Abstract
BACKGROUND: The Acapella device produces high-frequency oscillations and positive expiratory pressure to promote bronchial secretion clearance. Its performance during aerosol delivery has not been described. We evaluated the effect of nebulizer and Acapella configuration on pulmonary deposition of radio-tagged aerosol in healthy subjects.
METHODS: Ten healthy male subjects (mean age 24.4 ± 2.2 y) participated in a crossover study that compared pulmonary delivery of 4 mL of technetium-99m-labeled diethylene triamine penta-acetic acid (25 mCi) and 0.9% saline solution via jet nebulizer. We tested 3 configurations: nebulizer attached to the distal end of the Acapella; nebulizer placed between the mouthpiece and the Acapella; and nebulizer alone (control). With scintigraphy we measured radio-aerosol deposition in 6 lung regions: upper, middle, lower, central, intermediate, and peripheral.
RESULTS: Deposition was similar between the right and left lungs, with no significant differences between device configurations. Lung deposition was less with the nebulizer attached to the Acapella than with nebulizer between the mouthpiece and the Acapella (P = .001, for both lungs) or without the Acapella (P = .003 and P = .001 for the right and left lungs, respectively). There was no significant difference between the setup without Acapella and the setup with the nebulizer between the mouthpiece and the Acapella (P = .001, for both lungs). On the vertical axis, deposition was lower with the nebulizer attached to the distal end of the Acapella than with the nebulizer between the mouthpiece and the Acapella (upper region P < .001, middle region P = .001, lower region P = .003), and lower with the nebulizer attached to the distal end of the Acapella than with the setup without Acapella (upper and middle region both P = .001, lower region P = .002), with up to a 3-fold difference in the middle and lower regions. On the central-peripheral axis, deposition was lower with the nebulizer attached to the distal end of the Acapella than with the nebulizer between the mouthpiece and the Acapella (central region P < .001, peripheral region P < .001), and lower with the nebulizer attached to the distal end of the Acapella than with the setup without Acapella (central and peripheral regions both P = .002), with differences of 3–4-fold between the central and peripheral regions.
CONCLUSIONS: Placing the nebulizer distal to the Acapella, as recommended by the manufacturer, decreased intrapulmonary deposition, compared to placing the nebulizer between the Acapella and the patient airway, or delivering aerosol without the Acapella in the circuit. (ClinicalTrials.gov NCT01102166)
Introduction
The clearance of pulmonary secretions is directly affected by changes in cross-sectional area of the airways, the composition and production of mucus in the respiratory tract, ciliary function, and cough reflex.1 Inflammatory and infectious diseases or dyskinetic syndromes such as asthma, chronic bronchitis, and cystic fibrosis can involve hypersecretion and blockage of the airways by mucus.2 The retention of lung secretions can result in complete or partial obstruction of the airways, which results in atelectasis, lung hyperinflation, and increased respiratory muscle load.3
Inhalation of medical aerosols promotes hydration and reduction of secretions viscosity, optimizing mucociliary clearance.4 Thus, inhalation of drugs transported as inhalable particles may alter mucus rheology and has become clinically accepted as an adjunct to physiotherapy in the treatment of chronic lung conditions.5 In addition, inhalation increases mucociliary clearance by influencing ciliary function, increasing osmotic drive, and altering mucus viscoelasticity. This results in reduction of viscosity and augmentation of sputum clearance.6
Bronchial hygiene therapy involves the use of noninvasive and invasive techniques to assist the mobilization and clearance of secretions.7 The methods of bronchial hygiene reported in the literature include positive expiratory pressure (PEP) and high-frequency oscillation of the airway. The principle of PEP is based on generating positive pressure to splint open and stabilize airways, preventing bronchial collapse. PEP is also thought to improve collateral ventilation to allow better distribution of gas beyond occluded airways.7 Oscillatory PEP may have added benefit over the PEP mask alone in helping to dislodge thick secretions from the airway walls and decrease mucus viscoelasticity.8,9 High-frequency oscillation of both the chest wall and airway have been shown to facilitate secretion clearance.10
The Acapella high-frequency oscillatory PEP device (Smiths Medical, Watford, United Kingdom) produces oscillations during exhalation by intermittently occluding the air flow with a magnetic, adjustable valve that allows adjustment of frequency, oscillation amplitude, and mean pressure through 5 levels, via a dial on the device. It is suggested that expectoration can be optimized when the applied pressure frequency coincides with the ciliary movement, approximately 13 Hz.10 Acapella can be used in any position, does not depend on the slope or level of PEP, and allows concomitant aerosol administration.11
We hypothesized that the placement of the nebulizer distal to the Acapella, as recommended by the manufacturer, would reduce aerosol lung delivery, compared to more proximal placement (between device and mouthpiece) or compared to nebulizer alone. The aim of this study was to evaluate the effect of aerosol generator placement in 2 different positions during use of the Acapella device on delivery of radio-labeled aerosol to the lungs of healthy subjects. This is the first study to analyze radio-aerosol deposition in normal subjects with and without the Acapella.
QUICK LOOK
Current knowledge
Positive expiratory pressure (PEP) and oscillatory PEP devices are commonly used to aid bronchial secretion clearance. Delivering aerosol therapy during PEP may alter the aerosol and the aerosol delivery.
What this paper contributes to our knowledge
Placing the jet nebulizer distal to the PEP device decreased pulmonary deposition, compared to placing the nebulizer between the patient and the PEP device, or using the nebulizer alone.
Methods
This randomized crossover clinical trial received approval from the institutional ethics and human research committee at the Universidade Federal de Pernambuco, and subjects gave written informed consent. The study was performed at Português Hospital, Recife City, Brazil.
Subjects
The inclusion criteria were: no history of asthma or other lung disease; age 18–30 years; and FVC, FEV1, and peak expiratory flow all ≥ 80% of predicted. The exclusion criteria were history of smoking and signs or symptoms of illness (fever, tachycardia, and tachypnea) or respiratory disease.
Procedures
A jet nebulizer (ST3, NS, São Paulo, Brazil) operated at 7 L/min with compressed air was used in all configurations. The nebulizer's mass median aerodynamic diameter (MMAD, 3.6 μm) and geometric standard deviation of (2.2 μm) were previously determined with via cascade impactor, using standard United States Pharmacopoeia methods.
Device Configurations
The subjects were randomly allocated to receive aerosol with 3 device configurations (Fig 1):
A: With the nebulizer attached to the distal end of the Acapella, per the manufacturer's recommendation (distal)
B: With the nebulizer connected to a T-piece between the mouthpiece and the Acapella (proximal)
C: With the nebulizer attached to a standard mouthpiece, without the Acapella in the circuit (control)
A clinical assessment was performed and data measurements included age, weight, height, body mass index, spirometry, and maximal inspiratory pressure. FVC, FEV1, and peak expiratory flow were measured with a spirometer (2110, Vitalograph, Buckinghamshire, United Kingdom) according to the American Thoracic Society standards.12 Maximal inspiratory pressure was measured with a manometer (MV-150, Ger-Ar Comércio de Produtos Médicos, São Paulo, Brazil), with the subject seated and with a nose clip, and instructed to perform a maximal inspiration during a minimum of 1 second.13 Three measurements were performed and the highest value was retained for analysis.
The scintigraphy procedures were developed in the Department of Nuclear Medicine. The subjects performed aerosol inhalation based on the randomization schedule, and then underwent scintigraphy. Four milliliters of 0.9% physiologic saline solution containing technetium-99m-labeled diethylene triamine penta-acetic acid (25 mCi) was nebulized over 9 min. The aerosol was inhaled via mouthpiece, with the subject seated upright. All the subjects were previously trained for deep breathing and inspiratory pause.14
Immediately after aerosol administration scintigraphy was performed with a gamma camera (ADAC Forte, Philips, Eindhoven, The Netherlands), during a period of 300 seconds, using a matrix of 256 × 256 × 16 in the posterior position.
The lung images were divided into regions of interest vertically (upper, middle, and lower thirds) and on the central-peripheral axis (central, intermediate, and peripheral) (Fig. 2), and the radiation counts in the regions of interest were calculated.15 With each subject a wash-out period minimum of 7 days was used between radio-labeled aerosol exposure, to ensure clearance of the inhaled radioactive material.
Statistical Analysis
For statistical analysis the Kolmogorov-Smirnov test was used to analyze normality, followed by analysis of variance and the Tukey post hoc test, in statistics software (SPSS 18.0, SPSS, Chicago, Illinois). The results are expressed as mean ± SD, and we considered differences statistically significant at P < .05.
Results
Of the 14 healthy males we included, only 10 completed the protocol: 2 dropped out due to claustrophobia, and 2 retracted consent. The mean values in the 10 subjects who completed the protocol were: age 24.4 ± 2.2 years, body mass index 22.6 ± 2.6 kg/m2, peak expiratory flow 583 ± 41 L/min, FEV1 4.3 ± 0.5 L, FVC 4.7 ± 0.6 L, and maximal inspiratory pressure 110 ± 16 cm H2O.
Deposition in the right and left lungs was similar with all the configurations studied. There was less total deposition with configuration A than with B or C (Fig. 3). There was no significant difference between configuration B and C. The differences between the regions of interest were up 3-fold greater with B and C, compared to A.
All 3 vertical regions (upper, middle, and lower third) had less aerosol deposition, in both lungs, with A than with B and C (Fig. 4). The ratio of upper to lower deposition was marginally higher with A (0.70 ± 0.21) than with B or C (0.6 ± 0.11 and 0.58 ± 0.12, respectively). Similarly, the central-peripheral axis regions showed less deposition with configuration A than with B or C (Fig. 5). The ratio of central to peripheral deposition trended lower with configuration A (0.89 ± 0.11) than with B or C (1.05 ± 0.23 and 0.93 ± 0.12, respectively).
The distribution of radio-aerosol deposition inside the Acapella differed with the nebulizer at the inlet of the Acapella versus at the outlet (Fig. 6).
Discussion
Placing the nebulizer at the distal end of the Acapella significantly decreased pulmonary deposition of radio-aerosol, compared to placing the nebulizer at the proximal end of the Acapella, and compared to nebulizer alone, in both the right and left lungs. The primary reason for the lower pulmonary deposition of radio-aerosol with distal nebulizer placement is likely the inertial impaction of larger aerosol particles as they pass through the Acapella, which makes less aerosol available for inhalation. Internal mechanisms that create turbulent or transitional flow increase the deposition of larger particles.10 This is supported by the report of Berlinski and Hayden,16 who evaluated changes in aerosol particle size from 2 continuously operated jet nebulizers with several PEP and oscillatory PEP devices. The MMAD of aerosol produced by the Hudson nebulizer (4.1 μm) was reduced to 1.2 μm with both the Acapella Choice and Acapella Duet, when used as recommended by the manufacturer (distal and inferior, respectively). This supports our supposition that the internal mechanisms of an oscillatory PEP device filter larger aerosol particles, reducing the total dose of aerosol available for inhalation (however, this was not reported). This reduction in particle size was associated with an increase of fine-particle fraction (< 5 μm), from 59% to > 87%, which might be expected to have a greater peripheral deposition than the aerosol with the larger MMAD produced by the nebulizer alone. This is consistently greater peripheral deposition, or a lower central-to-peripheral deposition ratio. In testing of the Pari LC Plus nebulizer with its proprietary PEP valve at settings of 1.5 to 4.5, Berlinski and Hayden16 found no changes in MMAD or geometric standard deviation. The Pari PEP systems did not change the nebulized albuterol characteristics.
We did not find the hypothesized difference between proximal nebulizer placement (between the Acapella device and the patient) and the nebulizer alone in the normal subjects tested. This suggests that placing the nebulizer between the Acapella and the mouthpiece did not substantially change the fraction of deposited dose in the lung, versus control.
Laube et al17 compared pulmonary delivery of radio-labeled aerosol delivery from a jet nebulizer (Pari LC Plus) with and without PEP in patients with cystic fibrosis. They reported reduced pulmonary delivery with PEP, and a modestly lower central/peripheral ratio, compared to nebulizer alone. Our findings are consistent to the point that we also had a modest trend toward lower central/peripheral ratio and reduced pulmonary deposition with Acapella in configuration A, but not with B or C.
In contrast to Berlinski and Hayden,16 Laube et al17 reported a reduced MMAD of 3.3 μm with PEP and 4.07 μm without PEP, with similar geometric standard deviations (2.61 μm and 2.78 μm, respectively). This lower MMAD may account for the difference in both pulmonary deposition and distribution. Those authors used a controlled breathing pattern during PEP exhalation, to maintain a target during exhalation, which may have increased the expiratory time and reduced the percent of aerosol inhaled from continuous nebulization, resulting in the reduction versus nebulizer alone. They did not find the expected distribution difference to the periphery or upper regions of interest in their population of patients with severe cystic fibrosis lung disease.
In normal subjects we also did not find the expected level of increase in peripheral or upper-lobe distribution with Acapella, in either configuration. This might be due to the nature of the normal condition of the airways, so Acapella did not have “floppy” airways to splint or obstructed airways for collateral ventilation to impact distal distribution.
We believe that the lower deposition of radio-aerosol found with the manufacturer-recommended distal position can be attributed to particle impaction inside the Acapella,11 which has several small-diameter internal ducts that increase the likelihood of impaction.
We found higher deposition in the middle and lower thirds than in the upper third, with all 3 configurations. This distribution in normal subjects may be attributed mainly to the vertical difference in pleural pressure between the upper and lower lung regions.18
According to Alderson and Line,19 other factors also influence the distribution of particles into the lungs, such as minute ventilation, inspiratory pressure, and nebulization position during inhalation. Ventilation per unit volume decreases from the base to the apex of the lung, as a result of regional differences in intrapleural pressure resulting from gravitational influences.18
In the lung, the gravitational and retractable forces act in the same direction, making the subatmospheric intrapleural pressure more negative in the apex than in the lung base, where the gravitational and retractable forces act in opposite directions. As a result, alveoli in the apex become more expanded and less compliant, with a small change in volume during inspiration. Moreover, alveoli in the base are less expanded and more compliant and have a greater change in volume during inspiration.18 In addition, our findings corroborate a study of radio-aerosol deposition coupled with bi-level ventilation in normal subjects.14
We found more deposition in the intermediate and peripheral regions than in the central region with all 3 configurations. The flow favors deposition in the large airways, owing to the inertial impact and possibly the greater radio-aerosol deposition in those areas is related to the aerosol particle size.19–22 Another important consideration is that the intrapleural pressure change is relatively uniform on the central-peripheral axis, so there is no central-peripheral change in ventilation per unit alveolar volume in the upright position.18
We used normal subjects to have a homogenous population in which to determine the impact of device configuration on pulmonary deposition, without having to account for differences in disease severity. It was beyond the scope of this study to isolate variables in obstructive pulmonary diseases that might impact aerosol distribution during oscillatory PEP. As our objective was to determine differences in distribution in the lungs and across regions of interest, we did not perform a mass balance to express lung deposition as a percentage of nominal dose.
Conclusions
Placing the nebulizer distal to the oscillatory PEP device decreased intrapulmonary deposition, compared to proximal placement, and compared to nebulizer alone.
Acknowledgments
We are grateful to Maria José Clemente Menezes, Real Imagem, Hospital Português, for support during data collection.
Footnotes
- Correspondence: Armèle Dornelas-de-Andrade PhD, Department of Physical Therapy, Universidade Federal de Pernambuco, Avenida Jornalista Aníbal Fernandes, Recife, Pernambuco, Brazil, 50670–901. E-mail: armeledornelas{at}yahoo.com.
This study was partly supported by grants from Fundação de Amparo à Ciência e a Pesquisa and Conselho Nacional de Desenvolvimento Científico e Tecnológico.
Valdecir C Galindo-Filho PhD presented a version of this paper at the annual meeting of the European Respiratory Society, held September 15–19, 2007, in Stockholm, Sweden.
Dr Fink is a consultant to Aerogen, Bayer, Boehringer Ingelheim, Cubist, Dance Biopharm, Novartis, ONY, Parion, Aridis, and the WHO.
See the Related Editorial on Page 441
- Copyright © 2014 by Daedalus Enterprises