Abstract
INTRODUCTION: A mass-casualty respiratory failure event where patients exceed available ventilators has spurred several proposed solutions. One proposal is use of a single ventilator to support 4 patients.
METHODS: A ventilator was modified to allow attachment of 4 circuits. Each circuit was connected to one chamber of 2 dual-chambered, test lungs. The ventilator was set at a tidal volume (VT) of 2.0 L, respiratory frequency of 10 breaths/min, and PEEP of 5 cm H2O. Tests were repeated with pressure targeted breaths at 15 cm H2O. Airway pressure, volume, and flow were measured at each chamber. The test lungs were set to simulate 4 patients using combinations of resistance (R) and compliance (C). These included equivalent C and R, constant R and variable C, constant C and variable R, and variable C and variable R. Results. When R and C were equivalent the VT distributed to each chamber of the test lung was similar during both volume (range 428–442 mL) and pressure (range 528–544 mL) breaths. Changing C while R was constant resulted in large variations in delivered VT (volume range 257–621 mL, pressure range 320–762 mL). Changing R while C was constant resulted in a smaller variation in VT (volume range 418–460 mL, pressure range 502–554 mL) compared to only C changes. When R and C were both varied, the range of delivered VT in both volume (336–517 mL) and pressure (417–676 mL) breaths was greater, compared to only R changes.
CONCLUSIONS: Using a single ventilator to support 4 patients is an attractive concept; however, the VT cannot be controlled for each subject and VT disparity is proportional to the variability in compliance. Along with other practical limitations, these findings cannot support the use of this concept for mass-casualty respiratory failure.
Introduction
The threat of pandemic respiratory illness and mass-casualty respiratory failure has resulted in a number of potential solutions, ranging from ventilator stockpiling to triage.1–4 We and others have suggested characteristics for ventilators purchased for stockpiling, which include the ability to support patients with acute respiratory distress syndrome (ARDS). The most recent novel H1N1 flu, emerging from Mexico in early 2010 and resulting in frequent severe ARDS, validates this recommendation.
In 2007, Neyman and Irvin reported the use of one ventilator to support 4 passive, rubber test lungs.5 This paper did not subject the observations to any scientific measurement or critical evaluation. Paladino and colleagues followed that observation with an animal study using 4 sheep with normal lungs.6 In this trial there were significant differences in gas exchange, requiring frequent blood gas evaluations and manipulation of animal position to maintain homeostasis. While each of these investigations have been criticized at length,7,8 support for this idea continues.9
We designed a lung model investigation of the concept of one ventilator for 4 patients, using mechanical lung analogues at varying levels of airway resistance and lung compliance. The goal of the trial was to determine the distribution of tidal volume (VT) and end-expiratory lung volume in identical and varying “patient” characteristics.
QUICK LOOK
Current knowledge
Mass-casualty respiratory failure requires innovative solutions to meet demands of a surge of critically ill patients requiring ventilatory support. The use of a single ventilator to support more than 1 patient has been proposed as a solution.
What this paper contributes to our knowledge
The use of a single ventilator to ventilate more than 1 patient results in wide variations in tidal volume delivery. Variable compliance contributes to the inequity in tidal volume delivery more than resistance; pressure control ventilation exacerbates the disparity compared to volume control ventilation.
Methods
A Puritan-Bennett 840 ventilator was set up according to manufacturer specifications. A modified ventilator circuit, as described by Neyman and Irvin, was connected to the ventilator outlet.5 The outlet of the ventilator led to a T-piece that split the outlet in 2 and then into 2 additional T-pieces, thus creating 4 outlets for gas flow (Fig. 1). At each outlet a standard 60-inch disposable ventilator circuit was connected. The expiratory limb had the identical arrangement of T-pieces to allow a single flow of gas to the expiratory valve. The standard pre-use procedure was performed and all tests passed. The circuit compliance was measured at 9 mL/cm H2O. No humidification was used during the experiment. The ventilator was set at a VT of 2.0 L in the continuous mandatory ventilation mode. The respiratory rate was set at 10 breaths/min and the inspiratory time was 2.0 s, creating an inspiratory-expiratory ratio of 1:2. PEEP was 5 cm H2O and the FIO2 was 0.21. The breath type was then changed to pressure targeted, at a set pressure of 15 cm H2O, with all other settings remaining constant.
Each circuit was connected to a single chamber of 2 separate dual-chamber test lungs (Training and Test Lung, Michigan Instruments, Grand Rapids, Michigan). Compliance and resistance was varied according to Table 1. In this simulation we used endotracheal tubes (ETTs) of varying sizes (6.5–8.5 mm inner diameter) to control airway resistance. These ETTs were connected directly to the inlet of the test lung. There were 4 simulated conditions. In condition 1, ETT size and set lung compliance were equivalent for all 4 lungs. In condition 2, ETT sizes were fixed and compliance was varied. In condition 3, ETT sizes were varied and lung compliance was held fixed. In the final simulation both ETT size and lung compliance were varied.
Between each ventilator circuit and the ETT we placed a fixed orifice pneumotachograph that was connected to a COSMO+ respiratory monitor (Novametrix, Wallingford, Connecticut). Each monitor was connected to a portable computer for continuous recording of airway pressure, volume, and flow. The data were displayed using Analysis+ software (Novametrix, Wallingford, Connecticut) and downloaded to an Excel (Microsoft, Redmond, Washington) spreadsheet for later analysis. We also determined the test lung functional residual capacity (FRC) by monitoring the expired VT when PEEP was turned from 5 cm H2O to 0 cm H2O. The VT reported represent the mean ± SD of 10 breaths, and the FRC measurements represent the mean ± SD of 3 measurements.
Results
The delivered VT to each simulated patient under the 4 conditions of compliance and ETT size are shown in Figure 2 (volume targeted) and Figure 3 (pressure targeted). At each compliance and ETT size a pressure-targeted breath of 15 cm H2O resulted in a larger VT to each of the 4 simulated patients, compared to a set VT of 2.0 L. When the compliance and ETT size were varied, the differences in delivered VT to each simulated patient were exacerbated by the use of pressure targeted breaths. FRC measurements were similar between the breath types within each simulation (Table 2).
In volume targeted ventilation the peak inspiratory pressure was 14 cm H2O for simulations 1, 2, and 4, with the peak inspiratory pressure in simulation 3 being 15 cm H2O. In pressure targeted ventilation the peak inspiratory pressure was 16 cm H2O in each simulation, except in simulation 2 with patient 4, which had peak inspiratory pressures of 15 cm H2O.
Changes in compliance alone resulted in larger differences in delivered VT between “patients” than did changes in ETT size alone (see Figs. 2 and 3). The range of delivered VT in volume targeted ventilation was narrow under equivalent compliance and ETT size (simulation 1: 428– 442 mL) and widest when compliance alone was varied (simulation 2: 257–621 mL). VT variability with changes in ETT size alone was greater than simulation 1 but much smaller than simulation 2 (418–460 mL). Varying both compliance and ETT size predictably resulted in a wide variation in delivered VT (336–517 mL). Similar variability was seen under the 4 simulations with the use of pressure targeted ventilation.
Discussion
A potential ventilator shortage during a mass-casualty respiratory failure event has resulted in a number of real and potential remedies.1–6 While these remedies continue to be pursued, no mass-casualty respiratory failure event with too many patients and too few ventilators has thus far been encountered. The idea of using one ventilator to provide ventilation for more than one patient has been discussed by others.10 An initial report from 1994 attempted to use a single ventilator to care for 2 patients, but allowed completely separate circuits and independent adjustment of PEEP. This system would be superior to the one suggested by Neyman and Irvin.6
The major findings of our study are that, while equivalent ventilation can be provided when compliance and ETT size are identical, differences in compliance can result in a wide variation in delivered VT, and since PEEP is constant, FRC is variable as well. Changes in ETT size alone resulted in variable VT, but to a lesser extent, compared to compliance changes. We also found that when pressure targeted ventilation was used, the discrepancies in delivered VT between simulated patients was exacerbated.
The results of this study were anticipated based on ventilator operation and distribution of ventilation based on regional time constants. However, given the original description that did not provide any scientific measurements5 and the follow-on animal trial,6 this trial is justified to highlight limitations of a single ventilator for 4 patients concept.
The animal trial by Paladino et al used 4 normal, anesthetized sheep and observed marked differences in gas exchange, despite the absence of lung pathology. Both relative hypoxemia and hypercarbia were observed. Arterial blood gas analysis was obtained every hour and animals were re-positioned to provide more uniform ventilation. Inability to provide similar ventilation in 4 normal, anesthetized animals, combined with our findings and the heterogeneity of lung mechanics in patients with acute respiratory failure, would provide little support for the success of this technique in patients. This is further complicated by the importance of delivering VT based on patient height. Previous descriptions assume that all patients are the same height. Even if VT were equally distributed, this may result in inappropriate volume delivery to some patients. This study highlights another glaring limitation of this technique: the monitoring of airway pressure, volume, and flow represents the group, not individual subjects. Therefore, changes in individual subjects can be identified only by external monitoring, including pulse oximetry, stand-alone airway carbon dioxide analyzers, or invasive measurements.
Previously, investigators have designed techniques for ventilators to provide a constant respiratory rate at variable VT and PEEP for independent ventilation of the lungs in unilateral pulmonary pathology.11–13 Such techniques use a separate circuit for both lungs and allow for titration of ventilatory parameters to the individual pathology. Each of these techniques would allow for greater flexibility and reduced risk of cross contamination than the “one ventilator for 4 patients” concept.
Limitations of the concept are numerous. They include inability to monitor subjects individually; inability to provide separate inspired oxygen, VT, and PEEP based on pulmonary pathology; risk of cross contamination in febrile respiratory illness; and the need for staff training in a new methodology. We did not evaluate triggering formally, but noted that application of negative pressure to one of the 4 “patients” resulted in a reduction in the resting FRC of the other 3 “patients” without triggering the ventilator, suggesting sharing of gas from the circuit, thus cross contamination could be a major concern. In a scenario where 4 patients share a ventilator, the respiratory rate is determined by the patient with the greatest demand.
The study has a number of limitations. We tested only mandatory breaths in a single mode. We did not evaluate the effects of ventilator triggering, nor did we assess the issue of infection control, which is of major concern with this technique. We did not alter respiratory rate or inspiratory-expiratory ratio. We do, however, believe that this study demonstrates the multitude of limitations of using one ventilator to ventilate 4 patients simultaneously.
Conclusions
In conclusion, in a lung model evaluation of using a single ventilator to provide ventilatory support for 4 subjects, the data suggest that this technique has important hurdles to overcome. At present, the use of this technique should be avoided, given the potential hazardous and untoward complications.
Acknowledgments
The authors are grateful to Robert Campbell RRT FAARC for providing the additional COSMO+ monitors and for his technical expertise.
Footnotes
- Correspondence: Richard Branson MSc RRT FAARC, Division of Trauma and Critical Care, Department of Surgery, University of Cincinnati, 231 Albert Sabin Way, Cincinnati OH 45267-0558. E-mail: richard.branson{at}uc.edu.
Mr Branson presented a version of this paper at the Annual Congress of the Society of Critical Medicine, held January 9–13, 2010, in Miami, Florida.
Mr Branson has disclosed relationships with Ikaria, General Electric, Covidien, Hamilton Medical, and Breathe Technologies.
The other authors have disclosed no conflicts of interest.
- Copyright © 2012 by Daedalus Enterprises Inc.