Abstract
BACKGROUND: Use of negative pressure ventilation is neither well described nor widespread in pediatric critical care; existing data are from small, specialized populations. We sought to describe a general population of critically ill subjects with acute respiratory failure supported with negative pressure ventilation to find predictors of response or failure.
METHODS: We conducted a retrospective cohort study of subjects 0–18 y old admitted to a single (non-cardiac) pediatric ICU who received acute respiratory failure support via negative pressure ventilation from May 2015 through May 2016.
RESULTS: In 118 subjects, the most common causes of acute respiratory failure were viral bronchiolitis (86.4%) and pneumonia (15.3%). A majority of subjects (68.6%) stabilized with negative pressure ventilation and did not need a change of respiratory support; in those who failed with negative pressure ventilation, median time to respiratory support change was 5.1 h (interquartile range 1.9–11.0). Subjects stabilized with negative pressure ventilation did not differ from those needing a change of respiratory support in terms of age, comorbidities, or FIO2 at initiation of ventilation. Compared to those who did not respond to negative pressure ventilation, mean SpO2/FIO2 was higher at 1 h after start of negative pressure ventilation (218.8 vs 131.7) in those who did respond. Subjects with SpO2/FIO2 < 192 after 1 h on negative pressure ventilation support had 5-fold higher odds of needing a respiratory support change (odds ratio 5.143, 95% CI 1.17–22.7, P = .031). Analysis of SpO2/FIO2 was limited by 81.3% (96/118) of subjects who had an SpO2 > 97% at 1 h after the start of negative pressure ventilation.
CONCLUSIONS: Negative pressure ventilation successfully supported 69% of pediatric subjects with all-cause acute respiratory failure. Oxygen requirement was lower in subjects who were responsive to negative pressure ventilation within 1 h of initiation. Standardized negative pressure ventilation protocols should include weaning of supplemental oxygen to determine responsiveness.
- respiratory failure
- pediatrics
- negative pressure ventilation
- biphasic cuirass ventilator
- artificial respiration
Introduction
Negative pressure ventilation, a mode of noninvasive mechanical ventilation (NIV), applies sub-atmospheric pressures to the chest to augment the natural intrathoracic pressure generated during spontaneous breathing.1 Modern negative pressure ventilation devices (eg, the biphasic cuirass ventilator) are smaller, less obstructive to clinical assessments and care, and less detrimental to hemodynamics than the original negative pressure ventilator, the iron lung.
Negative pressure ventilators, such as the biphasic cuirass ventilator, apply negative pressure onto the chest wall to increase transpleural pressure and recruit areas of atelectasis while improving preload to the heart.2 In neonatal models, this led to dilation of the pulmonary capillaries and improved ventilation-perfusion matching.3 The biphasic cuirass ventilator can also apply positive pressure onto the chest wall to augment lung recoil and assist exhalation. This active exhalation is unique to the biphasic cuirass ventilator when compared to other forms of invasive or NIV currently in use.
Early studies found that the biphasic cuirass ventilator decreased FIO2 requirements and increased P22 clearance within hours, and its use was associated with lower likelihood of intubation in small pediatric populations.2,4 In the largest pediatric negative pressure ventilation study to date, 70% of 233 critically ill subjects treated with a biphasic cuirass ventilator in a single pediatric ICU had resolution of acute respiratory failure with negative pressure ventilation with a 3% complication rate.5
As proven in other forms of NIV, ways to detect treatment response are crucial before widespread application of negative pressure ventilation in critically ill patients can be recommended.6 For example, adults using bilevel positive airway pressure intubated within 12 h had lower hospital mortality than those whose intubation was delayed past that point.7 Noninvasive measures such as SpO2/FIO2 have been successful in other populations in predicting NIV failure.8
After more than 5 years of negative pressure ventilation use in our center, the biphasic cuirass ventilator is the most commonly used NIV device for pediatric acute respiratory failure. This robust, single-center experience provides an opportunity for research to inform other centers and protocol development. We performed this study to determine if early objective variables predicted response to negative pressure ventilation support. In addition, negative pressure ventilation use is not standardized in our institution because pediatric data to guide protocol development are lacking, specifically related to negative pressure ventilation settings and triggers for escalation or weaning. Therefore, this study also served to collect data for protocol development.
QUICK LOOK
Current knowledge
Negative pressure ventilation provides respiratory support by augmenting the strength of the chest wall. Newer devices are available but are not used widely in pediatric critical care despite possible advantages over other noninvasive or invasive respiratory modalities.
What this paper contributes to our knowledge
In a single-center retrospective cohort study of a heterogeneous pediatric population with acute respiratory failure, a majority of children supported with negative pressure ventilation did not require intubation. Children who responded to negative pressure ventilation had lower oxygen requirements within the first hour compared to those who went onto more support; however, supplemental oxygen was not weaned routinely.
Methods
This retrospective cohort study included all patients < 19 y of age admitted to Women and Children's Hospital, Buffalo, New York, a 20-bed pediatric ICU, with acute respiratory failure supported by negative pressure ventilation over 1 year (May 2015 to May 2016). The time frame was chosen to follow a previous cohort study as a validation dataset for a model to predict intubation.5 The only exclusion criterion was negative pressure ventilation use for chest physiotherapy only. The Institutional Review Board of the University at Buffalo approved this study with a waiver of consent.
Respiratory Support Modalities
Support for pediatric acute respiratory failure in our hospital during the study period began with high-flow nasal cannula (HFNC) in the emergency room or on the pediatric floor, and negative pressure ventilation was added if a subject were to clinically worsen. In all subjects, negative pressure ventilation was added to HFNC because negative pressure ventilation cannot deliver supplemental oxygen. Therefore, the data before initiation of negative pressure ventilation reflect the support provided with HFNC, whereas data after the addition of negative pressure ventilation reflect the effect of simultaneous support from negative pressure ventilation and HFNC. Because there was no protocol in place for weaning FIO2 while on HFNC in our institution, supplemental oxygen was titrated for low SpO2 values and often not weaned for excessive SpO2 values.
If a subject did not respond to negative pressure ventilation, other noninvasive modalities included nasal intermittent positive-pressure ventilation via a cannula, CPAP, or bilevel positive airway pressure via face mask or nasal interfaces. Subjects who failed negative pressure ventilation were placed more frequently on invasive mechanical ventilation instead of other NIV modalities.
Negative Pressure Ventilation Strategies
As previously mentioned, our institution did not have protocols dictating initiation, titration, or weaning of negative pressure ventilation. The only existing institutional protocol during this study relating to negative pressure ventilation was scheduled assessment for skin breakdown at contact sites with the chest cuirass, performed every 4 h by a respiratory therapist and bedside nurse. Practices related to sedation, nutrition, laboratory studies, and the biphasic cuirass ventilator pressure settings were dictated by the discretion of pediatric critical care providers.
The biphasic cuirass ventilator used was the Hayek RTX respirator (Hayek Industries, London, United Kingdom). It applies a plastic shell, or cuirass, over the anterior chest wall from the sternal notch to the top of the abdomen and across to each mid-axillary line. The cuirass seals to the chest with a disposable foam rim. It has two available modes. Continuous negative extrathoracic pressure, the most common mode, is similar to CPAP, holding a continuous negative pressure against the chest wall, typically ranging from −8 to −24 cm H2O. Control mode, which is akin to bilevel positive airway pressure, alternates negative pressure with positive pressure to force chest wall recoil, assisting in expiration. Control pressures are titrated in a 3:1 ratio typically (eg, −18 to +6 cm H2O. Control mode delivers these pressures at a mandatory frequency set higher than the patient's spontaneous frequency and cannot synchronize with patient efforts. Because there is no interface with the mouth or nose, the biphasic cuirass ventilator cannot provide supplemental oxygen. HFNC was left in place and flows and FIO2 were not routinely weaned at the time of biphasic cuirass ventilator start.
Data Collection
Data collected include demographics, past medical history, admission diagnosis, comorbidities (ie, history of asthma or reactive airways, chronic lung disease, or prematurity using physician documentation of these conditions), vital signs, any available blood gas results, and level of respiratory support 1 h prior to and 1, 4, 8, and 12 h after initiation of negative pressure ventilation. A previous study5 of negative pressure ventilation in similar pediatric populations reported median time to negative pressure ventilation failure of ∼ 7 h; therefore, data collection regarding predictor variables stopped at 12 h after start of negative pressure ventilation. The primary outcome was negative pressure ventilation failure, which was defined as a change in respiratory support from negative pressure ventilation as determined by providers at the time. Secondary outcomes included previously reported and theoretical complications of negative pressure ventilation, such as hypothermia, skin breakdown, pneumothorax, gastroesophageal reflux, delayed enteral nutrition (> 48 h), and agitation requiring sedation.5
Statistical Analysis
Appropriate measures of central tendency and comparison testing were used after Kolmogorov-Smirnov testing for normality of data. Regression modeling adjusted the measures of association for confounders and formed prediction models. Area under the receiver operating characteristic curves tested power to predict outcomes. Analyses were performed with SPSS 24 (IBM, Armonk, New York) with significance defined as P < .05.
Results
The study population consisted of 118 subjects aged 1–192 months who were supported with negative pressure ventilation for acute respiratory failure for any cause (Table 1). Eighty-one subjects (68.6%) stabilized with negative pressure ventilation and did not need intubation or a change in respiratory support. In those who did not stabilize with negative pressure ventilation, failure occurred at a median of 5.1 h (IQR 1–3, 1.9–11.0); 30 subjects required intubation (81% of subjects were did not respond to negative pressure ventilation; 25.4% of all subjects). The remaining 7 subjects were rescued with bilevel positive airway pressure or a nasal cannula. The only differences in baseline characteristics of subjects responsive to negative pressure ventilation compared to those who required a different mode of ventilation were proportion with upper airway pathology and PRISM score. Having a history of upper airway pathology or narrowing increased the odds of negative pressure ventilation failure 6-fold (odds ratio 6.016, 95% CI 1.109–32.629, P = .038). There were no reported cases of intrathoracic airway narrowing, such as tracheo- or bronchomalacia.
Characteristics for Subjects With Acute Respiratory Failure From Any Cause Supported With Negative Pressure Ventilation
Predictor Variables for Negative Pressure Ventilation Response
There were no differences in the baseline clinical variables between subjects who responded and those who did not respond to negative pressure ventilation, with the exception of initial venous pH (Table 1).
Within an hour of initiation of negative pressure ventilation, mean FIO2 was lower in those responsive to this ventilation than those who required escalation (0.47 ± 0.18 vs 0.56 ± 0.26, P = .02), and this value remained lower at 4, 8, and 12 h (Fig. 1). Having a FIO2 < 0.60 at 1 h carried 74.1% specificity, 32.4% sensitivity, and a negative predictive value of 72.3% for negative pressure ventilation response. If using FIO2 at 1 h to predict negative pressure ventilation failure, the area under the curve–receiver operating characteristic was 0.603, 95% CI 0.490–0.715, P = .08 (Fig. 2).
Trend in mean FIO2 after start of negative pressure ventilation over time (in hours) in subjects who did and did not require a change in respiratory support. *Differences reached statistical significance as defined by P < .05 via independent sample t tests comparing the mean FIO2 between groups. NPV = negative pressure ventilation.
Receiver-operator characteristic curve to predict negative pressure ventilation response by FIO2 as supplied by high-flow nasal cannula after the first hour that subjects were supported by negative pressure ventilation. Area under the curve = 0.603, 95% CI 0.490–0.715, P = .08.
Without a FIO2 weaning protocol, the majority of subjects had a SpO2 ≥ 97% at every time point measured. The proportion of subjects with a SpO2 ≥ 97% before negative pressure ventilation initiation was 67.8% (80/118), and this increased to 81.4% (96/118) at 1 h after starting negative pressure ventilation support. Calculating SpO2/FIO2 only in those subjects with SpO2 < 97% (n = 22) did show separation between negative pressure ventilation responders (213.8 ± 81.8) compared to those with negative pressure ventilation failure (181.9 ± 80.3) at 1 h after start of negative pressure ventilation; this did not reach statistical significance (P = .40).
Using SpO2/FIO2 at 1 h to predict negative pressure ventilation response did have better predictive power than FIO2 at 1 h, area under the curve–receiver operating characteristic = 0.702 (95% CI 0.503–0.900, P = .050) (Fig. 3). The optimal coordinate on this curve was SpO2/FIO2 = 192, with 72% specificity and 67% sensitivity for negative pressure ventilation failure with a negative predictive value of 81.8% for negative pressure ventilation failure. The unadjusted odds of negative pressure ventilation failure in those with SpO2/FIO2 < 192 at 1 h was 5.143 (95% CI 1.17–22.7, P = 0.031). After adjusting for age and sex, the odds of negative pressure ventilation failure with SpO2/FIO2 < 192 at 1 h was 3.953 (95% CI 0.844–18.521, P = .08).
Receiver-operator characteristic curve to predict response to negative pressure ventilation support by SpO2/FIO2 at 1 hour after start of negative pressure ventilation (including only those subjects with an SpO2 < 97%). Area under the curve = 0.702, 95% CI 0.503–0.900, P = .050.
Blood gas data were available for 29 (24.5%) subjects. Subjects responsive to negative pressure ventilation tended to have lower venous P22 values at a baseline (42 ± 6.6 mm Hg vs 49 ± 13.2 mm Hg, P = .066) and through 8 h after start of negative pressure ventilation. This difference was significant at 1 h after initiation when responsive subjects had a mean venous P22 of 42 ± 7.4 mm Hg, whereas those with negative pressure ventilation failure had a mean venous P22 of 55 ± 19.7 mm Hg, P = .03.
Subgroup Analysis of Subjects With Bronchiolitis
To determine if certain predictors existed for a more homogenous subject population, we performed a secondary analysis of subjects with bronchiolitis excluding those with a history of upper airway disease (n = 96) (Table 2). A larger proportion of subjects with negative pressure ventilation failure (6/29, 21%) had a history of chronic lung disease compared to those who responded to negative pressure ventilation (4/67, 6%) (P = .030). Of the viruses identified, respiratory syncytial virus was the most common (n = 51, 53.1%), followed by human metapneumovirus (n = 10, 10.4%). There was no difference in negative pressure ventilation response based on the viral source: respiratory syncytial virus (P = .69), human metapneumovirus (P = .54), adenovirus (P = .41), influenza (P = .49), or parainfluenza (P = .14).
Characteristics for Subjects With Acute Respiratory Failure From Viral Bronchiolitis Supported With Negative Pressure Ventilation
Blood gas data were available for 20 subjects in this secondary analysis: 19.4% (13/67) of subjects who responded to negative pressure ventilation, and 24.1% (7/29) of subjects who did not respond to negative pressure ventilation. Mean initial pH was higher in negative pressure ventilation responders than in non-responders (7.38 ± 0.05 vs 7.33 ± 0.05, P = .042). Initial venous P22 tended to be lower in responders, but the difference did not reach statistical significance (41 ± 6 mm Hg vs 50 ± 15 mm Hg, P = .062).
Initial FIO2 was similar between the groups but diverged at 1 h: 0.43 ± 0.15 in those who responded to negative pressure ventilation versus 0.51 ± 0.22 in non-responders (P = .048). In bronchiolitis, FIO2 at 1 h had fair predictive power for failure of negative pressure ventilation (ie, area under the curve–receiver operating characteristic = 0.612, 95% CI 0.495–0.729, P = .07). With similar baseline values before negative pressure ventilation, the mean SpO2/FIO2 at 1 h tended to be higher in those who responded to negative pressure ventilation (250.4 vs 218.4, P = .058).
Safety Data
In all subjects, the complication rate of negative pressure ventilation was 1.7% (ie, 2 cases of hypothermia in neonates). There were no instances of pneumothorax or breakdown in the skin from the foam ring around the cuirass. There was no reported gastroesophageal reflux or vomiting, including those receiving enteral nutrition while negative pressure ventilation was in place (5 subjects were fed by mouth, 3 via gastric tube).
Eighty percent of subjects on negative pressure ventilation support received continuous sedation (dexmedetomidine), with a lower proportion of those with negative pressure ventilation failure receiving sedation. This was likely because of the shorter duration of negative pressure ventilation. There was no difference in the proportion with delayed enteral nutrition between the groups.
Discussion
Our results show that the majority of pediatric subjects with all-cause acute respiratory failure responded to negative pressure ventilation without a need for intubation or transition to other forms of noninvasive support. The biphasic cuirass ventilator had a low complication rate and a response rate equivalent to other forms of NIV,9 with the exception of subjects with upper airway pathology. Because the interface allows for suctioning and oral intake and a rescue rate comparable to other forms of NIV, this modality should be considered a viable option for widespread use. Although we were unable to identify definitive objective predictors of negative pressure ventilation failure using this cohort, our results suggest these predictors exist.
As seen in published HFNC scoring systems that rely heavily on symptoms like retractions, signs of progressing pediatric acute respiratory failure have largely remained subjective.10 Our results imply that objective differences could exist if assessments of gas exchange were made more often in children with acute respiratory failure. Even with limited blood gas and SpO2/FIO2 data, there were differences in gas exchange in those responding to and those failing NIV within the first hour of support. These findings are consistent with our previous negative pressure ventilation data5 and other NIV studies.9,11 In a large, multi-center trial of 5,847 adults with acute hypoxemic respiratory failure, PaO2:FIO2 < 146 after 1 h of NIV was associated with 2-fold higher odds of NIV failure (odds ratio 2.51, 95% CI 1.45–4.35).9 In the subset of subjects in whom SpO2/FIO2 was calculable, we found double those odds for children with acute respiratory failure receiving negative pressure ventilation support using SpO2/FIO2 at 1 h. Other studies have reported a failure to reduce the PaCO2 in 1 h after NIV start as an important sign of NIV failure.12,13 Our limited blood gas data suggest similar is true for pediatric subjects supported by negative pressure ventilation.
Biologic plausibility exists for these findings. NIV, unlike HFNC, augments transpleural pressure to allow recruitment of areas of atelectasis, thus leading to optimal pulmonary blood flow and correction of ventilation/perfusion mismatch. Whether accomplished with positive pressure via face mask or negative pressure on the chest wall, both bilevel positive airway pressure and negative pressure ventilation restore functional residual capacity and improve lung function. As in other pulmonary disease states like ARDS, severity of lung disease is described by surrogates of lung function (ie, gas exchange). If negative pressure ventilation can improve lung function through recruitment, then gas exchange in the form of SpO2/FIO2 and P22 should improve. Our data, like others, indicate that this happens within 1 h of initiation of support.
Infrequent measurements of P22 and failure to wean supplemental oxygen in the majority of subjects restricted our sample size. Lack of blood gas data were not surprising because serial blood gas measurements are not standard of care in children with bronchiolitis due to its low disease acuity, excellent overall prognosis, and low mortality rate. This is likely why current scoring systems in this disease process remain subjective, as mentioned above. Our findings related to supplemental oxygen have revealed an opportunity for practice improvement that was not apparent at the onset of this study. Although only one fourth of the cohort had blood gas data, an even smaller proportion were weaned with high SpO2 levels to allow for accurate SpO2/FIO2 calculations. In addition to avoiding unnecessary oxygen exposure, making standardized oxygen weaning protocols could reveal the subset of patients who would benefit from more regular blood gas measurements or earlier escalation of respiratory support. Our results illustrate the ubiquitous practice of supplemental oxygen administration, with our pediatric ICU as an example for other centers. In an era of stewardship and conscious application of therapies to limit morbidity, our dataset emphasizes the importance to assess with reliable markers of clinical status, such as blood gas measurement, and to wean potentially dangerous therapies, such as supplemental oxygen.
Our findings are limited by the retrospective and single-center nature of this study. Without set protocols to obtain blood gases or to wean oxygen, clinical behaviors likely created a selection bias. Because FIO2 was likely to be increased and blood gas data were obtained in sicker subjects, this could have led to an overestimation of the true measure of association. Larger prospective studies with consistent sampling and protocolized weaning of oxygen in subjects supported with negative pressure ventilation are needed to validate these findings.
Conclusions
Negative pressure ventilation successfully supported 69% of pediatric subjects with all-cause acute respiratory failure with a complication rate < 2%. Oxygen requirement was lower in subjects who were responsive to negative pressure ventilation within 1 h of initiation. Standardized protocols for titration of negative pressure ventilation should include weaning of supplemental oxygen and routine blood gas measurements to guide prompt escalation of care.