with a CPF ≤ 160 L/min.1
Despite there being recommendations for commencing MI-E when CPF is < 160 L/min,1,3–6 there are no such recommendations for optimum settings in patients with NMD. Bench studies have shown that high insufflation and exsufflation pressures (+50 cm H2O to −50 cm H2O) are required to produce effective peak expiratory flows from artificial aiways.7 In a pediatric lung model, an insufflation time of > 1 s was required for equilibrium between insufflation pressure and alveolar pressure.8 Higher insufflation and exsufflation pressures both increased expiratory flow, but greater exsufflation pressure had a more substantial impact on mean expiratory flow.8 Long exsufflation times did not improve expiratory flow. In an adult lung model, high pressures (+40 to −40 cm H2O), with an insufflation time of 3 s and an exsufflation time of 2 s, were required to generate an exsufflation flow of 294 L/min.9 The authors concluded that increasing insufflation time more than exsufflation time was more important for optimum effect. A recent study found that mucus transport could be further enhanced by applying a slow insufflation, which created a greater expiratory flow bias.10 Despite this evidence base, some centers reported an increase in CPF with low pressures,11–14 whereas others have reported efficacy with high pressures.15–20 Hov and coworkers21 surveyed centers in Europe and found that younger children were set up with lower MI-E pressures than older children and that there was wide variation in settings prescribed. There has only been one study, a small survey of 37 subjects, to report that < 50% of subjects with an MI-E device used it daily.22 Optimum MI-E adherence in this group remains unknown.
Based on the British Thoracic Society Guidelines,3,4 subjects at the Royal Brompton Hospital were assessed for MI-E and other cough-augmentation techniques. MI-E was considered as a treatment option in subjects who had experienced recurrent respiratory tract infections (ie, > 3 in 12 months) and who remained unable to increase CPF to > 160 L/min with other cough-augmentation strategies. MI-E was considered in children (ie, < 12 y old) with recurrent respiratory tract infections when CPF was < 50% of predicted.23
There are no large reports of service provision for long-term use, safety, downloaded adherence, discontinuation rate, and prescription details of MI-E; such reports could inform clinical practice. We report our subject group with regard to CPF at initiation of MI-E, adherence to treatment recommendations, safety, settings, and discontinuation of MI-E.
QUICK LOOK
Current knowledge
Mechanical insufflation exsufflation (MI-E) devices improve cough efficacy. In children receiving home MI-E, wide variations in the settings prescribed and in adherence have been reported. No studies have reported long-term use with respect to adherence and safety.
What this paper contributes to our knowledge
Long-term use of MI-E was safe in subjects with neuromuscular disease with few side effects. Individualized settings and management plans were applied considering each subject's age and diagnosis. Greater exsufflation pressures than insufflation pressures were used. Adherence was variable; subjects with daily secretions or a tracheostomy were more likely to use MI-E daily.
Methods
See Figure 1 for subject flow process. Where possible, subjects performed an unassisted CPF. Subjects were asked to take a deep breath in and then cough into a pediatric flow meter (Pediatric Low Flow Wright Mini Peak Flow Meter, Clement Clarke, Harlow, Essex, United Kingdom) attached to a QuadraLite anesthetic full face mask (Intersurgical, Wokingham, Berks, United Kingdom). This measurement was repeated at least 6 times, and the best result was recorded.1,2
The Research Services team was contacted at the Royal Brompton Hospital to determine whether ethical approval was required for the project. All data were anonymized for analysis. The project was deemed to be service development and an evaluation project. This was registered on the clinical audit database (CIRIS for Healthcare, Project ID 001566).
Inclusion and Exclusion Criteria
Inclusion criteria were all adults and children with NMD who fulfilled criteria for MI-E use. Exclusion criteria were undrained pneumothorax present prior to initiation, pneumomediastinum, tracheoesophageal fistula, severe facial deformity, unstable upper airway, epistaxis, severe gastroesophageal reflux with risk of aspiration, severe esophageal and gastric varices, and inability to use a full face mask. No subject was excluded due to learning difficulties, severe bulbar impairment, or if they required nasopharyngeal or oropharyngeal suction.
MI-E Service Provision
If MI-E was deemed to be appropriate, subjects were placed within our personalized MI-E home care service, which consists of a 24-h emergency contact number for patients to call if they have any problems with their device or consumables, or to call if their device breaks down or becomes faulty. Devices were exchanged within 24 h if broken or faulty. Yearly servicing and consumables were sent directly from the home care company to the subject on a quarterly basis. Overall the service was part of patients' standard National Health Service care package.
MI-E Initiation
The need for a MI-E device was assessed by the respiratory physiotherapist. The initiating respiratory physiotherapist was also responsible for ensuring that the subject and their carers were trained in using the device and reached a level of competence to be able to use the device effectively and safely. Subjects were started on MI-E in both the in-patient and out-patient settings. During the initiation session, the subject's oxygen levels were monitored with pulse oximetry. Resuscitation equipment was readily available in the event that an overwhelming volume of secretions could not be cleared with MI-E.
Settings
Subjects were initiated on MI-E as per our starting settings algorithm (Fig. 2). All subject prescriptions for MI-E were then individualized. We did not have a maximum insufflation pressure that could be set, although insufflation pressure was limited to prevent mask leak. Feedback was provided by the subject as to whether they subjectively felt their cough was stronger. An increase in cough audibility was also used as an indicator of improvement in cough strength because the device in use did not measure CPF.24 Settings were optimized by the physiotherapist and not the home care service and maintenance team. Any changes in prescription or subject details were updated with the home care service and the maintenance team to ensure an up-to-date accurate prescription, should a home visit or device exchange be required.
Usage Recommendations
Subjects were advised to use the device on a daily basis with 3 aims: to ensure familiarity with the device, to ensure their chest was clear, and to maintain chest wall compliance. Subjects were encouraged to let the device fill up their lungs and then deflate them. Subjects were recommended to cough with exsufflation only when secretions were in the central airways.
All subjects after January 1, 2013, were provided with an MI-E device (Nippy Clearway, Breas Medical, Stratford Upon Avon, Warks, United Kingdom) from our home care service and maintenance team. From our database we extracted subject demographics: sex, age at initiation, diagnosis, whether they were able to perform a CPF maneuver, unassisted CPF, vital capacity. We evaluated MI-E modes, settings, and interface (mask or tracheostomy). We also examined subject adherence over a 12-month period and the number and the nature of calls to the emergency line and to the home care service and maintenance team.
Results
A total of 181 subjects (130 adults, 51 children; 0–16 y old; 115 male, 66 female) received a MI-E device between February 2014 and February 2018. Median (IQR) duration of MI-E use was 17 (8.5–32) months. Group median age (interquartile range [IQR]) was 27 (14–51) y. Adults were defined as > 16 y old, and the median age (IQR) was 33 (24–56) y. The median age (IQR) of pediatric subjects was 10 (4–13) y. The youngest subjects were 9 months old. In subjects with Duchenne muscular dystrophy (DMD), median (IQR) age was 27 (23–29) y; in subjects with spinal muscular atrophy (SMA), median (IQR) age was 14 (8–20) y, and in subjects with amyotrophic lateral sclerosis (ALS), median (IQR) age was 58 (51–73) y old, which were the most common diagnoses. See Table 1 for full subject diagnoses.
Device mode, settings, and interface are shown in Table 2. Significantly higher exsufflation pressures were used compared to insufflation (P < .001), along with a longer exsufflation to insufflation time (P < .001) for all subjects. Settings for the most common diagnoses are shown in Figure 3; subjects with DMD used significantly higher insufflation P = .02 and exsufflation pressures P = .009 compared to those with SMA. We found no correlation between age and settings.
CPF was measurable in 114 subjects, and median (IQR) CPF was 60 L/min (10–100 L/min); 16 subjects were unable to perform the maneuver. CPF was not measured in 51 subjects (17 had tracheostomies, 4 were initiated while acutely ill, and 30 were completely dependent on noninvasive ventilation. We found no correlation between pressure settings and unassisted CPF (P > .050).
During the study period, 29 subjects died. One device was removed because the subject was not using it. Another subject stopped due to pneumothorax related to pleuroparenchymal fibroelastosis; this subject was on noninvasive ventilation, and the pneumothorax was not thought to be directly related to MI-E use. Two subjects did not tolerate the device. The first of these 2 subjects stopped due to severe abdominal distention despite using a low insufflation pressure of 16 cm H2O and an exsufflation pressure of 25 cm H2O; this subject had longstanding issues with abdominal distention prior to the initiation of MI-E. The other subject discontinued due to feeling uncomfortable with an insufflation pressure of 15 cm H2O and a sensation of suffocation with exsufflation (−15 cm H2O), despite individualized settings and adjustments. Three subjects transferred care to other centers.
Information prospectively recorded by the home care service and maintenance team reported that our subjects made up < 1% of all after-hours calls received within a 2-y period. Calls received during office hours were to report change of address and changes to the device setting. Only 3 after-hours calls required intervention for circuit problems, device suction, and to see if the device was working. Two visits were made, and no issues with the device were found. There were no calls to the home care service regarding practical application of the device. No devices failed mechanically.
We received downloads for 76% of subjects (137 of 181 subjects). Download data were not available in 24% of subjects because not all subjects had their device for > 1 y, and therefore a download had not been performed. Results are shown in Table 3. Sixty-four percent of subjects with tracheostomies used MI-E on a daily basis versus 31% who did not have a tracheostomy. The only subject group to use the device on a daily basis was those with SMA type I. Conversely, 22% of subjects used the device for < 10% of the total days available on download. In this group of low users, only 1 subject had a tracheostomy; 4 subjects had a CPF of 100–160 L/min, whereas all of the others had a CPF < 100 L/min and 6 were unable to perform a CPF manuever. Download adherence was not related to a specific diagnostic group.
Discussion
We report the largest single-center experience of MI-E in the clinical setting. Although this is a diverse group of subjects with an ineffective cough, we believe that our results will be a record of practice to help other teams reach consensus. We generally used greater exsufflation pressures than insufflation pressures. Subjects with tracheostomies were more likely than those without to use the device on a daily basis. CPF did not correlate with the pressures required for an effective cough.
Limitations of the Study
Previous studies have reported improvements in SpO2, avoidance of hospitalizations, and invasive techniques for mucus removal, along with decreased work of breathing using MI-E.11,18,25,26 We did not capture any data on hospitalizations or avoidance of hospitalizations in our group, and therefore, it is impossible to know how many respiratory tract infections were prevented. We also were not able to measure any changes in CPF because the device we used does not have this feature.
Our center is a highly specialized sleep and respiratory care service. Many of our subjects were using ventilatory support. Therefore, our results may not be extrapolated to patients who are not using ventilatory support. That said, most patients who use MI-E devices are likely to need ventilatory support at some time. We hope that by providing detailed information on settings we can reach out to other teams and the wider community to help plan comprehensive care.
Significance of Findings
Settings.
Like Hov and co-workers21 we found the automatic mode was the most commonly used mode in our sample. We also used greater exsufflation pressures than insufflation pressures, and there was a tendency to increase pressures with age in the pediatric population. Our work adds to the literature by providing information about pressure settings in the adult population and comparing settings in 3 subject groups. Median insufflation and exsufflation settings were similar in subjects with DMD and ALS; subjects with SMA used significantly lower settings than those with DMD. One explanation for this may be that the subjects with SMA were younger at initiation than those with DMD. Our subjects' median (IQR) pressure settings were 25 (23–30) cm H2O for insufflation and −35 (−30 to −40) cm H2O for exsufflation. Pressures may seem lower than described in bench studies,7,9 but we used a NippyClearway device, which has a pressure line at the mask that delivers the set pressure on the machine14,27 and thus eliminates the typical pressure drop across circuitry reported by others.28 This is a more accurate way of ensuring that the pressure set on the machine is delivered to the subject at the mask.27,28 Knowledge regarding optimum time settings for insufflation, exsufflation, and pause is mainly experience-based. Our settings did have a greater insufflation time of 1 s as recommended for equilibrium between insufflation pressure and alveolar pressure.8 However, in contrast to bench studies, our insufflation time was shorter than our exsufflation time (1.5 [1.3–1.7] s vs 1.8 [1.5–2.0] s). Despite this, all of our subjects' settings were titrated so that they could clear secretions. Our insufflation time was slightly less than our exsufflation time, and we also used a greater exsufflation pressure. The combination of a high exsufflation pressure and shorter exsufflation time creates a high air-flow bias, and therefore did not cause secretion retention. All subjects' inspiratory flow profiles were set to high. This is also likely to contribute to an expiratory air-flow bias because our exsufflation pressures were greater than our insufflation pressure. However, with new evidence emerging from both a bench study,10 and flexible transnasal fiber-optic laryngoscopy during MI-E,29,30 it is clear that more attention needs to be paid to insufflation flow and insufflation pressures. A key implication for clinicians is that some bulbar ALS patients are unlikely to benefit from high pressures and flows and may not tolerate MI-E if they have spasticity of the upper airway.
Cough Peak Flow.
No study has evaluated CPF with respect to insufflation and exsufflation pressure settings. One possible explanation for this could be that the lower the unassisted CPF, the greater the insufflation and exsufflation pressures required to generate an effective cough. On the other hand, the individual may have a low CPF but a compliant chest wall, which would mean that lower driving pressures are required to deliver an effective CPF. Indeed, Mellies and Goebel31 showed that delivering a deep breath in to a maximum insufflation capacity did not produce the greatest CPF. Lacombe et al32 reported no benefit in the addition of MI-E when subjects could generate high CPF with a maximum insufflation combined with a manual assisted cough. It is difficult to have a protocol dictating one setting for all patients, thus we advocate for an individualized prescription. Further prospective studies are warranted to evaluate the relationship between CPF and settings.
Adherence.
Expert consensus reports suggest that use of MI-E in very weak individuals is a priority.3,4 When the CPF is ∼ 160 L/min, it is thought that the best device to improve cough efficacy is an MI-E device.14,33,34 The median CPF was 60 L/min in our subject group with an interquartile range of 10–100 L/min, and 97% of our subject group received ventilatory support. We therefore believe that we provided devices to the subjects who needed it most, based on current guidelines.1,3–6
We report 137 adherence downloads. This is the largest report of use. The median daily use was 60% of the days downloaded. Unlike others, we did not ask our subjects in detail why they used their devices as they did. Subjects with a tracheostomy or a diagnosis of SMA type I were more likely to use the device daily. One explanation for this is that individuals with tracheostomies are likely to produce secretions daily. The use of MI-E through a tracheostomy has been reported as less irritating, less painful, less tiring, and less uncomfortable,35 as well as more effective, than endotracheal suctioning, which can be uncomfortable.36 This may be one of the reasons why a higher daily adherence to the use of MI-E was noted in our tracheostomy group. In a survey of 37 subjects with NMD who used MI-E at home, 46% used MI-E daily and 27% weekly.22 We also found that 22% of subject downloads had low usage (< 10% of total days downloaded). From our data we cannot conclude a reason for low usage. The group had a mixture of diagnoses, and not all subjects had preserved pulmonary function; however, upon review, all had a low secretion burden. As a barrier to use, it has been suggested that domiciliary MI-E medicalizes the home,37 despite perceived benefits of MI-E for physical, social, and emotional aspects of living.38 Negative impacts were greatest for those parents who were sole operators of the device, who frequently reported disruption to lifestyle.37 Our view is that the daily presence of secretions is more likely to drive daily usage than low CPF. In individuals with low adherence, the barriers to use and the related consequences should be explored.
We asked subjects to use the device daily to check for secretions, to provide a chest wall stretch, and to remain familiar with the device. Some subjects may feel familiar with the device quickly, or they may not have secretions daily, or they may use a ventilator that provides a chest wall stretch. In this situation, perhaps the treatment plan should be different and less intensive, or the device should be available via a rapid-access program provision when needed, given that these individuals only require the device when secretions are a problem. Pragmatically, MI-E use when required is more sensible because we should avoid overburdening individuals with unnecessary medical interventions.39 However, it is important to have the right equipment available to individuals to prevent prolonged ineffective secretion-clearance sessions and respiratory tract infections. Requirements should be reassessed regularly, especially in those with progressive disorders.
Training and Support.
Lack of consistent advice and education has been seen as a barrier to use of MI-E in children.38 Another study investigating the home use of MI-E in subjects with ALS reported that health professionals need to be mindful of individual and social aspects of implementing an MI-E device in the individual's home environment. Proper instruction and practical training to instill confidence in how to use the device, plus trust and continuity among carers, are important factors for successful implementation.40 To facilitate this in our group, we had a comprehensive in-house training program.
Complications.
Reported adverse effects of MI-E include pneumothorax,26 nausea, bradycardia, tachycardia, and abdominal distention.41 In children, thoracic wall discomfort, crying, and agitation in response to treatment with MI-E have been reported.13 Like others,26 we have experienced complications with MI-E. Our process for provision of the device is highly selective. All subjects were assessed and trialed with the device prior to home use. Only 2 subjects returned the device because they could not tolerate it; one subject had existing abdominal symptoms, and these were exacerbated; the other could not tolerate pressures higher than their ventilator inspiratory positive airway pressure. One subject stopped due to a pneumothorax, which was managed medically, and 3 other subjects who developed pneumothorax continue to use the device with lower pressure settings and regular monitoring.
Conclusions
In this large, single-center review of MI-E practice, we have demonstrated that good adherence can be achieved with a comprehensive education program. MI-E devices were provided to the weakest subjects with a CPF < 160 L/min, which is in line with current guidelines. We used greater exsufflation pressures than insufflation pressures, along with insufflation times less than exsufflation times. Predicting which patients will exhibit good adherence is difficult, but our results indicate that patients who produce secretions daily are more likely to use MI-E daily.
Acknowledgments
The authors acknowledge Adam Rochester and Caroline Davies-Jones for their assistance in the initiation of subjects with MI-E devices.