To the Editor:
We read and appreciate the interesting article entitled “Impact of Early Passive Exercise With Cycle Ergometer on Ventilator Interaction” by Silva et al.1 The study aimed to assess the patient-ventilator interaction during passive mobilization performed with a cycle ergometer. This study included deeply sedated individuals who were mechanically ventilated for < 72 h. Parameters such as the asynchrony index and the types of asynchrony were evaluated for 20 min, followed by 10 min of recovery.
This issue is extremely relevant because patient-ventilator asynchrony is associated with worse clinical outcomes such as re-intubation, increased length of ICU and hospital stay, and mortality.2,3 Thus, we have some concerns regarding the study.
First, the sample was composed of 113 patients, and only 8 subjects were included. Despite the exclusion criteria, most individuals were not submitted to the passive mobilization protocol due to hemodynamic instability (ie, 45 patients). The authors described that the ICU was a clinical-surgical unit and used severity scales to characterize the sample (Sequential Organ Failure Assessment and Simplified Acute Physiology Score II); however, they de-scribed a significant number of exclusions in a study that was conducted over approximately 2 years. Thus, we believe that there is a need to characterize the enrolled subjects better. In this sense, we also believe that a small sample and the lack of sample size calculation can hardly infer results on the impact of mobilization on ventilatory asynchrony.
Second, although the authors used the Richmond Agitation-Sedation Scale, the variability allowed (ie, –3 to –5) associated with a small sample size may also interfere with both the results and the occurrence of asynchrony. Furthermore, it would be essential to standardize the analgesia assessment to exclude the possibility of asynchrony caused by pain or discomfort. The pain assessment and monitoring using appropriate scales (eg, Critical care pain observation tool or Behavioral pain scale)4 would also be essential to comprehend whether asynchrony was due to pain or discomfort during mobilization.
Third, only one (blinded) researcher evaluated asynchrony. It would be interesting to have more than one evaluator and compare the results to exclude bias. The sensitivity and positive predictive values of analyzing breath-by-breath waveforms are low (22% and 32%, respectively), and trained professionals recognize less than one third of asynchronies. It is also important to highlight that clinicians' ability to identify asynchronies visually is influenced by their ex-pertise and the type of asynchrony.5
Although the types of asynchrony assessed were described in the methodology, the authors did not include the reverse-triggering, a common asynchrony found in deeply sedated patients,6–9 and this type of asynchrony can be confounded with insufficient flow. Su et al8 observed that 30% of their subjects presented reverse-triggering in the late inspiratory and early expiratory phases.
The most frequent asynchronies observed during the passive exercise were ineffective triggering and insufficient flow. An inspiratory-flow mismatch is more common in constant modes such as volume controlled continuous mandatory ventilation10; however, in this study, all subjects received pressure controlled ventilation.
Therefore, the authors need to supplement the information concerning the admitted subjects' characterization, which pro-bably justifies the large number of excluded patients. We would like to know whether the authors calculated the sample size of this study. It is important to clarify the details of the pain monitoring to comprehend whether asynchrony was due to pain or discomfort during mobilization. It is also essential to clarify why only one reviewer evaluated asynchrony and why reverse-triggering was not added to the evaluation.
Footnotes
- Correspondence: Kelly Cattelan Bonorino MSc, R. Professora Maria Flora Pausewang, CEP 88036–800, Florianópolis, Brazil. E-mail: fisiocattelan{at}gmail.com
The authors have disclosed no conflicts of interest.
- Copyright © 2021 by Daedalus Enterprises