American Association for Respiratory Care
  • Subscribe
  • My alerts
  • Log in
American Association for Respiratory Care

Advanced Search

  • Home
  • Content
    • Current Issue
    • Editor's Commentary
    • Archives
    • Most-Read Papers of 2022
  • Authors
    • Author Guidelines
    • Submit a Manuscript
  • Reviewers
    • Reviewer Information
    • Create Reviewer Account
    • Reviewer Guidelines: Original Research
    • Reviewer Guidelines: Reviews
    • Appreciation of Reviewers
  • CRCE
    • Through the Journal
    • JournalCasts
    • AARC University
    • PowerPoint Template
  • Open Forum
    • 2023 Call for Abstracts
    • 2022 Abstracts
    • Previous Open Forums
  • Podcast
    • English
    • Español
    • Portugûes
    • 国语
  • Videos
    • Video Abstracts
    • Author Interviews
    • Highlighted Articles
    • The Journal
  • Twitter
  • Facebook
  • YouTube
Research ArticleOriginal Research

Heated Humidifier versus Heat-and-Moisture Exchanger During Positive Pressure Ventilation With a T-Piece Resuscitator in Rabbits

Agustín Bernatzky, María F Galletti, Santiago E Fuensalida and Gonzalo L Mariani
Respiratory Care September 2020, 65 (9) 1295-1300; DOI: https://doi.org/10.4187/respcare.06997
Agustín Bernatzky
Division of Neonatology, Department of Pediatrics, Hospital Italiano de Buenos Aires, Ciudad de Buenos Aires, Argentina.
Instituto Universitario, Hospital Italiano de Buenos Aires, Ciudad de Buenos Aires, Argentina.
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • For correspondence: [email protected]
María F Galletti
Division of Neonatology, Department of Pediatrics, Hospital Italiano de Buenos Aires, Ciudad de Buenos Aires, Argentina.
Instituto Universitario, Hospital Italiano de Buenos Aires, Ciudad de Buenos Aires, Argentina.
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Santiago E Fuensalida
Department of Anesthesiology, Faculty of Veterinary Science, Universidad de Buenos Aires, Buenos Aires, Argentina.
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Gonzalo L Mariani
Division of Neonatology, Department of Pediatrics, Hospital Italiano de Buenos Aires, Ciudad de Buenos Aires, Argentina.
Instituto Universitario, Hospital Italiano de Buenos Aires, Ciudad de Buenos Aires, Argentina.
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • Article
  • Figures & Data
  • Info & Metrics
  • References
  • PDF
Loading

Abstract

BACKGROUND: There are many proven benefits of the use of conditioned gases in mechanically ventilated patients. In spite of this, its use in the delivery room is limited, perhaps because of known difficulties with heated humidifiers (HH); moreover, there is no evidence regarding the use of heat-and-moisture exchangers (HME) in a delivery room setting. We sought to asess the airway's absolute humidity level using three diferents strategies: HH, HME and unconditioned gases.

METHODS: We conducted an experimental study in 12 intubated rabbits ventilated with a T-piece resuscitator. Absolute humidity levels in inspired gases were measured at baseline and at 5, 10, 15, and 20 min while using HH, HME, or no conditioning method (ie, unconditioned). The animals were initially randomized to one of the 3 interventions, and each animal underwent the other methods with at least 24 h between each test.

RESULTS: There were no differences in vital signs at baseline or at the end of the procedures. Mean absolute humidity at the end of the tests was 38.2 ± 1.7 g/m3 for HH, 28.9 ± 4.7 g/m3 for HME, and 13.9 ± 5.1 g/m3 for unconditioned gas (P = .003).

CONCLUSIONS: During ventilation with a T-piece resuscitator, the absolute humidity was the highest with HH. The absolute humidity with HME was lower, but it was still significantly more than that with unconditioned gas. Therefore, the use of a T-piece resuscitator with HME could be a good alternative to HH given that positive-pressure ventilation is used ideally for short periods of time in the delivery room.

  • gas conditioning
  • respiratory support
  • delivery room
  • newborn resuscitation
  • humidification

Introduction

The administration of cold and dry gases directly into the lower respiratory tract can cause lung damage.1–4 Mechanically ventilated neonates are particularly sensitive to lung injury. Furthermore, the use of unconditioned inspired gases has been associated with an increased severity of bronchopulmonary dysplasia.5–7

The respiratory loss of water and heat can cause hypothermia in neonates, and it has been reported that the use of active humidification associated with a T-piece resuscitator during initial stabilization of premature infants decreases the incidence of hypothermia on admission to the neonatal ICU.8–10 Admission hypothermia is an independent risk factor for mortality in preterm infants and neonatal morbidities.11–13

However, cold and dry gases are still used in many delivery rooms for the resuscitation of newborn patients.14 The heated humidifier (HH) system, which is the only device proposed for the delivery room, takes approximately 9 min to reach full operative functionality, which makes it inadequate for emergency situations.15 In addition, the HH requires a power source; if the neonatal ICU and the delivery room are far apart, absolute humidity levels would decrease during transport.

The heat-and-moisture exchanger (HME) could be a practical alternative to HH. Although there are studies that have tested the use of HME devices in mechanically ventilated neonatal and pediatric subjects, there is no evidence to date of the use of these devices associated with a T-piece resuscitator.16–18 Because HME devices require adequate humidification and temperature of the exhaled gases, it is difficult to evaluate the system in an in vitro model. An experimental model in animals was used as the initial stage for the evaluation of the system.

The objective of this study was to compare the absolute humidity levels in the airway of ventilated rabbits using a T-piece resuscitation device with active humidity (ie, HH), a passive system (ie, HME), or no conditioning at all.

QUICK LOOK

Current knowledge

The administration of cold and dry gases directly into the lower respiratory tract can cause lung damage in neonates. The use of active humidification via a heated humidifier associated with a T-piece resuscitator during initial stabilization of premature infants decreases the incidence of hypothermia on admission to the neonatal ICU.

What this paper contributes to our knowledge

In an animal model of ventilation with T-piece resuscitator, the absolute humidity obtained with the active humidification system was higher and within the recommended range. Although lower values of absolute humidity were obtained with the heat-and-moisture exchange system, these values were higher than without conditioning. Results of this study support the recommendation to discontinue the use of nonconditioned gases in the delivery room.

Methods

Study Design and Procedures

This randomized controlled study in animals was performed in December 2017 at the Institute of Basic Sciences and Experimental Medicine of the Hospital Italiano de Buenos Aires, Argentina. The research protocol was approved by both the Experimental Research Committee and the Ethics Committee of Research Protocols of the Hospital Italiano de Buenos Aires (Protocol E/131). The experiments were performed in adherence to the Canadian Council on Animal Care Guidelines on the Use of Laboratory Animals.

Adult New Zealand rabbits were anesthetized and intubated with neonatal endotracheal tubes size 3 and 3.5 mm internal diameter with and without balloon, without cutting the tube (Rusch, Teleflex, Wayne, Pennsylvania). The endotracheal tube size was determined by the veterinarian at the time of endotracheal intubation, and the same endotracheal tube size was used in each test for each animal. In all cases, animals were treated with intramuscular injections of 2 mg/kg midazolam and 10 mg/kg ketamine. Then we placed a venous access for induction with 5 mg/kg propofol. Once the endotracheal intubation was performed, the anesthetic plane was maintained with an infusion of 1% propofol at a dose of 15 mg/kg/h, which could be titrated to keep the animals in a deep anesthetic plane throughout the duration of the tests. Continuous monitoring of vital signs was performed with a multiparameter monitor, including rectal temperature, heart rate, breathing frequency, and oxygen saturation. The tests were performed on a radiant warmer set at a constant level. A neonatal T-piece resuscitator (NeoPuff, Fisher & Paykel, Auckland, New Zealand) was connected to a compressed air/oxygen blender with gas flow set at 10 L/min for all tests. The 3 options for interventions were HME (Humid-Vent Mini, Teleflex), active HH (MR850, Fisher & Paykel), and no device (Fig. 1). The HH was turned on in advance to achieve a full warm-up for each test.

Fig. 1.
  • Download figure
  • Open in new tab
  • Download powerpoint
Fig. 1.

Schematic picture of the circuits with heat-and-moisture exchanger (1) and heated humidifier (2). A: gas blender and flow meter; B: Neopuff infant T-piece resuscitator; C: T-piece; D: heat-and-moisture exchanger; E: thermo-hygrometer; F: endotracheal tube; G: rabbit; H: heated humidifier. The figure is not drawn to scale.

After intubation, adequate ventilation parameters were used for adult New Zealand rabbits, as indicated by the veterinary team. Starting settings were peak inspiratory pressure 15 cm H2O, PEEP 4 cm H2O, and breathing frequency 50 breaths/min. A metronome was used to ensure consistency in the breathing frequency.19 The animals were in a deep anesthetic plane during the tests, so all breaths were delivered actively through the T-piece resuscitator. Each test lasted 20 min. The initial parameters could be modified according to oxygen saturation and heart rate. During the tests, the relative humidity and temperature at the level of the endotracheal tube were recorded with a thermo-hygrometer (Testo 605-H1, Lenzkirch, Germany), and absolute humidity was calculated. These values were measured at the beginning of the test (ie, baseline) and then every 5 min up to 20 min, with 5 measurements per test. The thermo-hygrometer was placed on the edge of the radiant warmer, wrapped in thick plastic tubing to avoid receiving radiant energy. At the end of the test, sedation was suspended, and the animals were returned to the animal facility. Three tests were performed for each animal, one test for each arm of the study, with at least 24 h between each test. Each animal was identified and the order in which the tests were carried out was randomized using permuted blocks (ie, blocks of 4 or 6 opaque and sealed envelopes). Because the use of cold and dry gases could cause lung damage, randomization was limited to HH or HME, and the test without gas conditioning was performed last for each rabbit.

Study Outcomes

The primary outcome was the absolute humidity value measured at the beginning and every 5 min until the end of the test. Secondary outcomes variables included incidence of hypothermia (ie, rectal temperature < 38.5°C) and differences in oxygen saturation (Embedded Image), ventilator pressures, Embedded Image, and heart rate between the beginning and the end of each procedure.

Measurements

Temperature values of the exhaled air and relative humidity were assessed with a thermo-hygrometer. Absolute humidity was calculated from those values and registered at 0, 5, 10, 15, and 20 min of each test. Hypothermia was defined as rectal temperature < 38.5°C. Differences (Δ) in Embedded Image, peak inspiratory pressure, Embedded Image, and heart rate between the beginning and end of each procedure were recorded and compared.

Data Analysis

Continuous variables are expressed as mean (SD) or median (interquartile range [IQR]) according to their distribution and analyzed using the Shapiro-Wilk normality test. Differences among the means were analyzed using analysis of variance, and the Bonferroni test was performed for pairwise comparison between group means. For non-normal data distribution, the Kruskal-Wallis test was used. Absolute and relative frequencies were used to report categorical variables. Differences were calculated using chi-square test. Statistical analyses were performed using Stata Software 13 (StataCorp, College Station, Texas). Results with P values < .05 were considered statistically significant. Because there were no previous data to suggest the effect size, a sample size was not calculated for this research study. It was estimated that 12 rabbits were an adequate sample.

Results

A total of 36 tests were performed. The mean weight of the rabbits was 3,180 ± 400 g. One test from the HME arm was excluded from analysis due to a large leakage from the endotracheal tube, which could have altered the measurements. The absolute humidity values were different in the 3 arms at every measured time point. Mean absolute humidity values at the end of the tests were statistically different between the 3 studied arms: 38.2 ± 1.7 g/m3 for HH, 28.9 ± 4.7 g/m3 for HME, and 13.9 ± 5.1 g/m3 without conditioning (P = .003) (Fig. 2). The mean temperature in the airway was 34.3°C (IQR 33–34.8) with HH, 31.4°C (IQR 30–33) with HME, and 28.7°C (IQR 27.8–29.7) without conditioning (P < .001). Seven episodes of hypothermia were recorded: 3 episodes with HH, 2 with HME, and 2 with unconditioned gases. All of the episodes happened at the beginning of the test. There were no significant differences between the starting temperature and the ending temperature, for any of the arms of the study (Δ temperature). There were no significant differences in ambient temperature for any of the 3 arms: HH 25 ± 1.2°C, HME 24.7 ± 1.0°C, and no device 24.2 ± 1.2°C (P = .86). There were no significant differences neither in the vital signs recorded, nor in respiratory support at the beginning or at the end of the procedures (Table 1).

Fig. 2.
  • Download figure
  • Open in new tab
  • Download powerpoint
Fig. 2.

Mean and SD of absolute humidity as a function of time, according to the device used. HH = heated humidifier; HME = heat-and-moisture exchanger. Analysis of variance: Baseline P = .01; 5 minutes P = .01; 10 minutes P = .01; 15 minutes P = .006; 20 minutes P = .003.

View this table:
  • View inline
  • View popup
  • Download powerpoint
Table 1.

Vital Signs and Respiratory Support at the Beginning and at the End of the Procedures

Discussion

The absolute humidity measured using HH was not only higher but also adequate, according to recommended values for invasive ventilation. The use of HME resulted in lower values of absolute humidity compared to HH, but these values were significantly higher than those obtained in the tests without conditioning. These results could be reasonably interpreted as expected. However, to our knowledge, there are no published studies evaluating the use of HME devices for T-piece resuscitator ventilation.

A few studies have evaluated the absolute humidity obtained with the use of HME compared with HH in mechanically ventilated neonatal subjects. Fassassi et al16 compared the absolute humidity levels obtained with HH and HME, reporting similar results, although their HME absolute humidity values were somewhat higher than those obtained in our study. Schiffmann et al18 found no difference in the use of the 2 methods of humidification in 40 neonates and infants who needed mechanical ventilation. Luchetti et al17 compared 2 HME systems in mechanically ventilated children under 10 kg and reported values similar to ours. Chikata et al20 evaluated the effect of expiratory leakage in HME systems in an artificial lung model and reported absolute humidity values that were within the range of those observed in our study. Finally, Owen et al15 evaluated the use of an HH associated with a T-piece resuscitator, with higher absolute humidity values than those observed in our study. The variable values of absolute humidity obtained in all of these studies could be explained by the different measurement methods used, clinical conditions, ventilation methods, and the mixture of gases used. In all of these studies, the absolute humidity values obtained with HME devices were lower than or similar to those obtained with HH. However, the levels of absolute humidity were close to 30 mg/L, which is the minimum value recommended by the American Association for Respiratory Care.21

Despite lower absolute humidity values with HME devices, a recent systematic review of studies including both adult and pediatric subjects did not find significant differences in clinical outcomes.22

According to some published studies, one of the main limitations when using HME is increased endotracheal tube leaks of exhaled gases, which results in decreased effectiveness.20,23 In this study, we observed absolute humidity values below recommended values in a few rabbits, which improved when the airway was repositioned and the peritube leakage was reduced. Because respiratory function monitoring was not available for the study, the magnitude of the leakage could not be measured.

The efficiency of the HME device depends on the exhaled air temperature, so it is possible that the absolute humidity values are lower in hypothermic neonates. However, in the few HME-ventilated rabbits that experienced hypothermia, the obtained absolute humidity values were similar to the ones found in normothermic rabbits. Because all the hypothermia episodes were at the beginning of the test, they could not be related to the method of airway humidification, but rather to the ambient temperature and the anesthetic induction. It is important to mention that the experimental design of our study is different from real-life situations in a delivery room. Human neonates have no fur coat and are born wet. Although it is true that the animals were dry and warm, rabbits are particularly prone to develop hypothermia due to peripheral vasodilatation when under deep anesthesia. The latter was reflected in the fact that 7 animals had episodes of hypothermia. We evaluated these systems in a model in which the animals did not breathe spontaneously, so it may not reflect the way in which all newborns are ventilated.

One known issue with HME systems is the increased dead space, which can prevent adequate ventilation and can lead to unacceptably high CO2 levels. It has been reported that spontaneously breathing patients increase their breathing frequency to keep adequate minute ventilation, whereas paralyzed patients rely on ventilator parameters and may have some degree of hypercapnia.24 In our study, we did not analyze blood gases, so we cannot speculate about this issue, which may be important in actual clinical scenarios in neonates.

Both HH and HME devices achieved absolute humidity values above the minimum recommended by the American Association for Respiratory Care for each type of device; however, these results have to be considered carefully because the standards were developed for the adult population.21,25,26 As expected, with unconditioned gases, the recommended minimum humidity values were not reached at any time.

Randomized trials have indicated that gas conditioning in the delivery room using HH results in lower incidence of hypothermia in preterm infants on neonatal ICU admission. However, it has been stated that, considering the lack of effect on major outcomes and the cost of the equipment, the use of HH in the delivery room should be further evaluated before it is recommended for clinical practice.8–10,27 While awaiting more studies in neonates, alternative ways to provide adequate absolute humidity levels are being considered. Because the HME can be used immediately with no previous preparation, it could be particularly useful during emergency deliveries, when there is no time to prepare or set up more complicated equipment. Premature babies present a different kind of challenge: they have lower tidal volumes and high risk for CO2 retention and hypothermia. When there is a nonemergency premature delivery, the optimal gas-conditioning device could be prepared and preheated in advance.

Conclusions

In an animal model of ventilation with a T-piece resuscitator, the absolute humidity obtained with the active humidification system (ie, heated humidifier) was higher and within the recommended range for mechanical ventilation. Although lower values of absolute humidity were obtained with the heat-and-moisture exchange system, these values were higher than those observed in the tests without conditioning, and its use in the delivery room could be a feasible alternative, especially for brief periods of time. Human studies are needed to evaluate whether the difference between the 2 methods is clinically important. The absolute humidity levels obtained when using unconditioned gases were well below the recommended values, and its use should be discouraged. Our study adds to this recommendation.

Acknowledgments

We deeply thank Prof. Laura Konikoff and Prof. Nestor Vain for their critical advice and thorough review of the manuscript.

Footnotes

  • Correspondence: Agustín Bernatzky MD, Hospital Italiano de Buenos Aires, Department of Pediatrics, Neonatology Division, Tte. Gral. Juan Domingo Perón 4190, Buenos Aires, C1199ABB, Argentina. E-mail: agustin.bernatzky{at}hospitalitaliano.org.ar
  • Dr Bernatzky presented a version of this report at the Pediatric Academic Societies Annual Meeting, held May 8, 2018, in Toronto, Canada.

  • This work was supported by Foundation Dr Carlos A Gianantonio, Buenos Aires, Argentina.

  • The authors have disclosed no conflicts of interest.

  • Copyright © 2020 by Daedalus Enterprises

References

  1. 1.↵
    1. Horstmann G,
    2. Iravani J,
    3. Norris Melville G,
    4. Richter HG
    . Influence of temperature and decreased water content of inspired air on the ciliated bronchial epithelium a physiological and electron microscopical study. Acta Otolaryngol 1977;84(1-2):124–131.
    OpenUrlCrossRefPubMed
  2. 2.
    1. Jiang M,
    2. Song J-J,
    3. Guo X-L,
    4. Tang Y-L,
    5. Li H-B
    . Airway humidification reduces the inflammatory response during mechanical ventilation. Respir Care 2015;60(12):1720–1728.
    OpenUrlAbstract/FREE Full Text
  3. 3.
    1. Kilgour E,
    2. Rankin N,
    3. Ryan S,
    4. Pack R
    . Mucociliary function deteriorates in the clinical range of inspired air temperature and humidity. Intensive Care Med 2004;30(7):1491–1494.
    OpenUrlPubMed
  4. 4.↵
    1. Sottiaux TM
    . Consequences of under- and over-humidification. Respir Care Clin N Am 2006;12(2):233–252.
    OpenUrlPubMed
  5. 5.↵
    1. Greenspan JS,
    2. Wolfson MR,
    3. Shaffer TH
    . Airway responsiveness to low inspired gas temperature in preterm neonates. J Pediatr 1991;118(3):443–445.
    OpenUrlCrossRefPubMed
  6. 6.
    1. Schulze A
    . Respiratory gas conditioning and humidification. Clin Perinatol 2007;34(1):19–33.
    OpenUrlCrossRefPubMed
  7. 7.↵
    1. Tarnow-Mordi WO,
    2. Reid E,
    3. Griffiths P,
    4. Wilkinson AR
    . Low inspired gas temperature and respiratory complications in very low birth weight infants. J Pediatr 1989;114(3):438–442.
    OpenUrlCrossRefPubMed
  8. 8.↵
    1. McGrory L,
    2. Owen LS,
    3. Thio M,
    4. Dawson JA,
    5. Rafferty AR,
    6. Malhotra A,
    7. et al
    . A randomized trial of conditioned or unconditioned gases for stabilizing preterm infants at birth. J Pediatr 2018;193:47–53.
    OpenUrl
  9. 9.
    1. Meyer MP,
    2. Hou D,
    3. Ishrar NN,
    4. Dito I,
    5. Te Pas AB
    . Initial respiratory support with cold, dry gas versus heated humidified gas and admission temperature of preterm infants. J Pediatr 2015;166(2):245–250.
    OpenUrlCrossRefPubMed
  10. 10.↵
    1. Te Pas AB,
    2. Lopriore E,
    3. Dito I,
    4. Morley CJ,
    5. Walther FJ
    . Humidified and heated air during stabilization at birth improves temperature in preterm infants. Pediatrics 2010;125(6):e1427–e1432.
    OpenUrlAbstract/FREE Full Text
  11. 11.↵
    1. de Almeida MFB,
    2. Guinsburg R,
    3. Sancho GA,
    4. Rosa IRM,
    5. Lamy ZC,
    6. Martinez FE,
    7. et al
    . Hypothermia and early neonatal mortality in preterm infants. J Pediatr 2014;164(2):271–275.
    OpenUrlCrossRefPubMed
  12. 12.
    1. Laptook AR,
    2. Salhab W,
    3. Bhaskar B
    , Neonatal Research Network. Admission temperature of low birth weight infants: predictors and associated morbidities. Pediatrics 2007;119(3):e643–e649.
    OpenUrlAbstract/FREE Full Text
  13. 13.↵
    1. Lyu Y,
    2. Shah PS,
    3. Ye XY,
    4. Warre R,
    5. Piedboeuf B,
    6. Deshpandey A,
    7. et al
    . Association between admission temperature and mortality and major morbidity in preterm infants born at fewer than 33 weeks' gestation. JAMA Pediatr 2015;169(4):e150277.
    OpenUrl
  14. 14.↵
    1. Dawson JA,
    2. Owen LS,
    3. Middleburgh R,
    4. Davis PG
    . Quantifying temperature and relative humidity of medical gases used for newborn resuscitation. J Paediatr Child Health 2014;50(1):24–26.
    OpenUrlCrossRefPubMed
  15. 15.↵
    1. Owen LS,
    2. Dawson JA,
    3. Middleburgh R,
    4. Buttner S,
    5. McGrory L,
    6. Davis PG
    . Feasibility and practical considerations for heating and humidifying gases during newborn stabilisation: an in vitro model. Neonatology 2014;106(2):156–162.
    OpenUrl
  16. 16.↵
    1. Fassassi M,
    2. Michel F,
    3. Thomachot L,
    4. Nicaise C,
    5. Vialet R,
    6. Jammes Y,
    7. et al
    . Airway humidification with a heat and moisture exchanger in mechanically ventilated neonates: a preliminary evaluation. Intensive Care Med 2007;33(2):336–343.
    OpenUrlCrossRefPubMed
  17. 17.↵
    1. Luchetti M,
    2. Pigna A,
    3. Gentili A,
    4. Marraro G
    . Evaluation of the efficiency of heat and moisture exchangers during paediatric anaesthesia. Paediatr Anaesth 1999;9(1):39–45.
    OpenUrlPubMed
  18. 18.↵
    1. Schiffmann H,
    2. Rathgeber J,
    3. Singer D,
    4. Harms K,
    5. Bolli A,
    6. Züchner K
    . Airway humidification in mechanically ventilated neonates and infants: a comparative study of a heat and moisture exchanger vs. a heated humidifier using a new fast-response capacitive humidity sensor. Crit Care Med 1997;25(10):1755–1760.
    OpenUrlPubMed
  19. 19.↵
    1. Cocucci C,
    2. Madorno M,
    3. Aguilar A,
    4. Acha L,
    5. Szyld E,
    6. Musante G
    . A metronome for pacing manual ventilation in a neonatal resuscitation simulation. Arch Dis Child Fetal Neonatal Ed 2015;100(1):F47–F49.
    OpenUrlAbstract/FREE Full Text
  20. 20.↵
    1. Chikata Y,
    2. Sumida C,
    3. Oto J,
    4. Imanaka H,
    5. Nishimura M
    . Humidification performance of heat and moisture exchangers for pediatric use. Crit Care Res Pract 2012;2012:439267.
    OpenUrlPubMed
  21. 21.↵
    1. Restrepo RD,
    2. Walsh BK
    American Association for Respiratory Care, Restrepo RD, Walsh BK. Humidification during invasive and noninvasive mechanical ventilation: 2012. Respir Care 2012;57(5):782–788.
    OpenUrlAbstract/FREE Full Text
  22. 22.↵
    1. Gillies D,
    2. Todd DA,
    3. Foster JP,
    4. Batuwitage BT
    . Heat and moisture exchangers versus heated humidifiers for mechanically ventilated adults and children. Cochrane Database Syst Rev 2017;9:CD004711.
    OpenUrl
  23. 23.↵
    1. Tilling SE,
    2. Hayes B
    . Heat and moisture exchangers in artificial ventilation: an experimental study of the effect of gas leakage. Br J Anaesth 1987;59(9):1181–1188.
    OpenUrlCrossRefPubMed
  24. 24.↵
    1. Campbell RS,
    2. Davis K Jr.,
    3. Johannigman JA,
    4. Branson RD
    . The effects of passive humidifier dead space on respiratory variables in paralyzed and spontaneously breathing patients. Respir Care 2000;45(3):306–312.
    OpenUrlPubMed
  25. 25.↵
    International Organization for Standardization. Respiratory tract humidifiers for medical use – Particular requirements for respiratory humidification systems. ISO 8185. 2007.
  26. 26.↵
    International Organization for Standardization. Medical electrical equipment – Part 2–74: Particular requirements for basic safety and essential performance of respiratory humidifying equipment. ISO 80601-2-74. 2017
  27. 27.↵
    1. Carlo WA,
    2. Chatburn RL
    . Is it necessary to heat and humidify respiratory gases for resuscitation in preterm infants? J Pediatr 2018;193:10–11.
    OpenUrl
PreviousNext
Back to top

In this issue

Respiratory Care: 65 (9)
Respiratory Care
Vol. 65, Issue 9
1 Sep 2020
  • Table of Contents
  • Table of Contents (PDF)
  • Cover (PDF)
  • Index by author

 

Print
Download PDF
Article Alerts
Sign In to Email Alerts with your Email Address
Email Article

Thank you for your interest in spreading the word on American Association for Respiratory Care.

NOTE: We only request your email address so that the person you are recommending the page to knows that you wanted them to see it, and that it is not junk mail. We do not capture any email address.

Enter multiple addresses on separate lines or separate them with commas.
Heated Humidifier versus Heat-and-Moisture Exchanger During Positive Pressure Ventilation With a T-Piece Resuscitator in Rabbits
(Your Name) has sent you a message from American Association for Respiratory Care
(Your Name) thought you would like to see the American Association for Respiratory Care web site.
CAPTCHA
This question is for testing whether or not you are a human visitor and to prevent automated spam submissions.
Citation Tools
Heated Humidifier versus Heat-and-Moisture Exchanger During Positive Pressure Ventilation With a T-Piece Resuscitator in Rabbits
Agustín Bernatzky, María F Galletti, Santiago E Fuensalida, Gonzalo L Mariani
Respiratory Care Sep 2020, 65 (9) 1295-1300; DOI: 10.4187/respcare.06997

Citation Manager Formats

  • BibTeX
  • Bookends
  • EasyBib
  • EndNote (tagged)
  • EndNote 8 (xml)
  • Medlars
  • Mendeley
  • Papers
  • RefWorks Tagged
  • Ref Manager
  • RIS
  • Zotero

Share
Heated Humidifier versus Heat-and-Moisture Exchanger During Positive Pressure Ventilation With a T-Piece Resuscitator in Rabbits
Agustín Bernatzky, María F Galletti, Santiago E Fuensalida, Gonzalo L Mariani
Respiratory Care Sep 2020, 65 (9) 1295-1300; DOI: 10.4187/respcare.06997
del.icio.us logo Digg logo Reddit logo Twitter logo CiteULike logo Facebook logo Google logo Mendeley logo
  • Tweet Widget
  • Facebook Like
  • Google Plus One

Jump to section

  • Article
    • Abstract
    • Introduction
    • Methods
    • Results
    • Discussion
    • Conclusions
    • Acknowledgments
    • Footnotes
    • References
  • Figures & Data
  • Info & Metrics
  • References
  • PDF

Related Articles

Cited By...

Keywords

  • gas conditioning
  • respiratory support
  • delivery room
  • newborn resuscitation
  • humidification

Info For

  • Subscribers
  • Institutions
  • Advertisers

About Us

  • About the Journal
  • Editorial Board

AARC

  • Membership
  • Meetings
  • Clinical Practice Guidelines

More

  • Contact Us
  • RSS
American Association for Respiratory Care

Print ISSN: 0020-1324        Online ISSN: 1943-3654

© Daedalus Enterprises, Inc.

Powered by HighWire