Abstract
BACKGROUND: Although self-inflating bags are widely used for manual hyperinflation, they do not allow ventilation parameters, such as pressure or volume, to be set. We studied the ventilation performance of neonatal and pediatric self-inflating bags.
METHODS: We asked 22 physiotherapists to manually hyperinflate 2 lung models (neonatal and pediatric), using self-inflating bags from 3 manufactures (Hudson, Laerdal, and JG Moriya), with flows of 0, 5, 10, and 15 L/min. A pneumotachograph recorded tidal volume (VT), peak inspiratory pressure (PIP), peak inspiratory flow (PIF), peak expiratory flow (PEF), and inspiratory time.
RESULTS: The VT, PIP, and inspiratory time delivered by the Hudson, Laerdal, and JG Moriya bags, in both neonatal and pediatric self-inflating bags, were significantly different (P < .001). The PEF and PIF delivered were different only when using the neonatal self-inflating bags (P < .001). The VT, PIP, and PIF delivered with a flow of 0 L/min were lower than with 15 L/min (P < .05) with all the tested bags, in both the neonatal and pediatric sizes.
CONCLUSIONS: The performance of the tested neonatal and pediatric bags varied by manufacturer and oxygen flow. There was an increase in VT, PIP, and PIF related to the increase of oxygen flow from 0 L/min to 15 L/min. The neonatal bags showed higher ventilation parameters variation than the pediatric bags.
- resuscitation
- manual hyperinflation
- self-inflating bag
- pediatric
- respiratory therapy
- respiratory function monitor
Introduction
Manual hyperinflation is used to improve secretion clearance and alveolar expansion in mechanically ventilated patients, and the self-inflating bag is the main device used in manual hyperinflation, but the self-inflating bag does not allow ventilation parameters, such as pressure, volume, or FIO2, to be set.1–3
The American Society for Testing and Materials established standards4 for self-inflating bags. However, important physical and functional differences can be observed during cardiopulmonary resuscitation.5–13 These differences are among neonatal devices from different manufacturers. Several studies have confirmed that several factors might interfere with the performance of neonatal pulmonary ventilation.5,14,15 The factors are professional experience and training,12,16,17 lung compliance,18 visual feedback, pressure manometer,19 number of hands used in bag compression,20,21 and the operator's hand size.22 All these factors can affect the ventilation parameters of manual hyperinflation. Such variance may impact clinical outcomes.
Few studies have evaluated the functional performance of neonatal self-inflating bags according to oxygen flows, but they assessed only the FIO2.13,23,24 There have been no studies of the relationship between self-inflating bags' ventilation parameters with different oxygen flows applied. We studied the performance of neonatal and pediatric self-inflating bags from 3 manufacturers, at different oxygen flows, during manual hyperinflation.
QUICK LOOK
Current knowledge
Manual hyperinflation is used to improve secretion clearance and alveolar expansion in mechanically ventilated patients. The self-inflating bag is the primary device used in manual hyperinflation, but bagging does not allow good control of ventilation parameters such as pressure, volume, or FIO2.
What this paper contributes to our knowledge
The performance of neonatal and pediatric bags varied during manual hyperinflation, by manufacturer and by the oxygen flow applied. There was an increase in tidal volume, peak inspiratory pressure, peak inspiratory flow, and inspiratory time related to the oxygen flow, from 0 to 15 L/min. The neonatal bags showed greater ventilatory parameter variation than the pediatric bags.
Methods
Design and Ethics
The study had a randomized crossover design, and was approved by the University of Campinas ethics committee (408/2008).
Setting
The study was conducted at the Biomedical Engineering Center of the University of Campinas. We recruited 22 physiotherapists at the Hospital de Clínicas, Universidade Estadual de Campinas, Campinas, São Paulo, Brazil. Eleven of the physiotherapists recruited had clinical experience with bagging critically ill neonates and children, and 11 did not. No prior training was conducted with the subjects. They were instructed to manually hyperinflate with slow inflation and rapid inspiratory pause, followed by fast bag release.
The manual hyperinflation was simulated to set up the respiratory mechanics of a neonate and a pediatric intubated patient in a test lung model (Ventilator Tester 2, Biotek, Winooski, Vermont) (Fig. 1). We used a leak-free and intubated model. Measurements were recorded at different values of pulmonary compliance and resistance, to simulate 2 clinical situations: healthy, normal respiratory mechanics, and restrictive mechanics (decreased pulmonary compliance). The lung resistance was set at 50 cm H2O/L/s to simulate a newborn, and at 20 cm H2O/L/s to simulate a child. The pulmonary compliance was set at 3 m L/cm H2O to simulate normal compliance in a healthy newborn, at 1 m L/cm H2O to simulate reduced lung compliance in a newborn, at 10 m L/cm H2O to simulate normal compliance in a healthy child, and at 3 m L/cm H2O to simulate reduced lung compliance in a child. These compliance and resistance values were based on a previous study.20 The test lung apparatus was calibrated considering the environmental temperature, atmospheric pressure, and relative air humidity before the experiments.
The subjects used only their dominant hand. They were encouraged to ventilate the test lung as they would ventilate a patient, and to rest between tests to avoid fatigue. They were allowed to familiarize themselves with the equipment and environment. The tests were randomized. No visual or verbal feedback was provided during the tests.
After the tests, the subjects underwent a standardized interview about the bags' performance.
Self-Inflating Bags
Each subject used 6 new self-inflating bag units. The tested bags were obtained from 3 manufacturers: Hudson RCI/Teleflex Medical (Research Triangle Park, North Carolina), Laerdal Medical (Stavanger, Norway), and JG Moriya (São Paulo, Brazil) (Fig. 2). All the bags were tested with both the neonatal and pediatric models. The bags were connected to a 50-psig source of 100% oxygen, and flows of 0, 5, 10, and 15 L/min were delivered. An oxygen reservoir was attached to the unit when the oxygen flow meter was set above 5 L/min. The pressure relief valve was kept unlocked.
The volumes of the bags tested were: Hudson neonatal 280 mL, Hudson pediatric 500 mL, Laerdal neonatal 240 mL, Laerdal pediatric 500 mL, JG Moriya neonatal 130 mL, and JG Moriya pediatric 250 mL.
The bias was reduced by not telling the subjects what was being investigated. The order of the bag handling, compliance settings, and experience levels of the subjects were randomly assigned. The oxygen flows were used sequentially: 0, 5, 10, and 15 L/min, but the physiotherapists were unaware of this assignment.
Each physiotherapist performed 10 manual hyperinflations with each of the 6 bags (neonatal bag and pediatric bag from 3 manufacturers), with both normal and reduced compliance, and at all 4 flows.
Pneumotachograph
A sensor CO2 Capnostat (Novametrix, Wallingford, Connecticut) attached to the pneumotachograph (CO2SMO+, Novametrix, Wallingford, Connecticut) was fitted at the interface between the test lung and the bag, with no important increase in dead space. The signals were recorded by software (Analysis Plus, Novametrix, Wallingford, Connecticut). The physiotherapists were blinded to the pneumotachograph display.
The variables measured were tidal volume (VT), peak inspiratory pressure (PIP), peak expiratory flow (PEF), peak inspiratory flow (PIF), and inspiratory time.
Statistical Analysis
Statistical analysis was performed with statistics software (SAS 9.1.3, SAS Institute, Cary, North Carolina). The Kolmogorov-Smirnov test was used to assess data distribution, and the data did not have normal distribution, so all the variables were transformed into ranks. We calculated the mean of 10 inflations per test.
The ventilation measurements were compared for the different bag manufacturers and flows with repeated-measures analysis of variance. Then we applied within-subject contrast tests for post hoc analysis. There was no important period effect because the study was in an experimental setting. The results are reported as median and 95% CI. The significance level for the statistical analysis was P < .05.
Results
VT, PIP, and inspiratory time delivered by the Hudson, Laerdal, and JG Moriya bags, with both the neonatal and pediatric bags, were statistically different (P < .001). PEF and PIF were different only with the neonatal bags. These data were obtained independently of the flow applied (Table 1). Also, Table 1 presents the within-subjects contrasts.
The PIP delivered ranged from 0 to 46 cm H2O in 528 tests with the neonatal bags. Only one subject provided PIP higher than 40 cm H2O, with the Hudson neonatal bag. With the pediatric bags the PIP ranged from 12 to 52 cm H2O, and 16 of 22 subjects exceeded 40 cm H2O. All these measurements were obtained with the Hudson pediatric bag.
A minimum PIP of 20 cm H2O was not reached in 121 (22.92%) tests with the neonatal bags, and in 116 (21.97%) tests with the pediatric bags. Eighteen physiotherapists failed to deliver PIP greater than 20 cm H2O with the neonatal bags, and 19 with the pediatric bags. The PIP values are presented in Table 2.
VT, PIP, PIF, and inspiratory time increased with increasing oxygen flow (Fig. 3) and were significantly different at flows of 0 and 15 L/min (P < .001). This result was obtained independently of the bag's manufacturer.
The experienced and inexperienced physiotherapists were similar in their overall manual hyperinflation performance; the only difference was the highest PIF in the experienced group (P = .03 for neonatal bags, P = .03 for pediatric bags). The ventilation parameters were different between the normal and reduced compliance settings (P < .001). The subjects delivered a lower VT, a higher PIP, and a higher PIF in the low compliance setting than in the normal compliance setting (all P < .001). The comparison of ventilation parameters delivered by experience level and compliance was previously published.16
Discussion
To our knowledge, this is the first study to measure the influence of oxygen flow on physiotherapist performance with 3 neonatal and pediatric bags. We found functional and physical differences between the 3 tested bag models. These differences have been documented during neonatal resuscitation.1,7,25 In addition, during neonatal and pediatric manual hyperventilation, VT, PIP, and PIF increased with increased oxygen flow. A few studies of neonatal bag performance have found that FIO2 varies with the oxygen flow applied, but those studies did not measure VT, PIP, or PIF.13,23,24
In clinical practice the choice of oxygen flow varies according to the patient's oxygenation. However, the North American Neonatal Resuscitation Program recommends a flow of 5–10 L/min into the bag inlet when it is connected to a 100% oxygen source.26 Often the manufacturers' recommendations are unknown or not respected. The lack of difference in some ventilation parameters between the flows of 5 and 10 L/min demonstrate the safety of using this flow range in daily practice.
The heterogeneous PIPs we found in the present study are in agreement with other studies.9,27 The pressure-relief valve is supposed to open at 35 cm H2O with the Laerdal bag, and at 40 cm H2O with the JG Moriya and Hudson bags we tested; our measured PIPs often exceeded 40 cm H2O with the pediatric bags. This fact is even more important during preterm-infant ventilation, which requires more accurate VT and PIP. Furthermore, it is intriguing to notice how often the PIP cutoff of 20 cm H2O was not reached. This creates a risk of hypoventilation and inadequate alveolar recruitment during manual hyperinflation.
The bags we tested were suitable for each age group; however, the JG Moriya provided lower VT and PIP values than the bags from the other manufacturers. The JG Moriya neonatal bag's PIF was also the lowest among the neonatal bags, so the JG Moriya neonatal bag presumably provides less ventilation. Regardless of international consensus defining the self-inflating bag as the main instrument for manual ventilation, studies have shown that there is no unanimity in the guidelines for self-inflating bag use in neonatal ventilation.6,8,14 Lee et al28 reported that the volumes delivered range widely, and they do not encourage self-inflating bag use for careful and precise ventilation.
The ventilation parameter variation during neonatal manual hyperinflation could be explained by the fact that the neonatal bag (150 mL) is pressurized faster and also distended, when compared to the pediatric bags. This results in the higher pressures and volumes delivered by neonatal bags. Furthermore, an increased plateau pressure due to the low adjusted compliance at the infant setting could reflect in an increased PIP and VT.
There is a wide range of commercial neonatal and pediatric bags manufactured. The differences in the ventilation performance between those bags could be explained mainly by their physical characteristics, including bag and reservoir volume, valve type, valve resistance,29 manufacturing materials, elasticity, density, and bag shape and texture.30 To select the appropriate bag for each clinical setting, the clinician should be familiar with the equipment and check it before use. The choice of an inadequate bag for the patient's weight or the underlying disease could be dangerous.
The majority of our subjects (77%) reported increased resistance squeezing the JG Moriya bag, compared to the Laerdal and Hudson bags. The JG Moriya bag's silicone percentage could explain that higher resistance. Although the devices have similar bag formats, the subjects also described differences in the bags' textures. The Laerdal bag has a shape and texture that provide greater adherence between the bag and the clinician's hand. Mazzolini and Marshall30 indicated that differences in bag design and texture affect the FIO2 delivered by the bag, and that bag shape influences the gripping efficiency and, consequently, the ventilation performance. Bag texture and design can increase the clinician's capacity to perform inflations and prevent the bag from slipping when handled. The bag's manufacturing material can also affect the inflation pressure.18 Therefore, the clinician's ability to sense compliance changes could interfere with the ventilation parameters.
Our results were based on an experimental setting. The subjects could not be blinded to the bag used, and we were not able to test the bag's manufacturing materials. Further studies testing different bag manufacturers and evaluating pressures and volumes could contribute to the current results' application. To determine the optimal gas flow for safe manual hyperinflation will require clinical studies.
The self-inflating bag's use for manual ventilation still presents several advantages over other devices. It does not require a continuous gas flow or a power source to be connected,5 and bag is portable and easy to use.9,31 Based on our results, we recommend that the bag manufacturer and flow applied should be chosen carefully to avoid adverse impacts on clinical outcomes. Although there are several guidelines for resuscitation, recommendations about flow during manual ventilation from the European Respiratory Society or American Thoracic Society were not found.
In agreement with Jones et al,31 the present study suggests that bag selection should match the clinician's experience and skills. It is necessary to learn about a bag's physical and functional characteristics and to check the bag before using it.32 Until manual hyperinflation clinical evidence is available, we suggest that the clinician use a pneumotachograph or at least a manometer to ensure safe ventilation. Bag use by untrained hands can be dangerous.16
Conclusions
The performance of neonatal and pediatric bags varied during manual hyperinflation, by manufacturer and by oxygen flow applied. VT, PIP, PIF, and inspiratory time increased with increased oxygen flow. The neonatal bags had higher ventilation parameter variation than the pediatric bags. Our findings should alert clinicians about differences in manual hyperinflation ventilation parameters, in order to conduct safe manual ventilation.