Abstract
BACKGROUND: Office spirometers are now widely used to detect obstructive lung diseases. To test the technical characteristics of these devices, simulation of different forced expiratory maneuvers is performed, using computer generated waveforms. However, the tests with human subjects are also useful to detect technical flaws. The procedure used by some authors to test the accuracy of office spirometers is to compare measurements made by 2 spirometers connected in series.
OBJECTIVE: The aim of this study was to evaluate the accuracy of this latter procedure.
METHODS: Two sets of 2 spirometers connected in series were used: the PocketSpiro with the MicroLoop, and the PocketSpiro with the SpiroScout. Different standard American Thoracic Society curves were selected for both ambient temperature and pressure (ATP) and body temperature and pressure saturated (BTPS) conditions and generated with a waveform simulator. We compared lung function indices (FVC, peak expiratory flow, and FEV1) recorded by the PocketSpiro when it was placed respectively upstream or downstream in the assembly.
RESULTS: In ATP conditions, lung function indices were generally higher when the spirometer was placed downstream rather than upstream. The observed deviations reached up to 10%. In BTPS conditions, lung function indices were underestimated when the spirometer was placed downstream, as compared to the ATP procedure. The modification of the flow characteristics and the temperature drop are the 2 mechanisms that could explain our results.
CONCLUSIONS: Connecting the spirometers in series gives variable results, depending on the position of the spirometer in the assembly. Individual tests are therefore essential, as results are not interchangeable.
- spirometry
- instrumentation
- quality control
- COPD
- respiratory function tests
- technology assessment
- benchmarking.
Introduction
Obstructive lung diseases such as COPD are one of the leading causes of mortality in the world.1 The diagnosis of COPD should be confirmed by the measurement of FEV1, FVC, and FEV1/FVC,1 for which a spirometer is needed. Thanks to the improvement of micro-computers, a growing number of spirometers are currently in development and marketed worldwide. Those devices, called handheld spirometers or office spirometers, are relatively cheap, can be connected to a desktop computer, and are thus well suited for general practitioners to perform routine spirometry that was previously the domain of pulmonary specialists.2
To test those office spirometers, manufacturers generally use displacement pump systems to mechanically reproduce the American Thoracic Society (ATS) standard waveforms.3 Indeed, the ATS has published criteria describing adequate spirometer performances and suggested that these parameters should be verified before marketing the equipment.4 However, in order to validate spirometers for human subjects, some studies assessed spirometric accuracy by comparing one small spirometer with a conventional laboratory one, connecting those 2 devices in series.5–9 This procedure avoids inter-maneuver variability and thus reduces the number of subjects and time required for testing to validate a spirometer. The aim of this study was to evaluate the validity of the testing procedure with the spirometers connected in series.
QUICK LOOK
Current knowledge
Office spirometers are widely used to detect obstructive lung disease. Testing the technical characteristics of spirometers is accomplished using simulation of various forced expiratory maneuvers and computer generated waveforms.
What this paper contributes to our knowledge
Comparison of spirometers connected in series is an unacceptable method that gives inaccurate and variable results, depending on the position of the spirometer in the assembly. Individual testing remains the method of choice to evaluate the performance of new spirometers.
Methods
Materials
The study was carried out with a commercially available computer-controlled mechanical syringe (1120 flow/volume generator, Hans Rudolph, Shawnee, Kansas). According to the manufacturer, this device is able to generate flows up to 16 L/s and volumes up to 8.5 L. It can perform any kind of previously sampled and recorded flow-time curve downloaded in the waveform editor of the Hans-Rudolph simulator.
The flow-volume simulator was connected to the spirometers using a rubber connector of a total length of 100 mm. However, once connected, the distance between the simulator and the spirometers was only 55 mm. The internal and external diameters of the outlet of the simulator were 30 mm and 45 mm, respectively.
Three different spirometers, presented in Table 1, were tested. They all comply with the recommendations of the ATS, according to their manufacturers.3 If connected to a desktop computer provided with the appropriate software (see Table 1), these devices make spirometry readily accessible to the general practitioner. They measure instantaneous flows that are integrated electronically into volumes, expressed in BTPS conditions. Only the SpiroScout and the MicroLoop require a volume calibration involving calibration checks or recalibration factors.
Protocol
A total of 4 combinations of 2 spirometers each were assembled: the PocketSpiro (placed upstream) with the MicroLoop (placed downstream) and vice-versa, and the PocketSpiro (placed upstream) with the SpiroScout (placed downstream) and vice versa (Figure). To connect the spirometers in series, we used 4 different cone-shaped rubber connectors. The connectors were thin (1.2 mm) and they were positioned inside the outlet of the upstream spirometer and outside the mouthpiece (or inlet) of the downstream spirometer.
The length of the connectors was 50 mm. The internal diameters of the outlets and external diameters of the inlets of the spirometers were as follows:
A: 24 mm internal diameter → 30.5 mm external diameter
B: 24 mm internal diameter → 22 mm external diameter
C: 33 mm internal diameter → 27 mm external diameter
D: 19.5 mm internal diameter → 27 mm external diameter
The letters A through D correspond to the settings described in Figure 1.
Three of the 24 ATS standard volume-time curves, numbers 1, 17, and 23, and one of the 26 ATS flow-time curves, number 25, were selected according to differences in shape and complexity.3
Each combination of spirometers was submitted to the 4 curves, each curve being delivered by the Hans Rudolph simulator 3 times, under both ambient temperature and pressure (ATP), and body temperature and ambient pressure saturated with water vapor (BTPS) conditions. To simulate real biological conditions as much as possible, we used an external humidifier to obtain 100% relative humidity, and the heater included as an option in the lung simulator to warm up the air to 37°C.
The manufacturer's specifications and instructions were followed with the highest accuracy, in particular concerning the handling of the spirometers and their need for calibration checks before each series of tests.
Statistical Analysis
Three clinically relevant parameters were analyzed: peak expiratory flow (PEF), FVC, and FEV1. Since the aim of the study was to compare the measurements made by the same spirometer according to its position in the assembly, the absolute accuracy of the spirometers was not specifically assessed. For each assembly and each of the 2 test conditions we measured the deviation between the value given by a spirometer placed upstream and the value reported by the same spirometer placed downstream, using the following equation:
Results
Table 2 presents the observed deviations expressed in percentage according to the equation. A negative deviation means that the value was higher when the spirometer was placed downstream rather than upstream in the assembly. We observed deviations up to 10% when testing under ATP or BTPS conditions for FVC, and up to 4% for FEV1 in ATP conditions. In BTPS conditions we found, mainly for the assembly PocketSpiro-MicroLoop, that the measurements of the spirometer placed upstream were lower, as compared to the ATP procedure.
Particularly in the case of the SpiroScout placed in series with the PocketSpiro, the deviations were often positive and higher for the measurement of FVC. Figure 1 shows the values of FEV1 measured by the PocketSpiro when it was connected in series with the 2 other devices using curve number 1 of the standard volume-time curves (target FEV1 4.262 L). There was a substantial difference in FEV1 when the PockectSpiro was placed after the MicroLoop.
Discussion
Our data show that the position of the spirometer connected in series influences the results generated by the spirometer, with differences in magnitude that under certain circumstances and for some of the tested spirometers ranged between 5% and 10%. As shown in Table 2, these differences occasionally exceeded the recommended between-maneuver repeatability criteria for FEV1 and FVC, which should remain below 150 mL for FEV1 and FVC.4
It seems difficult to identify a general trend in the deviations observed, which do not appear to be related to one particular curve, one of the spirometric variables, or the degree of humidity or temperature. Possibly a number of parameters, such as resistance, dead space volume, and curve dynamics, may affect the FEV1, FVC, or PEF value generated by the spirometer. Likewise, the deviations observed in ATP conditions cannot be explained by the drop in temperature and humidity, as could be expected in BTPS conditions. We thus tend to attribute the observed deviations to the change in flow characteristics produced by the connection in series of 2 spirometers. One possible explanation could be that the geometry of the upstream spirometer disrupts the flow in the downstream spirometer, which might then become turbulent. Under these circumstances, the linearization table of the device is no longer adequate and the measurements are distorted.
Surprisingly, Barr et al previously reported that connecting spirometers in series gave acceptable results.6 However, careful analysis of the Bland-Altman plot of the FVC published in that paper showed a clear proportional error, the difference between the EasyOne spirometer and the reference spirometer became increasingly more prominent with higher FVC. These authors were unable to reproduce that proportional error if the tests were not performed in series, but in standard conditions.
We also observed that the deviations obtained for the PocketSpiro were more pronounced when it was in series with the MicroLoop than with the SpiroScout (see Table 2 and Fig. 1). This finding can certainly be ascribed to the characteristics of some of the spirometers under investigation. Indeed, the MicroLoop uses a turbine (moving parts) in the flow sensor, whereas the SpiroScout uses an ultrasonic sensor (no moving parts) and therefore a lower resistance. Such differences may largely explain why the flow downstream from the SpiroScout was only slightly disturbed, compared to the MicroLoop.
We also found important differences between data obtained in BTPS and ATP conditions for the PocketSpiro-MicroLoop assembly. This was particularly clear for spirometer placed upstream. This could possibly result from the drop in actual temperature between the simulator and the second spirometer, which could be larger than the expected value (3–4°C) programmed in the software of that spirometer. The lower observed volume might thus be the consequence of an underestimated expiratory correction factor and an insufficient conversion of the measured volume into BTPS conditions.
Conclusions
We conclude that comparison of spirometers connected in series is an inappropriate and unacceptable method that gives inaccurate and even variable results, depending on the position of the spirometer under investigation in the assembly. Since the results are not interchangeable, when the devices are tested in human subjects, individual spirometer testing remains the method of choice to evaluate performance of new spirometers.
Acknowledgments
We would like to thank Jean-Yves Moens, MEC (Medical Electronic Construction), Brussels, Belgium, for allowing us to use his Hans Rudolph flow/volume simulator to conduct our tests.
Footnotes
- Correspondence: Giuseppe Liistro MD PhD, Pneumology Department, Cliniques Universitaires Saint Luc, 10 Avenue Hippocrate, 1200 Brussels, Belgium. E-mail: giuseppe.liistro{at}uclouvain.be.
The authors have disclosed a relationship with MEC (Medical Electronic Construction) Group.
- Copyright © 2012 by Daedalus Enterprises Inc.