Abstract
BACKGROUND: Jet nebulizers constitute the aerosolization devices most frequently used during mechanical ventilation. Continuous nebulization can influence the delivered tidal volume (VT) and lead to significant medication loss during expiration. Ventilators thus provide integrated jet nebulization systems that are synchronized during inspiration and ostensibly keep VT constant.
METHODS: This was a bench study of systems integrated in the Evita XL, Avea, Galileo, and G5 ventilators. The VT delivered with and without nebulization, the inspiratory synchronization of nebulization, and the aerosol deposition were measured with 2 locations of the nebulizer.
RESULTS: Changes in VT with the nebulizer were below 20 mL and below 10% of set VT for all ventilators. Synchronization was good at the beginning of insufflation, but prolonged nebulization was observed with all ventilators at the end of insufflation, until up to 1 s during expiration: 5–80% of nebulization occurred during expiration with significant aerosol loss in the expiratory limb. Synchrony could be improved by (1) reducing gas compression/decompression phenomena proximal to the jet nebulizer and (2) increasing inspiratory time, which reduced the amount of nebulization occurring during expiration. Placing the nebulizer upstream in the inspiratory limb did not affect inspiratory synchrony but allowed reduction of the amount of aerosol lost in the expiratory limb.
CONCLUSIONS: Jet nebulizer systems integrated in the tested ventilators are reliable in terms of VT control. Gas compression in tubing driving gas to the nebulizer delays synchronization and reduces nebulization yield if the nebulizer is placed close to the Y-piece. Increasing inspiratory time with no end-inspiratory pause reduces the expiratory loss of medication if placement of the nebulizer upstream in the inspiratory limb is not feasible.
Introduction
Jet nebulizers are simple cost-effective devices for aerosolized medication delivery.1 They are designed for single use, thus obviating sterilization procedures and contamination-associated risks.2 Their operation requires a 6–8 L/min gas flow; the nebulizer's internal resistance (pinhole narrowing) induces a pressure drop within the device, resulting in aerosolization through a Bernoulli effect. Recent data indicate that jet nebulizers are the most common devices used for nebulization during mechanical ventilation in ICUs.3
During mechanical ventilation, the gas flow operating the jet nebulizer may interfere with the gas flow delivered by the ventilator.4 For instance, in the case of a ventilator set to deliver a 40 L/min square waveform flow and a jet nebulizer operated with an 8 L/min external gas source, a 20% increase in tidal volume (VT) may approximately be expected. As manual corrections of ventilator settings to take into account this extra gas flow are imprecise and cumbersome, manufacturers have implemented integrated jet nebulization systems into ventilators.4 Those systems comprise a jet nebulization port on the ventilator, providing the nebulizer's driving gas in parallel to the gas delivered through the inspiratory limb.5 This theoretically enables an automatic compensation of the insufflated volume to potentially maintain a constant VT and patient ventilation (Fig. 1). In addition to controlling VT, jet nebulization systems integrated into ventilators aim to synchronize nebulization with inspiration to avoid loss of medication aerosolized during expiration.6 Thus, ventilators aim to deliver the driving gas to the nebulizer only during inspiration.7 A study from 1993 evaluated the nebulizers of a preceding generation of ventilators and showed that some performed poorly.5 To the best of our knowledge, the possible benefits of the integrated jet nebulization systems, that is, control of VT and inspiratory synchronized aerosol delivery, have not been independently evaluated despite their frequent use in ICUs and regulatory approval.3
The aim of this bench study was to measure changes in VT and to assess inspiratory synchronization when operating ventilator-integrated jet nebulization systems.
QUICK LOOK
Current knowledge
Jet nebulizers are the most commonly used devices for delivery of aerosolized medications during mechanical ventilation. Continuous nebulization augments tidal volume (VT) and may induce some loss of medication. Ventilators may incorporate nebulizer drive systems to synchronize aerosol delivery with inspiration, maintaining constant VT and avoiding drug waste.
What this paper contributes to our knowledge
Jet nebulization systems integrated in ventilators provide control of VT and may be preferred to external gas sources or empirical ventilator setting changes. Placement of the nebulizer upstream in the inspiratory limb reduces the loss of medication in the expiratory limb due to nebulization during expiration.
Methods
Jet nebulization systems integrated in 4 ICU ventilators (Evita XL [Dräger, Lübeck, Germany], Avea [CareFusion, San Diego, California], and Galileo and G5 [Hamilton Medical, Reno, Nevada]) were evaluated in a bench model of mechanical ventilation (Fig. 1). Gas flow was measured twice (Fleisch No. 2 pneumotachograph, cleaned and dried between each experiment): first, immediately distal to the nebulizer, and second, distal to the Y-piece to measure the total flow entering the lungs. The airway pressure was measured at the Y-piece. The lung model (Training and Test Lung, Michigan Instruments, Grand Rapids, Michigan) compliance and resistance were set at 60 mL/cm H2O and 5 cm H2O/L/s, respectively. The delivered VT was measured through flow integration over time distal to the Y-piece. Inspiratory synchronization of nebulization was evaluated measuring the time lapse between beginning of insufflation and beginning of nebulization (beginning of insufflation synchrony: see Δ1 in Fig. 1) and the time lapse between end of insufflation and end of nebulization (end of insufflation synchrony: see Δ2 in Fig. 1). The beginning and end of insufflation and nebulization were determined based on flow and pressure tracings. Nine different volume controlled ventilator settings were evaluated varying VT (300, 500, and 700 mL) and square waveform inspiratory flow (30, 50, and 70 L/min) without an end-inspiratory pause. Breathing frequency was set at 12 breaths/min, FIO2 at 21%, and end-expiratory pressure at 5 cm H2O. All measurements were carried out at ambient temperature and pressure in dry conditions.
First, measurements were carried out with a standard jet nebulizer commonly used in the ICU (Micro Mist, Hudson RCI, Durham, North Carolina) filled with 0.9% sodium chloride and placed close to the Y-piece (Fig. 1). Second, to gain insight into the mechanisms ruling inspiratory synchrony/asynchrony of nebulization, measurements were also performed (1) without a nebulizer placed in the circuit; (2) with a nebulizer exhibiting a high internal resistance (Heart, WestMed, Tucson, Arizona); and (3) with the standard nebulizer but reducing the length of the nebulizer gas supply tubing to 2 cm (to reduce the volume of compressible gas), placing the nebulizer very close to the output of the ventilator. Indeed, we reasoned that the long tubing, driving gas from the ventilator to the nebulizer, was a source of gas compression that would continue to drive the aerosol after the end of the ventilator's insufflation. The Evita XL ventilator was also tested with nebulizer 84 12 935 (Dräger), as recommended by the ventilator manufacturer.
To distinguish between nebulization synchronized with insufflation as opposed to inspiration (insufflation + end-inspiratory pause), all measurements were repeated with a 1-s end-inspiratory pause.
Finally, to illustrate the link between inspiratory synchronization of nebulization and aerosol delivery (and conversely, aerosol loss in the expiratory limb), deposition experiments were performed using 0.9% sodium fluoride as a tracer as previously described.8,9 Briefly, filters were placed between the endotracheal tube and the lung model (to measure delivered aerosol) and between the Y-piece and expiratory limb (to quantify aerosol loss in the expiratory limb); the Micro Mist nebulizer (filled with 4 mL of 0.9% sodium fluoride) was run until no more nebulization could be visually observed, and filters were dismounted and desorbed overnight in 50 mL of 2% total ionic strength adjustment buffer (Reagecon, Shannon, Ireland). The sodium fluoride extracted was then assayed by electrochemical analysis (Orion combination fluoride electrode, Thermo Fisher Scientific, Waltham, Massachusetts) and expressed as a percentage of the total amount recovered on both filters. Measurements (triplicate) were made with the Evita XL ventilator in situations of poor synchrony (short inspiratory time of 0.6 s) and in situations of improved synchrony (long inspiratory time and/or reducing the volume of compressible gas proximal to the nebulizer), with the nebulizer being placed close to the Y-piece (Fig. 1). Finally, deposition measurements were also performed, placing the nebulizer in a position 25 cm proximal in the inspiratory limb.
Measurements were reported as mean ± SD. For each ventilator, comparisons were carried out using the Student t test and the Pearson correlation coefficient. Comparisons between ventilators were made using analysis of variances, followed, if significant, by two-by-two post hoc analysis. A P value below .05 was considered significant.
Results
Altogether, 2,880 respiratory cycles were analyzed.
Control of VT
When starting the nebulizer system, the change in VT was always below 20 mL in absolute value and below 10% of the set VT. Minimal changes observed in inspiratory flow and in the resulting VT and airway pressure, as well as the gas flow passing through the nebulizer, are detailed in Table 1.
Inspiratory Synchronization
Typical nebulizer and ventilator flow waveforms are depicted in Figure 2. At the beginning of insufflation, the time lapse between the start of insufflation and that of nebulization was always below 100 ms. Following this rapid start, nebulization flow gradually increased to a maximum during insufflation. At the end of insufflation, nebulization flow gradually decreased to zero. Due to this gradual decrease, prolonged nebulization persisted after the end of insufflation, until up to 1 s (Fig. 2 and Table 2). For all ventilators, this persisting expiratory nebulization (poor end-inspiratory synchronization) explained that up to 80% of nebulization could occur and be potentially lost during expiration. This proportion was negatively correlated with the insufflation time set on the ventilator (r2 = 0.45, P < .001) (Fig. 3).
For the Evita XL ventilator, results concerning VT control and synchronization were similar when using nebulizer 84 12 935 as recommended by the manufacturer; for all ventilators, synchrony was similar whatever the position of the nebulizer in the circuit, in particular placing the nebulizer 25 cm upstream in the inspiratory limb (data not shown).
Mechanism of Poor Synchronization
When evaluating synchronization without a nebulizer placed in the circuit (very low resistance opposed to the nebulization port gas flow), an excellent synchronization was observed at the beginning and end of insufflation for all ventilators, with time lapses below 60 ms whatever the ventilator setting. Conversely, using the high internal resistance nebulizer (12.7 ± 0.3 cm H2O/L/s measured with an air flow of 2 L/min), an extremely large delay between end of insufflation and end of nebulization, up to 2.5 s, was observed for all ventilators. When using the nebulizer with a standard internal resistance (0.67 ± 0.07 cm H2O/L/s), used for the rest of the experiments, or when reducing the tubing volume of the nebulizer gas supply (reducing the length of the tubing to 2 cm), end-insufflation synchronization improved for all ventilators. Those mechanisms of asynchrony are depicted in Figure 4.
Effect of an End-Inspiratory Pause
The Evita XL and Avea systems delivered nebulization synchronized with insufflation, whereas the Galileo and G5 systems were synchronized with inspiration (active nebulization persisting during the pause) (Fig. 5).
Impact of Poor Synchronization on Aerosol Loss and Delivery
When the nebulizer was placed close to the Y-piece (Fig. 1), ventilator settings associated with prolonged flow through the nebulizer during expiration (short inspiratory time related to high inspiratory flow and/or reduced VT) were associated with 80% aerosol deposition on the expiratory limb filter (lost aerosol produced by a sufficient flow through the nebulizer during expiration) and 20% at the tip of the endotracheal tube (Table 3). When improving inspiratory synchronization, through increasing inspiratory time to 1.40 s (reduced inspiratory flow and/or increased VT) or reducing the volume of compressible gas proximal to the nebulizer, only 40% and 47% of deposition, respectively, occurred on the expiratory limb filter (60% and 53% deposition on the filter placed at the tip of the endotracheal tube). Thus, when placing the nebulizer close to the Y-piece, changes in nebulization during expiration paralleled changes in the amount of aerosol lost in the expiratory limb.
Placing the nebulizer upstream in the inspiratory limb was associated with the greatest reduction in deposition on the expiratory limb filter whatever the synchronization and the degree of persistent nebulization during expiration (Table 3).
Discussion
The main finding of this experimental study is that the VT delivered by modern ICU ventilators remains well controlled when integrated jet nebulization systems are used. The synchronization with insufflation was poor, however, with a significant part of the nebulization occurring during the end-inspiratory pause and/or expiration. This aerosol generated during expiration may be lost if the nebulizer is placed close to the Y-piece. Given the frequent use of those systems, several implications concerning mechanical ventilation settings, nebulization performance, system design, and clinical use might be taken into consideration.
Nebulization Interaction With Ventilation
When using jet nebulizers during mechanical ventilation, a gas source external to the ventilator induces a significant increase in VT as 6–8 L/min gas are added to the inspiratory flow. In the present bench study, the change in VT observed when activating the integrated systems was always below 10% of the set VT, that is, within the precision range of the VT setting of modern ICU ventilators.10 Thus, those systems may be considered reliable in this regard during volume controlled mechanical ventilation. Previous work evaluating jet nebulization systems integrated in a previous generation of ventilators found that some ventilators delivered a low pressure and a low flow of gas to the nebulizer.5 A flow lower than 4 L/min has been linked to poor nebulization performance (in terms of nebulization rate and overall yield). For all ventilators, we observed flows through the nebulizer exceeding 6 L/min (Table 1). Thus, insufficient gas supply may no longer be a concern or potential factor for improvement in modern ventilators.
Nebulization During Expiration and End-Inspiratory Pause
Significant nebulization occurred during expiration in all the tested ventilators and may therefore constitute a factor for potential improvement. When the nebulizer is placed close to the Y-piece, medication nebulized during expiration is cleared mostly through the expiratory limb of the ventilator and thus is lost for the patient and has the potential to damage the expiratory flow meter and/or obstruct filters protecting this flow meter.11,12 To deliver a higher amount of drug to the patient, it may be desirable to reduce the drug loss during expiration.13,14
The Galileo and G5 ventilators are designed to synchronize nebulization with inspiration rather than with insufflation (Fig. 5). No standardization exists, and the value of such a design appears questionable, as medication nebulized during the pause is unlikely to reach the patient. The persistent nebulization during the pause also compromises proper plateau pressure measurement. The end-inspiratory pause used in clinical practice (300–500 ms) is usually shorter than the one used for illustrative purposes in this bench study (1 s). Because the flow through the nebulizer is ∼4–10 times lower than the inspiratory flow commonly set on ventilators (≈6–10 vs 40–60 L/min), the clinical impact of persistent nebulization during the plateau will be smaller than in our study.
Mechanisms of Asynchrony and Implication for System Design
The flow coming out of the nebulizer after insufflation has stopped was explained by the gas compressed proximal to the nebulizer. Indeed, testing the integrated nebulization system working without a nebulizer and with a reduced volume of compressible gas proximal to the nebulizer both improved synchronization. By contrast, using a nebulizer with high internal resistance resulted in poor synchronization. Therefore, gas compression and decompression proximal to the nebulizer are probably major determinants of persistent nebulization during expiration.
Synchronization Algorithm.
Albeit the ventilators per se did not appear to be at the origin of the asynchrony, a simple way to limit nebulization during expiration would be to stop delivering gas to the nebulizer earlier during inspiration. Actually nebulizing only during the very initial phase of inspiration would limit the amount of medication potentially lost during expiration while also increasing the amount of medication delivered to the distal parts of the lung. Indeed, the gas and thus the aerosol delivered at the end of insufflation do not reach the deep part of the lungs.7,15 An ideal algorithm would start nebulization immediately before the start of insufflation (this concerns only fully controlled ventilation) for the nebulizer flow to be highest at the beginning of insufflation. As the particle sizes produced by jet nebulizers decrease with increasing nebulizer flow, such an algorithm would allow production of the finest droplets at the beginning of inspiration, thus favoring distal lung deposition.16
Decrease Compressible Gas Volume Proximal to the Nebulizer.
Reducing the length of the nebulizer gas supply tubing is not feasible in the clinical setting for very practical reasons: the Y-piece and, as a consequence, the head of the patient would need to be placed within very short distance of the ventilator, which is, at best, very unpractical. Thus, reducing the inner tubing diameter may be an interesting approach to reduce the volume of compressible gas.
Clinical Implications
Internal Resistance of the Nebulizer.
Using high-resistance nebulizers leads to very poor synchronization and should probably be avoided. Conversely, a low-resistance nebulizer could be evaluated. However, the larger droplets produced by such devices may be counterproductive (by favoring impaction in the ventilator circuit).17
Nebulizer Position in the Circuit.
As inspiratory synchronization appeared poor in the present bench study, nebulizer placement may best be done as for continuous nebulizers, that is, proximal in the inspiratory limb, as this placement was associated with the least loss of medication in the expiratory limb and the highest amount of medication delivered at the tip of the endotracheal tube (Table 3).17 Precise optimal placement deserves further investigations since different values of bias flow may be associated with different optimal positions. Of note, similar results were observed previously with synchronized jet nebulization systems by O'Doherty et al18 and Hughes and Saez.19 Beyond confirming those results on modern ventilators, our results show that the benefit of positioning these synchronized systems upstream in the inspiratory limb is due to the observed persistent nebulization during expiration, that is, the nebulizer behaves like a continuous system due to poor synchronization. Thus, our results show that nebulizer placement close to the Y-piece, a practice acknowledged by a large number of intensivists,3 is far from optimal for the synchronized jet nebulization systems evaluated in the present study. Whether this placement is appropriate for other devices, potentially better synchronized with inspiration, deserves further evaluation.15
Inspiratory Time.
As placement of the nebulizer upstream in the inspiratory limb may be difficult in clinical practice, especially with disposable circuits (need for extra tubing and connection), a simple way to reduce the amount of medication potentially lost during expiration when using the integrated jet nebulization systems placed close to the Y-piece may be to increase the inspiratory time (increase in VT and/or decrease in inspiratory flow according to the patient's condition and tolerance). Of note, reducing inspiratory flow may be particularly interesting, as it may increase nebulization yield through other means (reduced droplet impaction in the ventilator circuit).16 However, placement upstream in the inspiratory limb had the greatest impact on delivery in the systems tested in the present study (Table 3).
Study Limitations
The results of the present bench study need to be confirmed in patients receiving aerosolized medication. However, the good results observed concerning VT control are unlikely to be very different in the clinical setting, as during volume controlled ventilation, the VT is independent of the patient. Pressure-regulated modes of mechanical ventilation were not evaluated in the present bench study; thus, no conclusion may be drawn concerning the function of the integrated jet nebulization systems in this situation. Of note, as VT partly depends on the patient's effort and respiratory mechanics, its control by the ventilator would be less relevant. Nebulization is unlikely to affect the delivered pressure, as this variable is very tightly controlled by modern inspiratory valves.
The different test conditions influencing expiratory nebulization also impacted nebulization during inspiration. Our findings strongly suggest that persistent nebulization during expiration is an important mechanism of drug loss. An associated change in nebulization characteristics during inspiration (mainly higher or lower flow through the nebulizer) cannot, however, be completely ruled out. Some variables such as the triggering type, breathing frequency, and pressure in the connecting tubing were not fully evaluated in the present study, although they may interact with some results.
Finally, only 0.9% sodium chloride/fluoride was used for nebulization in the present study. Behavior may be slightly different with other medications, particularly in the case of solutions exhibiting high viscosity.
Conclusions
Jet nebulization systems integrated in modern ICU ventilators allow a good control of VT and thus may be preferred to the use of external gas sources and empirical ventilator setting changes. Inspiratory synchronization sometimes appeared poor, due to persistent nebulization during the inspiratory pause and expiration, in part because of gas compression/decompression proximal to the nebulizer. Specific settings such as increasing inspiratory time might help to optimize the dose of drug delivered. Importantly, nebulizer placement upstream in the inspiratory limb is the most powerful means of reducing the loss of medication in the expiratory limb due to persistent nebulization during expiration. Delivering nebulization during the entire inspiration, that is, including the pause, appears questionable.
Footnotes
- Correspondence: Stephan Ehrmann MD PhD, Service de Réanimation Polyvalente, CHRU de Tours, F37044 Tours, France. E-mail: stephanehrmann{at}gmail.com.
Drs Ehrmann, Louis, Isabey, Brochard, and Apiou-Sbirlea were supported in part by Grant ANR-2010 BLAN 1119 05 from the Agence Nationale de la Recherche. Dr Ehrmann was provided with equipment for research purposes from La Diffusion Technique Française, Penn-Century, and Dräger. Dr Brochard's laboratory received research grants from Covidien, Dräger, Vygon, Philips Respironics, and GE Healthcare; none were related to this particular study. The other authors have disclosed no conflicts of interests.
Drs Ehrmann, Lyazidi, Brochard, and Apiou-Sbirlea presented a version of this paper at the 40th Congrès International de la Société de Réanimation de Langue Française, held January 18–20, 2012, in Paris, France, and at the American Thoracic Society 2012 International Conference, held May 18–23, 2012, in San Francisco, California.