BACKGROUND: A model system described in International Organization for Standardization 9360 is the standard method for estimating the humidifying performance of heat and moisture exchangers (HMEs). However, there are no reliable bedside methods for evaluating the ongoing humidification performance of HMEs. Therefore, this study aimed to develop 2 clinically applicable methods for estimating the ongoing humidifying performance of HMEs and to evaluate their reliability in a model system.
METHODS: Physiologically expired gas was simulated using a heated humidifier, and ventilation was delivered using a ventilator with constant flow through 3 different types of HMEs. Relative humidity (RH) was measured using a capacitive-type moisture sensor. Water content lost during expiration was calculated by integrating absolute humidity (AH), instantaneous gas flow measured at the expiratory outlet of the ventilator, and time. We also calculated the water content released and captured by the HMEs during tidal ventilation by integrating the difference in AH across the HMEs, instantaneous gas flow, and time.
RESULTS: We found that the RH, temperature, and AH were almost constant on the expiratory outlet of the ventilator but rapidly varied near the HMEs. The water content lost by the 3 HMEs was associated with the manufacturer-reported values and inversely correlated with the calculated values of the water content exchanged by the HMEs. The water content released and captured by HMEs was closely correlated with the difference in HME weight measured at the end of inspiration and expiration; however, the water content captured by HMEs seemed to be overestimated.
CONCLUSIONS: Our results demonstrated that our system was able to detect the differences in the performance of 3 models of HMEs and suggest that our method for calculating water loss is reliable for estimating the water retention performance of HMEs during mechanical ventilation, even in the presence of a constant flow.
- Heat and moisture exchanger
- humidification
- humidity
- mechanical ventilation
- water content
- water exchange
- water retention
Footnotes
- Correspondence: Tomio Andoh MD PhD, Department of Anesthesiology, Teikyo University School of Medicine Hospital Mizonokuchi, 5–1-1 Futako, Takatsu-ku, Kawasaki, Japan, 213–8507. E-mail: tandoh{at}med.teikyo-u.ac.jp
The authors have disclosed no conflicts of interest.
This study was supported by a research fund from the Department of Anesthesiology.
Supplementary material related to this paper is available at http://www.rcjournal.com.
- Copyright © 2024 by Daedalus Enterprises
Pay Per Article - You may access this article (from the computer you are currently using) for 1 day for US$30.00
Regain Access - You can regain access to a recent Pay per Article purchase if your access period has not yet expired.