Abstract
BACKGROUND: Although accurate quantification of oxygen consumption (V̇O2) and carbon dioxide elimination (V̇CO2) provides important insights into a patient's nutritional and hemodynamic status, few devices exist to accurately measure these parameters in children. Therefore, we assessed the accuracy and agreement of 2 devices currently on the market using a pediatric in vitro model of gas exchange.
METHODS: We utilized a Huszczuk simulation model, which simulates oxygen consumption and carbon dioxide production using gas dilution, to examine the accuracy of two FDA-cleared respiratory modules (E-COVX and E-sCAiOVX-00). V̇O2 and V̇CO2 were set at 20, 40, 60, and 100 mL/min, ranges typical for infant and pediatric patients. Bland-Altman analysis was used to calculate the bias and limits of agreement of each device relative to simulated values for V̇O2 and V̇CO2.
RESULTS: The E-COVX mean percentage bias (limits of agreement) was −26.3% (−36.1 to −16.6%) and −39.3% (−47.5 to −31.1%) for V̇O2 and V̇CO2, respectively. The mean bias (limits of agreement) for the E-aCAiOVX-00 was −0.5% (−13.3 to 12.3%) and −6.0% (−13.8 to 1.7%) for V̇O2 and V̇CO2, respectively.
CONCLUSIONS: The E-COVX demonstrated bias and limits of agreement that were not clinically acceptable; therefore, application of this module to pediatric patients would not be recommended. The new module, E-sCAiOVX, demonstrated acceptable bias and limits of agreement for the V̇O2 and V̇CO2 in the range 40–100 mL/min (which corresponds to patients in the range of ∼5–16 kg).
Introduction
The accurate quantification of oxygen consumption (V̇O2) and carbon dioxide elimination (V̇CO2) in critically ill, mechanically ventilated children is central to the assessment of nutritional status and hemodynamics.1–3 However, an accurate and broadly useful tool to quantify V̇O2, in particular, has remained elusive. Most devices available on the market are not specifically designed to measure gas exchange in children, and accuracy data are rarely reported.4 Gas exchange parameters are essential for understanding energy requirements in critically ill patients and in estimating cardiac output using the Fick equation and CO2 rebreathing.1,5–7 Recently, alternative methods for estimating energy expenditure have arisen that attempt to obviate the need to accurately measure V̇O2 by measuring V̇CO2 and utilizing a consistent assumed respiratory quotient.8,9 Further, due to the challenges of accurately quantifying V̇O2, many pediatric institutions estimate (rather than measure) this vital parameter based on healthy controls, but this practice is known to cause significant errors in cardiac output estimates.10,11
Although investigations have been conducted on previous generations of devices, recent improvements may affect the accuracy of measurements.12,13 We therefore sought to compare the accuracy of these 2 modules within the range of expected V̇O2 and V̇CO2 values for pediatric patients using an experimental model of simulated respiration.
QUICK LOOK
Current knowledge
Increasingly, technology to measure oxygen consumption and carbon dioxide elimination is being incorporated into mechanical ventilators. These devices can offer insights into cardiopulmonary performance and nutrition status. Importantly, the technological limitations of this technology must be understood to ensure that clinical application is appropriate and accurate, especially in the pediatric range, since most devices were designed primarily for adult use.
What this paper contributes to our knowledge
Of the two devices tested, only the newest module demonstrated a bias and limits of agreement that were within an a priori determined clinically acceptable range of ±20%. The limits of agreement were clinically acceptable from 40 to 100 mL/min. This range corresponds to approximate patient weights of 5-16 kg.
Methods
Gas Exchange Simulation
As described previously,14,15 we constructed a bench-top simulation model that allows precise control of respiratory mechanics and gas exchange parameters. The Huszczuk method is based on mass conservation and utilizes a calibrated gas injection system to simulate V̇O2 and V̇CO2.15 Briefly, a mass flow controller (FMA-2619A, Omega Engineering, Stamford, Connecticut) was utilized to control the injection of a specialized test gas composed of 21% CO2/balance nitrogen into a customized baffle and reservoir assembly; the accuracy of the mass flow controller is ±0.8% of the measurement reading. A precisely known volume of 21% CO2/balance nitrogen is pumped into the reservoir during exhalation. During inspiration, this volume of gas mixes with a measured (by the device being tested) volume of ambient air (21% oxygen/balance nitrogen); this dilutes the concentration of oxygen from inspired air by a known quantity, creating a precisely known lower quantity of expired oxygen than inspired oxygen, simulating oxygen consumption. Carbon dioxide elimination (V̇CO2) was calculated by multiplying the CO2 fraction of the test gas by the flow (0.21 × test gas flow). Since the expired oxygen concentration (FeO2) is diluted by the addition of CO2, the simulated V̇CO2 is equal to V̇O2 (this relationship is only true for a test gas whose carbon dioxide concentration is precisely 0.21/balance nitrogen, making this a convenient gas for this model). As an example, the calculation of the required test gas flow required to simulate 40 mL/min V̇CO2 is shown below: (1) (2) (3) where FCO2 is the fraction of CO2 in the test gas and k is a gas density correction factor. The correction factor is only required if the mass flow controller utilized is not specifically calibrated to the test gas (in this case, the k term can be ignored). Therefore, the set test gas flow would be 190 mL/min to simulate a V̇CO2/V̇O2 of 40 mL/min. In the present investigation, a correction factor, k = 0.9578 was used because the mass-flow controller was precalibrated for N2 (the test gas concentration is 0.9578 times as dense as N2). The QuickLung Breather (Ingmar Medical, Pittsburgh, Pennsylvania) was used to simulate tidal volumes, inspiratory times, breathing frequencies, and compliance. Compliance was set to 6 mL/cm H2O, resistance was set to 50 cm H2O/L/s, breathing frequency was set to 20 beats/min, and FIO2 was set to 33%.16
Gas Exchange Monitors
We compared the E-COVX and E-sCAiOVX-00 gas exchange modules (GE Healthcare, Helsinki, Finland) attached to a critical care monitor (Carescape B450, GE Healthcare, Helsinki, Finland). The device was calibrated according to the manufacturer's specifications. The device utilizes a fixed orifice pressure differential pneumotachometer that is preconfigured for pediatric use and does not require flow calibration. The sidestream paramagnetic O2 analyzer and near-infrared CO2 sensor were calibrated with a multipoint gas calibration. The appropriate pediatric airway adapters were used for the specified breathing frequency and tidal volumes.
Experimental Protocol
To examine the accuracy of these modules within the ranges expected for infants and children, we simulated V̇O2 and V̇CO2 values and tidal volumes according to Table 1. After each combination of gas exchange and tidal volumes was selected on the simulator, an equilibration period was instituted. This ensured that measured values had reached steady state before data collection began (this period lasted ∼5 min). For each combination, V̇O2 and V̇CO2 were measured for 10 min (no. = 20 replicates/combination).
Statistical Analysis
We computed bias and limits of agreement for each combination shown above, comparing set V̇O2 and V̇CO2 (as the reference standard) with measured values of each using a Bland-Altman analysis (GraphPad Prism 5.04, GraphPad Software, La Jolla, California). We decided a priori that a ±20% limit of agreement would be deemed clinically acceptable.19 For each device, we determined the minimum acceptable value that exhibited limits of agreement within the clinically acceptable range for both V̇O2 and V̇CO2.
Results
Differences between measured V̇O2 and V̇CO2 and set values for each were calculated for each experimental replicate, representing the bias of each measurement. Bias and limits of agreement for each set of conditions are depicted in Table 2, Figure 1, and Figure 2. Overall, the mean bias (95% CI of limits of agreement) for the E-COVX across the ranges tested were −26.3% (−36.1 to −16.6%) for V̇O2 and −39.3% (−47.5 to −31.1%) for V̇CO2. The newer module, E-aCAiOVX, exhibited a mean bias of −0.5 (−13.3 to 12.3) for V̇O2 and −6.0 (−13.8 to 1.7) for V̇CO2. In the E-aCAiOVX, but not the E-COVX, the limits of agreement for V̇O2 and V̇CO2 were within the clinically acceptable range of ±20% for all set values except for 20 mL/min.
Discussion
We report the accuracy and agreement of 2 available gas exchange monitoring modules from the same manufacturer that can be used alone or incorporated into a mechanical ventilator. The results demonstrate acceptable performance (mean bias and 95% limits of agreement) of the updated module (E-sCAiOVX) for V̇O2 and V̇CO2 values in the range of 40–100 mL/min with mean inspiratory flows ranging from 3.0 to 5.0 L/min. The older device (E-COVX) did not demonstrate agreement with the simulated values and therefore is not clinically acceptable for use in pediatric patients. The findings for the E-sCAiOVX are more favorable than those described in the manufacturer's specifications. In contrast, the older device (E-COVX) exhibited significantly greater negative bias across the spectrum of ranges tested.
It should be noted that our results of the E-COVX module differed from a prior report on the M-COVX module, which reported a <1% bias at simulated V̇O2 as low as 50 mL/min.10 However, we believe that there are important differences in the way in which set V̇O2 and V̇CO2 and respiratory parameters were controlled that account for these differences. However, since the modules themselves do function slightly differently, it is possible that actual differences in the performance of the modules account for some of the difference.
Recently, Wines et al20 studied the accuracy of the V-max Encore metabolic cart (CareFusion, San Diego, California). The investigators utilized a commercially available gas exchange simulator (VacuMed, Ventura, California). The principle of operation is similar to that employed in the present investigation. However, the VacuMed device was designed for adult applications, and therefore Wines et al20 simulated pediatric V̇O2 and V̇CO2 levels but with tidal volumes and respiratory parameters that were much larger than would be observed in children. The authors tested gas exchange from 20 to 204 mL/min. They reported strong correlations between set and measured gas exchange values and tight limits of agreement (±7 mL/min). Unlike the present investigation, which noted reduced accuracy of the E-sCAiOVX at the 20-mL/min gas exchange range with inspiratory flow ∼1.8 L/min, the maximum inspiratory flow demonstrated tight limits of agreement at this range. However, since maximum inspiratory flow data were collected with adult respiratory parameters, their utility is limited, especially since the performance of flow measurement (pneumotachographs) at low inspiratory and expiratory flows is known to adversely affect the accuracy of such a system. For this reason, we sought to construct our system to ensure that pediatric respiratory parameters were incorporated.
The etiology of the performance differences between the 2 modules is not clear. Both iterations of the device utilize a paramagnetic oxygen sensor and a non-dispersive near-infrared CO2 sensor. The technical specifications outlined by the manufacturer are consistent between the two, including an accuracy of V̇O2 and V̇CO2 each of ±10% or 10 mL total (whichever is larger); we found the accuracy to be superior to this, albeit under idealized conditions as described. The only clearly visible difference is that the sampling rate in the newer device has been decreased from 200 to 120 mL/min. This decrease could reduce the entrainment of ambient gas and may decrease the dilution of inspired versus expired gas difference, improving accuracy at low V̇O2 and V̇CO2 measurements. It is also possible that the measurement algorithm (which matches spirometry and gas measurements to calculate V̇O2 and V̇CO2) may have improved.
The putative benefits of a commercially available device capable of measuring V̇O2 and V̇CO2 accurately and continuously in children cannot be overstated. In children with congenital heart disease, for example, devices may overestimate V̇O2 (and cardiac output by a proportionate amount). This may significantly impact patient assessment, both in the cardiac catheterization laboratory and in the ICU.2,12,13 Thus, although estimates of V̇O2 and/or V̇CO2 for simulated values of 20 mL/min fell outside the 20% limit of agreement, it is still possible that estimations of V̇O2 and V̇CO2 provided by this device would represent an improvement in the estimation of select parameters compared with standard equations. Other benefits include the titration of respiratory support and accurate assessment of energy expenditure for the tailored prescription of nutrition.1,17,18 This is a subject that merits further study, although the lack of an accepted standard measure in patients makes determination of absolute accuracy difficult.
There are limitations to the present investigation. The experiment was not conducted on human subjects. Factors, including humidification, minute-to-minute variation of gas exchange values (such as occurs due to hemodynamic variability), and temperature of the circuit were not addressed. Humidification could not be incorporated into the model, since water condensate would adversely affect the performance of the mass-flow controller. Further, the effects of ventilator modes and changes in respiratory parameters are known to affect gas exchange parameters.21 We took important steps to minimize these limitations by selecting equipment that offered very good accuracy for controlling set simulated gas exchange flow, ensuring that leaks did not exist in the test apparatus and selecting combinations of pulmonary mechanics and set gas flows that corresponded to observed pediatric values. Importantly, there are advantages to conducting the investigation in a controlled laboratory environment, namely the ability to precisely control set V̇O2 and V̇CO2 and respiratory parameters, therefore obviating the effect that patient variability has on the interpretation of results if the experiment were to be conducted in humans. Further, since a human subject's V̇O2 and V̇CO2 cannot be precisely controlled, it is impossible to design a study that seeks to demonstrate the accuracy of a gas exchange monitor in vivo. This study is a comparison of the agreement between different devices. We selected a discrete range of test gas exchange values, and our conclusions are bound to this range. Further, we did not assess the affect of differences in FIO2 or of ventilator mode on device accuracy, either of which are known to impact these end points.11 In practice, none of these may be true, particularly in children who are weaning from the ventilator and are therefore breathing spontaneously; thus, these results must be applied to clinical practice with caution. Further investigation with methods able to incorporate increased gas exchange variability and humidification may be indicated to assess these effects on accuracy and confirm the findings of the present investigation. Nonetheless, these data provide an estimate of accuracy under idealized conditions.
Conclusions
The new module, E-sCAiOVX, demonstrated acceptable bias and limits of agreement for the V̇O2 and V̇CO2 in the range of 40–100 mL/min (which corresponds to patients in the range of ∼6–15 kg). The E-COVX demonstrated bias and limits of agreement that were not clinically acceptable, and therefore application of this module to pediatric subjects cannot be recommended in the ranges tested. Further investigations regarding the effects of patient factors, including humidity, tachypnea, and variations in inspired oxygen fractions, are warranted.
Footnotes
- Correspondence: Craig D Smallwood RRT, Division of Critical Care Medicine, Department of Anesthesiology, Pain and Perioperative Medicine, MSICU Office, Bader 634, Boston Children's Hospital, 300 Longwood Avenue, Boston, MA 02115. E-mail: craig.smallwood{at}childrens.harvard.edu.
The authors have disclosed no conflicts of interest.
- Copyright © 2017 by Daedalus Enterprises