Research ArticleOriginal Contributions
Accuracy of Physiologic Dead Space Measurements in Patients With Acute Respiratory Distress Syndrome Using Volumetric Capnography: Comparison With the Metabolic Monitor Method
Richard H Kallet, Brian M Daniel, Oscar Garcia and Michael A Matthay
Respiratory Care April 2005, 50 (4) 462-467;
Richard H Kallet
Cardiovascular Research Institute, University of California, San Francisco.
Respiratory Care Services, Department of Anesthesia, University of California, San Francisco at San Francisco General Hospital
MSc RRT FAARCBrian M Daniel
Cardiovascular Research Institute, University of California, San Francisco.
Respiratory Care Services, Department of Anesthesia, University of California, San Francisco at Moffitt-Long Hospital
RRTOscar Garcia
Cardiovascular Research Institute, University of California, San Francisco.
Respiratory Care Services, Department of Anesthesia, University of California, San Francisco at Moffitt-Long Hospital
RRTMichael A Matthay
Cardiovascular Research Institute, University of California, San Francisco.
Departments of Medicine and Anesthesia, University of California, San Francisco, San Francisco, California.
MDIn this issue
Respiratory Care
Vol. 50, Issue 4
1 Apr 2005
Accuracy of Physiologic Dead Space Measurements in Patients With Acute Respiratory Distress Syndrome Using Volumetric Capnography: Comparison With the Metabolic Monitor Method
Richard H Kallet, Brian M Daniel, Oscar Garcia, Michael A Matthay
Respiratory Care Apr 2005, 50 (4) 462-467;