Abstract
Controlled mechanical ventilation is characterized by a fixed breathing frequency and tidal volume. Physiological and mathematical models have demonstrated the beneficial effects of varying tidal volume and/or inspiratory pressure during positive-pressure ventilation. The addition of noise (random changes) to a monotonous nonlinear biological system, such as the lung, induces stochastic resonance that contributes to the recruitment of collapsed alveoli and atelectatic lung segments. In this article, we review the mechanism of physiological pulmonary variability, the principles of noise and stochastic resonance, and the emerging understanding that there are beneficial effects of variability during mechanical ventilation.
Introduction
Controlled mechanical ventilation is often used for the management of patients with respiratory failure and transiently during the intraoperative period. Mechanical ventilation employed during these periods is characterized by a fixed breathing frequency and monotonous tidal volume (VT) created by a change in airway pressure with either volume or pressure control mode. This is in sharp contrast to spontaneous breathing, which demonstrates small breath-to-breath variations in both VT and breathing frequency with interspersed, intermittent, deep-inflation breaths (sighs or augmented breaths). In 1964, Bendixen et al1 noted the significant short-term and long-term variability in ventilatory patterns in healthy adults.
Although it remains the mainstay of therapy for ARDS and other forms of lung injury, there is mounting evidence of surrogate injury associated with mechanical ventilation.2,3 Traditionally, ventilator-induced lung injury is mitigated by reducing VT and by the appropriate use of PEEP.2,3 However there is increasing preclinical evidence that mimicking the physiological variability during controlled and patient-triggered mechanical ventilation is a novel method to facilitate lung recruitment and reduce the risk of ventilator-induced lung injury.4–7 In this review article, we discuss normal physiological respiratory variability and the salutary effects on alveolar recruitment, ventilation/perfusion matching, and systemic oxygenation that occur when variability is applied to a variety of lung injury models.
Physiological Respiratory Variability
The characteristic pattern of spontaneous breathing is the result of integration of input from the higher central nervous system, brainstem control centers, chemoreception, and pulmonary mechanoreceptors (Fig. 1).8 The breathing pattern is adapted to meet the immediate metabolic demands while reducing the work of breathing by optimizing ventilation and perfusion.
The pattern of spontaneous breathing represents a complex interaction of structured time-dependent breath-to-breath changes coupled with random variations. These fluctuations are due to an inherent homeostatic mechanism and feedback loops that buffer changes to the respiratory system and attempt to return them to their dynamic steady state. This results in a highly variable, unpredictable breathing pattern.
Central and Peripheral Respiratory Control
The 4 major sites for respiratory control are the central respiratory control center, pulmonary mechanoreceptors, and central and peripheral chemoreceptors located in the brain and aorta, respectively. The central respiratory control center is located in the medulla and is functionally responsible for setting the rhythmic respiratory pattern through the ventilatory pattern generator. Modulation of the rate and amplitude of the respiratory pattern is accomplished via the integrator.8 The integrator receives multiple inputs from higher centers, including the hypothalamus, amygdala, cerebral cortex, and limbic system, and from the central and peripheral chemoreceptors (see Fig. 1). The central chemoreceptors are a group of specialized cells found on the ventrolateral surface of the medulla and respond to changes in the pH of the cerebrospinal fluid.8
In contrast, the peripheral aortic and carotid chemoreceptors respond to changes in PO2, PCO2, and pH. The final respiratory control mechanism is through the pulmonary mechanoreceptors. These include the rapidly adapting pulmonary stretch receptors, slowly adapting pulmonary stretch receptors, and juxta-alveolar receptors. The rapidly adapting pulmonary stretch receptors are located in the tracheobronchial superficial mucosal layer and result in an increase in airway resistance, reflex apnea, and coughing.8 The slowly adapting pulmonary stretch receptors are found in the tracheobronchial smooth muscle layer, are activated by lung inflation via the vagal afferent myelinated fibers, and result in termination of the inspiratory process. The juxta-alveolar receptors lie close to the pulmonary microcirculation and respond to mechanical and chemical irritation in the pulmonary interstitium. They are responsible for the subjective sensation of dyspnea and the respiratory pattern characterized by rapid shallow breathing.
Biological Systems and Noise
Stochastic Resonance
Many biological systems are characterized by a continuously variable response to changing intrinsic or extrinsic input (eg, sinus arrhythmia occurs in responses to changes in ventilation). Analysis of these biological responses demonstrates irregular random fluctuations of the output signal, characteristic of noise.9
In biological systems, noise traditionally refers to variable output with similar experimental designs or the inability to distinguish biological signals from background fluctuation.10 These are undesirable side effects of noise. However, when good noise (where both frequency and amplitude can be controlled) is added to a previously monotonous biological system, signal output can be improved. This paradoxical effect of noise is termed stochastic resonance.
Stochastic resonance was first described by Roberto Benzi while studying noise-enhancing signal processing.10 The primary feature of a nonlinear system with stochastic resonance is an increased quality or metric of the output signal with the addition of noise.11 The improved output quality is greatest at a single maximum non-zero point termed the stochastic resonance peak, equivalent to the resonance frequency of a mechanical system. The phenomenon of stochastic resonance has been quantified and reported in varied biological and nonbiological systems, such as climate and financial models.10
The static pressure-volume curve of the lung has classic nonlinear features with a steep pressure-volume relationship above the lower inflection point, which flattens above the upper inflection point (Fig. 2). Stochastic resonance can be observed theoretically by randomly varying the inspiratory pressure or VT in a gaussian distribution above the lower inflection point. The theoretical benefits are due to the marked nonlinearity above this critical point. The lung volume gained with the higher inspiratory pressures or VT will be significantly greater than the lung volume lost at the lower inspiratory pressures or VT.4,5 Additional potential benefits of stochastic resonance are suggested by the work of Suki et al,11 who demonstrated that once the critical opening pressure of a collapsed alveolus is superseded, adjacent alveoli with lower critical opening pressures will open in a series of avalanches. This is equivalent to the stochastic resonance peak of the pulmonary pressure-volume relationship and can potentially play an important role in the VT recruitment of collapsed lung units. The phenomenon of stochastic resonance can be observed by varying either inspiratory pressures or VT. When varying the VT, the overall effect is a reduction in mean airway pressure for the same exhaled minute volume.4
Limitations of Stochastic Resonance
Although addition of variability to previously monotonous biological systems can be beneficial, extremes of variability during mechanical ventilation may be detrimental. The one negative study of variable ventilation in a canine model of acute lung injury used a highly variable signal at or above the point on the pressure-volume curve where noise-related benefit can be derived.4 The addition of noise at or above the upper inflection point (where the pressure-volume curve flattens out) could have potentially deleterious effects due to increased mechanical strain on some alveoli with minimum gain in the VT.4
Under these circumstances, the risk of alveolar overdistention supersedes the beneficial effect of alveolar recruitment, resulting in volume-related, ventilator-induced lung injury. Therefore, the type of noise signal implemented in conjunction with the underlying pulmonary physiological state can potentially have a significant impact on outcomes.
Intermittent Deep Inflation: Sigh Breath
An intermittent deep-inflation breath provides an additional normal mechanism of creating respiratory variability. The physiological sigh or deep-inflation breath is characterized by a large VT breath (2–3 × VT) occurring at varying frequency (1–25 breaths/h) during spontaneous breathing.1,12 The sigh breath is triggered by the rapidly adapting pulmonary stretch receptors employing an inspiration-augmenting reflex from the vagal afferents.13–15 Central pacemaker control for the augmented breath appears to be integrated but autonomous to the control mechanism for eupnea.16 The augmented breath serves multiple physiological functions, including reaeration of collapsed alveoli, improved functional residual capacity, and reduction of the pulmonary shunt.1 The sigh breath also plays a vital role in resetting the breathing pattern. Vlemincx et al17 demonstrated that when breathing becomes either excessively random or lacks significant variability, the sigh breath is able to restore the deterministic nonrandom variability.
The frequency and pattern of sigh respirations can be altered under certain circumstances. Sigh respirations are eliminated following administration of opiates.18,19 However, Bell et al19 noted the return of augmented breath when an animal was exposed to a hypoxic mixture. Volatile and total intravenous anesthesia is not associated with any significant change in the frequency of sigh breaths.20,21 However, propofol anesthesia is associated with a reduction in mean inspiratory flow of the augmented breath and subsequent timing of the succeeding breath.20
In summary, VT and breathing frequency are constantly changing in the normal breathing pattern and are likely physiologically beneficial to normal lungs. The evidence for adding variability during controlled and patients-triggerred mechanical ventilation for ARDS and non-ARDS lung pathology is presented below.
ARDS and Variability in Mechanical Ventilation
ARDS is caused by either a primary injury to the lung or a remote extrapulmonary injury with associated lung involvement. With primary ARDS, the injury evolves mainly from the alveolar epithelial membrane with a cascade of changes, including increased alveolar edema, loss of surfactant, and subsequent healing with fibrosis.22 In contrast, extrapulmonary ARDS begins predominately with the pulmonary endothelium with increased vascular permeability. Both forms of ARDS are characterized by increased pulmonary edema and alveolar collapse that is heterogeneously distributed throughout the lung, with the dependent segments most affected.22 This results in worsening of the ventilation/perfusion mismatch characterized by increasing shunt and the development of progressive hypoxemia. Work of breathing increases dramatically, and respiratory exhaustion may result in acute mechanical respiratory failure.
Controlled mechanical ventilation remains the cornerstone of management of patients with ARDS. However, a careful ventilator management strategy is required to prevent ventilator-induced lung injury. Ventilator-induced lung injury may occur with inappropriately high alveolar distending pressure or volume, with inappropriate PEEP, or when a high ventilatory VT is delivered (> 8 mL/kg).3
Keeping the VT to ≤ 6 mL/kg as part of a strategy to prevent ventilator-induced lung injury has been shown to reduce mortality and ICU stay in both subjects with and without ARDS.2,23,24 However, applying a low-VT strategy without appropriate use of PEEP can result in alveoli collapse. In 15 subjects with ARDS, Richard et al25 demonstrated that low VT (6 mL/kg) with PEEP set at the lower inflection point was associated with significant alveolar de-recruitment, which improved with application of a recruitment maneuver. More importantly, Levin et al26 demonstrated an increased risk of 30-d mortality with low intraoperative VT and minimum PEEP in subjects without ARDS.
Controlled Variable Ventilation and ARDS
Variable ventilation was first described by Lefevre et al6 in a porcine model of oleic acid-induced acute lung injury. Oleic acid was infused in 17 animals until the systemic PaO2 was < 125 mm Hg with an FIO2 of 0.5. Animals were then randomly assigned to receive either conventional ventilation at a rate of 20 breaths/min or variable ventilation. The baseline pre-oleic acid-induced injury VT in the conventional ventilation group was 427 ± 64 mL versus 424 ± 46 mL in the variable ventilation group. The variable ventilation group had a breathing frequency of 20 ± 2.3 breaths/min (range of 15–27 breaths/min), with 369 different breathing frequencies and reciprocal changes in the VT over a 1,089-s period. The VT varied from 75 to 135% of the mean VT in this computer-controlled model of variable ventilation. After 4 h of mechanical ventilation, the variable ventilation group demonstrated a significant improvement in arterial oxygenation (140 ± 44 mm Hg vs 75 ± 20 mm Hg, P < .05) with a concomitant lower shunt fraction (Q̇S/Q̇T, 8.5 ± 3% vs 13.1 ± 3%, P < .05). The variable ventilation group also demonstrated a statistically significant higher respiratory compliance (0.62 ± 0.18 mL/cm H2O/kg vs 0.49 ± 0.16 mL/cm H2O/kg, P < .05) with lower postmortem wet/dry lung weight (8.1 ± 0.9 vs 9.3 ± 0.6, P = .01). This beneficial effect of reduced Q̇S/Q̇T and improved oxygenation with variable ventilation has been demonstrated with and without the use of PEEP.6,7
The improvement in respiratory compliance and systemic oxygenation and the reduced Q̇S/Q̇T are related to several salutary effects of variable ventilation on pulmonary alveolar mechanics. Arold et al27 demonstrated a nearly 2-fold increase in bronchoalveolar lavage phospholipid surfactant levels (160 ± 28 mg/kg vs 84 ± 29 mg/kg, P < .05) with reduced alveolar protein content (17.2 ± 3.9 mg/kg vs 12.0 ± 1.6 mg/kg, P < .05) with variable ventilation compared with conventional mechanical ventilation. Furthermore, variable ventilation also facilitates more uniform lung recruitment during acute lung injury. Using computed tomography in a porcine oleic acid-induced lung injury model, Graham et al28,29 demonstrated significantly improved recruitment of nonaerated and poorly aerated lung areas without a concomitant increase in hyperaerated lung areas. In addition, there is evidence that the recruitment associated with variable ventilation is more sustained than that with monotonous controlled mechanical ventilation when used in conjunction with PEEP.30 In an excised saline lavage bovine lung injury model, Bellardine et al30 demonstrated longer recruitment times with a combination of high VT and variable ventilation compared with high VT and conventional ventilation. These findings suggest that variable ventilation is able to provide continuous physiological recruitment without the associated risk of ventilator-induced lung injury.
There are limited studies of variable ventilation in human subjects. Boker et al31 conducted a prospective study with 40 subjects undergoing elective open abdominal aortic aneurysmectomy procedures. This procedure is associated with significant intraoperative atelectasis. Twenty subjects each were randomized to either conventional ventilation (VT of 10 mL/kg, breathing frequency of 10 breaths/min, zero PEEP) or variable ventilation (mean VT of 10 mL/kg, minimum VT of 6.4 mL/kg, maximum VT of 14.6 mL/kg). The variable ventilation group demonstrated higher arterial oxygenation and pulmonary compliance with lower PaCO2 and dead-space ventilation. The major limitation of this study was the absence of PEEP in the conventional ventilation group. In contrast, Kowalski et al32 performed a crossover trial in 8 subjects with ARDS undergoing protective lung ventilation. They reported a better oxygenation index (FIO2 × mean airway pressure × [100/PaO2], variable ventilation: 7.1 cm H2O/mm Hg vs conventional ventilation: 11.5 cm H2O/mm Hg, P = .034) and higher lung compliance (variable ventilation: 0.36 mL/cm H2O/kg vs conventional ventilation: 0.34 mL/cm H2O/kg, P = .049) when variable ventilation was instituted. Although the number of human studies on variable ventilation is limited, the aforementioned studies are congruent with the positive findings previously reported in animal models.
Patient-triggered Variable Ventilation and ARDS
The beneficial effects of noise are also seen with spontaneous ventilation modes. Conventional pressure support ventilation (PSV) is a spontaneous mode of ventilation that is either flow- or pressure-triggered, pressure-limited, and either flow- or time-cycled. Because PSV is pressure-limited, the VT can remain relatively constant if respiratory compliance does not change rapidly (min to h). Varying the inspiratory pressure randomly in a normal distribution will result in stochastic resonance as described previously. Noisy PSV was first described by Gama de Abreu et al33 in a porcine surfactant depletion model. They reported that noisy PSV decreased venous admixture compared with PSV alone (11.1% [5.5–20.3] versus 16.5% [7.9–29.8], P < .05). Using florescent, color-labeled microspheres, they also demonstrated that noisy PSV improved pulmonary blood flow to the nondependent areas of the lung, thus reducing the Q̇S/Q̇T. With noisy PSV, the amount of variability can influence both recruitment and patient-ventilator synchrony and comfort. Normally distributed inspiratory pressures with mean ± SD of 1 ± 0.075, 1 ± 0.15, 1 ± 0.30, and 1 ± 0.45 of the baseline mean pressure support required to maintain a VT of 6 mL/kg were studied by Spieth et al34 in a porcine surfactant depletion model. They found that a variability of 30% (mean ± SD of 1 ± 0.30) was associated with the highest PaO2/FIO2 with the lowest Q̇S/Q̇T.
Sigh Breaths and ARDS
As noted above, sigh breaths are normal physiological respiratory events that are thought to assist with re-recruitment of atelectatic lung units and to reset the breathing pattern when it becomes either excessively chaotic or monotonous.17 The sigh breath can be considered a limited form of noisy respiration. The first description of the sigh breath was reported in 1919 by Haldane et al35 in their study of shallow-breathing patterns in healthy adults. Subsequently, Laver et al36 demonstrated an improvement in pulmonary compliance in subjects undergoing general anesthesia when an intermittent deep-inflation breath was applied. In this early study of deep-inflation breaths, PEEP was not utilized, which may explain the marked improvement in pulmonary compliance with an intermittent augmented breath.37 Based on these earlier studies, sigh breaths were incorporated into mechanical ventilators. However, with the development and popularization of PEEP, the sigh breath has been relegated to antiquity.38
More recent studies of sigh respiration in ARDS have demonstrated consistent improvement in pulmonary parameters of gas exchange during both supine and prone ventilation.39–42 In these studies, protective lung strategies (low VT and PEEP) were utilized and subjects underwent controlled mechanical ventilation. The benefit of a sigh breath in these studies is likely related to the re-recruitment of atelectatic lung units when a low-VT strategy is used, and the PEEP values are insufficient to prevent atelectasis.
The beneficial effects of sigh-augmented ventilation are also evident with patient-triggered modes of ventilation. Patroniti et al43 demonstrated an improvement in arterial oxygenation (PSV: 91.4 ± 27.4 mm Hg vs PSV + sigh: 133 ± 42.5 mm Hg, P < .001), end-expiratory lung volume (PSV: 1,242 ± 507 mL vs PSV + sigh: 1,377 ± 484 mL, P < .01), and pulmonary compliance (PSV: 40.2 ± 12.5 mL/cm H2O vs PSV + sigh: 45.1 ± 15.3 mL/cm H2O, P < .01) when a sigh breath (1 breath/min) was applied to PSV in subjects with ARDS. There appear to be fewer and more variable benefits of an augmented breath in pulmonary and extrapulmonary ARDS. In 5 subjects with pulmonary ARDS and 5 subjects with extrapulmonary ARDS, Pelosi et al41 demonstrated significantly lower end-expiratory lung volumes after introduction of a sigh in subjects with pulmonary ARDS (0.16 ± 0.13 L vs 0.69 ± 0.34 L, respectively, P < .01).
There is limited evidence on the frequency or the volume/pressure of the sigh breath that optimizes recruitment with ARDS. However, an inappropriately high frequency of sighs may be associated with ventilator-induced lung injury, similar to that seen with high VT (>12 mL/kg). In a murine model of paraquat-induced acute lung injury, Steimback et al44 demonstrated that a high frequency of sighs (180 breaths/h vs 10 breaths/h) was associated with increased lung and kidney cell apoptosis and type III procollagen mRNA expression.
Variable Ventilation in Non-ARDS Models
The beneficial effects of variable ventilation have been found to extend beyond the acute lung injury model. Mutch et al45 reported improved oxygenation, lower PaCO2, higher dynamic and static compliance, and lower total respiratory system resistance in a porcine methacholine-induced bronchospasm model. The improvement in pulmonary gas exchange and airway resistance is related to the positive effect of noise on both terminal airway opening and alveolar recruitment, both of which can be severely compromised with bronchospasm. One-lung ventilation is commonly performed in thoracic surgery to optimize the operative field. Side effects of one-lung ventilation include a large Q̇S/Q̇T as a result of the controlled lung collapse and atelectasis of the dependent lung units and the gravitational effects of the lateral position on pulmonary blood flow. McMullen et al46 reported improved oxygenation and lower Q̇S/Q̇T and PaCO2 when variable ventilation was applied to the dependent lung in a porcine one-lung ventilation study. The positive impact on pulmonary compliance was seen 60 min after the initiation of one-lung ventilation and was sustained after return to 2-lung ventilation.
Practical Applications for Variable Ventilation
There are no commercially available ventilators capable of performing variable ventilation. This does limit the use of this novel mode of ventilation by the respiratory therapist managing patients requiring controlled mechanical ventilation. However, the application of intermittent sigh breaths offers the opportunity to introduce some limited variability. The sigh breath is set to either 2 or 3 times the baseline VT or to a predefined plateau pressure.40 The sigh breath frequency should be limited to 2–3 breaths/min to facilitate maximum recruitment without inducing volume-related lung injury.44
In the absence of severe ARDS requiring muscle paralysis and controlled mechanical ventilation, rapidly transitioning to patient-triggered or spontaneous modes of ventilation has multiple benefits.47 Both patient-triggered ventilation and spontaneous ventilation facilitate better ventilation/perfusion matching, reduce ventilator-induced lung injury, and preserve diaphragmatic function.48–50
Furthermore, patient-triggered ventilation modes can potentially yield a highly variable respiratory pattern, akin to the experimental variable ventilation studies reported previously. Multiple modes of patient-triggered ventilation are currently available, and the respiratory therapist should utilize the most familiar mode. When a patient-triggered mode of ventilation is instituted, no additional ventilator settings are required to establish variability, as this will be patient-driven.
Proportional assist and neurally adjusted ventilatory assist are 2 modes of patient-triggered ventilation. In the former, the pressure support is proportional to the instantaneous flow and volume requirement.47 In contrast with the neurally adjusted ventilatory assist mode, the level of inspiratory support is related to the electrical inspiratory activity of the diaphragm.47 Both modes of ventilation yield a highly variable VT pattern and can initiate the beneficial effects of noise and stochastic resonance.
Conclusions
The current ventilatory strategy for the management of ARDS is to improve oxygenation and reduce intrapulmonary shunt by recruitment of previously collapsed alveoli. In addition, uninjured lung units must be protected from the deleterious effects of positive-pressure ventilation. Based on the best evidence to date, this involves ventilating with low VT (6 mL/kg of predicted normal body weight) and limiting plateau pressures to < 30 cm H2O to prevent barotrauma while optimizing PEEP to reduce cyclical alveolar opening- and closing-related atelectrauma. There is less clear evidence regarding the optimum type of recruitment maneuver to perform when protective lung ventilation is employed.
Recruitment maneuvers include intermittent sustained inflation breaths, incremental increases in PEEP, prone positioning, and intermittent large VT breaths. The choice of the recruitment maneuver is influenced by the nature of the lung injury (primary vs secondary ARDS), extent of the lung injury, hemodynamic stability, and clinician experience. However, any recruitment maneuvers associated with high inspiratory pressures may result in deleterious cardiorespiratory changes.51
Variable ventilation offers a new physiological approach to lung recruitment without the negative hemodynamic effects from markedly elevated intrathoracic pressures. Although human data for variable ventilation are limited, animal data with variable ventilation in ARDS and non-ARDS models are encouraging. Larger prospective human studies utilizing this novel strategy are needed to confirm the benefits shown by early animal data.