Abstract
BACKGROUND: The placement of nasal or oral gastric tubes is one of the most frequently performed procedures in critically ill children; tube malposition, particularly in the trachea, is an important complication. Neurally adjusted ventilatory assist (NAVA) ventilation (available only on the Servo-i ventilator, Maquet Critical Care, Solna, Sweden) requires a proprietary-design catheter (Maquet Critical Care, Solna, Sweden) with embedded electrodes that detect the electrical activity of the diaphragm (EAdi). The EAdi catheter has the potential benefit of confirming proper positioning of a gastric catheter, based on and the EAdi waveforms.
METHODS: In a case series study, our multidisciplinary team used EAdi guidance for immediate, real-time confirmation of proper nasal or oral gastric tube placement in 20 mechanically ventilated pediatric patients who underwent 23 oral or nasal gastric tube placements. The catheters were placed with our standard practice, with the addition of a team member monitoring the EAdi waveforms. As the tube passes down the esophagus and posterior to the heart, a characteristic EAdi pattern is identified and the position of the atrial signal confirms correct placement of the gastric tube. If the EAdi waveforms indicate incorrect placement, the tube is repositioned until the proper EAdi waveform pattern is obtained. Then proper tube placement is reconfirmed via auscultation over the stomach while air is injected into the catheter, checking the pH of fluid suctioned from the catheter (gastric pH indicates correct positioning), and/or radiograph.
RESULTS: The group's median age was 3 years (range 4 d to 16 y). All 20 patients had successful gastric catheter placement. The EAdi catheter provided characteristic patterns for correctly placed tubes, tubes malpositioned above or below the gastroesophageal junction, and curled tubes. Proper catheter position was confirmed via radiograph and/or gastric pH in all 20 patients.
CONCLUSIONS: EAdi guidance helps confirm proper gastric catheter position, is equivalent to our standard practice for confirming gastric catheter placement, and may reduce the need for radiographs and improve patient safety by avoiding catheter malpositions.
- neurally adjusted ventilatory assist
- NAVA
- pediatric
- nasogastric tube
- enteral feeding tubes
- enteral nutrition
- catheter malposition
Introduction
Placement of a nasal or oral enteral tube, for feeding, medication administration, or gastric decompression, is one of the most frequently performed procedures in critically ill children.1 A key component of tube placement is confirming proper tube position. The incidence of improper tube position during placement is reported to be as high as 20–43.5% in pediatric patients.1–3 Methods to confirm proper tube placement include suctioning fluid from the tube and checking the fluid for gastric pH, auscultation over the stomach while air is injected into the tube, capnometry, and radiograph.4–7
A new mode of mechanical ventilation, neurally adjusted ventilatory assist (NAVA, available only on the Servo-i ventilator, Maquet Critical Care, Solna, Sweden) requires a proprietary-design catheter (Maquet Critical Care, Solna, Sweden) embedded with 9 electrodes that detect the electrical activity of the diaphragm (EAdi). NAVA uses the EAdi signal to deliver assisted ventilation in proportion to and in synchrony with that activity (Fig. 1).8 Recent trials found that NAVA improved patient-ventilator synchrony and reduced airway pressure in mechanically ventilated pediatric patients.9 To properly control NAVA, the EAdi catheter electrodes must be positioned at the level of the crura of the diaphragm. When properly positioned, the EAdi catheter tip lies within the stomach, and the electrodes align within the area of the diaphragm. To confirm proper placement the practitioner relies on the EAdi waveforms on the catheter-positioning screen on the ventilator (see Fig. 1). We studied EAdi guidance as an immediate, real-time confirmation of proper gastric tube placement in mechanically ventilated pediatric patients.
Methods
This study was part of a larger study10 on electrical activity of the diaphragm during extubation-readiness testing of intubated pediatric patients, in a 29-bed pediatric intensive care unit (ICU) in a tertiary-care pediatric hospital. The study was approved by our institutional review board, and written informed consent was obtained from all the patients' parents or guardian. A core group of nurses and respiratory therapists from the ICU had been trained in the proper positioning and maintenance of the EAdi catheter, and only those with this training were involved with the catheter placements.
All the patients were intubated and mechanically ventilated at the time of enrollment. Study patients were transitioned onto the Servo-i ventilator. An appropriate-size EAdi catheter was selected based on the patient's weight and the nurse's preference. Per the manufacturer's recommendation, we estimated the proper catheter insertion distance based on the measured distance from the nose to the ear to the xiphoid process (NEX) with each patient.11 We immersed the catheter's distal end in water for 5 min prior to insertion, to activate the self-lubricant embedded on the catheter. We conducted the NAVA module self-check, which shows 4 EAdi waveforms on the catheter-positioning screen, which correspond to 4 levels near the catheter's distal end. We then attached the proximal end of the tube to the ventilator module and inserted the catheter either orally or nasally, depending on the clinical indication and judgment per our current practice. We used NEX as an estimate of where the tube should be initially placed.
As the EAdi catheter was advanced, the respiratory therapist viewed the EAdi waveforms on the catheter-positioning screen, including the atrial signal as the tube passed posterior to the heart. When the catheter tip reached the proper position in the stomach, the lower EAdi waveforms lost the atrial signal (ie, no P waves) (Fig. 2), and the ventilator reported an EAdi waveform. Proper EAdi catheter placement was defined as: (1) no P waves in the distal waveform, (2) a decrease in overall voltage distally, and (3) correct highlighting (blue on the catheter-positioning screen) of the middle 2 EAdi waveforms. Occasionally a short (< 2 s) expiratory hold was required to encourage diaphragm contraction in over-ventilated patients.
After correct catheter positioning was indicated by the NAVA catheter-positioning screen, each patient had initial confirmation of catheter position according to our unit protocol, which is auscultation over the stomach while air is injected into the catheter, followed by suctioning of fluid from the catheter and checking that fluid's pH (which should be ≤ 5). If no fluid is obtained or the fluid does not have gastric pH, an immediate radiograph is obtained. All the patients had proper tube position reconfirmed via routine chest radiograph within 24 hours of initial tube placement.
Results
Table 1 describes the 20 subjects, who underwent a total of 23 catheter placements. Their median age was 3 y (range 4 d to 16 y). Their mean weight was 25.9 kg (range 3–80 kg). The most common etiologies contributing to the need for mechanical ventilation were acute respiratory distress syndrome and infectious etiologies (14 of the 20 patients). Two patients were neonates with congenital diaphragmatic hernia.
All 20 patients had successful gastric catheter placement with EAdi guidance. Eighteen catheter placements were confirmed via gastric pH, and 5 were confirmed via radiograph because no gastric fluid was obtained. With catheters whose position was initially confirmed via pH, 18 of 18 subsequently had proper position confirmed via routine chest or abdominal radiograph within 24 hours of catheter placement.
In 3 patients the original EAdi catheter was removed during the study period. Two patients had inadvertent catheter displacement, and those catheters were successfully replaced with EAdi guidance. One patient had the catheter electively removed for a magnetic resonance imaging (MRI) study (the EAdi catheter is not compatible with MRI), and the catheter was successfully replaced with EAdi guidance.
In the process of placing the catheters, the EAdi waveforms provided the characteristic patterns for EAdi catheters that were either correct or in malposition above the gastroesophageal junction, below the gastroesophageal junction, or curled. If the distal end of the catheter terminated in the oropharynx or in the esophagus before its passage posterior to the heart, the waveforms showed a complete absence of cardiac electrical activity. If the catheter was advanced further but continued to remain high relative to the optimal position, then P waves persisted in all 4 waveforms (see Fig. 2B). In contrast, if the EAdi catheter was advanced too distally, the P wave was absent from all of the waveforms (see Fig. 2C). If the catheter curled, 2 waveforms appeared identical, with a decreased voltage waveform in between (see Fig. 2D). There were no instances of tracheal catheter placement.
Discussion
In all 20 patients and all 23 catheter placements the catheter was correctly placed, as evidenced by the EAdi waveforms and confirmed via gastric pH or radiograph. No catheters were placed in the trachea. The catheter placement success in this series was probably related to the ability to make real-time adjustments based on the EAdi waveforms.
Methods to confirm proper gastric tube position include radiography, gastric pH of fluid suctioned from the tube, auscultation over the stomach while air is injected into the tube, fluoroscopy, and capnography.12–16 In our pediatric ICU, to confirm gastric placement the nursing staff rely on gastric pH and subsequent radiograph. In a previous study the gastric pH method was only 77% sensitive and 38% specific for proper placement.6 In an observational study of critically ill adults, the gastric pH method provided misinformation regarding tube placement in 64% of cases.17 Methods such as radiography and gastric fluid suctioning have the additional disadvantages of not providing real-time data.
The inadvertent placement of a gastric tube in the trachea contributes most substantially to the morbidities associated with tube misplacement, including instillation of food into the trachea, tracheal perforation, pneumothorax, and pulmonary hemorrhage.18–21 Factors associated with errors in enteral tube placement in children include younger age, decreased level of consciousness, abdominal distention, presence of emesis, and the use of the orogastric route.1 The EAdi waveforms can immediately notify the clinician that the tube is not in the esophagus/stomach and may be in the trachea. In the trachea the catheter would not pass posterior to the right atrium, because the carina is cranial to the right atrium, so no cardiac signal would be detected, which indicates misplacement.
Previous investigators have used ECG for positioning gastric catheters in adults. In critically ill patients, Diaz-Rodriguez et al22 used a standard nasoenteric tube, into which they placed an ECG lead. They placed other leads on the abdomen, and identified a characteristic ECG pattern that confirmed gastric placement in 95 of the 100 patients in the study.22
Another recent study, with 25 adult patients, compared NEX-guided placement to EAdi-guided placement.11 They evaluated the utility of NEX in approximating optimal catheter position with the distal end in the stomach, and in 72% of the patients they obtained correct catheter position with NEX alone.11 With EAdi-guided placement, all the patients had optimal catheter placement. The median difference between the positions obtained with NEX versus EAdi catheter was 2 cm, though that difference ranged from 3 cm too cranial to 12 cm too caudal. Though that study evaluated optimal catheter position to assure proper functioning of NAVA, and did not examine feeding-tube position, the study highlights the considerable variability in catheter positioning with NEX alone. In that adult population, EAdi guidance provided more accurate placement. The differences between the estimated tube-insertion depth and the depth as determined with EAdi guidance in that adult population are potentially of greater relative magnitude in infants and children, because of their smaller size.
We did not compare the initial NEX measurement to the catheter depth obtained with EAdi guidance. With the knowledge that the NEX method involves substantial variability in tube positioning, we used NEX only as a starting point, and subsequently made adjustments based on the EAdi waveforms. A critical component of our study was to determine whether the portion of the feeding tube distal to the electrodes is the proper length to allow the distal end of the tube to rest in the stomach. In our 23 catheter placements we found that, using the proper size tube and the EAdi waveforms, we were able to reliably place the tube in the stomach.
Limitations
To our knowledge, our series is the first to describe the use of EAdi to aid in enteral tube placement in critically ill children. Because this study is a case series, it lacks a control group of patients who received only the usual means of verification of tube placement. We conducted a relatively small number of catheter placements, and a larger number of catheter placements might have included some misplacements. Also, since this series describes the experience at only one center, with a multidisciplinary team well oriented to the use of NAVA and proper EAdi catheter placement, the results at other centers with different team compositions and levels of expertise might differ. In our pediatric ICU we rely on a core group of respiratory therapists and nurses with expertise in placing EAdi catheters. More experience using EAdi catheters in this manner at other centers is necessary before fully understanding the generalizability of our findings. Our series lacked blinding of the practitioners who placed the tubes, which adds the potential for bias. Finally, one major impediment to the general use of the EAdi catheter to verify gastric tube position is the cost of the catheter ($180), which is much more than a regular feeding catheter. In addition, the EAdi catheter requires a Servo-i ventilator equipped with NAVA.
Conclusions
Despite the limitations, we believe this case series highlights a novel method of confirming proper gastric tube placement that is equivalent to standard care in critically ill pediatric patients, and might improve patient safety by avoiding gastric tube malpositions. Further study of its use, including a trial comparing it to the current standard of care, along with an examination of the overall cost of the EAdi catheter compared to the cost of fewer chest radiographs, should be performed before this technology can be recommended as a routine means of confirming gastric tube placement.
Acknowledgments