Abstract
BACKGROUND: The rapid-shallow-breathing index (RSBI) is widely used to evaluate mechanically ventilated patients for weaning and extubation, but it is determined in different clinical centers in a variety of ways, under conditions that are not always comparable. We hypothesized that the value of RSBI may be significantly influenced by common variations in measurement conditions and technique.
METHODS: Sixty patients eligible for a weaning evaluation after ≥ 72 hours of mechanical ventilation were studied over 15 months in a medical intensive care unit. RSBI was measured while the patients were on 2 different levels of ventilator support: 5 cm H2O continuous positive airway pressure (CPAP) versus T-piece. RSBI was also calculated in 2 different ways: using the values of minute ventilation and respiratory rate provided by the digital output of the ventilator, versus values obtained manually with a Wright spirometer. Finally, RSBI was measured at 2 different times of the day.
RESULTS: RSBI was significantly less when measured on 5 cm H2O CPAP, compared to T-piece: the medians and interquartile ranges were 71 (52-88) breaths/min/L versus 90 (59-137) breaths/min/L, respectively (P < .001). There were no significant differences in the value of RSBI obtained using ventilator-derived versus manual measures of the breathing pattern. RSBI was also not significantly different in the morning versus evening measurements.
CONCLUSIONS: RSBI can be significantly affected by the level of ventilator support, but is relatively unaffected by both the technique used to determine the breathing pattern and the time of day at which it is measured.
Footnotes
- Correspondence: Michael P Young MD, Division of Pulmonary and Critical Care, University of Vermont College of Medicine, Smith 140, FAHC, 111 Colchester Avenue, Burlington VT 05401. E-mail: michael.young{at}vtmednet.org.
This research was partly supported by the American Lung Association of Vermont, and by National Institutes of Health grants R01 HL75593 and NCRR COBRE RR-15557. Trial registry name: University of Vermont College of Medicine. Trial registration number: CHRMS 04-016.
The authors have disclosed no conflicts of interest.
- Copyright © 2009 by Daedalus Enterprises Inc.