Abstract
Deaths and injuries related to accidental ventilator disconnection occur in complex ways. A death related to accidental ventilator disconnection is presented from a subacute ventilator facility, and corrective action is discussed. Single-limb ventilators are now equipped with a flow-bypass coupling that prevents patient-side occlusion during disconnect and therefore prevents low-pressure alarm malfunction. Physicians and respiratory therapists can consider this device to prevent partial or total occlusion of the ventilator tracheostomy adapter, thus allowing a low-pressure alarm in the event of a disconnect.
Footnotes
- Correspondence: William S Lowery MD, Pulmonary/Subacute Service, Alameda Hospital, 2070 Clinton Avenue, Alameda CA 94501. E-mail: slowery{at}pol.net.
Dr Lowery has a patent pending for the device described in the paper.