Abstract
BACKGROUND: Mixing of nebulized drugs is common in real life, but its consequences on aerosol output and granulometry are poorly known.
METHODS: In an in vitro study I evaluated the effects on aerosol output, drug output, and aerosol particle size characteristics of mixing either beclomethasone dipropionate or budesonide with albuterol and ipratropium bromide. I tested the SideStream and VentStream-Pro nebulizers, run with the AirClinic compressor. Using the same fill volume in all experiments, I nebulized and evaluated each studied drug alone, and 2 drug mixtures: beclomethasone plus albuterol plus ipratropium; and budesonide plus albuterol plus ipratropium. I measured aerosol output via gravimetrics. I measured drug delivery by collecting the aerosol on a filter in the inspiratory limb, and the residual solution in the reservoir and the circuit after nebulization, and assayed those liquids with chromatography. I measured particle size distribution via cascade impaction.
RESULTS: Mixing tended to reduce drug output and to increase mass median aerodynamic diameter with the SideStream, but not always with the VentStream-Pro. However, the drug output always remained satisfactory and the mass median aerodynamic diameters were within the respirable range. When nebulized alone, the respirable mass of bronchodilators ranged from 18% to 40% of the nominal dose; when mixed, it ranged from 13% to 37%. When nebulized alone, the respirable mass of corticosteroids ranged from 10% to 24% of the nominal dose; when mixed, it ranged from 10% to 17%.
CONCLUSIONS: Both the SideStream and VentStream-Pro have good aerosol performance in nebulizing budesonide or beclomethasone dipropionate alone, and when mixed with albuterol and ipratropium bromide.
- nebulizer
- drug administration
- inhalation
- aerosol therapy
- inhalation therapy
- COPD
- asthma
- corticosteroids
- beclomethasone dipropionate
- budesonide
- albuterol
- ipratropium bromide
Introduction
Aerosolized bronchodilators and corticosteroids are the mainstay of pharmacologic management for asthma and COPD. The production of therapeutic aerosol requires specific delivery devices, such as inhalers and nebulizers. Although nebulizers are cumbersome and time-consuming, a meta-analysis of randomized controlled studies indicated that they are an appropriate and effective alternative to inhaler.1 Effective delivery of inhaled drugs to infants and elderly patients may be challenging, mainly due to inhaler misuse. Likewise, some elderly patients prefer nebulizer, not only for acute severe episodes of bronchospasm, but for regular long-term home use.2–4 An exhaustive literature review revealed that there is a paucity of randomized controlled studies on nebulization of corticosteroids, as most research trials have employed metered-dose inhalers.1 Additionally, although in vitro experiments can be a surrogate for in vivo investigation of respiratory drug delivery,5 extensive information about nebulized corticosteroids from in vitro studies is still not available. We know only that corticosteroids are usually formulated as suspensions and that the choice of nebulizer system can strongly affect the outcome of aerosol therapy.6–8
Another complication arises when bronchodilators and corticosteroids are simultaneously prescribed for nebulization: combining the drugs in the nebulizer for simultaneous nebulization is not recommended but is often done.9,10 A nationwide study in Italy with 745 patients with chronic lung diseases who regularly used nebulized bronchodilator at home, found that 88% of the patients combined the bronchodilator with the corticosteroid.10 Although no serious incompatibility is known when bronchodilators and corticosteroids are mixed,11–14 to my knowledge no previous study has measured the particle size distribution and aerosol output of corticosteroids when mixed with albuterol and ipratropium bromide (ipratropium).
I chose albuterol and ipratropium because they are widely used short-acting bronchodilators; budesonide because it is a well studied corticosteroid; and beclomethasone dipropionate (beclomethasone) because it has proven clinical value15 and is the most commonly nebulized corticosteroid in Italy.10 I selected the SideStream (Artsana, Grandate, Como, Italy) and VentStream-Pro (Artsana, Grandate, Como, Italy) nebulizers because they are available worldwide and both can be powered by the double-piston AirClinic compressor (Artsana, Grandate, Como, Italy), which is known for high performance. The SideStream has a traditional continuous output, although enriched with an open-vent system through which extra air may be drawn into the nebulizer reservoir; VentStream-Pro is a breath-enhanced nebulizer.
The aim of this study was to evaluate the aerosol consequences of mixing budesonide or becomethasone with albuterol and ipratropium.
Methods
Drugs
Table 1 describes the tested drugs. One combined drug liquid consisted of 0.5 mL of the albuterol formulation, 2 mL of the ipratropium formulation, and 2 mL of the beclomethasone formulation. The other consisted of 0.5 mL of the albuterol formulation, 2 mL of the ipratropium formulation, and 2 mL of the budesonide formulation. The mixtures were prepared in accordance with our usual clinical practice and the product-information sheet. All the experiments used a fill volume of 4.5 mL, in accordance with common practice in Italy.10
Nebulizers
I unsystematically chose 6 SideStream and 6 VentStream-Pro nebulizers from standard commercial lots and water-tested them to determine intra-lot aerosol-output variability. I then selected the 2 units of each brand with the lowest aerosol-output variability at the mid-to-high aerosol-output range. Overall, there was less than 3% aerosol-output variation. Both nebulizers were driven with the double-piston AirClinic compressor. The compressor flow without external resistance was 17 L/min. The compressor flows through the operating SideStream and VentStream-Pro were, respectively, 9 L/min and 7 L/min, and the operating pressures were, respectively, 23 psi and 16 psi.
All experiments were carried out in triplicate, and the reported values are means from 3 experiments. The ambient temperature was 22 ± 2°C and the ambient relative humidity was 45–50%.
Aerosol Output
With each drug and drug combination I gravimetrically measured (balance model E154, Gibertini, Bologna, Italy, precision ± 0.01 mg) the aerosol output. I weighed the nebulizer with the reservoir empty, after charging the reservoir with 4.5 mL of the investigated formulation, and after nebulization, and calculated the aerosol output as the weight difference before versus after nebulization, assuming a liquid density of 1.0 g/mL.
After charging the fill volume, I closed the circuit, connected the vacuum pump to the nebulizer mouthpiece, turned on the pump (operated at 15 L/min), and after 10 seconds switched on the nebulizer compressor. I deemed nebulization complete when no visible aerosol had been produced for 60 seconds, without any shaking or tapping of the nebulizer.
Drug Output
Aerosol was captured on a filter (type A/E glass fiber filter, Pall, Port Washington, New York, lot 72303, which captures > 99% of aerosol particles at 15 L/min) in the inspiratory limb (Fig. 1) and eluted with a device designed for this purpose (DUSA [Dose Uniformity Sampling Apparatus], Copley Scientific, Nottingham, United Kingdom). After charging the nebulizer, I connected the nebulizer to the filter, and the filter to the vacuum pump, turned on the vacuum pump, and after 10 seconds switched on the nebulizer compressor. After nebulization was complete, the aerosol was eluted from the filter, and assayed with high-performance liquid chromatography (HP 1050, Hewlett Packard/Agilent Technologies, Palo Alto, California, with ChemStation REV.A.00.02 software, ChemStation, Houston, Texas) according to a standardized method.16
After nebulization, the residual drug liquid in the nebulizer reservoir and the circuit was also collected, mixed for homogeneity, and assayed. The medication concentrations were calculated based on standard calibration curves prepared with 8 known drug concentrations. The calibration curves for all the studied drugs had a correlation coefficient of r ≥ 0.999. The nominal dose was calculated based on the drug concentration indicated in the product-information sheet.
Particle Size Distribution
Aerosol particle size distribution was measured via cascade impaction, according to the Comité Européen Normalisé standard, prEN 13544-1, “Respiratory Therapy Equipment: Part 1: Nebulizing Systems and Their Components.”17 The nebulizer was held upright in a ring stand and connected with a tube (volume 170 mL) to an 8-stage cascade impactor (Mark II, Andersen Instruments, Atlanta, Georgia) with size cut-offs of 12.4, 8.0, 6.4, 4.5, 2.9, 1.5, 1.0, and 0.6 μm. The vacuum pump was connected to the impactor and run at 15 L/min, and the flow was periodically monitored with a calibrated flow meter, which minimizes temperature and relative humidity variations.18 Thirty seconds after the pump was switched on, the nebulizer compressor was switched on and run until nebulization was complete.
Preliminary experiments indicated that, to prevent sample overload on the impactor stages, the optimal nebulization times were 4 min and 9 min, respectively, when the drugs were nebulized separately and mixed. The impactor throat and each impactor stage plate were eluted with 0.1-M HCL solution to recover all the drug, and assayed with chromatography.16 I analyzed the mass median aerodynamic diameter (MMAD), geometric standard deviation, respirable fraction (percent of aerosol particles smaller than 5 μm), and respirable mass (expressed as a percent of the nominal dose) data with graphics software (KaleidaGraph 4.01, Synergy Software, Reading, Pennsylvania).
Statistical Analysis
I used 2-way analysis of variance to determinate differences between the nebulizers and between the different formulations. For the different formulations with the same nebulizer, and for the different nebulizers with same formulation, I compared aerosol output, nebulization time, drug output, and aerodynamic characteristics. A P value < .05 was considered statistically significant. The analysis was performed with statistics software (Stata 9, StataCorp, College Station, Texas).
Results
Table 2 shows the aerosol output and nebulization time data. The SideStream and VentStream-Pro aerosol output never significantly differed with the same formulation except for ipratropium when nebulized alone. The aerosol output with the 4 formulations that contained corticosteroids was greater than that of the other two (P < .001). Although always significant, a slightly lower difference was observed between albuterol alone and albuterol mixed with ipratropium and budesonide with VentStream-Pro (P = .04). Only the nebulization time of the different nebulizers for albuterol (P = .03) and ipratropium (P = .04) was statistically different.
Table 3 shows the drug output and residual liquid volume data. By assaying the drug captured on the filters and that remained in the nebulizer reservoir and the circuit after nebulization, I calculated the total drug recovery, running a mass-balance test throughout the study; the aerosol loss to the environment was in the range of 89% < total drug recovery < 100% of total drug in all cases. There was no difference between nebulizers in the amount of drug retained in the reservoir. Drug-mixing had a trend, although not always significant, of reduced drug output with the SideStream, whereas the results with VentStream-Pro were more variable.
Table 4 shows the mass median aerodynamic diameter and geometric standard deviation data. Drug-mixing increased MMAD with SideStream, but not with VentStream-Pro. Drug-mixing did not significantly change the geometric standard deviation, except for budesonide with SideStream. However, all the single-drug and mixed-drug nebulizations had heterodisperse aerosols (geometric standard deviation > 1.2 μm).
Figure 2 shows the respirable-fraction data. The only drug that when nebulized alone, versus in combination with other drugs, modified the respirable fraction was beclomethasone with SideStream (P = .02). The respirable-fraction range was 62–78% for the bronchodilators and 35–62% for the corticosteroids. Figure 3 shows the respirable-mass data. Overall, drug-mixing tended to reduce the respirable mass with SideStream, but not always with VentStream-Pro. When nebulized alone, the respirable mass of bronchodilators ranged from 18% to 40% of the nominal dose; when mixed, it ranged from 13% to 37%. When nebulized alone, the respirable mass of corticosteroids ranged from 10% to 24% of the nominal dose; when mixed, it ranged from 10% to 17%.
Discussion
The primary aim of this study was to evaluate whether mixing of either budesonide or beclomethasone with albuterol and ipratropium modified the drug output or aerosol particle size distribution. In the past, aerosol output was evaluated only by weighing the nebulizer before and after nebulization.19 This method is not accurate in measuring drug output. It was known that the gravimetric measurement, independently of volume fill or concentration of the solvent, overestimates the output of albuterol, compared to the filtering method.20 The present results confirm and extend that finding. By gravimetrics I found that the formulations containing corticosteroids, both alone and mixed, had a greater aerosol output, but it was only more solvent, not more drug, that was nebulized, as shown by the filtering-method results. So the bronchodilator and the corticosteroid output ranges, respectively, were 20–52% and 18–46% of the nominal dose.
I used continuous flow in these experiments, which is a study limitation because constant flow does not mimic the varying flow of the in vivo respiratory cycle. The importance of using a simulated breathing pattern rather than continuous flow in the assessment of nebulizer performances was clearly shown with budesonide, even if that difference was less evident with a breath-enhanced nebulizer such as VentStream-Pro.21 Nevertheless, the results for the corticosteroid output were similar to those from the literature using an adult breathing pattern. Nebulizing 2 mL of budesonide (0.25 mg/mL) with the Pari ProNeb Turbo/LC Plus and the PortaNeb/VentStream, Barry and O'Callaghan found drug outputs of 22% and 10% of the initial amount, respectively.22 With the Pari Inhaler Boy, O'Callaghan23 nebulized a lower-concentration brand of beclomethasone than I used in the present study, and found a drug output of about 30% of the initial charge. With the Pari LC Plus, Vaghi et al24 found an output of 27% of the nominal charge with both beclomethasone and budesonide. More importantly, drug-mixing did not severely compromise drug output, which always remained good.
Drug collection on the inspiratory filter represents all aerosol drug delivered to the patient, not just that deposited in the lungs. Only droplets less than about 5.5 μm are generally considered suitable for effective lung delivery. Mixing the drugs tended to increase droplet size, but not always with VentStream-Pro. However, the MMAD of the studied drugs, nebulized alone or mixed, was always acceptable. This finding is in accordance with another recent in vitro study, in which a mixture of beclomethasone and albuterol, nebulized with the BimboNeb and the Nebula nebulizer systems, had mean MMADs of 6.3 μm and 5.3 μm, respectively.25 Akapo et al,14 with the Pari LC Plus (and the Pari ProNeb II Ultra compressor), found that the MMAD of a formoterol-plus-budesonide mixture did not differ from that of single-drug nebulizations. Little other information is available about mixing of corticosteroids with bronchodilators. Berlinski and Waldrep26 observed a slightly smaller MMAD with albuterol alone (2.7 μm), compared to albuterol plus flunisolide (3.4 μm), with the Pari LC Plus. Di Berardino and Scaglione27 reported a substantial reduction in respirable fraction when albuterol was mixed with either beclomethasone or flunisolide, but they performed particle size analysis with laser diffraction, which is unsuitable for suspensions such as corticosteroids, because it cannot distinguish between droplets that contain drug and those constituted only of solvent.
The respirable-mass ranges of the bronchodilators and corticosteroids were 13–40% and 10–24% of the nominal dose, respectively, so although mixing can modify the respirable mass of the studied drugs, the VentStream-Pro and SideStream, driven by the AirClinic compressor, always provided a good respirable mass. It is noteworthy that mixing beclomethasone or budesonide with the bronchodilators did not substantially reduce the respirable mass, as was previously supposed.27 Akapo et al14 found that mixing could increase the respirable mass in single-drug nebulization, but that result was possibly due to the different fill volumes used in those experiments. This problem did not occur in the present study, as I always used the same fill volume. However, with regard to the respirable mass, because significant differences between nebulizers have been observed for the corticosteroid and bronchodilator mixtures, I suggest limiting these findings only to the SideStream and VentStream-Pro used with the AirClinic compressor.
Further in vitro studies should assess whether other nebulizers are suitable to nebulize drug mixtures that contain corticosteroids, and the consequences of drug-mixing on pH, osmolality, turbidity, and other physical and chemical properties. More importantly, further study is needed on whether the present study's in vitro findings are clinically relevant. Until clinical studies are available, physicians should be cautious about the prescription of nebulized mixtures.
Conclusions
The SideStream and VentStream-Pro offer good aerosol performance when nebulizing budesonide or beclomethasone alone, and when mixed with albuterol and ipratropium.
Footnotes
- Correspondence: Andrea S Melani MD, Fisiopatologia e Riabilitazione Respiratoria, Policlinico Le Scotte, Azienda Ospedaliera Universitaria Senese, Viale Bracci I-53100 Siena Italy. E-mail: a.melani{at}ao-siena.toscana.it.
-
Dr Melani has disclosed a relationship with Artsana.