Primary use of nasal CPAP in premature infants with respiratory distress syndrome has been associated with improved outcomes.1,2 Avoiding intubation and invasive ventilation may reduce injury, inflammation, pulmonary growth arrest, and development of chronic lung disease.3,4 However, nearly half of all neonates initially supported with CPAP will develop respiratory failure severe enough to require intubation for short-term invasive ventilation and surfactant administration.5
Noninvasive ventilation (NIV) is a form of respiratory assistance that provides greater support than CPAP and may prevent intubation in a larger fraction of neonates who would otherwise fail CPAP and receive invasive ventilation. NIV combines CPAP with positive pressure during inspiration to offload intrinsic efforts, and improve ventilation and lung recruitment.1 NIV breaths may be patient triggered and, thus, “synchronized,” or non-synchronized. NIV has been used to avoid intubation6–8 and facilitate extubation.9 NIV may be more lung protective and promote alveolar growth and development better than invasive ventilation.10,11 Compared with CPAP, NIV has been associated with greater tidal volume and minute ventilation, reduced thoraco-abdominal asynchrony,12 lower breathing frequency, improved gas exchange,13 and reduced work of breathing (WOB)14,15 as well as reduction in the magnitude and/or severity of apneic episodes9,16 and the need for invasive ventilation.17
Although these NIV findings seem promising, premature infants in many of these studies were primarily supported with synchronized NIV, not non-synchronized NIV, by using a ventilator that is no longer commercially available. The Infrasonics Infant Star ventilator with the StarSynch module was unique in that it permitted synchronized NIV based on pressure changes from diaphragm descent within an abdominal (Graseby) capsule attached below the xiphoid process. This method for synchronized NIV reportedly provided consistent triggering, independent of leak.18
The abdominal capsule continues to be used with synchronized NIV devices in Europe.19 After discontinuation of the Infant Star, clinicians have been left with very few options for providing neonatal synchronized NIV in the United States. Ventilator manufacturers have attempted to support the neonatal synchronized NIV niche market and address the need by incorporating synchronized NIV modes with advanced flow-triggering algorithms. However, these synchronized NIV modes may have difficulty with providing synchrony in premature infants.20 There is a delicate technologic balance between having a system that is sensitive enough to detect infants' weak efforts while still being able to avoid auto-triggering in the presence of variable nasal and/or oral leaks. Patient WOB, synchrony, and other outcomes with flow-triggered synchronized NIV have not been studied clinically.
Manufacturers continue to explore novel synchronized NIV triggering options with surface electromyography, inductance bands, hot-wire flow sensors, and other noninvasive signals. Clinicians are seeking a safe, accurate, and reliable synchronized NIV mode now, more than ever before, to prevent intubation and avoid lung injury. Although synchrony is preferred, it is difficult to measure and may not always be associated with improved outcomes when compared between synchronized and non-synchronized NIV. Early studies that enrolled hundreds of premature infants and that compared synchronized with non-synchronized invasive modes of ventilation, showed no differences in major outcomes or complications.21 Because of the technologic limitations and lack of synchronized NIV modes, many clinicians continue to have success with using non-synchronized NIV modes, and, thus, the superiority of synchronized NIV over non-synchronized NIV remains inconclusive.22 New research has left many clinicians wondering whether synchronized NIV modes are even necessary or if synchrony is needed to improve outcomes in premature infants. This debate has sparked national controversy across American neonatal ICUs over the past decade.
Currently, the most promising synchronized NIV mode for premature infants is NIV neurally adjusted ventilatory assist (NAVA). In theory, NIV-NAVA should result in optimal synchrony because it provides triggering and the timing and magnitude of the positive pressure breath is based on the electrical activity of the diaphragm acquired from an indwelling feeding tube.23,24 There is a paucity of clinical data related to NIV-NAVA but, based on findings from animal23 and human studies,25,26 it seems to be a reasonable and safe synchronized NIV approach. A common outcome or theme from these studies was that NIV-NAVA has been shown to improve synchrony. One study that compared NIV-NAVA with non-synchronized NIV in premature infants did not show notable differences in treatment failure or adverse events.27
In this issue of Respiratory Care, Jones et al28 added to the growing body of NIV-NAVA evidence by comparing gas exchange and estimated WOB in spontaneously breathing piglets before and after surfactant lavage between synchronized NIV and NIV-NAVA. They found no differences in gas exchange, but there were differences in WOB, which favored NIV-NAVA.28 Investigators evaluated pressure-time product (PTP) as a surrogate for WOB.28 This method has previously been described to evaluate patient-ventilator asynchrony and WOB based on observed airway pressure changes with airway graphics in adults.29 Other investigators described a similar PTP method for calculating imposed WOB.30,31 However, calculated values of respiratory effort when using airway pressure signals may not always correlate well with direct measurement of work (esophageal pressure) and may actually underestimate the work being done by the patient.32,33 Also, this PTP calculation28,29 has not been validated in premature infants undergoing NIV.
It is important to note that airway pressure changes during NIV may be related to ventilator performance and less so to patient effort. NIV modes use leak compensation algorithms that will rapidly add flow to the circuit during the respiratory cycle. In these cases, the ventilator's contribution to airway pressure may limit detection of a premature infant's efforts, and calculated PTP may be underestimated. Studies that compare this PTP calculation with airway pressure need to be compared with direct measurement of PTP when using esophageal pressure. Lastly, PTP from the current study was calculated only for synchronized NIV breaths that were triggered by the subjects.28 Because all of the non-synchronized or non-subject–triggered breaths were excluded from analysis, important information about the subjects' WOB and ability to make an effort and actually receive a supported breath with synchronized NIV or NAVA is unknown.28 These data would have added to the utility of the article. NIV-NAVA may have many physiologic advantages over synchronized NIV, which has been shown to have a high asynchrony index.20
According to Subirà et al,34 accurate interpretation of cycle asynchronies would require monitoring not only usual ventilator waveforms but also esophageal pressure or electrical activity of the diaphragm. Esophageal pressure monitoring is the accepted method for direct measurements of patient-ventilator interaction. PTP that uses esophageal pressure is a validated method as an index of WOB estimated from esophageal pressure and is calculated by integrating pressure and time for the duration of contraction of the respiratory muscles.31 This measurement approximates energy expenditure of the respiratory muscles and closely estimates the metabolic cost of breathing and quantifies inspiratory effort for both isometric and non-isometric contraction.
In several studies, PTP and pressure-rate product measured with esophageal pressure monitoring has been used to evaluate indices of WOB in infants on NIV.18,35–38 Subject synchrony and measured PTP could have easily been addressed in the current study to evaluate synchronized and non-synchronized breaths. It is unclear why most neonatal studies that involve NIV-NAVA focus on synchrony and WOB in the absence of esophageal pressure. Esophageal pressure changes can still be measured when using existing gastric tubes39 or when an NIV-NAVA catheter23 is being used.
An important consideration when interpreting the findings of the present study28 is that synchrony was evaluated only for those breaths that were synchronized in the subjects. As such, the most important and clinically meaningful findings related to the subjects' inability to effectively synchronize intrinsic efforts with corresponding mechanical breaths were not included. Also, the subjects were sedated, interfaced with an oronasal mask, and used pressure triggering for synchronized NIV with a ventilator that has been shown to have high asynchrony in a bench model.20 So, it is unclear how these findings may be translated clinically.
Based on the current study and other studies that evaluated NIV-NAVA, results overwhelmingly support improved synchrony. There currently is no clinical association or correlation between asynchrony and an increased risk for death, much less chronic lung disease or any other complications in premature infants. Based on these observations, it could be argued that there is not enough evidence to suggest that NIV-NAVA or any other form of synchronized NIV is superior to non-synchronized NIV. There are remaining questions: (1) should patient-ventilator interaction or synchronization even be considered an important outcome in infants during NIV, and (2) is synchronized NIV or NIV-NAVA needed to provide NIV support in premature infants, even when non-synchronized NIV has still been shown to be effective?
Findings from short-term physiologic studies on non-synchronized NIV in premature infants demonstrated that asynchrony is common and may result in missed triggers, auto-triggering, or poor volume delivery to the lungs.40,41 Although synchronized NIV has been shown to reduce respiratory effort, no differences in hypoxic events, apneas, tidal volume, minute ventilation, and abdominal girth have been demonstrated when compared with non-synchronized NIV.18 In the absence of patient effort, non-synchronized NIV in subjects with apnea, peak pressure has been shown to occasionally result in chest inflation and ameliorated oxygen desaturations.42 Patients are able to adapt to the ventilator within a short period, especially if the ventilator rate is set to at least 50% of their total breathing frequency.15
One clinical study showed lower incidence of chronic lung disease and length of support in premature infants with non-synchronized NIV than CPAP after extubation.43 A similar study showed a higher prevalence of re-intubation, post-extubation atelectasis, and mortality with premature infants supported with CPAP than with non-synchronized NIV.44 In a retrospective study that compared synchronized NIV with non-synchronized NIV in a cohort of 200 premature infants, after adjusting for disease severity, there were no differences in the combined outcome of bronchopulmonary dysplasia and death.45 Based on these findings with non-synchronized NIV, synchrony may not be important for predicting patient outcomes with NIV in premature infants.
Future studies are needed that use methods to measure esophageal pressure to determine whether NIV-NAVA reduces PTP and asynchrony in subjects. NIV-NAVA is likely to be very useful in the neonatal ICU setting but, until other clinically important outcomes, aside from synchrony, can be compared with non-synchronized NIV or other synchronized NIV modes, the technical complexity of use and cost related to this invasive NIV strategy will still need to be considered.