Abstract
BACKGROUND: Heat and moisture exchangers (HMEs) are commonly used for humidifying respiratory gases administered to mechanically ventilated patients. While they are also applied to tracheostomized patients with spontaneous breathing, their performance in this role has not yet been clarified. We carried out a bench study to investigate the effects of spontaneous breathing parameters and oxygen flow on the humidification performance of 11 HMEs.
METHODS: We evaluated the humidification provided by 11 HMEs for tracheostomized patients, and also by a system delivering high-flow CPAP, and an oxygen mask with nebulizer heater. Spontaneous breathing was simulated with a mechanical ventilator, lung model, and servo-controlled heated humidifier at tidal volumes of 300, 500, and 700 mL, and breathing frequencies of 10 and 20 breaths/min. Expired gas was warmed to 37°C. The high-flow CPAP system was set to deliver 15, 30, and 45 L/min. With the 8 HMEs that were equipped with ports to deliver oxygen, and with the high-flow CPAP system, measurements were taken when delivering 0 and 3 L/min of dry oxygen. After stabilization we measured the absolute humidity (AH) of inspired gas with a hygrometer.
RESULTS: AH differed among HMEs applied to tracheostomized patients with spontaneous breathing. For all the HMEs, as tidal volume increased, AH decreased. At 20 breaths/min, AH was higher than at 10 breaths/min. For all the HMEs, when oxygen was delivered, AH decreased to below 30 mg/L. With an oxygen mask and high-flow CPAP, at all settings, AH exceeded 30 mg/L.
CONCLUSIONS: None of the HMEs provided adequate humidification when supplemental oxygen was added. In the ICU, caution is required when applying HME to tracheostomized patients with spontaneous breathing, especially when supplemental oxygen is required.
- heat and moisture exchanger
- HME
- humidification
- respiratory
- mechanical ventilation
- tracheostomy
- spontaneous breathing
- supplemental oxygen
- CPAP
Introduction
During everyday spontaneous breathing, inspiratory gases are normally heated and humidified in the nasal cavity and pharynx: by the time the second bronchial bifurcation is reached, inspired gas temperature is 37°C and absolute humidity (AH) is 44 mg/L.1 In mechanically ventilated patients, because the anatomy that provides this natural conditioning is bypassed by an artificial airway, administered gases require artificial conditioning. If humidification is inadequate, inspissation of airway secretions, destruction of airway epithelium, and hypothermia2 may occur. The American Association for Respiratory Care recommends that inspiratory gas temperature should be > 30°C and AH more than 30 mg/L.3 Heat and moisture exchangers (HMEs) and heated humidifiers are commonly employed to warm and humidify medical gases, and are widely used with adult patients undergoing mechanical ventilation.4 Their performance has been thoroughly investigated, and the results suggest that performance depends on brand,5 location in the ventilator circuit, tidal volume (VT), and minute volume.6,7
HMEs are also applied to tracheostomized patients who are able to breathe spontaneously. The configuration of HMEs for tracheostomized patients with spontaneous breathing differs from applications for mechanically ventilated patients: some of the vapor in expired gas escapes to the atmosphere without being trapped by the HME, and room air that has bypassed the HME is inspired. Some HMEs have a port for delivering oxygen to hypoxemic patients, and administration of dry, cold oxygen may further decrease the temperature and AH of inspired gas. While the moisturizing ability of HMEs for daily use of active patients has been thoroughly evaluated,8,9 we found no published results showing how well HMEs provide humidification to patients in ICUs. Consequently, for spontaneously breathing tracheostomized patients we studied the effects of spontaneous breathing parameters and oxygen flow on the performance of humidifiers.
QUICK LOOK
Current knowledge
Humidification of inspired gases following tracheostomy can be accomplished with heated humidification or a heat and moisture exchanger designed for tracheostomy. The performance of these heat and moisture exchangers is variable and can be altered by air leaks or the addition of oxygen.
What this paper contributes to our knowledge
In a model of spontaneous breathing, increasing the tidal volume and adding supplemental oxygen reduced the delivered humidity. At an oxygen flow of 3 L/min the humidity fell below the minimum of 30 mg H2O/L recommended by the American Association for Respiratory Care.
Methods
We tested 11 models of HME, all commercially available in Japan, intended for use with spontaneously breathing tracheostomized patients. In addition we tested 2 other systems: an oxygen mask (Tracheomask, Intermed Japan, Tokyo, Japan) with a nebulizer heater (Aquatherm III, Hudson RCI, Durham, North Carolina), and a high-flow CPAP system (Airvo, Fisher & Paykel, Auckland, New Zealand) for tracheostomized patients (Table 1). Using a mechanical ventilator (Puritan-Bennett 840, Covidien, Carlsbad, California), a lung model (TTL 1601, Michigan Instruments, Grand Rapids, Michigan), and a heated humidifier (MR730, Fisher & Paykel, Auckland, New Zealand), we simulated real-life spontaneous breathing (Fig. 1). To simulate spontaneous breathing, the muscle and lung compartments of the lung model were connected, then the muscle compartment was inflated by the Puritan-Bennett 840 and the lung compartment inspired ambient air through an HME. Expired gas from the lung compartment went through a servo-controlled heated humidifier, and was warmed and humidified to 37°C and 100% relative humidity (RH). One-way valves prevented mixing of inspired and expired gases. Our model was similar to that of Lellouche et al.5 The compliance of the lung compartment was set at 0.05 L/cm H2O, and the ventilator was set in continuous mandatory ventilation mode with breathing frequencies of 10 and 20 breaths/min; VT of 300, 500, and 700 mL; inspiratory time of 1.0 s; and PEEP of 3 cm H2O. A ventilator self-test was performed before the study. Since the ventilator has compression volume correction, we did not measure the VT of the lung compartment. For testing, the oxygen mask with nebulizer heater was set at FIO2 of 0.33 and 0.98, and the high-flow CPAP system was set at 15, 30, and 45 L/min. The high-flow CPAP system had a port to deliver oxygen, as did 8 of the 11 HMEs: for each of these 9 systems, oxygen was delivered at 0 and 3 L/min (temperature 24 ± 0.2°C and RH 0%).
We measured AH, RH, and temperature using a capacitance-type moisture sensor (Moiscope, Senko Medical, Tokyo, Japan, response time 3 s in the range 40–100%) that was 2-point calibrated by a cooler/heater water source (HHC-51, Senkoika, Tokyo, Japan). After changing experimental settings, we waited for at least 1 hour. Because condensation on sensor surfaces compromises sensor performance, we did not insert sensors in the circuit during waiting. Usually we measured RH and temperature for 10 min, and, after confirming stabilization of RH and temperature, we recorded AH, RH, and temperature for the last 5 min. All experiments were performed in an air-conditioned room in which the ambient temperature was set at 24°C. The ambient temperature and humidity were monitored with the same sensor. All signals from the hygrometer were led to an analog/digital converter and recorded at 50 Hz/channel with data-acquisition software (Windaq, Dataq Instruments, Akron, Ohio). For each experimental setting we measured temperature and AH for 5 breaths. The results are expressed as mean ± SD.
Analysis of variance and the Bonferroni test for multiple comparisons were used for overall comparisons between with and without supplemental oxygen, and among 300, 500, and 700 mL of VT. All statistical analysis was performed with statistical software (SPSS 11.01, SPSS, Chicago, Illinois) and P values of < .05 were considered statistically significant.
Results
Room temperature was 23.7 ± 0.3°C, and the ambient RH was 57.2 ± 4.0%. The AH of the simulated expired gas was 34.7 ± 1.4 mg/L (RH 100% at 32.5 ± 0.8°C). Among the tested HMEs, AH ranged from 25.3 to 30.7 mg/L. With high-flow CPAP, AH was 37.7 ± 2.8 mg/L, and with the oxygen mask with nebulizer heater AH was 31.9 ± 2.2 mg/L (Table 2). When the FIO2 of the oxygen mask with nebulizer heater was changed from 0.33 to 0.99, AH changed from 30.9 ± 2.2 mg/L to 32.9 ± 2.7 mg/L. Adding oxygen at 3 L/min decreased the AH, and none of the HMEs continued to maintain AH above 30 mg/L. In the high-flow CPAP system, with 3 L/min oxygen (Fig. 2), AH was 37.8 ± 2.5 mg/L.
As VT increased, AH significantly (P < .01) decreased in all the HMEs and in the system delivering high-flow CPAP via oxygen mask and nebulizer heater (Fig. 3). For all tested devices, as breathing frequency increased, AH increased: the percentage difference between breathing frequencies of 10 and 20 breaths/min was around 5%.
Discussion
Taking into account the effects of supplemental dry oxygen, as well as of variations in VT and breathing frequency, we assessed the humidifying capabilities of 11 HMEs for tracheostomized patients with spontaneous breathing, of a system delivering high-flow CPAP, and of an oxygen mask with nebulizer. While AH was not evidently related to breathing frequency, it was profoundly affected by VT and oxygen supplementation. In each instance, when 3 L/min of oxygen was supplied through HMEs equipped with oxygen ports, AH declined to below 30 mg/L. Meanwhile, with all devices, as VT increased, AH decreased.
With 48 commercially available HMEs for mechanically ventilated patients, Lellouche et al found that only one in 3 devices maintained AH at ≥ 30 mg/L.5 HMEs for mechanically ventilated patients are configured differently than those for tracheostomized spontaneously breathing patients. All the HMEs we tested have a suction port through which the patient inspires ambient air. When VT increases, more ambient air is present in the inspired mixture. At VT of 300 mL, 8 of the 11 HMEs maintained AH at ≥ 30 mg/L. At VT of 700 mL, only one of the HMEs maintained 30 mg/L. VT greatly affected the performance of the tested HMEs for spontaneously breathing patients.
Assessing HMEs for mechanically ventilated patients, Lucato et al found that HMEs are more efficient with low-VT ventilation.7 HMEs for spontaneously breathing tracheostomized patients are applied differently than are those for mechanically ventilated patients. The use of a suction port causes the escape of some expiratory gas that does not pass through the HME, resulting in the loss of the vapor contained in that expired gas. HME performance is impaired by the volume loss to the atmosphere.10 These results are not directly comparable to the results we obtained for HMEs for mechanically ventilated patients.
With no supplemental oxygen, AH differed among the investigated HMEs. These findings are in line with those in the Lellouche study.5 One of the great advantages of a bench study is that it enables investigation and comparison of medical devices under the same conditions. Even so, it was not possible to maintain all environmental and simulated spontaneous breathing conditions. Ambient temperature and humidity, and the temperature and AH of the gas expired from the lung model differed between the tested HMEs. Since these values are important determinants of HME performance, care is required when using these results to compare the performance of HMEs.
When 3 L/min of oxygen was added, none of the HMEs with an oxygen port maintained AH at ≥ 30 mg/L. As far as we know, this is the first study to evaluate the ability of HMEs to return vapor to spontaneously breathing tracheostomized patients. Supplemental oxygen also significantly (P < .01) affected AH (see Fig. 3). ICU patients have diverse individual needs, but, after liberation from mechanical ventilation, supplemental oxygen is often required. Moreover, VT also differs between patients. To avoid complications from insufficient humidification, our findings suggest that it is prudent to consider use of a Tracheomask or a high-flow CPAP system, rather than an HME, in patients who require supplemental oxygen or who have spontaneous breathing with high VT.
To accurately measure the AH of inspired gas, we inserted one-way valves to prevent overestimation owing to admixture, in the ventilator tubing, of water vapor from expired gases. As evaluated thoroughly in Zuur et al's study, when measuring the AH of breathing gas, it is important to control the durations of the inspiratory and expiratory phases.9 In our protocol the inspiratory time was 1 s and the expiratory time was either 5 s (at 10 breaths/min) or 3 s (at 20 breaths/min). If we had measured AH at the beginning of inspiration, we might have underestimated the actual values. We traced raw data and measured the highest value to evaluate HME performance. We cannot discount the possibility of underestimation, but, according to Zuur et al's study, any differences are likely to have been small. We suspect that the previously published higher AH (34 ± 2.6 mg/L) readings for the tested HMEs probably reflect contamination with fully humidified expiratory gas. On the other hand, using a servo-controlled heated humidifier to simulate physiological effects, our model maintained the expired gas temperature at 37°C, resulting in slightly higher than previously reported AH of expired gas.5
Our study has several limitations. Obviously, with a lung-model simulation the findings do not directly correspond to clinical situations. Moreover, because our observation period was relatively short, our findings offer no evidence that HMEs can be used reliably and safely for prolonged periods in spontaneously breathing patients. Also, it was not possible to keep the ambient temperature and humidity precisely constant, which may have influenced our data.
Conclusions
In conclusion, with supplemental oxygen, humidification decreased below the American Association for Respiratory Care recommended value. When supplemental oxygen is required, humidification is insufficient with HME for spontaneously breathing tracheostomized patients with in the ICU.
Footnotes
- Correspondence: Masaji Nishimura MD, Emergency and Critical Care Medicine, The University of Tokushima Graduate School, 3-18-15 Kuramoto, Tokushima, Japan 770-8503. E-mail: nmasaji{at}clin.med.tokushima-u.ac.jp.
Mr Chikata presented a version of this paper at the Open Forum of the 58th AARC Congress, held November 10–13, 2012, in New Orleans, Louisiana.
The authors have disclosed no conflicts of interest.