Abstract
Background: Aerosol delivery during invasive mechanical ventilation is significantly affected by multiple factors related to the patient, device, and ventilator settings. Thus, we aimed to assess the impact of different ventilator settings, nebulizer positions, bias flows, circuit adaptors, and humidification on aerosol delivery during invasive ventilation.
Methods: An experimental setup was used to evaluate aerosol delivery using a vibrating mesh nebulizer (VMN) during invasive ventilation (Figure 1). Albuterol at 2.5 mg/3 mL was used under different ventilator settings, nebulizer placements, bias flows, and circuit connectors in the dry circuit. The exhaled humidity and the use of heated humidified circuits were only used in some settings. Each condition was repeated five times. The drug was eluted from the collecting filter and assayed using ultraviolet spectrophotometry at 276 nm.
Results: The inhaled dose was higher with high settings compared to low settings regardless of nebulizer position, bias flow, heated humidification, and presence or absence of exhaled humidity (Table 1) (all P < .05) except at 2 L/min bias flow with exhaled humidity and VMN placed proximal to the endotracheal tube in dry circuits. The presence of exhaled humidity or heated humidification tended to diminish the differences between high and low settings in the inhaled dose. In dry circuits, no significant differences were observed in the inhaled dose when comparing Y-shaped and V-shaped connectors, except when the nebulizer was placed proximally at a bias flow of 2 L/min. Also, in dry circuits, proximal placement of the nebulizer at high settings generated higher inhaled doses than distal placement, regardless of the bias flow and connector type used. At low settings, with a Y-connector, no significant differences in the inhaled dose were observed between the two nebulizer placements.
Conclusions: In dry circuits, high ventilator settings generated a greater inhaled dose than low settings, with the nebulizer placed 30 cm proximal to the circuit connector showing higher inhaled doses at high settings. Exhaled humidity or the use of heated humidified circuits reduced the differences between high and low settings.
Footnotes
Commercial Relationships: Dr. Li discloses relationships with Fisher & Paykel Healthcare Ltd, Aerogen Ltd, the Rice Foundation Ltd, MEK ICS Co. Ltd, Vincent Ltd, the American Association for Respiratory Care, and Heyer. Dr. Li is the section editor for RESPIRATORY CARE. Albuainain, Man and Alamoudi have no conflict to disclose.
- Copyright © 2024 by Daedalus Enterprises