In reply:
The study by Weaver and colleagues1 demonstrated that, while the RAD-57 CO-oximeter operated within the manufacturer's specifications, with 68% of SpCO measurements falling within ± 3% of the laboratory carboxyhemoglobin (COHb) measurements of < 40%, in several cases the SpCO reported by the RAD-57 underestimated the COHb. This raises concerns about the utility of the RAD-57 in identifying cases of occult CO poisoning, which is one of the primary potential benefits of a point-of-care noninvasive carboxyhemoglobin screening test. Furthermore, Weaver et al's findings were consistent with other studies of the RAD-57.2–4
In response to Weaver et al's study, Dr McEvoy opines that these failures of the RAD-57 to report a SpCO consistent with the laboratory COHb measurement could have been due to technician technique. The concern has previously been raised by another industry representative,5 in response to a prospective study that demonstrated wide limits of agreement and poor sensitivity of the RAD-57.2 While technique may have been a contributor to the discrepancy between SpCO and COHb in Weaver et al's study, this does not excuse the failure of the RAD-57 to identify elevated COHb levels. The use of any medical device is not isolated from user technique or user error, and dismissing false negative results described by Weaver et al and others as being due to poor technique ignores the potential consequences of broadening the clinical use of the RAD-57. If false negative values were obtained under relatively idealized settings (technicians were trained by industry representatives and were obtaining measurements in the setting of a research study), it is reasonable to assume that the rate of false negatives will not be lower in the non-idealized setting of real world clinical medicine, where attention to technique may be less meticulous than in a research study.
Given these considerations, the poor sensitivity, rate of false negatives, and the inaccuracy of the RAD-57 should be a warning to medical personnel that SpCO is not definitive, and that a normal SpCO should not be reassuring. Ultimately, the false negatives obtained by the RAD-57, whether due to technician technique, intrinsic device inaccuracies, or patient-level factors, demonstrates that the RAD-57 is not suitable as a screening device and that there is potential for measurement inaccuracies and patient harm in real world clinical settings.
We agree that further work to develop an accurate, precise, user-friendly, and noninvasive SpCO monitor is warranted, and that a rapid, accurate, point-of-care carbon monoxide monitor would be extremely valuable. The RAD-57 monitor, however, does not meet these criteria, based on the available clinical data, and there is insufficient evidence for its broad clinical use.1–4
Footnotes
The authors have disclosed no conflicts of interest.
- Copyright © 2013 by Daedalus Enterprises