Abstract
BACKGROUND: Noninvasive ventilation (NIV) is commonly used in neonates. A mode of NIV called neurally adjusted ventilatory assist (NAVA) offers patient-ventilator interactions by using electrical activity of the diaphragm to control mechanical breaths. We hypothesized that the work of breathing (WOB) would decrease with NIV-NAVA. Secondary objectives evaluated the impact of NIV-NAVA on arterial blood gases and respiratory parameters.
METHODS: We compared WOB between synchronized breaths in NIV-NAVA and NIV in piglets with healthy lungs and then with surfactant-depleted lungs. Neonatal pigs (median, 2.0 [range, 1.8–2.4] kg) with healthy and then surfactant depleted lungs were sedated and ventilated with NIV-NAVA and NIV in random order. Airway flow and pressure waveforms were acquired. Waveforms were analyzed for the pressure-time product that reflected WOB. The primary outcome between modes was assessed with repeated measurement analysis of variance.
RESULTS: The pressure-time product was significantly decreased for NIV-NAVA in both healthy and injured lungs (P < .001). PaO2, PaCO2, inspiratory tidal volume, and peak inspiratory flow were not different in either model.
CONCLUSIONS: Synchronized breaths during NIV-NAVA resulted in decreased WOB compared with synchronized breaths during NIV.
- neurally adjusted ventilatory assist
- NAVA
- noninvasive ventilation
- work of breathing
- NIV
- respiratory physiology
Introduction
Intubation and mechanical ventilation have significantly improved the survival rate of neonates. These treatments, however, are not without risk. Acute complications include air-leak syndromes, airway trauma, infection, and subglottic stenosis.1,2 Intubation and mechanical ventilation have also been shown to be associated with long-term risks, for example, bronchopulmonary dysplasia,3 and have been indicated as risk factors associated with retinopathy of prematurity.4,5 Currently, there is agreement that invasive mechanical ventilation should be minimized or avoided completely to reduce the risk of ventilator-induced lung injury.6 Use of noninvasive ventilation (NIV) in the preterm infant has been reported to reduce the need for intubation, decrease lung injury, and decrease the risk of long-term complications, for example, bronchopulmonary dysplasia.7,8
Commonly used NIV strategies in the neonatal ICU include nasal CPAP and ventilation (NIV). Nasal CPAP applies a constant distending pressure level (above the atmospheric pressure level) during inhalation and exhalation to support spontaneously breathing neonates. NIV has a constant distending pressure with intermittent superimposed inflations. A recent meta-analysis compared NIV with CPAP and found that NIV decreases the need for intubation and reduces the incidence of extubation failure.9
Neurally-adjusted ventilator assist (NAVA) is a mode of assisted ventilation that uses the electrical activity of the diaphragm (EAdi) to control the ventilator. The EAdi represents the neural effort, both with respect to timing and amplitude.10,11 With NAVA ventilation, the EAdi signal can trigger and cycle a positive-pressure breath independent of airway leak.10–12 NAVA can be used as an invasive form of ventilation as well as noninvasively (NIV-NAVA).10,13 Improved synchrony with NAVA over NIV has been documented.14–16 However, to our knowledge, there are no data that indicate whether synchronized NIV-NAVA breaths may result in less work of breathing (WOB) than synchronized NIV breaths.
Our study used a piglet model. A number of studies have shown that the relative maturity of the piglet brain, lungs, and cardiovascular system is comparable with the human neonate.17–21 Eiby et al21 were able to further define similarities between piglets and humans at specific gestations. We hypothesized that WOB during synchronized breaths would be decreased with NIV-NAVA compared with NIV in both healthy and surfactant-depleted lungs in a piglet model. Our secondary objectives were to compare the effects of NIV-NAVA and NIV on arterial blood gases and respiratory parameters.
QUICK LOOK
Current knowledge
Because of the risks associated with invasive ventilation in a neonatal population, clinical practice is shifting to an emphasis on noninvasive ventilation (NIV) strategies in this subset of patients. NIV and neurally-adjusted ventilatory assist (NAVA) improve patient-ventilator synchrony when compared with other commonly used forms of neonatal NIV.
What this paper contributes to our knowledge
In a neonatal pig model, NIV-NAVA decreased the work of breathing during synchronous breaths when compared directly with synchronous breaths given with NIV. Decreased work of breathing was demonstrated in both healthy neonatal pig lungs and surfactant-depleted neonatal pig lungs.
Methods
The animal review committee of the University of Arkansas for Medical Sciences approved this study. The work was conducted at Arkansas Children's Hospital Research Institute in Little Rock, Arkansas. The animals were cared for in accordance with the standards for care and use of laboratory animals set forth by the university. Twelve neonatal pigs were used for this evaluation. The pigs were male, domestic porcine anesthetized with a mixture of tiletamine, ketamine, and xylazine (0.04 mL/kg) intramuscularly and Forane gas (Baxter Healthcare Corp, Deerfield, Illinois) (3–4%) via mask (Canine Anesthesia Mask, Kent Scientific, Torrington, Connecticut). Dissection to the external jugular vein was performed and a central venous line placed for drug administration (dexmedetomidine 6–10 μg/kg/h plus fentanyl 3–5 μg/kg/h) during the experiments. The level of sedation was clinically monitored to maintain the animal in a light sleep, defined as not moving but breathing spontaneously and with a consistent EAdi tracing. In addition, an arterial catheter was placed in the carotid artery for blood gas and blood pressure monitoring.
Heart rate, oxygen saturation, and temperature were monitored with a physiologic monitor (SC 9000 XL, Siemens, Dräger, Germany). Each pig was randomized to NIV-NAVA or NIV, and evaluated first in the healthy state and then again after lung injury. Thus, each pig was its own control for comparison between NIV-NAVA and NIV, and between healthy and surfactant depleted states so to minimize variability (Fig. 1). An EAdi catheter was placed in all the pigs and remained in place throughout the experiment. The position of the EAdi catheter was verified by using the position screen on the ventilator as well as the observation of the EAdi signal itself. This position was monitored routinely throughout the study. Before any data collection, the position was confirmed in the same manner.
The pigs were ventilated (via mask) by using the Servo-i ventilator (Maquet, Getinge Group, Rastatt, Germany) with randomly applied NIV-NAVA and NIV with 3 cm H2O PEEP and FIO2 of 1.0. In healthy porcine lungs, PEEP was optimized at 3 cm H2O to avoid suppression of the respiratory drive. For NIV-NAVA, the median (range) set NAVA level was 0.4 (0.3–0.5) cm H2O/μV for the healthy lung and 0.5 (0.3–1.0) cm H2O/μV for the partially injured lung. The NAVA level was adjusted according to the EAdi signal, similar to a clinical setting. According to the manufacturer of NAVA, the usual NAVA level is 0.5–2.0 cm H2O/μV, with EAdi signals between 5 and 15 μV. These parameters were used for adjustment of our study subject's NAVA level.
The set EAdi trigger (median, range) was 0.5 (0.2–0.5) μV for the healthy lung and 0.5 (0.3–0.6) μV for the partially injured lung. For NIV, pressure control was set to achieve a target tidal volume (VT) of 8–10 mL/kg, which is optimal in the piglet mode. For each mode of ventilation, a stabilization period of 30 min was followed by a 5-min data collection period. The pigs were then intubated with a 3.0 cuffed endotracheal tube and were ventilated on baseline settings; pressure control mode, target VT of 8–10 mL/kg, PEEP of 3 cm H2O and FIO2 of 1.0. After 10 min of ventilation, an arterial blood gas sample was obtained to document PaO2, added together with PCO2 in the arterial blood (PaCO2) of > 400 mm Hg; dynamic compliance was also documented.
To produce a partially injured lung model, the pigs were subject to a saline solution washout.22 The saline solution washout was completed with volumes of warm normal saline solution (30 mL/kg) administered in 1–3 instillations through a tight tracheal cannula via gravity drainage into and out of the lungs. Pigs were then ventilated with baseline settings for 5 min and dynamic compliance documented. Lungs were considered partially injured when compliance was 50–70% of baseline. The pigs were then extubated and ventilated (via mask) with randomly applied NIV-NAVA and NIV as with the healthy lung. We used a published method to set PEEP with a computerized conventional ventilation recruitment tool available on the Servo-i ventilator.23 The median (range) set PEEP was 5 (3–8) cm H2O for the partially injured lung.
Measurements of respiratory flow and pressure waveforms were acquired by using the Biopac MP-100 System (Biopac Systems, Santa Barbara, California). A heated 0–35 L/min pneumotachograph (PNT) (Hans Rudolph, Shawnee, Kansas) was placed in line between the mask and the ventilator. Volume measurements were obtained through the computer by integrating the flow signal. Volume was verified with a calibrated syringe (Hans Rudolph). To monitor pressure, the pneumotachograph was equipped with a pressure hose, barb-type port that allowed airway pressure sampling. Pressure was calibrated with a manometer (Air Flow Developments, Wycombe, United Kingdom). All output signals were routed via an analog channel box into the Biopac MP100 data acquisition unit, which converted them into digital signals. Signals were obtained at a rate of 500 samples per second. Gas humidification was maintained by an in-line Fisher & Paykel MR850 humidification system (Fisher & Paykel, Irvine, California). Data were collected for each mode (NIV-NAVA and NIV) for both healthy and partially injured lungs. For each measurement period, there was a stabilization period of 30 min, followed by data collection for 5 min.
In the NIV mode, all breaths are pneumatically triggered. Flow and pressure waveforms obtained with the pneumotachograph at the airway were analyzed to identify evidence of subject-ventilator asynchrony. Flow asynchrony was determined by identifying the presence of concavity (≥1 cm H2O) in the pressure-time waveform during inspiration.24 Only synchronous breaths were analyzed. In the NIV-NAVA mode, breaths are either pneumatically triggered, which results in a pressure support breath, or neurally triggered. NAVA is triggered on a first-come-first-served basis between a neural signal, which is determined by a deflection in the EAdi signal, and a pneumatic trigger, which is caused by a deflection in airway pressure. Maximum pressure that can be achieved in NIV-NAVA is 30 cm H2O. Classification of breaths was determined via video recording of the ventilator screen during experiments. Trigger indication is displayed in the message/alarm field of the screen of the ventilator and is displayed as a pneumatic or neural trigger for each breath. Breaths from the video recording of the ventilator screen were matched timewise to breaths collected from the pneumotachograph at the airway opening and documented as pneumatically or neurally triggered for 1 min. Only neurally triggered synchronous breaths were analyzed.
Pressure waveforms were analyzed for pressure-time product (PTP), as a measure of WOB. PTP area A was defined as the area of the pressure curve (integration of pressure with respect to time) from initiation of a breath (zero flow) to the beginning of ventilator pressurization (most negative deflection of pressure). PTP area A reflects the subject's WOB or the work required for the subject to trigger a breath. PTP area B was defined as the area of the pressure curve from the beginning of ventilator pressurization (most negative deflection of pressure) to return to baseline pressure. PTP area B reflects both subject and ventilator work, or the work needed for the subject and the ventilator to complete a breath that has already been initiated. PTP area 1 is area A and area B combined. PTP area 1 reflects the effort of the subject to trigger a breath as well as the initial work performed by both the subject and the ventilator. A graphic depiction can be seen in Figure 2.
Flow and pressure waveforms were analyzed for inspiratory VT, peak inspiratory flow, and peak pressure. Inspiratory VT was recorded rather than expiratory VT due to the inherent nature of leaks with NIV. Inspiratory VT, peak inspiratory flow, peak pressure, and PTP were measured for 10 neurally triggered breaths for NIV-NAVA and 10 synchronous breaths for NIV. Asynchronous breaths for any mode were excluded from the analysis. Arterial blood gases were analyzed by using i-STAT (Abbott, Princeton, New Jersey), a blood gas analyzer system. Primary outcomes of interest were PaO2 (mm Hg) and PaCO2 (mm Hg). Blood gas samples were collected for both NIV and NIV-NAVA modes after the 5-min data collection period with both the healthy and partially injured lung. Mean arterial blood pressure (mm Hg) and heart rate (beats/min) were monitored continuously with a physiologic monitor (SC 9000XL, Siemens, Dräger Medical, Telford, Pennsylvania). Breathing frequency (f), EAdi peak (μV), mean EAdi (μV), leak fraction (%), and dynamic compliance (mL/cm H2O) were downloaded from the ventilator.
Each outcome was summarized as mean ± SD separately by lung status and mode of ventilation. PTP areas between the modes of ventilation were assessed by repeated measurement analysis of variance models in healthy and injured lung models separately. Secondary outcomes (f, leak fraction, peak EAdi, mean EAdi mean, heart rate, PaO2, PaCO2, inspiratory VT, peak inspiratory flow, and peak pressure) were similarly assessed. Statistical significance was defined as P < .05. All statistical analyses were performed with Stata v 14.2 (StataCorp, College Station, Texas).
Results
The 12 pigs weighed a mean (range) of 2.0 (1.8–2.4) kg. There was a significant difference in PTP area A, area B, and area 1 between NIV-NAVA and NIV in both lung models, as seen in Table 1 and Figure 3. There was no difference in PaO2, PaCO2, inspiratory VT, and peak inspiratory flow. Peak pressure was statistically higher in the NIV-NAVA group in both healthy and injured lungs. The EAdi peak was not different between the groups, although it was nearly significant in the healthy lung. The median (range) dynamic compliance was 1.9 (1.3–2.2) mL/cm H2O for healthy lung (baseline) and 1.0 (0.7–1.4) mL/cm H2O for the partially injured lung. There was no difference in f, heart rate, or mean arterial blood pressure between the groups. There was also no difference in leak fraction or mean EAdi. These results are shown in Table 2.
Discussion
Our study demonstrated that WOB in synchronous breaths, as measured by PTP, was significantly less for NIV-NAVA breaths versus NIV breaths in both the healthy and injured lung. WOB, in general terms, represents an assessment of the afterload on the respiratory muscles or opposition to their contraction.25 One common way of measuring WOB uses an esophageal balloon to measure esophageal pressure. This method of measuring WOB relies on pressure and volume changes but cannot accurately measure WOB when there is ineffective respiratory effort. PTP more closely approximates energy expenditure of the respiratory muscles.26 PTP estimates the metabolic cost of breathing and quantifies inspiratory effort.27–29 PTP evaluates energy expenditure during both isometric and non-isometric contraction.27,29 PTP is calculated by integrating pressure and time for the duration of contraction of the respiratory muscles.
Lee et al14 showed that NIV-NAVA improved patient-ventilator synchrony and diaphragmatic unloading in preterm infants in comparison with NIV, despite large air leaks. Chen et al30 demonstrated that NIV-NAVA improves patient-ventilator synchrony and reduces diaphragmatic muscle load and WOB in preterm infants in comparison with NIV.30 Beck et al16 summarized 15 studies in pediatric subjects and demonstrated improved patient-ventilator interaction with NAVA, whether invasive or noninvasive, compared with conventional modes. To our knowledge, our study was the first to show that the WOB with the synchronous breaths of NIV-NAVA was less than that of synchronous breaths that may occur with NIV.
These findings were likely explained because EAdi signal allows more or less support with each breath, depending on patient demand. For NIV, support is not patient driven; breaths are set by the operator with regard to timing and pressure, and are consistent with each breath. Thus, the operator makes a “best guess” as to the support required by the patient. With NIV, increased patient demand results in more flow delivery but not more pressure. With NIV-NAVA, the patient can demand and receive higher pressures, and also has the option to take sigh breaths with variable inspiratory times, which allow better lung recruitment. These factors likely decrease WOB in synchronized NIV-NAVA breaths compared with NIV breaths, even if the NIV breaths are synchronized with the baby's breathing pattern. An elevated EAdi signal represents patient demand for increased respiratory support.
Lee et al showed a decrease in peak pressure with NIV-NAVA compared with non-invasive pressure support.14 In our study, we found peak pressure to be statistically higher in the NIV-NAVA group; however, the difference was very small and likely not clinically important. In addition, there were no decreases in f, heart rate, or mean arterial blood pressure, whereas significant differences in the WOB were found. Our study time period may have been too short to detect decreases in peak pressure that may have occurred over time, or changes in f, heart rate, or mean arterial blood pressure. WOB is difficult to assess during NIV. Use of esophageal balloons and calibrated pneumotachography is cumbersome and time consuming. PTP is an estimate of the WOB but is not possible if interaction with the ventilator does not occur. In our study, it was very important to note that WOB was not analyzed in asynchronous breaths. Thus, overall WOB may vary and must be evaluated by other means. Clearly, it is important to know that the synchronized breaths that occur with NIV-NAVA result in decreased WOB compared with NIV under the best of circumstances (synchrony).
Study Limitations
This was an animal model study performed in a controlled laboratory environment and, as such, may not directly translate to the neonatal population and ICUs. Also, as noted, we did not analyze the WOB in asynchronous breaths. If patients have long stretches of backup ventilation during NIV-NAVA, which accompanied increases in asynchronous breathing, could result in increased WOB even if NIV-NAVA is the set mode. Piglets needed to be lightly sedated to perform the study; sedation may have influenced the results, although using each piglet as its own control should have provided consistency.
Conclusions
In the piglet model, synchronous breaths delivered with NIV-NAVA resulted in decreased WOB in both healthy and surfactant depleted lungs when compared with synchronous breaths during NIV. Further trials of NIV-NAVA in neonates are warranted.
Footnotes
- Correspondence: Sherry E Courtney MD, Arkansas Children's Hospital, One Children's Way, Slot 512-5 Little Rock, AR 72202. E-mail: scourtney{at}uams.edu.
Financial support was provided by a Children's University Medical Group institutional grant.
Dr Jones presented the data at the American Academy of Pediatrics National Conference and Exhibition, held October 22–25, 2016, in San Francisco, California; Hot Topics in Neonatology Conference, held December 5–7, 2016, in Washington DC; Diamond Conference, held April 28, 2017, in Little Rock, Arkansas.
The authors have disclosed no conflicts of interest.
See the Related Editorial on Page 1579
- Copyright © 2018 by Daedalus Enterprises