Abstract
BACKGROUND: Although the self-inflating bag is widely used in the hospital setting, variability of delivered ventilatory parameters is usually high, which might result in both hypoventilation and lung injury. The aims of this study were to assess possible sources of the high variability and to evaluate the adequacy of obtained values in relation to the recommended values for neonatal resuscitation.
METHODS: This was an experimental study in which 172 health professionals (physicians, resident physicians, physiotherapists, nurses, and nursing technicians) who work with neonatal intensive care manually ventilated a test lung (adjusted to simulate the lungs of an intubated term newborn) with a self-inflating bag in 5 different handling techniques, using 10, 5, 4, 3, and 2 fingers. Delivered values of peak inspiratory pressure (PIP), tidal volume (VT), and ventilatory frequency (f) were compared, taking into account the different handling modalities and professions by analysis of variance for repeated measures. Chi-square, the Friedman test and the Fisher exact tests were performed to compare the delivered and standard values.
RESULTS: PIP and VT were significantly affected by the handling technique, with higher values for a greater number of fingers used for ventilation. Profession also influenced VT and f significantly: physiotherapists tended to deliver higher volumes and lower rates. Nevertheless, we observed high variability of all studied ventilatory parameters and overall inadequacy of obtained values. Most volunteers delivered excessive pressures and volumes at insufficient ventilatory frequency.
CONCLUSIONS: Delivered values seem to depend on operators' individual and professional differences, as well as on the number of fingers used to compress the bag. However, from the clinical point of view, it is important to point out the high occurrence of inadequate delivered values, regardless of handling technique and profession.
- cardiopulmonary resuscitation
- newborn infant
- barotrauma
- positive pressure respiration
- pulmonary ventilation
- tidal volume
Introduction
Self-inflating bags are widely used in the hospital setting, especially for emergency ventilation during cardiopulmonary resuscitation.1–3 When used adequately, positive-pressure ventilation is sufficient to resuscitate almost all of apneic or bradycardic newborns.2 However, it is well known that manual ventilation with self-inflating bags has limitations, and the literature has not yet established whether or not the benefits are worth the risks.2 There is evidence that the ventilatory parameters delivered with this device may be influenced by several factors such as circuit configuration and brand,4,5 the use of 1 or 2 hands to ventilate,4,6–11 and operator's hand size,4,7,12 profession,6,13 and experience.14 This may cause high variability of the absolute values of these parameters,11,15–18 causing delivery of either excessive or insufficient values. In the former case the risk for pulmonary injury (barotrauma and volutrauma) is increased, and in the latter there is a higher risk for hypoventilation.11 Both situations are potentially harmful to the newborn and might lead to a worse prognosis, longer hospital stay, and development of chronic lung disease.
The aim of this study was to evaluate how peak inspiratory pressure (PIP), tidal volume (VT), and ventilatory frequency (f) are affected by the number of fingers used to compress the self-inflating bag (handling technique) and the operator's profession. We also compared delivered values with those recommended for neonatal resuscitation, and examined the influence of profession and handling technique in the adequacy of the delivered values.
QUICK LOOK
Current knowledge
Manual ventilation using a self-inflating bag is a common procedure in neonatal intensive care. The optimum technique and monitoring of manual ventilation is unknown.
What this paper contributes to our knowledge
Regardless of technique or the operator's profession, high airway pressures and high volumes are delivered that were far outside the recommended range. During manual ventilation there is a high variability of delivered tidal volume, peak inspiratory pressure, and respiratory frequency.
Methods
This is an experimental study involving health professionals of 5 different groups (physicians, resident physicians, physiotherapists, nurses, and nursing technicians) who work at the Neonatal Intensive Care Unit of the Centro de Atenção Integral à Saúde da Mulher (Center for Integral Attention to Women's Health, CAISM) of the University of Campinas, São Paulo, Brazil. After reading and signing a written consent form, each subject was instructed to manually ventilate a test lung (Adult/Infant Ventilator Tester model VT-2, Bio-Tek, Winooski, Vermont) set with compliance (0.003 L/cm H2O) and resistance (200 cm H2O/L/min) levels to simulate the lungs of an intubated term newborn of approximately 3 kg. The test lung complies with the requirements of the American Society for Testing and Materials, F920–93.19 A new silicone neonatal self-inflating bag resuscitator (JG Moriya, São Paulo, Brazil) with maximal capacity of 300 mL was used for ventilation. This self-inflating bag was chosen for this study because it is currently the model most commonly used in our hospital. In these experiments, compressed gas source and reservoir bag were not used. The resuscitator is equipped with a pressure-relief valve set to activate at 40 cm H2O (according to the manufacturer), to prevent the administration of excessive pressures.
The trial setup was based on the method reported by Bassani et al11 In summary, all volunteers ventilated manually the test lung with all of the 5 different handling techniques, using 10 fingers (both hands), or only the dominant hand with either 5, 4, 3, or 2 fingers to compress the bag. A 1 min interval was allowed between each test condition. All volunteers were blind to their results during the trials, and the sequence of handling techniques was predetermined with a random number table.20 Delivered values of PIP, VT, and f recorded after approximately 5 ventilatory cycles were provided by each subject.
For each ventilatory parameter, 3 categories (high, adequate, and low) were defined based on standard guidelines for neonatal resuscitation. The ranges of values considered as adequate were: PIP 20–25 cm H2O,3,21 VT 24–30 mL (8–10 mL/kg considering a newborn of approximately 3 kg),22 and f 40–60 compressions/min.23 This study protocol was approved by the institutional review board of the Faculty of Medical Sciences, Universidade Estadual de Campinas, São Paulo, Brazil.
Statistical analysis was performed using software (SAS System for Windows, version 9.1.3, SAS Institute, Cary, North Carolina). The effects of handling technique and profession on PIP, VT, and f were evaluated by analysis of variance for repeated measures, followed by contrast profile and the Tukey post hoc test. Chi-square, the Friedman test, and the Fisher exact test were used to compare delivered and recommended values, considering the different handling techniques and professions. A P value ≤ .05 was considered statistically significant.
Results
One hundred seventy-two volunteers of 5 different professional groups were recruited for this study. Fifteen (8.7%) were attending physicians, 28 (16.3%) were resident physicians in pediatrics or neonatology, 43 (25.0%) were physiotherapists, 30 (17.4%) were nurses, and 56 (32.6%) were nursing technicians.
The handling technique significantly affected PIP (P < .001) and VT (P = .043), but not f (P = .39). Post hoc analysis showed significant increase and decrease of PIP when using 10 fingers and 2 fingers, respectively, compared with the other techniques. Similar results were found for VT, for which the use of 2 hands (10 fingers) produced significant higher volumes, when compared to the use of one hand.
Profession had an isolated influence on VT (P < .001) and f (P < .001) (Table 1). Physiotherapists delivered significantly higher volumes than attending and resident physicians, and nursing technicians (P < .001), although the f values were lower, compared to all other professional groups (P < .001). Significant interaction between profession and handling technique was not observed for any of the ventilatory parameters.
More importantly, a large discrepancy between the delivered and the current guideline values was observed for all parameters. Regardless of profession or handling technique, 155 out of the 172 participants (ie, 88.4%) delivered excessive pressures (> 25 cm H2O), whereas 127 (73.8%) exceeded the recommended range of volume (> 30 mL) (Fig. 1). On the other hand, 85 volunteers (49.4%) ventilated at insufficient rates (< 40 breaths/min) (see Fig. 1). There were no significant effects of the handling techniques on the inadequacy of PIP (P = .55), VT (P = .09), and f (P = .13), regarding the categorized variables. The same was observed with relation to the effects of profession on PIP (P = .98) and VT (P = .051). This indicates that the great majority of participants of all professional groups delivered excessive pressures and volumes, independently of the number of fingers used for ventilation. However, a significant effect of profession was observed for f (P < .001). An insufficient number of delivered breaths was generated by most physiotherapists (81.4%), as opposed to nurses (53.3%), resident physicians (46.4%), and nursing technicians (35.7%), who achieved adequate rates of ventilation in a higher proportion. Among physicians we observed an equal proportion (26.7%) of participants who delivered adequate and low f values.
Despite the high pressures and volumes delivered by most volunteers, we observed a high variability of both ventilatory parameters. PIP ranged from 2.5 to 103.6 cm H2O, and VT from 4 to 96 mL. Ventilatory frequency also showed high variability: 4.3 to 54.05 compressions/min.
Discussion
Over the past 3 decades, evaluation of the self-inflating bag resuscitator has been the goal of a number of published studies, some of which focused on how the resuscitator handling may affect its performance and the potential clinical implications. However, most of these studies were directed to adult patients,4,6–9 and only a few of them addressed this aspect in newborns.11,24 Ganga-Zandzou et al24 and Bassani et al11 demonstrated that delivered peak pressures are higher when more fingers are used to compress the bag. Bassani et al11 observed that this relationship also occurs for VT, as shown in earlier studies that compared the use of 1 or 2 hands to simulate ventilation in adults.4,6–9 These observations are confirmed in the present study with a larger volunteer sample.
Rezende et al evaluated PIP, VT, and f, during ventilation with self-inflating bag in an analogue neonatal lung model,17 as well as in term and preterm newborn sheep.18 Both studies showed wide variability of delivered pressures and volumes, which frequently exceeded standard values for safe ventilation.17,18 The present study shows that the high variability does not seem to be an artifact due to restricted sample size, because it was still observed even with a considerably greater number of volunteers (n = 172). This is in agreement with the results of Roehr et al,25 in which high variability of PIP (2.7–58.9 cm H2O) and VT (1–14.6 mL) was observed in a sample of 120 volunteers during ventilation of a neonatal mannequin with a self-inflating bag.
We observed that approximately 88% and 74% of the participants delivered pressure (> 25 cm H2O) and volume (> 30 mL) values that are excessive for ventilation of a newborn. We also found that nearly half of the participants delivered breaths at insufficient rates (< 40 breaths/min), in contrast with the findings by Rezende at al,17,18 who showed good adequacy for f. This discrepancy is possibly due to differences in the volunteer population. Rezende et al17,18 studied a small group of experienced physicians, whereas we studied a larger sample of subjects from 5 different professions who work with neonatal intensive care, with variable experience in resuscitation.
It is important to point out the possible implications of the present findings, obtained in a heterogeneous sample, especially regarding the high values of PIP and VT delivered by over 75% of the participants, which represent a potential risk for barotrauma and volutrauma, respectively. If the poor adequacy of these isolated variables increases the risk of morbidity and mortality, the combination of excessive ventilatory pressure and volume might be even more harmful. On the other hand, due to the wide variability of the 3 variables, insufficient values of PIP, VT, and f might also be delivered. Low pressures and/or volumes may lead to hypoventilation and, consequently, hypoxemia and hypercapnia.
To decrease the variability of ventilatory parameters delivered by the self-inflating bag, some authors suggest the use of a manometer,26 which was not tested in our trials. The pop-off valve is another safety item designed to reduce delivered pressures and, consequently, the risk of lung injury. However, some studies showed that the valve may be activated at an inconsistent, wide range of pressures,27 frequently exceeding safe values.25,27 In the present study, even though the pop-off valve was kept open, we observed peak pressures over 40 cm H2O (the activation pressure limit indicated by the manufacturer) in 52% of the cases. Kain et al10 observed higher levels of PIP when the pop-off valve was kept closed; however, the authors tested different brands of resuscitators and found high variability in PIP and VT independently of the position of the valve (opened or closed).
Although the physiotherapists in the present study tended to deliver higher volumes at lower rates, most participants of all professions delivered excessive pressures and volumes, as well as insufficient f values for the newborn lung model proposed for this study. While our results are in agreement with some previous reports,6,13 some studies did not find influence of profession on ventilatory parameters delivered by the self-inflating bag.5,11,25 Nevertheless, it is important to stress that it is difficult to compare the results of the several available studies mainly because of differences in the professional profile of the volunteers.
There are limitations to our study. First of all, our tests were conducted in a laboratory environment with a test lung that does not resemble a newborn and does not simulate chest expansion, which is a situation far from the actual clinical setting and may possibly influence volunteers' performance. Additionally, compressed gas source or reservoir bag, which are commonly used for enhancing FIO2, especially during resuscitation, were not used in the tests. There is evidence that the use of compressed gas in some models of self-inflating bags may influence delivered pressures and volumes.28 It has been demonstrated, in T-piece studies, that there is a wide variability of VT due to variable inspiratory time.29 This indicates that lack of monitoring inspiratory time could contribute to wide ranges of delivered VT. This may also happen during manual ventilation with the self-inflating bag and may be considered as another limitation to this study.
An important issue raised in this study is whether or not the self-inflating bag should be used to ventilate a newborn. The answer to this question is quite difficult because, despite the disadvantages that this resuscitator carries regarding patient safety, this equipment is convenient due to its simplicity of operation and transportation, and relatively low cost. Possibly, these advantages contribute to the widespread use of the self-inflating bag in most hospitals all over the world.
Based on the results of this study, one cannot conclude that there is an ideal way of handling the self-inflating bag; however, it is important to stress that this equipment should be used only by well trained personnel, highly aware of its limitations, and in emergency situations.
Conclusions
The self-inflating bag resuscitator permits a wide range/high variability of delivered pressures, volumes, and ventilatory frequencies, frequently insufficient or excessive for neonatal resuscitation. This may contribute to the occurrence of iatrogenic complications such as barotrauma and volutrauma, as well as hypoventilation, which can increase hospitalization time and costs and, consequently, unfavorable prognosis. An important, clinically relevant finding of this study was that, independently of the profession or the number of fingers used to ventilate, in most instances the target value ranges of ventilatory pressure, volume, and frequency were not attained.
Acknowledgments
The authors thank Osvaldo Ribeiro for technical support, Helymar Machado and Dr André Morcillo for statistical analysis, and Dr Rosana Bassani for the valuable suggestions.
Footnotes
- Correspondence: Mariana Almada Bassani RPT MSc, Centro de Atenção Integral à Saúde da Mulher (CAISM), Universidade Estadual de Campinas (UNICAMP), Rua Dr Alexander Fleming, 1101 Campinas, São Paulo 13083970, Brazil. E-mail: bassanimariana{at}hotmail.com.
The authors have disclosed no conflicts of interest.
See the Related Editorial on Page 655
- Copyright © 2012 by Daedalus Enterprises Inc.