Abstract
BACKGROUND: Colorimetric end-tidal carbon dioxide (ETCO2) detectors can identify airway obstruction during noninvasive ventilation and successful intubation during newborn resuscitation. The resistance of these devices is not well described, and the information provided by manufacturers is incomplete.
METHODS: We compared the resistance of 3 colorimetric ETCO2 detectors (Neo-StatCO2, Pedi-Cap, and Mini StatCO2,) and 2 mainstream capnograph sensors (EMMA infant airway adapter 17449 and neonatal/infant airway adapter YG-213T). Endotracheal tubes, 2.5–4.0-mm inner diameter (Portex) were measured as a reference range. A differential pressure transducer was placed between the device and a T-piece resuscitator. The other side of the device was open to air. Resistance to flow was tested at 1–10 L/min. Resistance was calculated as the change in pressure over change in flow and expressed as cm H2O/L/s.
RESULTS: There was a significantly higher mean resistance across all flows tested for the Neo-StatCO2 compared with the other ETCO2 devices (P < .001). There was a 6-fold difference between the least and most resistive colorimetric detectors. At the commonly utilized flow of 10 L/min, the resistance of the Neo-StatCO2 was 61.1 cm H2O/L/s, comparable with that of a 3.0 endotracheal tube, which we measured at 62.7 cm H2O/L/s. The resistance values of the Pedi-Cap and Mini StatCO2 were 9.9 and 8.4 cm H2O/L/s, respectively. Those of the EMMA and YG-213T were 7.1 and 2.6 cm H2O/L/s, respectively.
CONCLUSIONS: We found significant differences in resistance between devices used to detect ETCO2 during resuscitation of premature infants. Future trials are needed to determine the effects of this resistance on work of breathing, particularly on very premature newborns receiving mask CPAP.
- resistance
- work of breathing
- continuous positive airway pressure
- capnography
- carbon dioxide detector
- infant
- newborn
- neonatal intensive care
- resuscitation
- colorimetric
Introduction
Colorimetric end-tidal carbon dioxide (ETCO2) detectors are inexpensive, portable, lightweight, single-use devices. They have a pH-sensitive chemical indicator, which changes color in the presence of exhaled carbon dioxide. When color change occurs, it indicates that there is both adequate ventilation and adequate blood flow circulating through the lungs. They are the most frequently utilized ETCO2 device in the neonatal resuscitation environment and have been demonstrated to be effective in determining successful endotracheal intubation.1,2 We have shown that colorimetric ETCO2 detectors are also effective in recognizing airway obstruction, with bag-mask ventilation or CPAP application, during delivery room resuscitation of premature infants.3 At the time of these findings, we utilized a detector that was labeled for use in infants >1 kg. Since that time, a detector has been approved for infants <1 kg that has a reduced dead space. There are no published data on the resistance of this newer device with the flows commonly utilized clinically, 6–10 L/min. Because colorimetric ETCO2 detectors have historically been used to determine the success of intubation, which can be done in just a few breaths, little attention has been paid to their resistance profile.
It has long been accepted that the resistance of the endotracheal tube (ETT) is a factor in gas exchange and the work of breathing in neonatal patients.4,5 The filling and emptying of the lung is dependent on the time constant (airway resistance × respiratory system compliance). Imposed resistance through any or all airways can lead to ventilation-perfusion mismatch, increased work of breathing, ineffective gas exchange, and hyperinflation.6
The purpose of this study was to test colorimetric ETCO2 detectors to determine their resistance, which would play an important role in their use over a longer period of time in spontaneously breathing infants.
QUICK LOOK
Current knowledge
Previous studies have demonstrated that colorimetric ETCO2 detectors are helpful in confirming successful endotracheal intubation. Current practice is evolving to also utilize these devices during neonatal resuscitation with a mask interface. Colorimetric ETCO2 detectors are effective at identifying airway obstruction in spontaneously breathing premature neonates. One device, the Neo-StatCO2, is labeled for neonates <1 kg, but the resistance in the normal operating range is not available in publications or the product literature.
What this paper contributes to our knowledge
In an in vitro model, the Neo-StatCO2 had a significantly greater resistance than 2 other colorimetric ETCO2 detectors used in neonatal resuscitation. The resistance of the Neo-StatCO2 ETCO2 detector was comparable with that of a 3.0 ETT.
Methods
We compared the resistance of 3 colorimetric ETCO2 detectors, Neo-StatCO2 and Mini-StatCO2 (Mercury Medical, Clearwater, Florida) and Pedi-Cap (Covidien, Mansfield, Massachusetts), and 2 mainstream capnograph sensors, the EMMA infant airway adapter 17449, a portable ETCO2 sensor (Masimo Corporation, Irvine, California) and the neonatal/infant airway adapter YG-213T (Nihon Kohden America, Foothill Ranch, California). Four standard Portex neonatal ETTs, size 2.5–4.0-mm inner diameter (Smiths Medical, Hythe, United Kingdom), which were cut to a length appropriate for clinical use, were also tested to help establish a reference range. The manufacturer's specifications for the ETCO2 devices can be found in Figure 1. Three samples of each ETCO2 detector were tested and compared. The devices were placed in a flow circuit distal to the transducer with the opposite end of the device open to air (Fig. 2). All 3 samples of each device were tested once at flows of 1–10 L/min in 1-L increments of dry, room temperature air, using a flow meter (Precision Medical, Northampton, Pennsylvania). The flow meter has an accuracy of ±0.5 L/m from 0 to 5 L/min and ±1.0 L/min from 5 to 15 L/min. The flow meter was not calibrated because flow was set before each run and was constant across all measurements. Hence, the accuracy of the flow signal was not critical to the findings. Three measurements were taken of proximal airway pressure, one from each of the 3 device samples, and the resistance was calculated as the change in pressure over the change in flow divided by 60 and expressed as cmH2O/L/s. Pressure was measured using a differential pressure transducer and a universal interface module attached to an MP150 data acquisition system (Biopac Systems, Goleta, California). The TSD-160C transducer is a ±25 cm H2O differential transducer, with accuracy of ±0.5%. A 2-point calibration was done at 0 and 20 cm H2O before recording. All data were sampled at 200 Hz. Post-collection analysis was completed using the accompanying Acqknowledge software.
Data Analysis
Data are expressed as mean ± SD. Data analyses were performed using SPSS 22.0 for Windows (IBM, Armonk, New York). The level of statistical significance was established a priori as P < .05. Descriptive analyses were used to evaluate frequencies and distributions of variables and potential outlier values. Scatterplots were examined to evaluate assumptions of normality, linearity, and homogeneity of variance. The mean and median values were comparable at each level and consistency among standard deviations, suggesting the use of parametric methods for subsequent analyses. Additional post hoc confirmation regarding appropriate use of the measure of central tendency for the 3 values was conducted using non-parametric testing (Kruskall-Wallis and median tests), and these tests led to similar statistical conclusions regarding acceptance or rejection of null hypotheses. Two-tailed Student t tests were used to compare the overall mean values obtained for each device as compared with the Neo-StatCO2 device. Repeated measures analysis of variance was used to test whether the means obtained at the 10 observations of the 5 devices were equal. Bonferroni post hoc adjustment was used to reduce the likelihood of type-1 error.
Results
The mean ± SD values of all ETCO2 devices across all flows are provided in Table 1. The mean resistance of the Neo-StatCO2 compared with the other 4 devices was different (P < .001), and these differences persisted at variable flows (1–10 L/min). These differences also persisted (P < .001) after adjusting for each flow rate by repeated measures analysis of variance.
We found that the Neo-StatCO2 had the greatest resistance of all of the ETCO2 devices tested. At the commonly utilized flow of 10 L/min, the resistance of the Neo-StatCO2 was 61.1 cm H2O/L/s, comparable with that of a 3.0 ETT, which we measured at 62.7 cm H2O/L/s. The Neo-StatCO2 detector had approximately 5–6 times greater resistance to air flow, on average across all flows, than the other 2 calorimetric ETCO2 detectors. It had approximately 8 times the resistance of the EMMA airway adapter and 22 times the resistance of the Nihon Kohden airway adapter. As flow increased, so did resistance for all of the devices tested (Fig. 3).
Discussion
Colorimetric ETCO2 detectors are primarily used to recognize whether an ETT is accurately placed. Current practice is evolving to also utilize these devices with a mask interface during CPAP. Although the effects of artificial airways on spontaneous breathing during mechanical ventilation are well studied, their effects on work of breathing during bag mask ventilation or applied CPAP are not.7 Resistance to flow is a critical factor when providing ventilation or CPAP to the spontaneously breathing infant. Imposed work of breathing can occur when a patient must breathe spontaneously through an artificial apparatus placed between the lungs and the air, such as ETTs, ETCO2 detectors, flow sensors, and expiratory valves.6 With an increase in the work of breathing, a neonate may experience diaphragmatic fatigue, apnea, and respiratory failure.8,9 The resistance profiles of neonatal ETTs are well documented, and as a result, ETT CPAP is not routinely performed in neonates.4,5,10 One of the devices we measured, if used in a non-intubated infant, may cause similar imposed work of breathing as if the infant were intubated with a 3.0-mm ETT and breathing spontaneously.
A colorimetric ETCO2 detector can be a valuable tool to assist the clinician with identifying when the airway is not patent and airway repositioning is necessary. However, for color change to occur on a colorimetric ETCO2 detector, the infant's exhaled gas must be in contact with the CO2-sensitive paper for a sufficient amount of time. This is done by either creating a longer very shallow path through which gas must flow, as in the Pedi-Cap, or by passing the gas through tiny holes in the detector, as in the Neo-StatCO2 (Fig. 4).
In addition to enough contact time, adequate tidal volume is necessary to create a recognizable color change in colorimetric ETCO2 detectors. A device with a larger dead space volume may require larger tidal volumes to show a color change than one with a smaller dead space. Small volumes may get washed out in the dead space of the device, producing a false negative result.1,11 The Neo-StatCO2 is the first colorimetric detector labeled for infants <1 kg. The product literature from the manufacturer on the resistance of this new detector is limited to 5 L/min. We compared the Neo-StatCO2 with other colorimetric ETCO2 devices in the flow range in which they are most widely utilized, 6–10 L/min. To our knowledge, there are no prior publications validating the manufacturer's specifications (Fig. 1). The dead space has been decreased and the resistance to flow has increased in the Neo-StatCO2 compared with the Mini StatCO2. There are no studies demonstrating that these design changes have improved the sensitivity to the presence of ETCO2, improved response time, or decreased the possibility of a false negative reading.
We previously studied both the Pedi-Cap and the Mini StatCO2 in an artificial lung model to determine the minimal tidal volume necessary to generate a color change.12 The Mini StatCO2 tidal volume threshold was 0.83 mL, and the Pedi-Cap tidal volume threshold was 1.08 mL, both amply sensitive for the expected tidal volume of a 400-g infant. Both detectors are appropriate for use with any premature neonate. In another study of 45 neonates, 19 subjects were <1 kg; the Pedi-Cap correctly identified 18 tracheal intubations with only one false negative. The authors recommended the Pedi-Cap for all neonates, including babies <1 kg.13 All 3 of the false negatives in the overall study population had Apgar scores of <2 or cardiac arrest. ETCO2 detectors are susceptible to false negative readings secondary to inadequate circulation as well as false positive readings due to contact with medications such as epinephrine.13–15
There are currently no recommendations for any one method of ETCO2 monitoring for neonates.14 There are limited data regarding the perceived benefits of one colorimetric ETCO2 device over another. Hawkes et al16 found that two thirds of neonatal and pediatric trainees preferred the Neo-StatCO2 because of its horizontal positioning, but they also felt that the visual indicator was smaller and more frequently covered by the hand of the operator when compared with the Pedi-Cap. They found no significant differences in efficacy between devices.
Our study was an in vitro design, and as such we did not test our devices on spontaneously breathing newborns. Further testing in the clinical environment will be helpful in determining the exact effect of these high-resistance devices in the spontaneously breathing neonate.
Conclusion
Detection of ETCO2 in newborns after delivery has been critical for advancing resuscitative techniques and improving the efficiency of responses to airway emergencies. We have found that some detectors have significant resistance, equivalent to ETTs. It is important for clinicians to recognize the increased resistance caused by ETCO2 detectors and to discontinue their use when they are no longer required. Clinicians also need to be aware that there is potential impact on work of breathing from utilizing ETCO2 detectors in spontaneously breathing newborns on mask CPAP. Future trials are needed to determine the effects of this resistance on work of breathing, particularly on very premature newborns receiving mask CPAP.
Footnotes
- Correspondence: Melissa K Brown RRT-NPS, Neonatal Research Institute, Sharp Mary Birch Hospital for Women and Newborns, 3003 Health Center Drive, San Diego, CA 92123. E-mail: melissa.brown{at}sharp.com.
Mr Rich has disclosed a relationship with Discovery Laboratories. The other authors have disclosed no conflicts of interest.
Ms Gonzales presented an abstract of this paper at the Pediatric Academic Societies Annual Meeting, held April 25–28th, 2015, in San Diego, California.
See the Related Editorial on Page 1129
- Copyright © 2016 by Daedalus Enterprises